K212740 · Puravita M, LLC · FXX · Dec 2, 2021 · General, Plastic Surgery
Device Facts
Record ID
K212740
Device Name
Puravita Medical Fold Flat Surgical Mask
Applicant
Puravita M, LLC
Product Code
FXX · General, Plastic Surgery
Decision Date
Dec 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The PuraVita Medical Fold Flat Surgical Mask is intended to be worn to protect both patients and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. The PuraVita Medical Fold Flat Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device that is provided non-sterile.
Device Story
4-layer, two-panel fold-flat surgical mask; protects patients and healthcare personnel from microorganisms, body fluids, and particulates. Composed of polypropylene spunbond outer/inner layers and two polypropylene meltblown filter layers; includes nylon/spandex earloops and malleable metal-core nosepiece. Used in clinical settings for infection control; operated by healthcare personnel or general public (OTC). Provides physical barrier against blood/body fluids; reduces potential exposure to contaminants. Non-sterile, single-use, disposable.
Clinical Evidence
No clinical or animal testing performed. Bench testing only. Performance validated via ASTM F2101 (BFE >99.1%), ASTM F2299 (PFE >98%), ASTM F1862 (synthetic blood penetration at 160 mmHg), and EN 14683 (differential pressure <6.0 mmH2O/cm2). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (sensitization and intracutaneous reactivity).
Indicated for patients and healthcare personnel to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices. Single-use, disposable, non-sterile device.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 2, 2021
Puravita M, LLC % Rhonda Alexander Sr. Consultant, Regulatory Strategy IUVO Consulting, LLC P.O. Box 56436 Virginia Beach, Virginia 23456
Re: K212740
Trade/Device Name: Puravita Medical Fold Flat Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: August 25, 2021 Received: August 30, 2021
Dear Rhonda Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K212740
Device Name Puravita Medical Fold Flat Surgical Mask
#### Indications for Use (Describe)
The Puravita Medical Fold Flat Surgical Mask is intended to be worn to protect both patients and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. The Puravita Medical Fold Flat Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a singleuse, disposable device that is provided non-sterile.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|--------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K212740
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
Date of Summary: November 24, 2021
#### Contact Information
| Submitter: | Submitter Contact: |
|------------------|------------------------|
| PuraVita M, LLC | Jefferson Nemelka |
| 1287 W 300 S | Plant Manager |
| Lindon, UT 84042 | 801-836-7940 |
| | jefferson@puravita.com |
| Consultant: | Dr. Rhonda Alexander |
|-------------|-------------------------------------|
| | Sr. Consultant, Regulatory Strategy |
| | IUVO Consulting, LLC |
| | (757) 582-4337 |
| | ralexander@iuvoconsulting.com |
#### Device Information
| Trade Name: | PuraVita Medical Fold Flat Surgical Mask |
|----------------------|------------------------------------------|
| Common Name: | Surgical Face Mask |
| Classification Name: | Mask, Surgical |
| Classification: | Class II per 21 CFR 878.4040 |
| Review Panel: | General Hospital |
| Product Code: | FXX |
### Predicate Device:
| Manufacturer: | Guangdong Kingfa Sci. & Tech. Co., Ltd, |
|----------------|-----------------------------------------|
| Product: | Medical Protective Mask |
| 510(k) Number: | K202107 |
#### Intended Use Statement:
The PuraVita Medical Fold Flat Surgical Mask is intended to be worn to protect both patients and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. The PuraVita Medical Fold Flat Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device that is provided non-sterile.
