← Product Code [MOS](/productcode/MOS) · K212864

# dS TorsoCardiac 1.5T, dS MSK S 1.5T, dS MSK M 1.5T (K212864)

_Philips Healthcare (Suzhou) Co., Ltd. · MOS · Dec 1, 2021 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212864

## Device Facts

- **Applicant:** Philips Healthcare (Suzhou) Co., Ltd.
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Dec 1, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Magnetic Resonance (MR) coil is used with a Philips 1.5T MR scanner. A trained physician interprets the diagnostic images (of the anatomy of interest) produced.

## Device Story

Phased-array receive-only MR coils; used with Philips 1.5T scanners. Input: RF signals from patient anatomy (torso, cardiac, neck, wrist, foot, ankle, elbow, shoulder, hip). Operation: Coils capture MR signals; transmit to scanner via interface; no internal energy source. Construction: PU foil, flexible PCBA, EVA foam; bendable design for patient fit. Output: Diagnostic images for physician interpretation. Used in clinical imaging environments; operated by trained technicians/physicians. Benefits: High-resolution imaging; improved flexibility and serviceability compared to rigid predecessors.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including SNR (NEMA MS 1), image uniformity (NEMA MS 3), phased array characterization (NEMA MS 9), and RF coil heating (NEMA MS 14).

## Technological Characteristics

Phased-array receive-only coils. Materials: PU foil, EVA30 foam, flexible PCBA. Energy: Derived from MR scanner. Connectivity: DCI connector to MR system. Standards: AAMI/ANSI ES60601-1, IEC 60601-2-33, IEC 60601-1-2, ISO 14971, IEC 62366.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- SENSE Torso 16 ([K122897](/device/K122897.md))
- SENSE MSK S 8CH 1.5T ([K120122](/device/K120122.md))
- 1.5T 8-CHANNEL MEDIUM GENERAL PURPOSE FLEX COIL ([K111673](/device/K111673.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 1, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left, and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Philips Healthcare (Suzhou) Co., Ltd. % Seaman Shao Associate Regulatory Affairs Manager No.258, ZhongYuan Road, Suzhou Industrial Park Suzhou, Jiangsu 215024 CHINA

Re: K212864

Trade/Device Name: dS TorsoCardiac 1.5T, dS MSK S 1.5T, dS MSK M 1.5T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: August 26, 2021 Received: September 8, 2021

Dear Seaman Shao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212864

Device Name

dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T

Indications for Use (Describe)

The Magnetic Resonance (MR) coil is used with a Philips 1.5T MR scanner. A trained physician interprets the diagnostic images (of the anatomy of interest) produced.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.

| Date Prepared:                  | August 26, 2021                                                                                                                                                             |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:                   | Philips Healthcare (Suzhou) Co., Ltd.<br>No. 258, ZhongYuan Road, Suzhou Industrial Park,<br>Suzhou Jiangsu, CHINA, 215024<br>Establishment Registration Number: 3009529630 |
| Primary<br>Contact Person:      | Seaman Shao<br>Associate Regulatory Affairs Manager<br>Phone: +86-13913116320<br>E-mail: seaman.shao@philips.com                                                            |
| Secondary<br>Contact Person     | Erhong Wang<br>Senior Regulatory Affairs Manager<br>Phone: +86-13021019589<br>E-mail: erhong.wang@philips.com                                                               |
| Device Name:                    | dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T                                                                                                                       |
| Classification:                 | Classification Name: Magnetic resonance diagnostic device                                                                                                                   |
|                                 | Classification<br>Regulation: 21CFR §892.1000                                                                                                                               |
|                                 | Classification Panel: Radiology                                                                                                                                             |
|                                 | Device Class: Class II                                                                                                                                                      |
|                                 | Primary Product code: MOS                                                                                                                                                   |
| Primary<br>Predicate<br>Device: | Trade Name: (1) SENSE Torso 16<br>(2) SENSE MSK S 8CH 1.5T<br>(3) 1.5T 8-CHANNEL MEDIUM GENERAL<br>PURPOSE FLEX COIL                                                        |
|                                 | Manufacturer: Invivo Corporation                                                                                                                                            |
|                                 | 510(k) Clearance: (1) K122897, 21-12-2012<br>(2) K120122, 10-02-2012<br>(3) K111673, 23-12-2011                                                                             |

