EchoSK and EchoSGyn modules for EchoS Family devices
Applicant
Echolight S.P.A
Product Code
IYO · Radiology
Decision Date
Mar 16, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
1) EchoSK and EchoSGyn Modules are additional software/hardware modules for EchoS Family devices and they are additions in terms of indication for use of the legally marketed EchoS Family. 2) EchoSK optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately trained professional in a clinical setting for musculosketal applications. The EchoSK module can be used both with a convex or a linear probe. 3) EchoSGyn optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately-trained professional in a clinical setting for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications. The EchoSCyn module can be used both with a convex or with an endocavitary probe.
Device Story
EchoSK and EchoSGyn are software modules for the EchoS Family ultrasound system. The device uses convex, linear, or endocavitary probes to transmit ultrasound waves into tissue. Reflected signals are detected, filtered, amplified (18dB), and digitized (16-bit, 40MHz) via an internal electronic board. Data is transferred to a PC via USB for visualization. Operated by trained professionals in clinical settings, the system provides real-time ultrasound imaging for musculoskeletal and OB/GYN diagnostics. Clinicians use these images to visualize anatomical structures, aiding in clinical assessment and decision-making. The device benefits patients by enabling non-invasive diagnostic imaging.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993), electrical safety/EMC (EN 60601-1/1-2), software validation (ISO 62304), and usability (IEC 62366-1).
Technological Characteristics
Portable ultrasound system; ABS plastic enclosure; Echo Blaster 128 CEXT-1Z beamformer; 16-bit/40MHz ADC; 18dB gain amplifier; bandpass filtering. Probes: convex (3.5MHz), linear (7.5MHz), endocavitary (6.5MHz). Connectivity: USB. Power: 12VDC external medical grade transformer. Software: EchoStudio (moderate level of concern).
Indications for Use
Indicated for trained professionals in clinical settings to provide ultrasound images/signals for musculoskeletal applications (EchoSK) and Fetal/OB/GYN applications including visualization of ovaries, follicles, uterus, and pelvic structures (EchoSGyn).
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
ACUSON Sequoia Diagnostic Ultrasound System (K202683)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
Echolight S.p.A Giuseppe Criscenti Quality, Certification and Regulatory Manager Via Raffaello Sanzio, 18 Lecce, LE 73100 ITALY
March 16, 2022
### Re: K212851
Trade/Device Name: EchoSK and EchoSGyn Modules for EchoS Family devices Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: IYO, MUA Dated: February 2, 2022 Received: February 8, 2022
### Dear Giuseppe Criscenti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D. Assistant Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212851
Device Name
EchoSK and EchoSGyn modules for EchoS Family devices
Indications for Use (Describe)
1) EchoSK and EchoSGyn Modules are additional software/hardware modules for EchoS Family devices and they are additions in terms of indication for use of the legally marketed EchoS Family.
2) EchoSK optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately trained professional in a clinical setting for musculosketal applications. The EchoSK module can be used both with a convex or a linear probe.
3) EchoSGyn optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately-trained professional in a clinical setting for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications. The EchoSCyn module can be used both with a convex or with an endocavitary probe.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K212851
# Section 5 510(k) Summary
#### I. SUBMITTER
| Applicant Information | Echolight S.p.A. |
|-----------------------|---------------------------------------------------------------|
| | Via Raffaello Sanzio, 18 |
| | 73100 Lecce, Italy |
| Contact | Dr. Giuseppe Criscenti |
| | (Quality, Certification and Regulatory Manager) |
| Date Prepared | 1st September 2021 |
| II. | DEVICE |
| Device Trade Name | EchoSK and EchoSGyn Modules for EchoSFamily devices |
| Common Name | Musculoskeletal and gynecologic modules for Ultrasound System |
| Classification Name: | Ultrasonic pulsed echo imaging system |
| Regulation Number: | 21 CFR 892.1560 |
| Classification Name: | Bone Sonometer |
| Regulation Number: | 21 CFR 892.1180 |
| Product Code | IYO, MUA |
#### III. PREDICATE DEVICE
## Legally Marketed Primary Predicate Device
| 510(k)<br>Number | Pro Code | Trade Name | Applicant |
|------------------|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| K202683 | 21 CFR 892.1550 | ACUSON Sequoia Diagnostic Ultrasound System,<br>ACUSON SC2000 Diagnostic Ultrasound System,<br>ACUSON Freestyle Diagnostic Ultrasound System,<br>ACUSON S1000, S2000, S3000 Diagnostic<br>Ultrasound Systems, ACUSON P200 Diagnostic<br>Ultrasound System, ACUSON P500 Diagnostic<br>Ultrasound System, ACUSON NX3, NX3 Elite<br>Diagnostic Ultrasound System | Siemens Medical<br>Solutions, USA, Inc. |
### Legally Marketed Additional Predicate Device
| 510(k)<br>Number | Pro Code | Trade Name | Applicant |
|------------------|-----------------|--------------|------------------|
| K202514 | 21 CFR 892.1180 | EchoS Family | Echolight S.p.a. |
{4}------------------------------------------------
#### IV. Device Description
EchoSK and EchoSGyn are optional modules for EchoS Family devices.
