The VenSure™ LightGuide is used to locate, illuminate within, and transilluminate across, nasal and sinus structures in adults.
Device Story
VenSure LightGuide is a sterile, single-use accessory for the VenSure Balloon Sinus Dilation System. It consists of a flexible light fiber, protective sheath, and integrated battery-powered LED light source. Operated by a clinician, the device is loaded into the sinus dilation system and attached via a luer lock fitting. Pulling an activation tab triggers the distal tip to emit red light (625nm) for up to two hours. The light allows the physician to visualize and transilluminate nasal and sinus structures during surgical procedures, aiding in the accurate placement of the balloon dilation device. This visualization helps ensure proper positioning within the sinus anatomy, potentially improving procedural outcomes and patient safety.
Clinical Evidence
Bench testing only. Testing included dimensional and slider characterization, light output and run time, slider separation force, advancement and retraction force, light fiber tensile strength, and comparative light output. Biocompatibility was evaluated per ISO 10993-1 (cytotoxicity, irritation, sensitization, acute systemic toxicity). Sterilization validation performed per ISO 11135-1. Electrical safety and EMC testing conducted per IEC 60601-1 and IEC 60601-1-2. All tests met predefined acceptance criteria.
Technological Characteristics
Flexible light fiber (0.5mm OD, 304mm length) with protective sheath and integrated battery-powered LED light source. Energy source: Lithium Manganese Dioxide battery (CR2, 3V). Light output: 625nm (Red), 1W max. Sterilization: Ethylene Oxide. Complies with IEC 60601-1 (Electrical Safety) and IEC 60601-1-2 (EMC). Biocompatible per ISO 10993-1. Type BF applied part.
Indications for Use
Indicated for locating, illuminating, and transilluminating nasal and sinus structures in adults.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
February 4, 2022
Intersect ENT Yoko Enrile Advisor, Regulatory Affairs 1555 Adams Dr. Menlo Park, California 94025
Re: K212774
Trade/Device Name: VenSure LightGuide Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose, And Throat Manual Surgical Instrument Regulatory Class: Class I Product Code: LRC Dated: January 7, 2022 Received: January 10, 2022
Dear Yoko Enrile:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212774
Device Name VenSure™ LightGuide
Indications for Use (Describe)
The VenSure™ LightGuide is used to locate, illuminate within, and transilluminate across, nasal and sinus structures in adults.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Intersect ENT. The logo consists of a colorful, abstract symbol resembling an infinity sign or a stylized flower, followed by the word "intersect" in bold, black letters. Below "intersect" is the abbreviation "ENT" in smaller, black letters.
# 510(k) Summary
510(k) Number: K212774
Date prepared: February 2, 2022
# 1.Submitter Information
| Submitter: | Intersect ENT |
|------------|---------------------------------------------|
| Address: | 1555 Adams Dr.<br>Menlo Park, CA 94025, USA |
| Telephone: | (650) 681 3143 |
| Fax: | (650) 641 2065 |
| Contact: | Yoko Enrile<br>Advisor, Regulatory Affairs |
# 2.Device Information
| Trade Name: | VenSure™ LightGuide |
|----------------------|-------------------------------------------------|
| Common Name: | Sinus Guidewire |
| Classification: | Class I per 21 CFR 874.4420 |
| Classification Name: | Ear, nose and throat manual surgical instrument |
| Product Code: | LRC |
# 3.Predicate Device Information
The VenSure™ LightGuide is substantially equivalent to PathAssist LED Light Fiber (K141916).
# 4.Device Description
VenSure™ LightGuide is a sterile, single-use device designed to emit light from the distal end. The device is an accessory to the VenSure™ Balloon device and used to locate, illuminate within and transilluminate across nasal and sinus structures. The device consists of a flexible light fiber, a protective sheath and an integrated battery-powered LED light source. When activated, the light fiber will emit red light from the distal tip for two (2) hours. It has a fiber nominal working length of 20mm (0.79 inch) with an outer diameter of 0.5mm (0.02 inch). The VenSure™ LightGuide is packaged separately.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Intersect ENT. The logo consists of a colorful, abstract shape resembling four interconnected loops or petals, with a black dot in the center. To the right of the shape, the word "intersect" is written in a bold, sans-serif font, and below that, the letters "ENT" are written in a smaller font.
# 5. Intended Use
The VenSure™ LightGuide and the predicate device have the same Intended Use, which is to locate, illuminate within, and transilluminate across, nasal and sinus structures.