{4}------------------------------------------------

| Classification Regulation: | 21CFR §892.1000                      |
|----------------------------|--------------------------------------|
| Classification Name:       | Magnetic resonance diagnostic device |
| Classification Panel:      | Radiology                            |
| Device Class:              | Class II                             |
| Product Code:              | MOS                                  |

#### Device The proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T description: are designed to be used in conjunction with MR Scanner to produce diagnostic images of the anatomy of interest that can be interpreted by a trained physician.

The proposed dS TorsoCardiac 1.5T is a phased array receive-only coil for high resolution diagnostic imaging of torso, cardiac, and neck. The coil foam is composed of PU foil, flexible PCB and EVA to provide sufficient flexibility along Left-Right direction for patient body scan. The layers from outside to patient side are: PU foil (outer surface), EVA30 foam, PCBA of the coil and PU foil (inner surface). The foam looks flat at the top surface. A few parts, two feed-board boxes, cable housing and a small connector placed across the central Head-Feet axis are also at the top surface. Inner surface is naturally flat and is bendable along slots to fit well to the patient body.

The proposed dS MSK S 1.5T is a phased array receive-only coil for high resolution diagnostic imaging of wrist, foot-ankle, and elbow. The coil foam is composed of PU foil, flexible PCB and EVA to provide sufficient flexibility along Left-Right direction for patient body scan. The layers from outside to patient side are: PU foil (outer surface), EVA30 foam, PCBA of the coil, EVA 30 foam, and PU foil (inner surface). A few parts, feed-board boxes, cable housing and a small connector placed across the central Head-Feet axis are also at the top surface. Inner surface is naturally flat and is bendable along slots to fit well to the patient body.

The proposed dS MSK M 1.5T is a phased array receive-only coil for high resolution diagnostic imaging of shoulder, knee, hip, foot, and elbow. The coil foam is composed of PU foil, flexible PCB and EVA to provide sufficient flexibility along Left-Right direction for patient body scan. The layers from outside to patient side are: PU foil (outer surface), EVA30 foam, PCBA of the coil, EVA30 foam, and PU foil (inner surface). A few parts, feed-board boxes, cable housing and a small connector placed across

{5}------------------------------------------------

|                                    | the central Head-Feet axis are also at the top surface. Inner surface is naturally flat and is bendable along slots to fit well to the patient body.                               |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                    | The proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T are designed to be used with the Philips 1.5T MRI Scanners.                                                     |
| Indications for Use:               | The indications for use for the proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T are as follows:                                                                     |
|                                    | The Magnetic Resonance (MR) coil is used with a Philips 1.5T MR scanner. A trained physician interprets the diagnostic images (of the anatomy of interest) produced.               |
| Fundamental Scientific Technology: | The 510(k) summary contains a summary of the technological characteristics of the proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T compared to the predicate devices |

