K212803 · 3D Global Biotech, Inc. · NXC · May 23, 2022 · Dental
Device Facts
Record ID
K212803
Device Name
DailyMate Orthodontic Aligner System
Applicant
3D Global Biotech, Inc.
Product Code
NXC · Dental
Decision Date
May 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DailyMate Orthodontic Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition. The aligner system repositions teeth by way of continuous gentle force.
Device Story
DailyMate Orthodontic Aligner System consists of custom-made, clear, removable plastic aligners; fabricated from medical-grade polyurethane. Input: dental impressions or digital scans. Processing: 3Shape Ortho System software used by dental professionals to design treatment plans and generate 3D-printed molds; aligners fabricated via thermoforming. Output: series of sequential aligners. Usage: clinic setting; operated by dental practitioners. Mechanism: aligners worn by patient as prescribed; exert continuous gentle force to achieve progressive tooth realignment. Benefit: correction of malocclusion. Healthcare provider reviews and approves treatment plans before manufacturing.
Clinical Evidence
No clinical data provided. Bench testing performed comparing DailyMate to Invisalign, including pH change, water absorption, and tensile strength tests. Biocompatibility testing conducted per ISO 10993 (cytotoxicity, sensitization, oral mucosa irritation).
Technological Characteristics
Material: medical-grade polyurethane. Principle: sequential thermoformed plastic aligners applying continuous force. Connectivity: uses 3Shape Ortho System software for design. Sterilization: none. Form factor: custom-fitted dental trays.
Indications for Use
Indicated for treatment of tooth malocclusion in patients with permanent dentition.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
3D Global Biotech Inc % Diana Lam Regulatory Affairs Specialist 370 W Grand Blvd #110 Corona, California 92882
Re: K212803
Trade/Device Name: DailyMate Orthodontic Aligner System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: February 22, 2022 Received: February 22, 2022
Dear Diana Lam:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212803
Device Name DailyMate Orthodontic Aligner System
Indications for Use (Describe)
The DailyMate Orthodontic Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition. The aligner system repositions teeth by way of continuous gentle force.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for 3D Global Biotech. The logo consists of the number "3D" in green on the left, and the words "GLOBAL BIOTECH" in gray on the right. The words "GLOBAL" and "BIOTECH" are stacked on top of each other.
# 510(k) Summary K212803
# Applicant:
3D Global Biotech Inc.
- Address: 21F.-3, No.99, Sec. 1, Xintai 5th Rd., Xizhi Dist., New Taipei City 221, Taiwan .
- Contact: Dr. Keng-Liang Ou, President .
- Telephone: +886-2-26971270
# Correspondent Contact:
Diana Lam,
370 W. Grand Blvd, Suite 110, Corona, CA 92882
# Date Summary Prepared:
May 6th, 2022
DEVICE NAME: DailyMate® Orthodontic Aligner System
TRADE NAME: DailyMate® Orthodontic Aligner System
COMMON NAME: Aligner, Sequential
DEVICE CLASSIFICATION Name: Orthodontic Plastic Bracket
CLASSIFICATION REGULATION NUMBER: 21 CFR 872.5470 DEVICE CLASSIFICATION: CLASS II CLASSIFICATION PRODUCT CODE: NXC
#### Predicate Device
| Primary | SureCure Orthodontic<br>Aligner System | K182329 | Digital Orthodontic<br>Care |
|---------------------|----------------------------------------|---------|-----------------------------|
| Predicate | | | |
| Reference<br>Device | Invisalign System | K143630 | Align Technology, Inc. |
| Reference<br>Device | 3Shape Ortho SystemTM | K152086 | 3Shape A/S |
#### Indications for Use
The DailyMate® Orthodontic Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition. The aligner system repositions teeth by way of continuous gentle force.
#### Description of Device
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Global Biotech. The logo consists of a green stylized "3D" on the left, followed by the words "GLOBAL" and "BIOTECH" stacked on top of each other in gray. The "3D" is a stylized design, with the "3" and "D" connected.
