ComfortPro Blue Nitrile Examination Gloves Powder Free
K212850 · Thai Rubber Industry Company Limited · LZA · Dec 10, 2021 · General Hospital
Device Facts
Record ID
K212850
Device Name
ComfortPro Blue Nitrile Examination Gloves Powder Free
Applicant
Thai Rubber Industry Company Limited
Product Code
LZA · General Hospital
Decision Date
Dec 10, 2021
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
ComfortPro Blue Nitrile Examination Gloves Powder free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Disposable nitrile (NBR) examination gloves; worn on examiner's hands; barrier protection to prevent cross-contamination between patient and examiner. Used in clinical or medical settings. Non-sterile, powder-free, ambidextrous design. Available in sizes XS to XL. Provides physical barrier; protects both patient and clinician during examinations.
Clinical Evidence
No clinical data. Bench testing only, including physical property testing (tensile strength, elongation), water-tightness (ASTM D5151-19), residual powder (ASTM D6124-06), and biocompatibility (ISO 10993-5, 10, 11).
Technological Characteristics
Material: 100% Nitrile (NBR). Standards: ASTM D6319-19, ASTM D5151-19, ASTM D6124-06. Biocompatibility: ISO 10993-5, 10, 11. Form factor: Ambidextrous, powder-free, non-sterile. Sizes: XS, S, M, L, XL.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Palm Care Blue Nitrile Examination Gloves (K202384)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 10, 2021
Thai Rubber Industry Company Limited % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE 114, Aurora, Illinois 60504
Re: K212850
Trade/Device Name: ComfortPro Blue Nitrile Examination Gloves Powder Free Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: August 13, 2021 Received: September 7, 2021
Dear Manoj Zacharias:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212850
#### Device Name
ComfortPro Blue Nitrile Examination Gloves Powder free
Indications for Use (Describe)
ComfortPro Blue Nitrile Examination Gloves Powder free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable):
| <input type="checkbox"/> | Commercial |
|--------------------------|-------------|
| <input type="checkbox"/> | Residential |
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K212850 510(K) SUMMARY AS REQUIRED BY: 21CFR§807.92(C)
## A. APPLICANT INFORMATION
| 510(K) Owner's Name | THAI RUBBER INDUSTRY COMPANY LIMITED |
|---------------------|--------------------------------------------------------------|
| Address | 738 MOO 5, MANAM KOO,<br>PLUAKDAENG<br>RAYONG 21140 THAILAND |
| Phone | +66-098-2520209 |
| Fax | - |
| E-mail | vanida@thairubberindustry.com |
| Contact Person | Ms. Supawadee Phoungthong |
| Designation | Managing Director |
| Contact Number | +66-098-2520209 |
| Contact Email | vanida@thairubberindustry.com |
| Date Submitted | December 9, 2021 |
### B. DEVICE IDENTIFICATION
| Name of the device | ComfortPro Blue Nitrile Examination Gloves<br>Powder Free |
|-----------------------------------|-----------------------------------------------------------|
| Product proprietary or trade name | ComfortPro |
| Common or usual name | Nitrile Examination Gloves |
| Classification name | Polymer Patient Examination Glove |
| Device Classification | Class-1 |
| Product Code | LZA |
| Regulation Number | 21 CFR 880.6250 |
| Review Panel | General Hospital |
### C. PREDICATE DEVICE
| Predicate Device | Hi-Care Thai Gloves Co. Ltd. |
|------------------|------------------------------|
| 510(k) Number | K202384 |
| Regulatory Class | Class 1 |
| Product code | LZA |
### D. DESCRIPTION OF THE DEVICE:
ComfortPro Blue Nitrile Examination Gloves Powder Free is Class I patient examination gloves bearing the product code LZA (21CFR880.6250). The gloves are made from Nitrile (NBR)100%. These gloves are blue in color, non-sterile, ambidextrous, powder free and single use only. Blue Nitrile Examination Gloves Powder Free are available in sizes X-small, Small, Medium, Large and X Large.