EchoSK is an optional module of the EchoStudio software that allows the operator to visualize echographic images for musculoskeletal applications. The EchoSK module can be used both with a convex or a linear probe.
EchoSGyn is an optional module of EchoStudio that allows the operator to visualize echographic images for gynecological/obstetric applications. EchoSGyn module can be used both with a convex or with an endocavitary probe.
#### V. Indication for Use
1) EchoSK and EchoSGyn Modules are additional software modules for EchoS Family devices and they are additions in terms of indication for use of the legally marketed EchoS Family.
2) EchoSK optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately trained professional in a clinical setting for musculoskeletal applications. The EchoSK module can be used both with a convex or a linear probe.
3) EchoSGyn optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately-trained professional in a clinical setting for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications. The EchoSGyn module can be used both with a convex or with an endocavitary probe.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The subject device and the primary predicate device K202683 are both ultrasound devices intended to provide images of inside the body by an appropriately trained professional in a clinical setting for Fetal, OB/GYN and musculoskeletal applications.
EchoSK optional module is intended to provide images of, or signals from, inside the body by an appropriately trained professional in a clinical setting for musculoskeletal applications as the
{5}------------------------------------------------
analogous function performed by the primary predicate K202683 "ACUSON Sequoia Diagnostic Ultrasound System".
EchoSGyn optional module is intended to provide images of, or signals from, inside the body by an appropriately trained professional in a clinical setting for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications as the analogous function performed by the primary predicate K202683 "ACUSON Sequoia Diagnostic Ultrasound System".
Considering the technological characteristics, the subject device and the primary predicate device K202683 are both portable and software controlled diagnostic ultrasound systems that transmit, receive and display ultrasound echo data.
Both subject device and ACUSON Sequoia Diagnostic Ultrasound System are powered by, and communicates with a computer. A transducer is used to generate and transmit ultrasound waves into the tissue. The reflected sound waves are detected and digitized before the data is transferred to the computer. Both devices, use an echographic probe that generates and transmits ultrasound waves into the tissue and the detected analog signals are converted into digital ones by means of an analogto-digital converter.
Both the subject device and the ACUSON Sequoia Diagnostic Ultrasound System, respectively through "EchoSK" and ACUSON proprietary softwares, are used as similar device modules for musculoskeletal applications.
Both the subject device and the ACUSON Sequoia Diagnostic Ultrasound System, respectively through "EchoSGyn" and ACUSON proprietary softwares, are used as similar device modules for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications.
In addition, the technological characteristics of the subject device are the same of the additional predicate EchoS Family K202514 because they share the same hardware (EchoS and EchoStation devices) and software (EchoStudio).
EchoSK and EchoSGyn can be considered as additional modules of EchoS Family devices meaning that the subject device includes also the bone sonometer (it is the same of the additional predicate device) and it is intended to be used as bone sonometer and for muscoloskeletal and gynecological ultrasound applications.
{6}------------------------------------------------
Considering the hardware, subject and additional predicate devices share identical hardware. In particular, the beamformer is equipped with three SMA connectors for the transmission of the analog radio frequency (RF) signal and the track synchronization signals (Line Trigger) and frame (Frame Trigger). The model used is the Echo Blaster 128 CEXT-1Z, size 62x215x165 (mm), weighing 1.6 kg and is powered by an external medical grade transformer that provides 12VDC, 4 A. This device is already CE certified, KFDA and FDA 510 (k).
