The Jada® System is intended to provide control and treatment of abnormal postpartum uterine bleeding or hemorrhage when conservative management is warranted.
Device Story
Jada System is a sterile, single-use, silicone intrauterine device designed for postpartum hemorrhage control. Device consists of an intrauterine loop with 21 vacuum pores and an inflatable cervical seal. Operated by clinicians in clinical settings; device is inserted transvaginally into the uterus. Cervical seal is inflated with 60-120 mL of sterile fluid to secure the device at the external cervical os. Vacuum is applied (max 90 mmHg) via external vacuum source to induce uterine collapse, creating a tamponade effect on bleeding vessels. Device includes a drain tube to evacuate blood and fluids. Output is the mechanical tamponade of the uterus; healthcare providers monitor bleeding control and fluid evacuation. Benefits include non-invasive hemorrhage management compared to more aggressive surgical interventions.
Clinical Evidence
Bench testing only. No clinical data presented. Performance testing included mechanical verification (vacuum flow rates, seal integrity, bond strength, clotted blood evacuation), biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity), and packaging/sterilization validation (ASTM D4169-16, F2096-11, F88/F88M-15, F1980-16).
Technological Characteristics
Materials: Silicone, polycarbonate, ABS. Principle: Vacuum-induced uterine tamponade. Form factor: 41 cm intrauterine loop with inflatable cervical seal. Connectivity: Connects to external hospital vacuum line. Sterilization: Gamma radiation (SAL 10^-6). Standards: ISO 10993-1 (biocompatibility), ISO 11137-2:2013 (sterilization), ASTM D4169-16 (transportation), ASTM F2096-11/F88/F88M-15 (packaging).
Indications for Use
Indicated for control and treatment of abnormal postpartum uterine bleeding or hemorrhage in patients where conservative management is warranted.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 30, 2021
Alydia Health % Cindy Domecus, R.A.C. Principal Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough, CA 94010
Re: K212757
> Trade/Device Name: Jada System Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: OQY Dated: August 30, 2021 Received: August 31, 2021
Dear Cindy Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212757
Device Name Jada System
Indications for Use (Describe)
The Jada® System is intended to provide control and treatment of abnormal postpartum uterine bleeding or hemorrhage when conservative management is warranted.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary - K212757
#### I. SUBMITTER
### 510(k) Owner
Colby Holtshouse Alydia Health 3495 Edison Way Menlo Park, CA 94025 Phone: 650-275-3772 Fax: 415-354-3473 Email: colby(@alydiahealth.com
Submission Correspondent Cindy Domecus, R.A.C. (US & EU) Domecus Consulting Services LLC Phone: (650) 343-4813 Fax: (650) 343-7822 Email: Cindy(@)DomecusConsulting.com
Date Prepared September 29, 2021
| II. DEVICE | |
|-----------------------|-----------------------------------------------------|
| Name of Device: | Jada System |
| Common or Usual Name: | Vacuum-induced Hemorrhage Control System |
| Regulation Name: | Obstetric-Gynecologic Specialized Manual Instrument |
| Regulation Number: | 21 CFR § 884.4530 |
| Regulatory Class: | II |
| Product Code: | OQY |
#### III. PREDICATE DEVICE
The predicate device is the Jada System, K201199. This predicate has not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
The subject of this submission is the Jada System kit, which includes 1) the Jada System, 2) a commercially available pre-sterilized vacuum tubing, and 3) a commercially available pre-sterilized luer lock syringe.
The Jada System is a 41 cm long intrauterine device made of silicone. The device consists of an intrauterine loop on the distal end of a translucent tube. The proximal end of the tube has a vacuum connector for connection to the vacuum tubing. Proximal to the connection of the intrauterine loop is a donut-shaped cervical seal. The cervical seal is filled with and emptied of 60-120 mL of sterile fluid by attaching a luer lock syringe to the seal valve. The intrauterine loop consists of a loop tube with 21 vacuum pores oriented toward the inside diameter of the intrauterine loop. On the outer surface of the intrauterine loop is a shield which overhangs the vacuum pores to protect tissue from vacuum and to prevent the vacuum pores from plugging with tissue and blood clots.