{4}------------------------------------------------
### Device Description:
The PuraVita Medical Fold Flat Surgical Mask is a 4-layer, two panel fold-flat surgical mask. It is comprised of polypropylene spunbond inner and outer layers and two inner polypropylene meltblown filter layers. The mask is white with an extended chamber. The dimensions of each mask are length 16.3cm ± 0.5cm and width 10.6cm ± 0.5cm. The mask has nylon/spandex earloops (length 19.5cm ± 1cm) and a malleable, metal-core nosepiece (0.4cm x 8.2cm x 0.1cm) to provide a functional fit over the nose and mouth of the user. It is a non-sterile, single use, disposable device. This device is not made with natural rubber latex.
| Device | Predicate Device: K202107<br>Guangdong Kingfa Sci. & Tech. Co.,<br>LTD, Medical Face Mask, Model<br>KF-A F02(N) | Subject Device: K212740<br>PuraVita Medical Fold Flat<br>Surgical Mask | Comparison |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use/Indications for<br>Use | The Medical Protective Mask is<br>intended to be worn to protect both the<br>patient and healthcare personnel from<br>transfer of microorganisms, body<br>fluids and particulate material. The<br>Medical Protective Mask is intended<br>for use in infection control practices to<br>reduce the potential exposure to blood<br>and body fluids. This is a single use,<br>disposable device, provided non-sterile | The PuraVita Medical Fold Flat<br>Surgical Mask is intended to be worn<br>to protect both patients and healthcare<br>personnel from transfer of<br>microorganisms, body fluids, and<br>particulate material. The PuraVita<br>Medical Fold Flat Surgical Mask is<br>intended for use in infection control<br>practices to reduce the potential<br>exposure to blood and body fluids.<br>This is a single-use, disposable device<br>that is provided non-sterile. | Similar |
| Materials | | | |
| Outer Layer | Polypropylene spunbond | Polypropylene spunbond | Same |
| Middle Filter Layer<br>#1 | Polypropylene Meltblown | Polypropylene Meltblown | Same |
| Middle Filter Layer<br>#2 | Non-woven Polypropylene | Polypropylene Meltblown | Different. Although<br>the 2nd filter layer of<br>the subject device is<br>different than the<br>predicate device, the<br>device met all<br>standards when tested. |
| Inner Layer | Polypropylene spunbond | Polypropylene spunbond | Same |
| Ear Loops | Spandex and Polyester | Spandex and Nylon | Different. Although<br>the material used in the<br>subject device is<br>slightly different than<br>the predicate device, it<br>passed all<br>biocompatibility test<br>requirements. |
| Nose Piece | Iron core polypropylene strip | Dual Iron core polypropylene strip | Similar |
| Latex | Not made with natural rubber latex | Not made with natural rubber latex | Same |
| Description | | | |
| Color | White | White | Same |
| Style | Extended chamber flat-folded | Fold Flat | Similar |
| Single Use | Yes | Yes | Same |
| Sterility | Non-Sterile | Non-sterile | Same |
| Length | $16.2cm ± 0.5cm$ | $16.3cm ± 0.5cm$ | Different. Although<br>the specifications and<br>dimensions of the<br>subject device are<br>slightly different than<br>the predicate device. |
| | | | the predicate device,<br>the differences are<br>minimal. |
| Width | $10.2cm \pm 0.5cm$ | $10.6cm \pm 0.5cm$ | Different. Although<br>the specifications and<br>dimensions of the<br>subject device are<br>slightly different than<br>the predicate device,<br>the differences are<br>minimal. |
| Performance Testing | | | |
| ASTM<br>Performance Level | ASTM Level 2 | ASTM Level 3 | Different. Although the<br>performance of the<br>subject device differs<br>from that of the<br>predicate device, the<br>subject device performs<br>at least as well as the<br>predicate. |
| Fluid/Blood<br>Penetration F1862 | Pass at 120 mmHg | Pass at 160mmHg | Different, however the<br>subject device filters as<br>well as the predicate<br>device. |
| Particulate<br>Filtration F2299 | 99.1%, filtration efficiency | >99.4% filtration efficiency | Different; however, the<br>subject and predicate<br>devices both pass the<br>test at ≥98% filtration<br>efficiency. |
| Bacterial Filtration<br>F2101 | 99.9% | >99.9% | Same |
| Diff. Pressure<br>(Delta-P)<br>MIL-M-36954C | On average of 5.04 mmH2O/cm2 | On average 5.3 mmH2O/cm2 | Different, both the<br>subject device and<br>predicate device pass<br>the test with a<br>differential pressure<br><6.0mmH2O/cm2. |
| Flammability<br>16 CFR 1610 | Class 1 | Class 1 | Same |
| Biocompatibility Testing | | | |
| Cytotoxicity | Non-Cytotoxic | Non-Cytotoxic | Same |
| Skin Irritation | Non-Irritating | Non-Irritating | Same |
| Skin Sensitization | Non-Sensitizing | Non-Sensitizing | Same |
### Comparison to predicate device:
{5}------------------------------------------------
- Middle Layer #2 the subject device is constructed using meltblown polypropylene, . while the predicate device is constructed with non-woven polypropylene. The mask meets the same testing requirements as the subject device.
