Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs
K212827 · Jr Engineering & Medical Technologies (M) Sdn. Bhd. · LZC · Dec 6, 2021 · General Hospital
Device Facts
Record ID
K212827
Device Name
Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs
Applicant
Jr Engineering & Medical Technologies (M) Sdn. Bhd.
Product Code
LZC · General Hospital
Decision Date
Dec 6, 2021
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
JR Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.
Device Story
Disposable nitrile examination glove; worn by clinicians to prevent cross-contamination between patient and examiner. Manufactured from acrylonitrile-butadiene copolymer; blue color; powder-free; non-sterile. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05 (2019). Provides barrier protection; breakthrough times vary by drug; contraindicated for use with Carmustine (BCNU) and Thiotepa due to low permeation times. Used in clinical settings.
Clinical Evidence
Bench testing only; no clinical data provided. Performance verified via ASTM D6319-2019 (physical properties, dimensions, thickness, powder residue) and ASTM D5151-2019 (watertightness). Biocompatibility testing performed per ISO 10993 series (irritation, sensitization, cytotoxicity, systemic toxicity, pyrogenicity).
Indicated for use as a disposable medical examination glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for use with specific chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Harbour Health Powder Free Nitrile Examination Glove, Blue (Tested for use with Chemotherapy Drugs) (K210944)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2021
JR Engineering & Medical Technologies (M) SDN.BHD. Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE 114 Aurora, Illinois 60504
Re: K212827
Trade/Device Name: Jr Medic Blue Nitrile Examination Powder Free Gloves tested for use with chemotherapy drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ Dated: November 2, 2021 Received: November 4, 2021
Dear Manoj Zacharias:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212827
#### Device Name
JR Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs
#### Indications for Use (Describe)
JR Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs is a disposable device intended for medical purpose that is worn on the examination between patient and examiner.
Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.
The tested chemotherapy drugs and their breakthrough detection times are as follows:
| Tested Chemotherapy Drug Name & Concentration | Minimum Breakthrough Detection Time |
|-----------------------------------------------|-------------------------------------|
| Carmustine (BCNU) (3.3 mg/ml) | 35.0 Minutes |
| Carboplatin (10 mg/ml) | >240 Minutes |
| Cisplatin (1 mg/ml) | >240 Minutes |
| Cyclophosphamide (Cytoxan) (20 mg/ml) | >240 Minutes |
| Dacarbazine (10.0 mg/ml) | >240 Minutes |
| Doxorubicin HCl (2 mg/ml) | >240 Minutes |
| Etoposide (20 mg/ml) | >240 Minutes |
| Fluorouracil (50 mg/ml) | >240 Minutes |
| Ifosfamide (50 mg/ml) | >240 Minutes |
| Methotrexate (25 mg/ml) | >240 Minutes |
| Mitomycin C (0.5 mg/ml) | >240 Minutes |
| Mitoxantrone (2 mg/ml) | >240 Minutes |
| Paclitaxel (6 mg/ml) | >240 Minutes |
| Thiotepa (10 mg/ml) | 64.9 Minutes |
| Vincristine Sulfate (1 mg/ml) | >240 Minutes |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY K212827 As required by: 21CFR§807.92(c)
### A. APPLICANTINFORMATION
| 510(K) Owner's Name | JR Engineering & Medical Technologies (M) SDN.BHD. |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Address | Lot 8 &10, Jalan Zurah 3 & Lot 1&3, Jalan Zurah 3A/1,<br>Pusat Perindustrian 2, 44200 Rasa, Hulu Selangor, Selangor<br>Darul Ehsan, Malaysia. |
| Phone | +603-60572081 |
| Fax | +603-60572181 |
| E-mail | ganeshjrmt@gmail.com |
| Contact Person | Mr. Ganesan Subramaniam |
| Designation | Managing Director |
| Contact Number | +6012 224 6677 |
| Contact Email | ganeshjrmt@gmail.com |
| Date Submitted | 02 NOV 2021 |
# B. DEVICE IDENTIFICATION
| Name of the device | JR Medic Blue Nitrile Examination Gloves Powder<br>Free tested for use with Chemotherapy drugs |
|-----------------------------------|------------------------------------------------------------------------------------------------|
| Product proprietary or trade name | JR MEDIC |
| Common or usual name | Exam Gloves |
| Classification name | Patient Examination Gloves, Specialty |
| Device Classification | Class I |
| Product Code | LZC, LZA, OPJ |
| Regulation Number | 21 CFR 880.6250 |
| Review Panel | General Hospital |
### C. PREDICATE DEVICE
| Predicate Device | Harbour Health<br>Powder Free Nitrile Examination Glove, Blue (Tested for<br>use with Chemotherapy Drugs) |
|------------------|-----------------------------------------------------------------------------------------------------------|
| 510(K) Number | K210944 |
| Regulatory Class | 1 |
| Product code | LZA, LZC, OPJ |
### D. DESCRIPTION OF THEDEVICE:
The subject device in 510(K) notification is JR Medic blue nitrile examination gloves powder free tested for use with Chemotherapy drugs.