Before placing the Jada System device inside the uterus, the intrauterine loop is compressed. The compressed loop is inserted into the uterus transvaginally. The cervical seal is placed within the vagina, at
{4}------------------------------------------------
the external cervical os, and inflated and filled with 60-120 mL of sterile fluid. The vacuum tubing is attached to the vacuum connector on the Jada System and vacuum is then applied to a maximum value of 90 mmHg until bleeding is controlled. The Jada System should be fixed to the thigh along the tube.
#### V. INDICATIONS FOR USE
The Jada® System is intended to provide control and treatment of abnormal postpartum uterine bleeding or hemorrhage when conservative management is warranted.
### VI. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Attribute | K212757<br>Subject Device:<br>Jada System | K201199<br>Predicate Device:<br>Jada System | Comparison |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Alydia Health | Alydia Health | N/A |
| Product Code | OQY | OQY | Same |
| Indications for<br>Use | The Jada System is intended<br>to provide control and<br>treatment of abnormal<br>postpartum uterine bleeding<br>or hemorrhage when<br>conservative management is<br>warranted. | The Jada System is intended<br>to provide control and<br>treatment of abnormal<br>postpartum uterine bleeding<br>or hemorrhage when<br>conservative management is<br>warranted. | Same |
| Mechanism of<br>Action | Inserted into the uterus and<br>establishes a vacuum to<br>cause the uterine walls to<br>press against one another,<br>producing a tamponade<br>of the bleeding vessels. | Inserted into the uterus and<br>establishes a vacuum to<br>cause the uterine walls to<br>press against one another,<br>producing a tamponade<br>of the bleeding vessels. | Same |
| Design | Inflatable cervical seal and<br>intrauterine loop with<br>vacuum pores. The<br>intrauterine loop tube<br>vacuum pore count is<br>increased from 20 to 21 and<br>the seal valve features luer<br>syringe attachment. | Inflatable cervical seal and<br>intrauterine loop with<br>vacuum pores. Seal Valve<br>features the friction syringe<br>attachment. | Different |
| Rx/OTC | Rx | Rx | Same |
| Materials | Silicone, polycarbonate,<br>Acrylonitrile- Butadiene-<br>Styrene (ABS) | Silicone, Polyvinylchloride<br>(PVC), Acrylonitrile-<br>Butadiene-Styrene (ABS) | Different |
| Sterile | SAL 10-6 | SAL 10-6 | Same |
| Single Use | Yes | Yes | Same |
The subject and predicate device have identical indications for use statements and have the same intended use - the treatment of abnormal uterine bleeding when conservative management is warranted.
The subject Jada System is an updated version of the predicate Jada System with modifications to 1) improve manufacturability, 2) streamline packaging, and 3) improve user experience and aesthetics of the device. Following is a list of technological characteristics that remain the same between the subject and predicate devices:
{5}------------------------------------------------
- Vacuum induced collapse of the uterus causing the uterine walls to press against one another, producing tamponade of the bleeding vessels.
- Intrauterine loop to transmit vacuum to the uterine space.
- Drain pores within the intrauterine loop to evacuate blood, body fluids, and residual clots from the ● uterine space.
- Filled and expanded cervical seal to enable the maintenance of vacuum within the uterine space.
- Vacuum connector for attachment of vacuum tubing.
- Seal valve with automatic shut-off for filling the cervical seal.
- In-dwelling treatment within the uterus and genital tract <24 hours
- Drain tube to remove blood and fluid from postpartum uterus
- All patient contacting materials are silicone
The following technological differences exist between the subject and predicate devices:
- . The seal valve on the subject device features a luer syringe attachment instead of the friction syringe attachment of the predicate device, to provide a more secure attachment of the syringe.
- The inflation lumen that is used to fill the cervical seal on the subject device has a larger crosssectional area, to allow easier filling and removal of fluid from the Cervical Seal during use.
- The seal valve orientation of the subject device was changed from 12 o'clock to 9 o'clock/3 o'clock, to achieve a lower profile of the device in the device packaging.
- The seal valve body on the subject device was changed to clear polycarbonate from the white ABS (acrylonitrile butadiene styrene) of the predicate device. The polycarbonate valve enables a luerlock syringe fitting with higher flow.
- The vacuum connector on the subject device was changed to white ABS from the clear PVC (polyvinyl chloride) of the predicate device, to improve the aesthetics of the device.
- The intrauterine loop tube vacuum pore count was increased from 20 to 21 on the subject device, to improve manufacturability.