| Attribute                                                                                                      | Predicate Device,<br>SENSE Torso 16                                                                                                                                                                                                                                                                       | Proposed dS TorsoCardiac 1.5T                                                                                                                                                                                                                                                                                                                                                                                          |
|----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fundamental Scientific Technology                                                                              |                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Type of coil                                                                                                   | Phased-array Receive<br>only coil                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                                                                                                                                                                                                                              |
| Magnetic<br>Field<br>Orientation<br>(B0)                                                                       | Head-Feet oriented                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                                                                                                                                                                              |
| Frequency<br>range                                                                                             | 63.87MHz+/-0.75MHz                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                                                                                                                                                                              |
| Housing<br>Material                                                                                            | PC and PU                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                                                                                                                                                                                                                                                                                                              |
| Decoupling<br>method                                                                                           | Overlap, pre-amp<br>decouple, Active/Passive<br>PIN diode decoupling                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                              |
| Energy<br>source                                                                                               | Derived from MR<br>scanner, no internal<br>energy source                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                                                              |
| Workflow to<br>support the<br>connection<br>and signal<br>transmission<br>between the<br>coil and MR<br>system | The coil cable connects<br>to MR system tabletop<br>via coaxial and DC pins<br>mixed analogue<br>connector to fit any<br>intended coil position on<br>tabletop according to the<br>needs of MR scan.                                                                                                      | The coil cable connects to<br>dS Interface S 1.5T or dS<br>Interface L 1.5T, which is<br>connected into DCI of MR<br>system tabletop, via a DCI<br>connector with coaxial<br>pins to fit any intended<br>coil position on tabletop<br>according to the needs of<br>MR scan.<br>No impact to device safety<br>and effectiveness since it<br>only provides a better<br>workflow.                                         |
| Flexibility of<br>coil to ensure<br>well fit to<br>patient body                                                | The coil foam is<br>composed of PU foil,<br>flexible PCBA and EVA<br>to provide sufficient<br>flexibility along the Left-<br>Right direction for<br>patient body scan,<br>whereas it is rigid in<br>Head-Feet direction due<br>to the existing of long<br>rigid PCBAs and baluns<br>along this direction. | The coil foam is composed<br>of PU foil, flexible PCBA<br>and EVA to provide<br>sufficient flexibility along<br>the Left-Right direction<br>for patient body scan, and<br>the coil offers a degree of<br>flexibility along the Head-<br>Feet direction since the<br>baluns are moved to the<br>coil cable and the size of<br>rigid PCBAs are<br>decreased.<br>No impact to device safety<br>and effectiveness since it |
|                                                                                                                |                                                                                                                                                                                                                                                                                                           | only provides a better<br>flexibility and workflow.                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                                                |                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Easily<br>removeable<br>structure to<br>support the<br>serviceability<br>of coil                               | The baluns and one of<br>rigid PCBA are<br>centralized in the middle<br>of coil, and a removeable<br>plastic cover is designed<br>to cover them, whereas<br>the other rigid PCBAs<br>can be repaired or<br>replaced only after the<br>entire coil foam is<br>disassembled.                                | All rigid PCBAs are<br>placed at the left and right<br>side of coil, and two<br>removeable plastic covers<br>are designed to cover<br>them, meanwhile the<br>baluns are covered by<br>plastic housings in the coil<br>cable, and they can be also<br>repaired or replaced by<br>removing housings.<br>No impact to device safety<br>and effectiveness since it<br>only provides a better<br>serviceability.            |

{6}------------------------------------------------

{7}------------------------------------------------

| Attribute                                                                                                       | Predicate Device,<br>SENSE MSK S 8CH 1.5T                                                                                                                                                                                                                                                                 | Proposed dS MSK S 1.5T                                                                                                                                                                                                                                                                                                                                                         |
|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fundamental Scientific Technology                                                                               |                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                |
| Type of coil                                                                                                    | Phased-array Receive<br>only coil                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                                                                                                                                                                                      |
| Magnetic<br>Field<br>Orientation<br>(B0)                                                                        | Head-Feet oriented                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                                                                                                                                      |
| Frequency<br>range                                                                                              | 63.87MHz+/-0.75MHz                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                                                                                                                                      |
| Housing<br>Material                                                                                             | PC and PU                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                                                                                                                                                                                                                                                                      |
| Decoupling<br>method                                                                                            | Overlap, pre-amp<br>decouple, Active/Passive<br>PIN diode decoupling                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                      |
| Energy<br>source                                                                                                | Derived from MR<br>scanner, no internal<br>energy source                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                      |
| Workflow to<br>support the<br>connection<br>and signal<br>transmission<br>between the<br>coil and MR<br>system  | The coil cable connects<br>to MR system tabletop<br>via coaxial and DC pins<br>mixed analogue<br>connector to fit any<br>intended coil position on<br>tabletop according to the<br>needs of MR scan.                                                                                                      | The coil cable connects to<br>dS Interface S 1.5T or dS<br>Interface L 1.5T, which is<br>connected into DCI of MR<br>system tabletop, via a DCI<br>connector with coaxial<br>pins to fit any intended<br>coil position on tabletop<br>according to the needs of<br>MR scan.<br>No impact to device safety<br>and effectiveness since it<br>only provides a better<br>workflow. |
| Flexibility of<br>coil to ensure<br>well fit to<br>patient body                                                 | The coil foam is<br>composed of PU foil,<br>flexible PCBA and EVA<br>to provide sufficient<br>flexibility along the Left-<br>Right direction for<br>patient body scan,<br>whereas it is rigid in<br>Head-Feet direction due<br>to the existing of long<br>rigid PCBAs and baluns<br>along this direction. | The coil foam is composed<br>of PU foil, flexible PCBA<br>and EVA to provide<br>sufficient flexibility along<br>the Left-Right direction<br>for patient body scan, and<br>the coil offers a degree of<br>flexibility along the Head-<br>Feet direction since the<br>baluns are moved to the<br>coil cable and the size of<br>rigid PCBAs are decreased<br>significantly.       |
|                                                                                                                 |                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                |
| No impact to device safety<br>and effectiveness since it<br>only provides a better<br>flexibility and workflow. |                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                |