The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I).
{5}------------------------------------------------
The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)
### E. INTENDED USE OF THE DEVICE:
JR Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.
| Tested Chemotherapy Drug Name &<br>Concentration | Minimum Breakthrough Detection<br>Time (Minutes) |
|--------------------------------------------------|--------------------------------------------------|
| Carmustine (BCNU) (3.3 mg/ml) | 35.0 Minutes |
| Carboplatin (10 mg/ml) | >240 Minutes |
| Cisplatin (1 mg/ml) | >240 Minutes |
| Cyclophosphamide (Cytoxan) (20 mg/ml) | >240 Minutes |
| Dacarbazine (10.0 mg/ml) | >240 Minutes |
| Doxorubicin HCl (2 mg/ml) | >240 Minutes |
| Etoposide (20 mg/ml) | >240 Minutes |
| Fluorouracil (50 mg/ml) | >240 Minutes |
| Ifosfamide (50 mg/ml) | >240 Minutes |
| Methotrexate (25 mg/ml) | >240 Minutes |
| Mitomycin C (0.5 mg/ml) | >240 Minutes |
| Mitoxantrone (2 mg/ml) | >240 Minutes |
| Paclitaxel (6 mg/ml) | >240 Minutes |
| Thiotepa (10 mg/ml) | 64.9 Minutes |
| Vincristine Sulfate (1 mg/ml) | >240 Minutes |
The tested chemotherapy drugs and their breakthrough detection times are as follows:
Please note that the following drugs have low permeation times: Carmustine (BCNU) (3.3 mg/ml) 35.0 Minutes Thiotepa (10 mg/ml) 64.9 Minutes
Warning: Do not use with Carmustine (BCNU) & Thiotepa
{6}------------------------------------------------
# F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
| CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | REMARKS |
|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| | | PREDICATE | PROPOSED<br>DEVICE | |
| 510(K) Number | - | K210944 | K212827 | |
| Name of device | - | Harbour Health<br>Powder Free<br>Nitrile<br>Examination<br>Glove, Blue<br>(Tested for use<br>with<br>Chemotherapy<br>Drugs) | JR Medic<br>Blue Nitrile<br>Examination Gloves<br>Powder Free tested<br>for use with<br>Chemotherapy drugs | Similar |
| Product Code | - | LZA, LZC, OPJ | LZC, LZA, OPJ | Similar |
| Intended use | - | Harbour Health<br>Powder Free<br>Nitrile<br>Examination<br>Glove, Blue<br>(Tested for use<br>with<br>Chemotherapy<br>Drugs) disposable<br>device intended for<br>medical purpose that<br>is worn on the<br>examiner's hand to<br>prevent<br>contamination<br>between patient<br>and examiner.<br>The proposed<br>device was tested<br>for use with<br>chemotherapy<br>drugs as per<br>ASTM D6978-05<br>(2019), Standard<br>Practice for<br>Assessment of<br>Medical Gloves to<br>Permeation by<br>Chemotherapy<br>Drugs | JR Medic Blue<br>Nitrile Examination<br>Gloves Powder Free<br>tested for use with<br>Chemotherapy drugs<br>is a disposable<br>device intended for<br>medical purpose that<br>is worn on the<br>examiner's hand to<br>prevent<br>contamination<br>between patient and<br>examiner.<br>Additionally, the<br>gloves were tested<br>for use with<br>chemotherapy drugs<br>in accordance with<br>ASTM D6978-05<br>(2019) Standard<br>Practice for<br>Assessment of<br>Medical Glove to<br>Permeation by<br>Chemotherapy<br>Drugs. | Similar |
| CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | REMARKS |
| | | PREDICATE<br>K210944 | PROPOSED<br>DEVICE<br>K212827 | |