The RF signal is suitably filtered and amplified, before being converted into a digital signal. The designed filter is a bandpass type: the lower cutoff frequency eliminates the noise introduced by the beamformer, the higher cutoff frequency eliminates any signal aliasing phenomena. The amplifier has a gain of 18dB.
The analog signal is converted into digital by means of an analog-to-digital converter (ADC) with a resolution of 16bit and a sampling frequency of 40MHz.
The data transmission block consists of an FPGA and a microcontroller: the first appropriately recomposes the data received from the converter and the second transfers them to the PC using the USB protocol.
The scanned frames are available in proprietary format on the PC.
The filtering, amplification, digitization and data transmission phases have been implemented on a single electronic board, with reduced dimensions. The connection between the beamformer and the designed electronic board takes place by means of coaxial cables with SMA connectors. The voltage from the external medical power supply is used to power both the beamformer and the electronic boards. Inside the ABS cover there is also a USB HUB with 4 ports, on 3 of which are connected: the USB Type A- Type B cable connected to the beamfomer, the USB type B - Mini USB cable connected to the data acquisition card, the encryption USB pendrive.
All the electronic components and the cables are enclosed in a plastic ABS case.
The ultrasound transducer is a convex probe (CE marked) and it has a nominal central operating frequency of 3.5MHz and a radius of curvature of 60 mm (field of view 59°).
The newly added transducers are:
The linear probes (CE marked) with a nominal central operating frequency of 7.5 MHz and a length of 4 cm (field of view of 39 mm) or 6 cm (field of view of 59 mm).
The endocavitary probe (CE marked) with a nominal central operating frequency of 6.5MHz and a radius of 10 mm (field of view 156°).
{7}------------------------------------------------
Considering the software, subject and additional predicate devices share identical software named EchoStudio. The subject device adds two additional software modules:
- EchoSK optional module for EchoS Family devices is intended to provide images of, or . signals from, inside the body acquired by an appropriately trained professional in a clinical setting for musculoskeletal applications. The EchoSK module can be used both with a convex or a linear probe.
- . EchoSGyn optional module for EchoS Family devices is intended to provide images of, or signals from. inside the body acquired by an appropriately-trained professional in a clinical setting for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications. The EchoSGyn module can be used both with a convex or with an endocavitary probe.
#### VII. Performance Data
The part in contact with patient has been tested and it is compliant with:
Biocompatibility (according to ISO 10993)
The battery of testing included the following tests:
- Cytotoxicity -
- । Sensitization
- Irritation/Intracutaneous -
Electrical Safety, Electromagnetic Compatibility EMC (according to EN 60601-1 and EN 60601-1-2)
The battery of testing included the following tests:
- Electrical Safety Test:(see attachment 15_A) -
- Electromagnetic Compatibility: Test Report (see attachment 15_B and attachment 15_C) -
Software Validation (According to ISO 62304)
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could not directly result in serious injury or death to the patient or operator.
{8}------------------------------------------------
| 510(k) Premarket Notification | Echolight S.p.A. |
|-------------------------------|------------------|
| EchoS Family | rev.06 |
| Section V | pag. 6 of 6 |
The documentation of Software Validation can be found in the annexes of Section 14
Usability (According to IEC 62366-1:2015)
The usability of the product was conducted in accordance with IEC 62366-1:2015 and concerns various aspects of the device use.
The documentation can be found in the annexes of Section 16:
Usability File (see Attachment 16 A) -
# ANIMAL STUDIES
None
### PERFORMANCE TESTING - CLINICAL
None
#### VIII. Conclusions
Echolight S.p.A. has determined, by using comparisons and tests, that EchoSK and EchoSGYN modules for Echostudio software and EchoS Family devices are substantially equivalent to the listed predicate devices in terms of intended use, typical clinical use, operational characteristics, and fundamental technological characteristics. Any differences are considered minor and do not raise new issues of the safety and effectiveness of EchoSK and EchoSGYN modules for Echostudio software and EchoS Family devices device when compared to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.