{8}------------------------------------------------

{9}------------------------------------------------

| Attribute                                                                                                      | Predicate Device,<br>1.5T 8-CHANNEL MEDIUM<br>GENERAL PURPOSE FLEX COIL                                                                                                                                                                                                                                   | Proposed dS MSK M 1.5T                                                                                                                                                                                                                                                                                                                                                                                                               |
|----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fundamental Scientific Technology                                                                              |                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Type of coil                                                                                                   | Phased-array Receive<br>only coil                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Magnetic<br>Field<br>Orientation<br>(B0)                                                                       | Head-Feet oriented                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Frequency<br>range                                                                                             | 63.87MHz+/-0.75MHz                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Housing<br>Material                                                                                            | PC and PU                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Decoupling<br>method                                                                                           | Overlap, pre-amp<br>decouple, Active/Passive<br>PIN diode decoupling                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Energy<br>source                                                                                               | Derived from MR<br>scanner, no internal<br>energy source                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Workflow to<br>support the<br>connection<br>and signal<br>transmission<br>between the<br>coil and MR<br>system | The coil cable connects<br>to MR system tabletop<br>via coaxial and DC pins<br>mixed analogue<br>connector to fit any<br>intended coil position on<br>tabletop according to the<br>needs of MR scan.                                                                                                      | The coil cable connects to<br>dS Interface S 1.5T or dS<br>Interface L 1.5T, which is<br>connected into DCI of MR<br>system tabletop, via a DCI<br>connector with coaxial<br>pins to fit any intended<br>coil position on tabletop<br>according to the needs of<br>MR scan.<br>No impact to device safety<br>and effectiveness since it<br>only provides a better<br>workflow.                                                       |
| Flexibility of<br>coil to ensure<br>well fit to<br>patient body                                                | The coil foam is<br>composed of PU foil,<br>flexible PCBA and EVA<br>to provide sufficient<br>flexibility along the Left-<br>Right direction for<br>patient body scan,<br>whereas it is rigid in<br>Head-Feet direction due<br>to the existing of long<br>rigid PCBAs and baluns<br>along this direction. | The coil foam is composed<br>of PU foil, flexible PCBA<br>and EVA to provide<br>sufficient flexibility along<br>the Left-Right direction<br>for patient body scan, and<br>the coil offers a degree of<br>flexibility along the Head-<br>Feet direction since the<br>baluns are moved to the<br>coil cable and the size of<br>rigid PCBAs are decreased<br>significantly.<br>No impact to device safety<br>and effectiveness since it |