| Regulation Number | - | 21 CFR<br>880.6250 | 21 CFR<br>880.6250 | Same |
| Material | - | Nitrile | Nitrile | Same |
| Color | - | Blue | Blue | Same |
| Texture | - | Finger Texture | Finger texture | Same |
| Size | ASTM D6319-<br>2019 | Small, Medium,<br>Large, Extra<br>Large | Small, Medium,<br>Large, Extra<br>Large | Same |
| Single Use | Medical Glove<br>Guidance<br>Manual<br>- Labeling | Single Use | Single Use | Same |
| Sterile/non sterile | - | Non sterile | Non sterile | Same |
| Dimensions | ASTM D6319-<br>2019 | Length: Small-<br>Min 220 mm &<br>Medium, Large<br>& Extra large-<br>Min 230 mm<br>Width Min<br>95+/-10 mm(<br>Medium Size) | Length > 230<br>mm<br>Width Min<br>95+/-10 mm(for<br>medium size) | Similar |
| Physical Properties | ASTM D6319-<br>2019 | Before Ageing<br>Tensile Strength<br>min 14 Mpa<br>Ultimate<br>Elongation<br>Min 500%<br>After Ageing<br>Tensile Strength<br>min 14 Mpa<br>Ultimate<br>Elongation<br>Min 400% | Before Ageing<br>Tensile Strength<br>> 14 Mpa<br>Ultimate<br>Elongation<br>>500%<br>After Ageing<br>Tensile Strength<br>>14 Mpa<br>Ultimate<br>Elongation<br>> 400% | Same |
| Thickness | ASTM D6319-<br>2019 | Palm<br>min 0.05 mm<br>Finger<br>min 0.05 mm | Palm >0.05 mm<br>Finger > 0.05<br>mm | Same |
| Powder Free<br>Residue | ASTM D6319-<br>2019 | <2 mg/glove | <2 mg/glove | Same |
| Watertight<br>(1000 ml) | ASTM D5151-<br>2019 | Passes AQL-2.5 | Passes AQL-1.5 | Similar |
| Label and Labeling | FDA Label<br>requirements | Meets FDA's<br>requirements | Meets FDA's<br>requirements | Same |
| CHARACTE<br>RSTICS | STANDARDS | DEVICE PERFORMANCE | | REMARKS |
| | | PREDICATE<br>K210944 | PROPOSED<br>DEVICE<br>K212827 | |
| Bio-<br>compatibility | Primary Skin<br>Irritation-ISO 10993-<br>10:2010<br>(E) | Under the condition of<br>study not an irritant | Under the condition<br>of study not an<br>irritant | Same |
| | Dermal Sensitization-<br>ISO 10993-10:2010<br>(E) | Under the conditions of<br>the study not a<br>sensitizer | Under the conditions<br>of the study not a<br>sensitizer | Same |
| | In vitro<br>cytotoxicity<br>ISO10993-5 :2009(E) | Under the<br>conditions of the<br>study, potentially<br>cytotoxic | Under the<br>conditions of the<br>study, cytotoxic | Similar |
| | Material<br>Mediated<br>Pyrogenicity<br>ISO 10993-<br>11:2017(E) /<br>USP 41<151> | No Data Available | Under the<br>conditions of the<br>study non<br>pyrogenic | ---- |
| | Acute Systemic<br>Toxicity Test ISO<br>10993-11:2017(E) | Under the conditions of<br>the study, the device<br>does not elicit a<br>systemic toxicity<br>response in the model<br>animal | Under the condition<br>of study does not<br>induce any systemic<br>toxic concern | Similar |
| Chemotherapy Drugs Tested with Minimum Breakthrough Detection Time as tested per ASTM D6978-<br>05 (2019) | | | | |
| Busulfan (6mg/ml) | | >240 Minutes | Not Tested | Different |
| Carmustine (BCNU) (3.3 mg/ml) | | 14.5 Minutes | 35.0 Minutes | Different |
| Carboplatin (10 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Cisplatin (1 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Cyclophosphamide (Cytoxan) (20 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Cytarabine (100 mg/ ml) | | >240 Minutes | Not Tested | Different |