- Ear Loops the subject device's ear loops are constructed with spandex and nylon, while ● the predicate device's loops are made with spandex and polyester. Despite this difference in material composition, the ear loops of the subject device meet all biocompatibility requirements; the difference does not negatively affect performance of the device.
- The dimensions (length and width) of the subject and predicate devices differ by 0.1cm. . This is within normal limits of variation (± 0.5cm).
{6}------------------------------------------------
- . Performance Testing (Fluid/Synthetic Blood Penetration ASTM F1862) – the subject and predicate devices differ in performance of fluid penetration testing. The subject device resists penetration by blood at the same level as the predicate device.
- Performance Testing (Particulate Filtration Efficiency ASTM F2299) the subject . performed differently from the predicate device; however, subject and predicate devices both passed the test at ≥ 98% filtration efficiency, meeting the requirements for levels 2 and 3.
- Performance Testing (Differential Pressure EN 14683:2019) While the average . differential pressures differ between the subject and predicate devices, both have average differential pressure values that pass the test at <6.0mmH2O/cm2.
- There were no differences between the biocompatibility of the subject and predicate . devices.
| Test | Device Description/<br>Sample Size | Test<br>Method/Applicable<br>Standards | Acceptance Criteria | Unexpected<br>Results/<br>Significant<br>Deviations | Results | Test | Predicate | Subject | Comparison |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------|--------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Bacterial<br>Filtration<br>Efficiency<br>ASTM F2101-19<br>Standard Test<br>Method for<br>Evaluating the<br>Bacterial Filtration<br>Efficiency (BFE)<br>of Medical Face<br>Mask Materials,<br>Using a Biological<br>Aerosol of<br>Staphylococcus<br>aureus<br><br>FDA Recognition<br>Number: 6-427<br><br>Objective of the<br>test:<br>Evaluate the<br>bacteria filtration<br>efficiency (BFE)<br>of face masks<br>samples, using a<br>biological aerosol<br>of Staphylococcus<br>aureus. | Test articles used in<br>the test<br>32 samples of PuraVita<br>Medical Fold Flat<br>Surgical Masks from 3<br>non-sequential lots<br>(total of 96 samples)<br>were tested; final<br>finished devices | Test methods and<br>procedures<br>(including any<br>specific test<br>conditions)<br>ASTM F2101-19<br>method was<br>followed under<br>standard conditions<br>with no deviations | Study endpoint, i.e., the<br>specific parameter<br>measured<br>Bacterial filtration<br>efficiency<br><br>Pre-defined acceptance<br>or pass/fail criteria.<br>Level 2 Barrier: ≥ 98%<br>filtration efficiency<br>Level 3 Barrier: ≥ 98%<br>filtration efficiency | None | All samples tested at least<br>98% filtration efficiency.<br><br>All 96 samples meet the<br>"pass" criteria of the test<br>for levels 2 and 3.<br><br>Comparative testing of the<br>predicate device was not<br>performed, since<br>information regarding its<br>performance can be<br>obtained from FDA's<br>public database.<br><br>The predicate device<br>passed at 99.9% filtration<br>efficiency.<br><br>When compared to the<br>information in the public<br>summary, the subject<br>device and predicate<br>device are Similar. | Bacterial Filtration<br>Efficiency | $> 99.1% FiltrationEfficiencyPass$ | $\ge 98% FiltrationEfficiencyPass$ | Similar, both the subject<br>device and predicate<br>device passed the test at<br>the same level ( $\ge$ 98%). |
| Differential<br>Pressure (Delta-<br>P) | Test articles used in<br>the test | Test methods and<br>procedures<br>(including any | Study endpoint, i.e., the<br>specific parameter | None | All 96 samples measured<br>an average differential | Differential Pressure<br>(Delta-P) | $< 6.0 mmH_2O/cm^2$<br>Pass | $< 6.0 mmH_2O/cm^2$<br>Pass | Same |