The DailyMate® Orthodontic Aligner System is a series of dental aligners fabricated of clear, plastic removable aligners are designed and custom-made by dental professionals. The aligners are made of aesthetic and medical grade materials. Corrective force to straighten the teeth is delivered via minor changes into a modified position in each subsequent aligner.
# Substantial equivalence
The DailyMate® Orthodontic Aligner System is substantially equivalent to the predicate device and the reference device with respect to indications for use, mechanism of action and materials...... as demonstrated in the comparison table below.
| Item Name | Subject device | Predicate Device | Predicate Device | Substantial<br>equivalence<br>Analysis |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Device name | DailyMate®<br>Orthodontic<br>Aligner System | SureCure<br>Orthodontic Aligner<br>System | Invisalign<br>System | |
| Manufacturer | 3D Global<br>Biotech Inc. | Digital Orthodontic<br>Care | Align<br>Technology, Inc. | |
| 510(K) No. | K212803 | K182329 | K143630 | - |
| Regulation<br>No. | 21 CFR 872.5470 | 21 CFR 872.5470 | 21 CFR 872.5470 | Same |
| Classification<br>Name | Orthodontic<br>Plastic Bracket | Orthodontic Plastic<br>Bracket | Orthodontic<br>Plastic Bracket | Same |
| Regulatory<br>Class | Class II | Class II | Class II | Same |
| Product Code | NXC | NXC | NXC | Same |
| Indications<br>for<br>use/Intended<br>use | The DailyMate<br>Orthodontic<br>Aligner System is<br>indicated for the<br>treatment of tooth<br>malocclusion in<br>patients with<br>permanent<br>dentition. The<br>aligner system<br>repositions teeth<br>by way of<br>continuous gentle<br>force. | The SureCure<br>Orthodontic Aligner<br>System is indicated<br>for the treatment of<br>tooth malocclusion<br>in patients with<br>permanent dentition.<br>The aligner system<br>repositions teeth by<br>way of continuous<br>gentle force. | The Invisalign<br>System is indicated<br>for the alignment of<br>teeth during<br>orthodontic<br>treatment of<br>malocclusion | Same as the<br>predicate<br>device |
| Mechanism of<br>Action | Based on the<br>clinician's<br>treatment plan,<br>each aligner is<br>used for a<br>defined period of<br>time to exert | Based on the<br>clinician's treatment<br>plan, each aligner is<br>used for a defined<br>period of time to<br>exert gentle force to<br>achieve progressive | Sequential aligners<br>apply continuous<br>gentle force to the<br>teeth | Same |
| | gentle force to<br>achieve<br>progressive<br>realignment of<br>the teeth. This<br>occurs over time<br>until the final<br>correction has<br>been achieved. | realignment of the<br>teeth. This occurs<br>over time until the<br>final correction has<br>been achieved. | | |
| Method of<br>Use | Each preformed<br>plastic tray is<br>worn by the<br>patient as<br>prescribed by the<br>dental<br>practitioner,<br>usually a few<br>weeks prior to<br>using the next<br>sequential aligner<br>tray. | Each preformed<br>plastic tray is worn<br>by the patient as<br>prescribed by the<br>dental practitioner,<br>usually a few weeks<br>prior to using the<br>next sequential<br>aligner tray. | Aligners are worn for<br>approximately 2<br>weeks of 20-22<br>hours of wear per<br>day, after which it is<br>replaced by the next<br>stage aligners. This<br>is<br>repeated for duration<br>as prescribed by the<br>Dental<br>Practitioner | Same as the<br>predicate<br>device |
| Method of<br>Manufacture | Using software,<br>molds/models are<br>3D printed based<br>on the treatment<br>plan. Aligners<br>are fabricated on<br>the molds using a<br>thermoforming<br>machine. | Using software,<br>molds/models are<br>3D printed based on<br>the treatment plan.<br>Aligners are<br>fabricated on the<br>molds using a<br>thermoforming<br>machine. | The Align 3-D<br>Software uses a<br>scan of a PVS<br>impression<br>or a digital scan<br>(which represents<br>an untreated<br>state) to<br>generate the<br>image of a final,<br>treated state and<br>then interprets a<br>series of images<br>that represent<br>intermediate teeth<br>states. The dental<br>practitioner then<br>reviews these<br>images and has<br>the option to<br>reject or request<br>modifications to<br>the set-up prior to<br>approving it for<br>aligner<br>fabrication. Once<br>the dental<br>practitioner | Same as the<br>predicate<br>device |
