← Product Code [OHS](/productcode/OHS) · K212771

# LightStim Elipsa (K212771)

_Led Intellectual Properties, LLC · OHS · Jan 26, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212771

## Device Facts

- **Applicant:** Led Intellectual Properties, LLC
- **Product Code:** [OHS](/productcode/OHS.md)
- **Decision Date:** Jan 26, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The LightStim Elipsa is a LED light therapy device which uses specific wavelengths of light emitting diodes (LEDs). It is intended to emit energy in the red and infrared region of the light spectrum to provide treatment for full face wrinkles. The blue and red light spectrum is intentent of mild to moderate inflammatory acne.

## Device Story

LightStim Elipsa is an over-the-counter LED light therapy device for home use. It utilizes four LED wavelengths (612nm, 645nm, 655nm, 850nm) for wrinkle treatment and two wavelengths (410nm, 645nm) for acne treatment. The device is operated by the patient; no user settings or adjustments are required. The system includes the LED module, power supply, goggles, and travel case. By emitting specific light energy, the device targets dermatological conditions; patients apply the device to the skin for prescribed treatment durations. The output is intended to reduce wrinkles and treat inflammatory acne, providing a non-invasive therapeutic benefit.

## Clinical Evidence

Bench testing only. The device underwent electrical safety testing (AAMI/ANSI ES60601-1), electromagnetic compatibility testing (IEC 60601-1-2), and performance testing per IEC 60601-1-6, IEC 60601-1-11, and IEC 60601-2-83. Usability and label comprehension studies were conducted to evaluate self-selection, training effectiveness, and device operation by intended users, confirming that instructions were effective for primary operating functions.

## Technological Characteristics

LED-based light therapy device. Emits red/infrared (612nm, 645nm, 655nm, 850nm) and blue/red (410nm, 645nm) light. Powered by external supply. Tested to AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, and IEC 60601-2-83. No user-serviceable components. Over-the-counter use.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- DemarkQ ([K203214](/device/K203214.md))
- Poly Clear ([K183708](/device/K183708.md))
- LightStim Professional 2-Panel Light ([K150098](/device/K150098.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 26, 2022

LED Intellectual Properties LLC Chase Marchese VP of Manufacturing 16552 Von Karman Ave Irvine. California 92606

Re: K212771

Trade/Device Name: LightStim Elipsa Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP, OHS Dated: December 1, 2021 Received: December 1, 2021

Dear Chase Marchese:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K212771

Device Name LightStim Elipsa

#### Indications for Use (Describe)

The LightStim Elipsa is a LED light therapy device which uses specific wavelengths of light emitting diodes (LEDs). It is intended to emit energy in the red and infrared region of the light spectrum to provide treatment for full face wrinkles. The blue and red light spectrum is intentent of mild to moderate inflammatory acne.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K212771

510(k) Summary LED Intellectual Properties, LLC. Device: LightStim Elipsa K212771

## 1. General Information

Submitter: LED Intellectual Properties, LLC 16552 Von Karman Ave. Irvine, Ca. 92606

Contact Person: Steve Marchese & Chase Marchese Phone: Chase (714) 924-0492 Date Prepared: 2022/01/19 Emails: Chase@lightstim.com & Steve@lightstim.com

## 2. Device name and code

Device Proprietary Name: LightStim Elipsa

Product Code and name:

OHS - Light Based Over-the-Counter Wrinkle Reduction OLP - Over-the-Counter Powered Light Based Laser For Acne

Classification Name: Laser Surqical Instrument For Use in General And Plastic Surgery And in Dermatology. 21 CFR 878.4810

Common or Usual Name: Light-Based over-the-counter wrinkle reduction and acne treatment.

Requlatory Class: II

## 3. Predicate Devices

SHENZHEN KAIYAN MEDICAL CO LTD, DemarkQ (K203214) Medtek Skincare, LLC, Poly Clear (K183708) LED Intellectual Properties, LLC, LightStim Professional 2-Panel Light (K150098)

## 4. Device Description

The LightStim Elipsa is an over-the-counter light emitting diode (LED) device that emits energy for use in dermatology for the treatment of acne and wrinkles. The device uses four types of LEDs for Wrinkles: 612nm amber, 645nm red, 655nm red, and 850nm infrared and two types of LEDs for Acne: 410nm and 645nm. The treatment time is controlled by the user. There are no user settings or adjustments required.