- The sterile packaging of the subject device was changed to a Tyvek lidded tray from a pouch with a ● device backing card of the predicate device, to improve the appearance of the packaging, remove the complexity of the backing card and to achieve a smaller form factor for the sterilized unit.
- . The subject device will be provided for the user in a kit carton that contains the Jada System, a commercially available pre-sterilized 60 mL syringe and a commercially available pre-sterilized 12' vacuum tubing for customer convenience.
These differences in technological characteristics do not raise different questions of safety and effectiveness and can be addressed through performance testing. Non-clinical data provided by Alydia Health were used to address the differences related to design and packaging to demonstrate substantial equivalence to the predicate device, as discussed in Section VII.
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Mechanical Testing
Mechanical testing on the subject Jada System was identical to the testing performed on the predicate Jada System. Specifically, the following mechanical tests were performed at baseline (T=0) and on aged samples (T=1 year):
{6}------------------------------------------------
- . Cervical Seal and Tube Dimensions: Verification of tube and seal dimensions
- Intrauterine Loop Portion Dimensional Test – Verification of intrauterine loop dimensions
- Vacuum Pore Diameter – Verification of vacuum pore size
- No Sharp Edges Verification of smooth edges and surfaces of device ●
- Attaining Pressure Drop Verification that cervical seal withstands pressure differential of 180 ● mmHg vacuum
- Static Load Test – Verification that the cervical seal withstands a static load of 1 lb applied axially along the tube without failure
- Overfill Capacity - Verification that cervical seal does not fail when filled with 180 mL water.
- Cervical Seal Inflation - Verification that cervical seal can be filled with 60 mL of water within 30 seconds with 10 lbs of force on syringe
- Impact Load Test – Verification that the cervical seal withstands an impact test of dropping a 1 1b weight 2 ft axially along the tube without failure
- Connection Tube Junction Impact Load Test - Verification that the intrauterine loop withstands an impact test of dropping a 1 1b weight 2 ft axially along the Tube without failure
- Flow Rate Verification that the device with vacuum is able to evacuate 400 mL of simulated . blood in 1 min or less
- Device Integrity Leak Test – Verification that the joints of the device do not leak when 180 mmHg of vacuum is applied
- Integration to Hospital Vacuum Line – Verification that the device connects to a vacuum tubing
- Inflation Tube Geometry – Verification that the cervical seal inflation lumen is functional
- Syringe Accommodation - Verification that a luer tapered syringe can be attached to the seal valve
- Cervical Seal Deflation Verification that cervical seal can be emptied of 60 mL of water within ● 30 seconds with 10 lbs of force on syringe
- . Cervical Seal Diameter and Bond Stability – Verification that the seal maintains a diameter of 70 mm and maintains integrity after 48 hours
- Clotted Blood Test – Verification that the device can evacuate simulated blood in the presence of clotted blood without occluding
- Vacuum Connector Bond Test – Vacuum connector bond remains intact after a proof load of 8.8 lbf is applied
- Seal Valve Bond Test - Seal valve bond remains intact after a proof load of 3.7 lbf is applied
## Packaging Testing
Testing of the kit packaging for the subject device consisted of transportation per ASTM-D4169-16 followed by gross leak detection per ASTM F2096-11 and for seal strength per ASTM F88/F88M-15. Furthermore, the Tyvek lidded tray of the subject Jada System was tested for aging (ASTM F1980-16) followed by packaging integrity testing for gross leak detection per ASTM F2096-11 and for seal strength per ASTM F88/F88M-15.
### Biocompatibility Testing
The Jada System is a surface device in contact with a breached surface, with limited duration (< 24 hours).
The biocompatibility evaluation for the subject Jada System was conducted in accordance with the FDA June 2016 guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff. The battery of testing included the following tests:
- Cytotoxicity
- Maximization Sensitization
{7}------------------------------------------------
- Vaginal Irritation
- Systemic toxicity
- Material Mediated Pyrogenicity
### Sterilization and Shelf-Life Testing
The subject and predicate Jada Systems are both sterilized using gamma radiation to a SAL of 10° according to ISO 11137-2: 2013. A shelf-life of 1 year has been established based on accelerated aging for the subject device.
#### VIII. CONCLUSIONS
The nonclinical performance data described above demonstrate that the Jada System is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.