Indications for Use for VenSure™ LightGuide and Predicate Device
| Device | Indications for Use |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| VenSureTM LightGuide – Subject device | The VenSureTM LightGuide is used to locate, illuminate within, and transilluminate across, nasal and sinus structures in adults. |
| PathAssist LED Light Fiber (K141916) –<br>Predicate device | To locate, illuminate within, and transilluminate across nasal and sinus structures in patients aged 18 and over. |
## 6. Comparison of Technological Characteristics
The VenSure™ LightGuide is substantially equivalent to PathAssist LED Light Fiber (K141916). The VenSure™ LightGuide has the same intended use and fundamental scientific technology as the predicate device (K141916). The subject device has the same technological characteristics; in particular, basic design, energy source, performance, and principle of operation.
| Description | VenSure™ LightGuide - Subject Device | PathAssist LED Light Fiber (K141916) - Predicate Device |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Device Classification | Class I, LRC, 21 CFR 874.4420 | Class I, LRC, 21 CFR 874.4420 |
| Indications for use | See above | See above |
| Light fiber dimensions | OD: 0.5mm<br>Length: 304mm | OD: 0.375mm<br>Length: 276mm |
| Device Design | Device consists of a flexible light fiber, a protective sheath/housing and an integrated battery-powered LED light source | Device consists of a flexible light fiber, a protective sheath and an integrated battery-powered LED light source |
| Biocompatibility | Biocompatible per ISO 10993-1 | Biocompatible per ISO 10993-1 |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
| Battery Type | Lithium Manganese Dioxide, CR2, 3 Volts | Lithium Manganese Dioxide, CR2, 3 Volts |
| Light Source Type | LED (Red) | LED (Red) |
| Maximum LED Power Output | 1 W | 1 W |
| Light Source Wavelength / Color | 625nm / Red | 625nm / Red |
| Medical Safety Standards | Complied with IEC 60601-1 | Complied with IEC 60601-1 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Intersect ENT. The logo consists of a colorful abstract symbol on the left, followed by the word "intersect" in black font. Below the word "intersect" is the word "ENT" in a smaller black font. The abstract symbol is made up of four overlapping circles in different colors: pink, yellow, blue, and purple.
| Description | VenSure™ LightGuide - Subject Device | PathAssist LED Light Fiber (K141916) - Predicate Device |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Medical EMC Standard | Complied with IEC 60601-1-2; Type BF applied part | Complied with IEC 60601-1-2; Type BF applied part |
| Illumination and activation time | Emits red LED light from the distal end of light fiber for 2 hours | Emits red LED light from the distal end of light fiber for over 60 minutes |
| Use | Single use, disposable | Single use, disposable |
| Operation Method | Load the light fiber into Sinus Dilation Device (VenSure Balloon Sinus Dilation System) and attach the light fiber housing to the luer lock fitting on the VenSure Balloon device. Pull the tab to activate the LED light. | Load the light fiber into Sinus Dilation Device (XprESS Multi-Sinus Dilation System) and attach the LED light fiber housing to the barbed fitting of XprESS device. Pull the tab to activate the LED light. |
The performance testing including mechanical integrity, functionality, EMC and Electrical Safety testing / evaluation support that the difference in the fiber dimensions and activation time between the subject device and the predicate device do not raise any new concerns of safety or effectiveness.
# 7. Performance Data
# Bench Testing
Bench testing was conducted to ensure that the VenSure™ LightGuide met the predefined acceptance criteria to demonstrate safety and performance. Testing included the following:
- Dimensional and Slider Characterization
- Light Output and Run Time Test .
- Slider Separation Force
- Advancement and Retraction Force ●
- Light Fiber Tensile ●
- Comparative Light Output Testing against the predicate device ●
All tests met the predefined acceptance criteria. The test results demonstrated that differences in device characteristics between the subject device and predicate device do not raise any new questions of safety or effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Intersect ENT. The logo features a colorful, abstract symbol to the left of the word "intersect" in black, with a small circle above the "t". Below "intersect" is the word "ENT" in a smaller font size, also in black. The symbol is made up of four overlapping circles in different colors: pink, yellow, green, and blue.
# Biocompatibility
The biocompatibility evaluation for the VenSure™ LightGuide was conducted in accordance with ISO 10993-1. Biocompatibility testing included cytotoxicity, irritation, sensitization and acute systemic toxicity testing. All tests successfully met the required acceptance criteria, demonstrating that the patient contacting materials used in the subject device are biocompatible.
# Sterilization & Stability
Sterilization validation testing was performed to demonstrate compliance with ISO 11135-1. Shelf life and packaging testing were performed to support the labeled shelf life. All tests were successfully completed.
# Electromagnetic compatibility and Electrical Safety
Electromagnetic Compatibility (EMC) and Electrical Safety testing was conducted per IEC 60601-1 and IEC 60601-1-2.
# 8. Conclusion
Based on the indications for use, technological characteristics, performance testing, and comparison to the predicates, it is concluded that the VenSure™ LightGuide performs as its intended use, is substantially equivalent to the predicate device identified in this submission and does not present any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.