{4}------------------------------------------------

#### 510(k) Summary LED Intellectual Properties, LLC. Device: LightStim Elipsa K212771

The LightStim Elipsa system components include the device containing the LED module, power supply, goggles, and travel case.

The LightStim Elipsa does not contain any user serviceable components. The device is sold as Over-the-counter (OTC)

#### 5. Indications for Use

The LightStim Elipsa is a LED light therapy device which uses specific wavelengths of light, produced by light emitting diodes (LEDs. It is intended to emit energy in the red and infrared region of the light spectrum to provide treatment for full face wrinkles. The blue and red light spectrum is intended for the treatment of mild to moderate inflammatory acne.

## 6. Comparison of Technological Characteristics

| Company                   | SHENZHEN KAIYAN MEDICAL<br>CO LTD                                                                                       | Medtek Skincare, LLC                                                                                                                                                                                                                                                    | LED Intellectual Properties, LLC                                                                                                                                                                                      | LED Intellectual Properties, LLC                                                                                                                                                                                                                                                                                                                                                                       |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name                | DemarkQ                                                                                                                 | Poly Clear                                                                                                                                                                                                                                                              | LightStim Professional 2-Panel Light                                                                                                                                                                                  | LightStim Elipsa                                                                                                                                                                                                                                                                                                                                                                                       |
| 510(k) Number             | K203214                                                                                                                 | K183708                                                                                                                                                                                                                                                                 | K150098                                                                                                                                                                                                               | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                        |
| Product Code              | OLP                                                                                                                     | OLP                                                                                                                                                                                                                                                                     | OHS                                                                                                                                                                                                                   | OLP & OHS                                                                                                                                                                                                                                                                                                                                                                                              |
| FDA Device Classification | Class II                                                                                                                | Class II                                                                                                                                                                                                                                                                | Class II                                                                                                                                                                                                              | Class II                                                                                                                                                                                                                                                                                                                                                                                               |
| FDA Clearance             | Over the Counter                                                                                                        | Over the Counter                                                                                                                                                                                                                                                        | Over the Counter                                                                                                                                                                                                      | Over the Counter                                                                                                                                                                                                                                                                                                                                                                                       |
| Indications for use       | The DemarkQ is an Over-the-<br>Counter (OTC) device intended<br>for treatment of mild to<br>moderate inflammatory acne. | The Poly Clear combination of<br>Red (633nm±10nm) and Blue<br>(417nm±10nm) is intended to<br>emit energy in the red, blue<br>regions of the light spectrum to<br>treat dermatological conditions,<br>specifically indicated to treat<br>mild to moderate acne vulgaris. | Two (2) interchangeable 2-Panel LED Systems,<br>each mounted on a hands-free, fully articulating<br>arm:<br>System #1 = intended for the use in the<br>treatment of full-face wrinkles.<br>System #2 = Not applicable | The LightStim Elipsa is a LED light<br>therapy device which uses specific<br>wavelengths of light, produced by light<br>emitting diodes (LEDs). It is intended to<br>emit energy in the red and infrared<br>region of the light spectrum to provide<br>treatment for full face wrinkles. The blue<br>and red light spectrum is intended for<br>the treatment of mild to moderate<br>inflammatory acne. |
| Target Population         | Women and Men with mild to<br>moderate acne.                                                                            | Women and Men with mild to<br>moderate acne.                                                                                                                                                                                                                            | System #1 = People with full-face wrinkles.<br>System #2 =Not applicable                                                                                                                                              | People with Full-face wrinkles.   Women<br>and Men with mild to moderate<br>inflammatory acne.                                                                                                                                                                                                                                                                                                         |
| Wavelengths (nm)          | 630nm +/-10nm<br>415nm -/-10nm                                                                                          | 633nm +/-10nm<br>417nm -/-10nm                                                                                                                                                                                                                                          | System #1 = 605nm, 630nm, 660nm, 855nm,<br>System #2 = Not applicable                                                                                                                                                 | RED: 612nm, 645nm, 655nm, 850nm<br>BLUE: 410nm & 645nm                                                                                                                                                                                                                                                                                                                                                 |
| Output in milliwatts      | DemarkQ WOW:<br>Red: 5<br>Blue: 25<br>DemarkQ POP:<br>Red: 25<br>Blue: 25 +/- 5                                         | Combo Red/Blue head:<br>Red 30mW/cm^2 +/-10nm<br>Blue 20mW/cm^2 +/-10nm                                                                                                                                                                                                 | System #1 = 65mW cm2<br>System #2 = Not applicable                                                                                                                                                                    | RED: 11.8 mW/cm2<br>BLUE: 12.8 mW/cm2                                                                                                                                                                                                                                                                                                                                                                  |
| Treatment Time            | 3 minutes per treatment                                                                                                 | Blue: 13minutes.<br>Red: 15 minutes                                                                                                                                                                                                                                     | System #1 = 3min<br>System #2 = Not applicable                                                                                                                                                                        | RED: 16 minutes per area<br>BLUE: 12 minutes per area                                                                                                                                                                                                                                                                                                                                                  |
| Dose                      | RED: 0.9-4.5 J/cm2<br>Blue: 4.5 J/cm2                                                                                   | Red: 52 J/cm2<br>Blue 26 J/cm2                                                                                                                                                                                                                                          | System #1 = 11.7 j/cm^2<br>System #2 = Not applicable                                                                                                                                                                 | RED: 11.3 j/cm^2<br>BLUE: 9.2 j/cm^2                                                                                                                                                                                                                                                                                                                                                                   |
| Total dose/Treatment      | 9 J/cm2                                                                                                                 | 52 J/cm2                                                                                                                                                                                                                                                                | 11.7 J/cm2                                                                                                                                                                                                            | Red 11.3 J/cm2<br>Blue 9.2 J/cm2                                                                                                                                                                                                                                                                                                                                                                       |
| Total dose/week           | 63 J/cm2                                                                                                                | 78 J/cm2                                                                                                                                                                                                                                                                | 58.5 J/cm2                                                                                                                                                                                                            | Red 56.5 J/cm2<br>Blue 46 J/cm2                                                                                                                                                                                                                                                                                                                                                                        |