### E. INDICATION FOR USE OF THE DEVICE:
ComfortPro Blue Nitrile Examination Gloves Powder Free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
{4}------------------------------------------------
# 510(K) SUMMARY AS REQUIRED BY: 21CFR§807.92(C)
## F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
| CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | | REMARKS |
|--------------------------|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|---------|
| | | PREDICATE | SUBJECT | | |
| 510(K) Number | | K202384 | K212850 | | |
| Name of device | -- | Palm Care Blue Nitrile | ComfortPro Blue Nitrile | | |
| | | Examination Gloves | Examination Gloves | | ---- |
| | | Powder free | Powder Free | | |
| Dimensions- Length | ASTMD 6319-19 | Length Min 230 mm | Length > 230 mm | | Similar |
| | | | Size | Average | |
| | | | X-Small | 250 | |
| | | | Small | 241 | |
| | | | Medium | 245 | |
| | | | Large | 242 | |
| | | | X- Large | 242 | |
| Dimensions- Width | ASTMD 6319-19 | Width Min 95+/-10mm | Width Min 95+/-10mm | | Similar |
| | | (for medium size) | (for medium size) | | |
| | | | Size | Average | |
| | | | X-Small | 80 | |
| | | | Small | 82 | |
| | | | Medium | 93 | |
| | | | Large | 103 | |
| | | | X- Large | 110 | |
| Physical Properties- | ASTMD 6319-19 | Before Ageing | Before Ageing | | |
| Tensile Strength | | Tensile Strength | Tensile Strength > | | |
| | | min 14 Mpa<br>After Ageing | 14 Mpa<br>Size | Average | |
| | | Tensile Strength | | | Similar |
| | | min 14 Mpa | X-Small | 27.2 | |
| | | | Small<br>Medium | 23.2 | |
| | | | | 24 | |
| | | | Large | 26.2 | |
| | | | X- Large | 27.4 | |
| | | | | After Ageing | |
| | | | Tensile Strength > | | |
| | | | 14 Mpa | | |
| | | | Size | Average | |
| | | | X-Small<br>Small | 24.2<br>23.8 | |
| | | | Medium | 27.6 | |
| | | | Large | 27.6 | |
| | | | X- Large | 27.9 | |
| Physical Properties- | ASTMD 6319-19 | Before Ageing | | | Similar |
| Ultimate Elongation | | Ultimate Elongation | Before Ageing<br>Ultimate Elongation ><br>500% | | |
| | | > 500% | | | |
| | | | X-Small | 620 | |
| | | | Small | 290 | |
| | | | Medium | 570 | |
| | | | Large | 570 | |
| | | | X- Large | રરા | |
| | | | | | |
| | | | | | |
| | | | | | |
| CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | REMARKS | |
| | | PREDICATE | SUBJECT | | |
| | ASTMD 6319-19 | After Ageing<br>Ultimate<br>Elongation >400% | After Ageing<br>Ultimate Elongation > 400%<br>X-Small 600<br>Small 560<br>Medium 540<br>Large 540<br>X- Large 550 | Similar | |
| Thickness | ASTMD 6319-19 | Palm min 0.05 mm<br>Finger min 0.05 mm | Palm > 0.05 mm Finger > 0.05 mm<br>Size Palm (Actual value) mm Finger (Actual value) mm<br>X-Small 0.101 0.121<br>Small 0.081 0.097<br>Medium 0.091 0.108<br>Large 0.092 0.108<br>X- Large 0.091 0.111 | Similar | |
| Powder Residue | ASTMD 6319-19 | <2 mg/glove | <2 mg/glove<br>Actual value :0.49 mg/glove | Similar | |
| Biocompatibility | Primary Skin Irritation-<br>ISO 10993-10:2010(E) | Under the<br>condition of study<br>not an irritant | Under the condition of study not<br>an irritant | Same | |
| | Dermal Sensitization-<br>ISO 10993-<br>10:2010(E) | Under the conditions<br>of the study not a<br>sensitizer | Under the conditions of the<br>study not a sensitizer | Same | |
| | In vitro cytotoxicity<br>ISO10993-5 2009(E) | Under the conditions<br>of the study,<br>noncytotoxic | Under the conditions of the study<br>cytotoxic for 100% test item extract<br>and non-cytotoxic at 50%, 25%,<br>12.5% and 6.25% test item extracts.<br>Moreover, under the conditions of<br>the study, non acute systemic toxic. | Similar | |
| | Acute Systemic<br>Toxicity Test ISO<br>10993-11:2017(E) | Under the<br>conditions of study,<br>the device extracts<br>do not pose a<br>systemic toxicity<br>concern. | Under the condition of study, the<br>device extracts did not reveal any<br>systemic toxicity. | Same | |
| Water Tight<br>(1000 ml) | ASTM D 5151-19<br>AQL 2.5 | Passes | Passes | Similar | |