{10}------------------------------------------------

{11}------------------------------------------------

|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |  | only provides a better<br>flexibility and workflow. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------------------------|
| Based on the information provided above, dS TorsoCardiac 1.5T is<br>considered substantially equivalent to the primary currently marketed and<br>predicate device SENSE Torso 16 (K122897, 21-12-2012) in terms of<br>fundamental scientific technology<br><br>Based on the information provided above, dS MSK S 1.5T is considered<br>substantially equivalent to the primary currently marketed and predicate<br>device SENSE MSK S 8CH 1.5T (K120122, 10-02-2012) in terms of<br>fundamental scientific technology<br><br>Based on the information provided above, dS MSK M 1.5T is considered<br>substantially equivalent to the primary currently marketed and predicate<br>device 1.5T 8-CHANNEL MEDIUM GENERAL PURPOSE FLEX COIL<br>(K111673, 23-12-2011) in terms of fundamental scientific technology |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |  |                                                     |
| Summary of<br>Non-Clinical<br>Performance<br>Data:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | The proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T<br>complies with the following international and FDA-recognized consensus<br>standards:<br>AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012,<br>C1:2009/(R) 2012 and A2:2010/(R) 2012 (Consolidated Text)<br>Medical Electrical Equipment - Part 1: General Requirements. For<br>Basic Safety And Essential Performance (IEC 60601-1:2012,<br>MOD)<br>FDA/CDRH recognition number 19-4 IEC60601-2-33 Ed. 3.2:2015 Medical electrical equipment - Part 2-<br>33: Particular requirements for the basic safety and essential<br>performance of magnetic resonance equipment for medical<br>diagnosis<br>FDA/CDRH recognition number 12-295 IEC60601-1-2 Ed. 4.0:2014 Medical electrical equipment - Part 1-<br>2: General requirements for basic safety and essential performance<br>- Collateral standard: Electromagnetic disturbances - Requirements<br>and tests<br>FDA/CDRH recognition number 19-8 ISO 14971 Ed. 2.0:2007 Medical devices – Application of risk<br>management to medical devices.<br>FDA/CDRH recognition number 5-40. |  |                                                     |

{12}------------------------------------------------

- Magnetic Resonance (MR) Receive only Coil Performance ● Criteria for Safety and Performance Based Pathway - issued on December 11, 2020
- . Guidance for Industry and FDA Staff - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process" (issued June 16, 2016 - document number 1811)
- IEC 62366 Edition 1.1: 2014-01 Medical devices Application of . usability engineering to medical devices FDA/CDRH recognition number 5-87
- NEMA MS 1-2008(R2020) ● Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging
- NEMA MS 3-2008 (R2020) ● Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
- . NEMA MS 9-2008 (R2020) Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images
- NEMA MS 14-2019 Characterization of Radiofrequency (RF) Coil . Heating in Magnetic Resonance Imaging Systems

There are no risks identified in risk management documentation that require clinical data for the purpose of clinical evaluation: Risk Management Plan as Appendix 001, Risk Management Report as Appendix 002, and Risk Management Matrix as Appendix 003 of dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T.

Sufficient evidence is available to demonstrate the ability of dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T achieve the intended performances during normal condition of use;

Full consistency exists between the state-of-the-art, the evaluated data, the risk management documentation and the information materials supplied.

Therefore, the proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T are substantially equivalent to the primary currently marketed and predicate device SENSE Torso 16 (K122897, 21-12-2012), SENSE MSK S 8CH 1.5T (K120122, 10-02-2012) and 1.5T 8-CHANNEL MEDIUM GENERAL PURPOSE FLEX COIL (K111673, 23-12-2011) in terms of safety and effectiveness.

{13}------------------------------------------------

The proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T Summary of Clinical Data: do not require clinical study since substantial equivalence to the predicate device was demonstrated with the following attributes:

- . Design features.
- . Fundamental scientific technology
- Indications for use; ●
- Safetv and effectiveness ●
- Non-clinical performance testing; .

### Substantial The dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T are substantially equivalent to the primary currently marketed and predicate Equivalence device SENSE Torso 16, SENSE MSK S 8CH 1.5T, 1.5T 8-CHANNEL Conclusion: MEDIUM GENERAL PURPOSE FLEX COIL in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Additionally, substantial equivalence was demonstrated with non-clinical performance tests, which complied with the requirements specified in the international and FDA-recognized consensus standards. The results of these tests demonstrate that dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T met the acceptance criteria and are adequate for this intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K212864](https://fda-staging.innolitics.com/device/K212864)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).