{5}------------------------------------------------

### 7. Performance Testing

The LightStim Elipsa device results in patient contact with an electrically powered component: therefore, it was tested for conformance to AAMI/ANSI ES60601-1:2005(R) 2012 and A1:2012 4th Edition: Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance

The LightStim Elipsa device includes an electronic component, and it was tested for conformance to IEC 60601-1-2:2014 4th Edition Medical Electrical Equipment – Part 1-2: General Requirements for Safety and Essential Performance- Collateral standard: Electromagnetic Compatibility-Requirements and tests.

The LightStim Elipsa device was also tested for conformance to IEC 60601-1-6:2010. IEC 60601-1-11:2015, IEC 60601-2-83:2019 including gap analysis demonstrating conformance to IEC 60601-2-57:2011.

Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software in Medical Devices. The minor (A) level of concern was determined because malfunction of, or a latent design flaw in the software could not lead to a patient injurv

### 8. USABILITY TESTING:

The sponsor conducted a Usability/Label Comprehension study to acquire data in order to evaluate and measure labeling comprehension, device usability, self-selection component, and training effectiveness by "intended users" of the Elipsa device.

The results of this testing on the final version of the labeling and instruction plan showed that the instructions were effective and intended users were able to understand the labeling and apply this information to device use for the primary operating functions of the device.

#### 9. Conclusions

After analysis of safety, indications, intended uses, dose rates, performance, features, design materials, power output, technological properties, treatment areas, treatment reqimens and methods of operation, the manufacturer asserts that no significant differences exist between the subject device and predicate, and no different questions of safety and effectiveness arise. Therefore, the subject device is as safe, as effective, and performs as well as the referenced predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K212771](https://fda-staging.innolitics.com/device/K212771)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com