| Indication for Use | -- | Blue Nitrile<br>Examination Gloves<br>Powder Free is<br>disposable devices<br>intended for medical<br>purpose that are<br>worn on the<br>examiner's hand to<br>prevent contaminatio<br>between patient and<br>examiner. | ComfortPro Blue Nitrile<br>Examination Gloves Powder<br>Free is disposable device<br>intended for medical purpose<br>that is worn on the examiner's<br>hand to prevent contamination<br>between patient and examiner. | Similar | |
| CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | REMARKS | |
| | | PREDICATE | SUBJECT | | |
| Material | ASTMD 6319-19 | Nitrile (NBR) | Nitrile (NBR) | Same | |
| Color | -- | Blue | Blue | Same | |
| Size | ASTMD 6319-19 | Extra Small, Small,<br>Medium, Large, Extra<br>Large | Extra Small, Small,<br>Medium, Large, Extra<br>Large | Same | |
| Single Use | Medical Glove<br>Guidance Manual<br>- Labeling | Single Use | Single Use | Same | |
| Sterile/non sterile | -- | Nonsterile | Nonsterile | Same | |
| Powder/Powder free | -- | Powder free | Powder free | Same | |
| Label and Labeling | Meets FDA's label<br>and labeling<br>requirements | Meets FDA's label and<br>labeling requirements | Meets FDA's label and<br>labeling requirements | Same | |
| Manufacturer(s) | -- | Hi-Care Thai Gloves Co.<br>Ltd. Thailand. | THAI RUBBER<br>INDUSTRY CO., LTD.. | ----- | |
Page 2 of 6
{5}------------------------------------------------
# 510(K) SUMMARY
AS REQUIRED BY: 21CFR§807.92(C)
{6}------------------------------------------------
# 510(K) SUMMARY AS REQUIRED BY: 21CFR§807.92(C)
There are no significant differences between the two products and are identical in terms of intended use, materials, design, manufacturing methods. Both devices meet the ASTM standard D 6319-19.
# G. COMPARISON BASED ON ASSESSMENT OF NON-CLINICAL PERFORMANCE DATA
## BENCH TEST DATA
| TEST METHOD | PURPOSE | ACCEPTANCE<br>CRITERIA | RESULT |
|-------------------------------------------------------------------------------------------------------|------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| ASTM D6319-19<br>Standard Specification for nitrile<br>examination gloves for medical<br>Application. | To determine the<br>length of the gloves | Min 230 mm for all<br>sizes | X-Small: 250mm<br>Small:- 241mm<br>Medium: - 245 mm<br>Large:- 242 mm<br>X-Large- 242 mm |
| ASTM D6319-19<br>Standard Specification for nitrile<br>examination gloves for medical<br>Application. | To determine the<br>width of the gloves | X-Small:-70+/-10 mm<br>Small:-80+/-10mm<br>Medium:-95+/-10 mm<br>Large:-110+/-10 mm<br>X Large:- 120+/-10 mm | X-Small:- 80 mm<br>Small:- 82mm<br>Medium:-93mm<br>Large:- 103mm<br>X-Large-110mm |
{7}------------------------------------------------
# 510(K) SUMMARY AS REQUIRED BY: 21CFR§807.92(C)
| TEST<br>METHOD | PURPOSE | ACCEPTANCE<br>CRITERIA | RESULT | | |
|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|----------------------|---------------------|
| ASTM D6319-19<br>Standard<br>Specification for<br>nitrile examination<br>gloves for medical<br>Application. | To determine the<br>thickness of the<br>gloves | Palm<br>0.05 mm min<br>Finger<br>0.05 mm min<br>for all sizes | Size | Palm | Finger |
| | | | X-Small | 0.101 mm | 0.121 mm |
| | | | Small | 0.081mm | 0.097mm |
| | | | Medium | 0.091 mm | 0.108mm |
| | | | Large | 0.092mm | 0.108mm |
| | | | X Large | 0.091 mm | 0.111mm |
| ASTM D6319-19<br>Standard<br>Specification for<br>nitrile examination<br>gloves for medical<br>Application. | To Determine the<br>physical properties-<br>Tensile strength | <b>Before Ageing</b><br>Tensile Strength 14<br>Mpa Min for all<br>sizes<br><b>After Ageing</b><br>Tensile Strength<br>14Mpa Min for all<br>sizes | Size | <b>Before ageing</b> | <b>After ageing</b> |
| | | | X-Small | 27.2 Mpa | 24.2 Mpa |
| | | | Small | 23.2 Mpa | 23.8 Mpa |
| | | | Medium | 24 Mpa | 27.6 Mpa |
| | | | Large | 26.2 Mpa | 27.6 Mpa |
| | | | X Large | 27.4 Mpa | 27.9 Mpa |