| | | | approves the<br>treatment plan,<br>the software<br>converts the files<br>to produce the<br>series of custom-<br>made aligners | |
| Raw Material<br>Used | Thin<br>thermoformed<br>polyurethane | Thin thermoformed<br>polyurethane | Thin<br>thermoformed<br>polyurethane | Similar |
| OTC or Rx | Rx | Rx | Rx | Same |
| Software Use | 3Shape Ortho<br>System | 3Shape Ortho<br>System | Invisalign<br>System 3-D<br>software | Same as the<br>predicate<br>device |
| Sterilization | No | No | No | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains a logo with the text "3D GLOBAL BIOTECH". The "3D" is in green, while "GLOBAL BIOTECH" is in gray. The "3D" is a stylized graphic, and the text is aligned horizontally.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Global Biotech. On the left side of the logo is a green graphic that appears to be the number 3 with a circle attached to the right side of the number. To the right of the graphic is the text "GLOBAL BIOTECH" in gray. The text is stacked, with "GLOBAL" on top of "BIOTECH".
The DailyMate® Orthodontic Aligner System is substantially equivalent to the predicate device and the reference device. Both the subject and predicate devices use the same 510(k) cleared dental software 3Shape Ortho System (K152086) to translate tooth movements in developing the model schemes and allow the dental practitioner to review and approve the model schemes before aligner fabrication.
The bench tests using both the DailyMate® Orthodontic Aligner System and the Invisalign System were performed, the results of pH change test, water absorption change test and tensile strength change test showed that the DailyMate® Orthodontic Aligner System and the Invisalign System were substantial equivalent.
The differences among the subject device, the predicate device and the reference device lie in the variances in the aligner treatment plans which were customized for each individual patient. The associated risks have been mitigated by (1) A 510(k) requlated broadly recognized dental software, 3Shape Ortho System, is used for aligner design and review. (2) Every treatment plan receives a certified dental practitioner' approvals before aligners were manufactured. (3) The manufacturer is an ISO13485 certified facility which follows GMP standards during the manufacturing processes.
# Non-Clinical performance Data
The DailyMate® Orthodontic Aligner System uses polyurethane as the raw material, same as the predicate device and reference device. The DailyMate® Orthodontic Aligner System is a device contacts mucosal membrane for a duration of greater than 30 days. The biocompatibility testing for the aligner materials was completed in accordance with ISO 10993 per the following:
- Part 5: Cytotoxicity Test .
- . Part 10: Sensitization Test
- Part 10: Oral Mucosa Irritation Test ●
The testing results demonstrate that the material of DailyMate® Orthodontic Aligner System is biocompatible for the intended use.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Global Biotech. The logo consists of a green stylized "3D" on the left, followed by the words "GLOBAL" and "BIOTECH" stacked on top of each other in gray. The logo is simple and modern, and the use of green and gray gives it a clean and professional look.
The laboratory testing data was provided to confirm the quality of the DailyMate® Orthodontic Aligner System. A manufacturing validation report was also included to assure the substantial equivalence of the DailyMate® Orthodontic Aligner System.
# Clinical performance Data
No clinical data was provided in this submission
# Conclusion
The data included in this submission demonstrates that the DailyMate® Orthodontic Aligner System is substantially equivalent to the predicate device and the reference device in indications for use, mechanism of action, aligner design process, raw material and manufacturing fabrication process.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.