| EN 14683:2019 +<br>AC 2019 (E)<br>Annex C: Medical<br>Face Masks -<br>Requirements and<br>Test Methods;<br>Method for<br>determination of<br>breathability<br>Objective of the<br>Test: Determine<br>the breathability of<br>masks by<br>measuring the<br>differential air<br>pressure on either<br>side of the test<br>article using a<br>manometer, at a<br>constant flow rate<br>of 8 L/min. | 32 samples of PuraVita<br>Medical Fold Flat<br>Surgical Masks from 3<br>non-sequential lots<br>(total of 96 samples)<br>were tested; final<br>finished devices | specific test<br>conditions)<br>EN 14683:2019 +<br>AC 2019 (E) Annex<br>C method was<br>followed under<br>standard conditions<br>with no deviations | Pressure in mmH2O/cm2<br>Pre-defined acceptance<br>or pass/fail criteria.<br>Passing differential<br>pressure must be<br>< 6.0mmH2O/cm2. | | (The average pressure was<br>5.3mmH2O/cm2.)<br>Comparative testing of the<br>predicate device was not<br>performed, since<br>information regarding its<br>performance can be<br>obtained from FDA's<br>public database. The<br>predicate device had an<br>average differential<br>pressure of<br>5.04mmH2O/cm2.<br>When compared to the<br>information in the public<br>summary, the subject<br>device and predicate<br>device were found to be<br>Similar. | Sub-Micron<br>Particulate Filtration<br>Efficiency at 0.1um<br>of Polystyrene Latex<br>Spheres | 99.9% filtration<br>efficiency<br>Pass | $\ge$ 98% filtration<br>efficiency<br>Pass | Similar; both the subject<br>device and predicate<br>device passed the test at<br>the same level ( $\ge$ 98%). |
| Sub-Micron<br>Particulate<br>Filtration<br>Efficiency at<br>0.1um of<br>Polystyrene Latex<br>Spheres<br>ASTM F2100-19<br>Standard<br>Specification for<br>Performance of<br>Materials Used in<br>Medical Face<br>Masks<br>(Parent Standard<br>of ASTM<br>F2299/F2299M-03<br>(Reapproved 2017)<br>- Test Method for<br>Determining the<br>Initial Efficiency<br>of Materials used<br>in Medical Face<br>Masks to<br>Penetration by<br>Particulates Using<br>Latex Spheres)<br>Objective of Test:<br>Determine the<br>Initial Efficiency<br>of Materials Used<br>in Medical Face<br>Masks to<br>Penetration by<br>Particulates Using<br>Latex Spheres | Test articles used in<br>the test<br>32 samples of PuraVita<br>Medical Fold Flat<br>Surgical Masks from 3<br>non-sequential lots<br>(total of 96 samples)<br>were tested; final<br>finished devices | Test methods and<br>procedures<br>(including any<br>specific test<br>conditions)<br>ASTM<br>F2299/F2299M-03<br>(reapproved 2017)<br>method was<br>followed under<br>standard conditions<br>with no deviations | Study endpoint, i.e., the<br>specific parameter<br>measured<br>Sub-micron particulate<br>filtration at 0.1 micron<br>Pre-defined acceptance<br>or pass/fail criteria.<br>Level 1 Barrier: ≥ 95%<br>filtration efficiency<br>Level 2 Barrier: ≥ 98%<br>filtration efficiency<br>Level 3 Barrier: ≥ 98%<br>filtration efficiency | None | All samples filtered at a<br>level ≥ 98% filtration<br>efficiency<br>All samples meet the<br>standard's requirements for<br>Level 2 filtration and met<br>the same criteria as the<br>predicate.<br>Comparative testing of the<br>predicate device was not<br>performed, since<br>information regarding its<br>performance can be<br>obtained from FDA's<br>public database.<br>The predicate device's<br>particulate filtration<br>efficiency was 99.1%.<br>When compared to the<br>information in the public<br>summary, the subject<br>device and predicate<br>device were found to be<br>Similar, as they meet the<br>requirements of the same<br>level of filtration<br>efficiency. | Resistance to<br>Penetration by<br>Synthetic Blood | Fluid resistant claimed<br>at 120mmHg<br>Pass | Fluid resistance at<br>160mmHg<br>Pass | Similar- the subject<br>device resisted blood<br>penetration at least as<br>well as the predicate<br>device. |