| Dacarbazine (10.0 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Docetaxel (10 mg/ml) | | >240 Minutes | Not Tested | Different |
| Doxorubicin HCl (2 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Etoposide (20 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Fluorouracil (50 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Ifosfamide (50 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Mechlorethamine<br>HCL (1 mg/ml) | | >240 Minutes | Not Tested | Different |
| Methotrexate (25 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Mitomycin C (0.5 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Mitoxantrone (2 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Paclitaxel (6 mg/ml) | | >240 Minutes | >240 Minutes | Same |
| Thiotepa (10 mg/ml) | | 47.4 Minutes | 64.9 Minutes | Different |
| Vincristine Sulfate (1 mg/ml) | | >240 Minutes | >240 Minutes | Same |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| Test Method | Purpose | Acceptance Criteria | Result |
|--------------------------------------------------------------------------------------------------------|------------------------------------------|------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| ASTM D6319-2019 Standard<br>Specification for Nitrile<br>Examination Gloves for<br>Medical Application | To determine the<br>length of the gloves | Min 220 mm for Size Small<br>& Min 230 mm for all<br>other sizes | Small:- 404 mm<br>Medium:- 405 mm<br>Large:- 405 mm<br>X-Large:- 406 mm |
| ASTM D6319-2019 Standard<br>Specification for Nitrile<br>Examination Gloves for<br>Medical Application | To determine the<br>width of the gloves | Small:- 80+/-10 mm<br>Medium:- 95+/-10mm<br>Large:- 110+/-10 mm<br>X-Large:- 120+/-10 mm | Small:- 84 mm<br>Medium:- 94 mm<br>Large:- 105 mm<br>X-Large:- 115 mm |
# G. NON-CLINICAL TESTING SUMMARY PERFORMANCE DATA
| Test Method | Purpose | Acceptance Criteria | Result |
|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM D6319-2019 Standard<br>Specification for Nitrile<br>Examination Gloves for<br>Medical Application | To determine the<br>thickness of the<br>gloves | Palm 0.05 mm min<br>Finger 0.05 mm min<br>for all sizes | Size Palm Finger<br>Small 0.19mm 0.21mm<br>Medium 0.19mm 0.21mm<br>Large 0.19mm 0.21mm<br>X-Large 0.19mm 0.21mm |
| ASTM D6319-2019 Standard<br>Specification for Nitrile<br>Examination Gloves for<br>Medical Application | To Determine the<br>physical properties-<br>Tensile strength | <b>Before Ageing</b><br>Tensile Strength<br>14Mpa Min for all<br>sizes<br><b>After Ageing</b> Tensile<br>Strength 14Mpa Min<br>for all sizes | Size <b>Before<br/>ageing After<br/>ageing</b><br>Small 22.77Mpa 20.50Mpa<br>Medium 24.46Mpa 21.81Mpa<br>Large 24.51Mpa 21.95Mpa<br>X-Large 24.59Mpa 22.05Mpa |
| | | <b>Before Ageing</b><br>Ultimate Elongation<br>500% Min for all<br>sizes<br><b>After Ageing</b><br>Ultimate Elongation<br>400% Min for all<br>sizes | Size <b>Before<br/>ageing After<br/>ageing</b><br>Small 885% 868%<br>Medium 888% 870%<br>Large 891% 872%<br>X-Large 892% 875% |
{10}------------------------------------------------
| Test Method | Purpose | Acceptance Criteria | Result…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.