| | To Determine the<br>physical properties-<br>Ultimate Elongation | <b>Before Ageing</b><br>Ultimate<br>Elongation<br>500% Min for all<br>sizes<br><b>After Ageing</b><br>Ultimate<br>Elongation 400%<br>Min for all sizes | Size | <b>Before ageing</b> | <b>After ageing</b> |
| | | | X-Small | 620% | 600% |
| | | | Small | 590% | 560% |
| | | | Medium | 570% | 540% |
| | | | Large | 570% | 540% |
| | | | X Large | 550% | 550% |
| ASTMD 5151-19<br>Standard Test<br>Method for<br>Detection of<br>Holes in Medical<br>Gloves | To determine the<br>holes in the gloves | AQL 2.5 | Gloves Passes AQL 2.5 | | |
| ASTM D6124-06<br>(Reapproved<br>2017) Standard<br>Test Method for<br>Residual Powder<br>on Medical<br>Gloves | To determine the<br>residual powder in<br>the gloves | ≤2 Mg/Glove | 0.49 mg/glove | | |
# BIOCOMPATIBILITY DATA
| TEST METHOD | PURPOSE | ACCEPTANCE<br>CRITERIA | RESULT |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|----------------------------------------------------|--------------------------------------------------|
| ISO 10993-10:2010( E)<br>Biological Evaluation Of<br>Medical Devices - Part 10:<br>Tests For Irritation And Skin<br>Sensitization.<br>Test done for irritation. | To evaluate the test item,<br>for skin irritation test in<br>New Zealand White<br>rabbits. | Under the condition<br>of study not an<br>irritant | Under the condition of<br>study, not an irritant |
{8}------------------------------------------------
# 510(K) SUMMARY AS REQUIRED BY: 21CFR8807.92(C)
| TEST METHOD | PURPOSE | ACCEPTANCE<br>CRITERIA | RESULT |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993 10:2010(E)<br>Biological Evaluation of<br>Medical Devices - Part 10:<br>Tests for Irritation and Skin<br>Sensitization.<br>Test done for skin sensitization | To evaluate the test item,<br>for the Skin Sensitization<br>in Guinea pigs by<br>maximization test. | Under the conditions<br>of the study, not a<br>sensitizer | Under the conditions of the<br>study, not a sensitizer |
| ISO10993-5:2009(E)<br>Biological evaluation of<br>medical devices - part 5, tests<br>for In vitro cytotoxicity. | To evaluate the test item,<br>for its ability to induce<br>cytotoxicity using L-929<br>mouse fibroblast cells by<br>Elution Method. | Under the conditions<br>of the study, non-<br>cytotoxic | Under the conditions of the<br>study cytotoxic for 100% test<br>item extract and non-<br>cytotoxic at 50%, 25%,<br>12.5% and 6.25% test item<br>extracts. Moreover, under the<br>conditions of the study, non<br>acute systemic toxic. |
| ISO 10993-11:2017(E)<br>Biological evaluation of<br>medical devices - part 11, tests<br>for systemic toxicity. | To evaluate the test item,<br>for Acute Systemic<br>Toxicity in Swiss Albino<br>Mice. | Under the conditions<br>of study, the device<br>extracts do not pose<br>a systemic toxicity<br>concern | Under the condition of<br>study, the device extracts<br>did not reveal any<br>systemic toxicity. |
The performance test data of the non-clinical tests demonstrates that the subject device meet the acceptance criteria and the specifications found in the standards and test methodology.
The performance test data of the non-clinical tests meet following standards:
ASTMD 6319-19 Standard Specification for Nitrile examination gloves for medical Application.
ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
ISO 10993-5:2009 (E) Biological evaluation of medical devices - part 5, tests for In vitro cytotoxicity.
ISO 10993-11:2017(E) biological evaluation of medical devices - part 11, tests for systemic toxicity.
#### H. COMPARISON BASED ON ASSESSMENT OF CLINICAL PERFORMANCE DATA
Not applicable - Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
#### I. CONCLUSION
The conclusions drawn from the non-clinical tests demonstrate that the subject device in 510(K) submission, ComfortPro Blue Nitrile Examination Gloves Powder Free is as safe, as effective, and performs as well as or better than the legally marketed predicate device K202384.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.