| Flammability<br>16 CFR 1610<br>Standard for the<br>flammability of<br>clothing textiles.<br>73 FR 62187, Oct.<br>20, 2008<br>Objective of the<br>Test: Evaluate the<br>flammability of<br>face mask samples,<br>prohibiting the use<br>of any dangerous<br>flammable<br>materials | Test articles used in<br>the test<br>32 samples of PuraVita<br>Medical Fold Flat<br>Surgical Masks from 3<br>non-sequential lots<br>(total of 96 samples)<br>were tested; final<br>finished devices | Test methods and<br>procedures<br>(including any<br>specific test<br>conditions)<br>The method<br>described in 16 CFR<br>1610 was followed<br>under standard<br>conditions with no<br>deviations | Study endpoint, i.e., the<br>specific parameter<br>measured<br>Textile's burn time (the<br>time elapsed from<br>ignition until the stop<br>thread is severed as<br>measured by the timing<br>mechanism of the test<br>apparatus.)<br>Pre-defined acceptance<br>or pass/fail criteria.<br>Class I criteria:<br>The sample does not<br>ignite (DNI) or ignites<br>but extinguishes (IBE).<br>Or burn time, as defined<br>by the standard is ≥ 3.5<br>seconds | None | All samples either did not<br>ignite (DNI) or ignited but<br>extinguished (IBE).<br>All samples met the<br>requirements for Class I<br>Flammability.<br>Comparative testing of the<br>predicate device was not<br>performed, since<br>information regarding its<br>performance can be<br>obtained from FDA's<br>public database. The<br>predicate device met the<br>requirements for Class I<br>flammability.<br>When compared to the<br>information in the public<br>summary, the subject<br>device and predicate<br>device were found to be<br>Similar. | Flammability | Class 1 | Class 1 | Same |
| Resistance to<br>Penetration by<br>Synthetic Blood<br>F1862/F1862M-17<br>Standard Test<br>Method for<br>Resistance of<br>Medical Face<br>Masks to<br>Penetration by<br>Synthetic Blood<br>(Horizontal<br>Projection of Fixed<br>Volume at a<br>Known Velocity)<br>FDA Recognition<br>Number: 6-406<br>Objective of the<br>Test:<br>Measure the<br>resistance of<br>medical face<br>masks samples to<br>penetration by<br>synthetic blood. | Test articles used in<br>the test<br>32 samples of PuraVita<br>Medical Fold Flat<br>Surgical Masks from 3<br>non-sequential lots<br>(total of 96 samples)<br>were tested; final<br>finished devices | Test methods and<br>procedures<br>(including any<br>specific test<br>conditions)<br>The method<br>described in<br>F1862/F1862M-17<br>was followed under<br>standard conditions<br>with no deviations | Study endpoint, i.e., the<br>specific parameter<br>measured<br>Visual detection of<br>synthetic blood<br>penetration.<br>Pre-defined acceptance<br>or pass/fail criteria.<br>Pass criteria: No visual<br>blood penetration | None | None of the tested samples<br>showed penetration of<br>synthetic blood up to<br>160mmHg.<br>All samples passed the<br>requirements of the test<br>and performed as well as<br>the predicate.<br>Comparative testing of the<br>predicate device was not<br>performed, since<br>information regarding its<br>blood penetration<br>performance can be<br>obtained from FDA's<br>public database.<br>The predicate device<br>resisted penetration at<br>120mmHg.<br>When compared to the<br>information in the public<br>summary, the subject<br>device and predicate<br>device were found to be<br>Similar. | | | | |
### Summary of Bench Performance Testing
{7}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{8}------------------------------------------------
## Comparison of Performance Testing
When compared to FDA's public information regarding the predicate's performance, the subject device was found to perform similarly to or exactly the same as the predicate.
{9}------------------------------------------------
{10}------------------------------------------------
## Summary of Biocompatibility Testing
The nature of body contact for the subject device is Surface Device category, Skin Contact and duration of contact is limited (≤24h). The following tests were conducted on the final, finished form of the subject device to demonstrate that the subject device is biocompatible and safe for its intended use:
| Test | Device Description/<br>Sample Size | Test Method/Applicable<br>Standards | Acceptance Criteria | Unexpected<br>Results/<br>Significant<br>Deviations | Results |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity -<br>MEM Elution<br>Test<br><br>FDA Recognition<br>Number: 2-245<br><br>Objective of the<br>test:<br>The purpose of this<br>test was to<br>evaluate the<br>cytotoxic potential<br>of extracts of<br>polymeric<br>materials or any<br>other materials<br>intended to be<br>implanted in the<br>human body or<br>that may come into<br>contact with bodily<br>fluids or injectable<br>solutions. | Test articles used in<br>the test<br>1 Puravita Medical<br>Fold Flat Surgical<br>Mask; final finished<br>device; all components<br>tested together as<br>composite.<br><br>(Lot No. 12142020A) | Test methods and<br>procedures<br>(including any<br>specific test<br>conditions)<br>The testing standard<br>ANSI/AAMI/ISO<br>10993-5:2009 was<br>followed under<br>standard conditions<br>with no deviations<br>or amendments. | Study endpoint, i.e.,<br>the specific parameter<br>measured<br>The morphology of cells<br>was observed at the 24<br>and 48 hour examination<br>points according to the<br>criteria stated in the ISO<br>10993-5 guidelines. The<br>average score for the<br>three test wells and<br>controls at the 48-hour<br>point was used to<br>determine the cytotoxic<br>response.<br><br>Pre-defined acceptance<br>or pass/fail criteria.<br>Cell morphology graded<br>greater than 2 is<br>considered to have a<br>cytotoxic effect. | None | Pass.<br><br>Based on qualitative<br>evaluation of the cells<br>exposed to the test article<br>extract, the test article was<br>not considered to have a<br>cytotoxic effect.<br><br>Comparative testing of the<br>predicate device was not<br>performed, since<br>information regarding its<br>biocompatibility can be<br>obtained from FDA's<br>public database.<br><br>When compared to the<br>information in the public<br>summary, the subject<br>device and predicate device<br>were both found to be non-<br>cytotoxic. |
| Sensitization -<br>Maximization<br>Test for Delayed-<br>Type<br>Hypersensitivity<br>in Hartley Guinea<br>Pigs (ISO 10993-<br>10:2010)<br><br>FDA Recognition<br>Number: 2-174<br><br>Objective of the<br>test:<br>The purpose of this<br>test was to<br>determine to what<br>extent the test<br>article has the<br>potential to act as a<br>contact sensitizer<br>in guinea pigs. | Test articles used in<br>the test<br>6 Puravita Medical<br>Fold Flat Surgical<br>Masks; final finished<br>devices; all<br>components tested<br>together as composite.<br><br>(Lot no. 12142020A) | Test methods and<br>procedures<br>(including any<br>specific test<br>conditions)<br>The study was<br>performed<br>according to ISO<br>10993-10<br>guidelines, with no<br>deviations or<br>amendments. | Study endpoint, i.e.,<br>the specific parameter<br>measured<br>In the final analysis of<br>data, consideration was<br>given to the overall<br>patterns, intensity,<br>duration, and the nature<br>of reactions of the test as<br>compared with the<br>control.<br><br>Pre-defined acceptance<br>or pass/fail criteria.<br>Any skin reaction scores<br>received by the test<br>group, which were<br>greater than the scores<br>received by the negative<br>control group, were<br>considered to represent<br>sensitization. | None | Pass.<br><br>No sensitization reactions<br>or patterns were noted in<br>animals exposed to test<br>article extracted in either<br>saline or cottonseed oil.<br>The test animals did not<br>receive scores higher than<br>those of the negative<br>control animals.<br><br>Comparative testing of the<br>predicate device was not<br>performed, since<br>information regarding its<br>biocompatibility can be<br>obtained from FDA's<br>public database.<br><br>When compared to the<br>information in the public<br>summary, the subject<br>device and predicate device<br>were both found to be non-<br>sensitizing. |
| Intracutaneous<br>(Intradermal)<br>Reactivity Test<br>ISO 10993-<br>10:2010<br>Intracutaneous<br>(Intradermal)<br>Reactivity Test in<br>New Zealand<br>White Rabbits<br><br>FDA Recognition<br>Number: 2-174<br><br>Objective of the<br>test:<br>The purpose of this<br>test was to<br>evaluate local<br>responses to<br>extracts of the test<br>article following<br>intracutaneous<br>injections into<br>rabbits. | Test articles used in<br>the test<br>2 Puravita Medical<br>Fold Flat Surgical<br>Masks final, finished<br>devices; all<br>components tested<br>together as composite.<br><br>(Lot No 12142020A) | Test methods and<br>procedures<br>(including any<br>specific test<br>conditions)<br>The study was<br>performed<br>according to ISO<br>10993-10<br>guidelines, with no<br>deviations or<br>amendments. | Study endpoint, i.e.,<br>the specific parameter<br>measured<br>Injection sites were<br>observed for erythema,<br>eschar formation,<br>edema, and necrosis, and<br>scored at $24 \pm 2$ hours,<br>$48 \pm 2$ hours, and $72 \pm 2$<br>hours. The average<br>scores for the test sites<br>were calculated and<br>compared to the average<br>scores for the control<br>sites.<br><br>Pre-defined acceptance<br>or pass/fail criteria.<br>According to ISO<br>10993-10, the<br>requirements of the test<br>are met if the difference<br>between the test article<br>extract average score<br>and the control average<br>score is 1.0 or less and<br>the test does not fail at<br>any observation period. | Pass.<br>Based on erythema and<br>edema scores, the test<br>article extracted in saline<br>or cottonseed oil did not<br>elicit biologically<br>significant irritation<br>reactions when compared<br>to the control after being<br>injected intracutaneously.<br><br>Both the subject and<br>predicate devices were<br>found to be non-irritating.<br><br>Comparative testing of the<br>predicate device was not<br>performed, since<br>information regarding its<br>biocompatibility can be<br>obtained from FDA's<br>public database.<br><br>When compared to the<br>information in the public<br>summary, the subject<br>device and predicate device<br>were both found to be non-<br>irritating. | |
{11}------------------------------------------------
## Comparison of Biocompatibility Results Between the Subject & Predicate Devices
| | Predicate Device<br>(K202107) | Subject Device<br>PuraVita Medical<br>Fold Flat Surgical<br>Mask | Comparison |
|----------------------------------------------------|------------------------------------|------------------------------------------------------------------|------------|
| Cytotoxicity | Not Cytotoxic | Not Cytotoxic | Sam…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.