← Product Code [LZA](/productcode/LZA) · K212735

# Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs) (K212735)

_Xingyu Medical Tech Co., Ltd. · LZA · Jan 11, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212735

## Device Facts

- **Applicant:** Xingyu Medical Tech Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jan 11, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs) are a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Summarized in Table 1 below, the proposed device was tested for use with chemotherapy drugs per ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

## Device Story

Disposable, non-sterile, powder-free nitrile examination gloves; worn by healthcare examiners to prevent cross-contamination. Designed for single use; available in sizes S, M, L, XL. Tested for permeation resistance against various chemotherapy drugs per ASTM D6978-05(2019). Provides a physical barrier between patient and examiner; intended for use in clinical settings. Does not protect against all chemotherapy agents (e.g., do not use with Carmustine).

## Clinical Evidence

Bench testing only. Performance evaluated per ASTM D6319-19 (physical dimensions, tensile strength, elongation, freedom from holes, residual powder) and ASTM D6978-05(2019) (chemotherapy drug permeation). Biocompatibility testing performed per ISO 10993-10 (skin irritation, sensitization) and ISO 10993-11 (acute systemic toxicity). All tests passed.

## Technological Characteristics

Material: Nitrile. Form factor: Disposable, powder-free, non-sterile glove. Standards: ASTM D6319-19 (physical properties/dimensions), ASTM D6978-05(2019) (chemotherapy permeation), ASTM D5151-19 (freedom from holes), ASTM D6124-06 (residual powder). Biocompatibility: ISO 10993-10, ISO 10993-11.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Medline Nitrile Powder-Free Dark Blue Examination Gloves (Tested for use with Chemotherapy Drugs) ([K200960](/device/K200960.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

January 11, 2022

Xingyu Medical Tech Co., Ltd. % Eva Li Consultant Shanghai Sungo Management Consulting Company Limited Room 1309, Dongfang Building, 1500#Century Ave Shanghai, Shanghai 200122 China

Re: K212735

Trade/Device Name: Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: December 16, 2021 Received: December 16, 2021

Dear Eva Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212735

#### Device Name

Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs)

#### Indications for Use (Describe)

The Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs) are a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Summarized in Table 1 below, the proposed device was tested for use with chemotherapy drugs per ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

| Chemotherapy Drug                  | Concentration          | Breakthrough time    |
|------------------------------------|------------------------|----------------------|
| Carmustine (BCNU)                  | 3.3mg/ml (3,300ppm)    | 47.9(68.2,47.9,69.3) |
| Cyclophosphamide (Cytoxan)         | 20.0mg/ml (20,000ppm)  | >240 min             |
| Doxorubicin HCI                    | 2.0 mg/ml (2,000ppm)   | >240 min             |
| Etoposide                          | 20.0 mg/ml (20,000ppm) | >240 min             |
| Fluorouracil                       | 50.0 mg/ml (50,000ppm) | >240 min             |
| Paclitaxel                         | 6.0 mg/ml (6,000ppm)   | >240 min             |
| ThioTepa                           | 10.0 mg/ml (10,000ppm) | >240 min             |
| Bleomycin Sulfate(Blenoxane)       | 15mg/ml (15,000ppm)    | >240 min             |
| Carboplatin                        | 10 mg/ml (10,000ppm)   | >240 min             |
| Cisplatin                          | 1.0 mg/ml (1,000ppm)   | >240 min             |
| Cytarabine                         | 100mg/ml(100,000ppm)   | >240 min             |
| Dacarbazine                        | 10.0mg/ml (10,000ppm)  | >240 min             |
| Idarubicin HCL                     | 1mg/ml (1,000ppm)      | >240 min             |
| Ifosfamide                         | 50mg/ml (50,000ppm)    | >240 min             |
| Mechlorethamine HCL                | 1mg/ml (1,000ppm)      | >240 min             |
| MESNA                              | 100mg/ml (100,000ppm)  | >240 min             |
| Methotrexate                       | 25mg/ml (25,000ppm)    | >240 min             |
| Mitomycin C                        | 0.5 mg/ml (500ppm)     | >240 min             |
| Mitoxantrone                       | 2mg/ml(2,000ppm)       | >240 min             |
| Trisenox                           | 1 mg/ml(1,000ppm)      | >240 min             |
| Vincristine Sulfate                | 1 mg/ml (1,000ppm)     | >240 min             |
| Do Not Use with Carmustine (BCNU). |                        |                      |

#### Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

XINGYU MEDICAL TECH CO.,LTD.

NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE,

CHINA

# 510K Summary

The assigned 510(k) number is: K212735

Premarket Notification [510(k)] Summary

### 1. Submitter / 510(k) Sponsor:

Submitter's name : XINGYU MEDICAL TECH CO.,LTD. Submitter's Address: NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE, CHINA Phone number: 0086-18263618867 Contact person: Cathrine Luan Email: cathrine@xingyugloves.com

Submission Correspondent Primary contact: Ms. Eva Li Shanghai SUNGO Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: eatereva@hotmail.com Secondary contact: Mr. Raymond Luo Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

Date of Preparation: 2021-12-14

### 2. Proposed Device

Device Name / Classification Trade Name: Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs) Common Name: Patient examination glove Classification Name: Medical Gloves with Chemotherapy Labeling Claims – Test For Use with Chemotherapy Drugs Product Code: LZA. LZC Classification Panel: General Hospital Regulatory Class: Class I Regulation Number: 21 CFR 880.6250

#### 3. Predicate Device

Device Name: Medline Nitrile Powder-Free Dark Blue Examination Gloves (Tested for use with

{5}------------------------------------------------

NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE,

CHINA

Chemotherapy Drugs) Company Name: Medline Industries, Inc. 510(K) Number: K200960

## 4. Device Description

The Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs) are nonsterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the patient and the examiner. The gloves are blue, powder free, nitrile gloves. The gloves are offered in sizes small, medium, large, extra large, and packaged in a color paper box.

The gloves are designed and manufactured in accordance with the ASTM D6319-19 standard and are tested for use with chemotherapy drugs per ASTM D6978-05(2019).

# 5. Indications for Use

The Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs) are a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Summarized in Table 1 below, the proposed device was tested for use with chemotherapy drugs per ASTM D6978-05(2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

| Chemotherapy Drug                | Concentration          | Breakthrough time        |
|----------------------------------|------------------------|--------------------------|
| Carmustine (BCNU)                | 3.3mg/ml (3,300ppm)    | 47.9<br>(68.2,47.9,69.3) |
| Cyclophosphamide<br>(Cytoxan)    | 20.0mg/ml (20,000ppm)  | > 240 min                |
| Doxorubicin HCI                  | 2.0 mg/ml (2,000ppm)   | > 240 min                |
| Etoposide                        | 20.0 mg/ml (20,000ppm) | > 240 min                |
| Fluorouracil                     | 50.0 mg/ml (50,000ppm) | > 240 min                |
| Paclitaxel                       | 6.0 mg/ml (6,000ppm)   | > 240 min                |
| ThioTepa                         | 10.0 mg/ml (10,000ppm) | > 240 min                |
| Bleomycin<br>Sulfate (Blenoxane) | 15mg/ml (15,000ppm)    | > 240 min                |
| Carboplatin                      | 10 mg/ml (10,000ppm)   | > 240 min                |
| Cisplatin                        | 1.0 mg/ml (1,000ppm)   | > 240 min                |
| Cytarabine                       | 100mg/ml(100,000ppm)   | > 240 min                |
| Dacarbazine                      | 10.0mg/ml (10,000ppm)  | > 240 min                |
| Idarubicin HCL                   | 1mg/ml (1,000ppm)      | > 240 min                |
| Ifosfamide                       | 50mg/ml (50,000ppm)    | > 240 min                |
| Mechlorethamine HCL              | 1mg/ml (1,000ppm)      | > 240 min                |
| MESNA                            | 100mg/ml (100,000ppm)  | > 240 min                |
| Methotrexate                     | 25mg/ml (25,000ppm)    | > 240 min                |
| Mitomycin C                      | 0.5 mg/ml (500ppm)     | > 240 min                |
| Mitoxantrone                     | 2mg/ml(2,000ppm)       | > 240 min                |

{6}------------------------------------------------

XINGYU MEDICAL TECH CO.,LTD.

NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE,

K212735(Page 3 / 7)

CHINA

| Trisenox            | 1 mg/ml(1,000ppm)  | > 240 min |
|---------------------|--------------------|-----------|
| Vincristine Sulfate | 1 mg/ml (1,000ppm) | > 240 min |

Do Not Use with Carmustine (BCNU).

### 6. Summary of Technological Characteristics

Table2: Comparison of Proposed and Predicate Devices

| Device<br>Characteristic         | Proposed Device                                                                                                                                                                                                                                            | Predicate Device                                                                                                                                                                                                                                           | Comparison<br>Analysis |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Product Name                     | Nitrile disposable examination<br>gloves (Tested for use with<br>Chemotherapy Drugs)                                                                                                                                                                       | Medline Nitrile Powder Free<br>Dark Blue Examination Gloves<br>(Tested for use with<br>Chemotherapy Drugs)                                                                                                                                                 |                        |
| 510(k)<br>Reference              | K212735                                                                                                                                                                                                                                                    | K200960                                                                                                                                                                                                                                                    |                        |
| Product Owner                    | XINGYU MEDICAL TECH CO.,LTD.                                                                                                                                                                                                                               | Medline                                                                                                                                                                                                                                                    |                        |
| Product Code                     | LZA, LZC                                                                                                                                                                                                                                                   | LZA, LZC                                                                                                                                                                                                                                                   | Same                   |
| Intended Use                     | A patient examination glove is a<br>disposable device intended for<br>medical purposes that is worn on<br>the examiner's hand to prevent<br>contamination between patient<br>and examiner. These gloves were<br>tested for use with<br>chemotherapy drugs. | A patient examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiner's hand to<br>prevent contamination between<br>patient and examiner. These<br>gloves were tested for use with<br>chemotherapy drugs. | Same                   |
| Regulation<br>Number             | 21 CFR 880.6250                                                                                                                                                                                                                                            | 21 CFR 880.6250                                                                                                                                                                                                                                            | Same                   |
| Design<br>Configurations         | Blue                                                                                                                                                                                                                                                       | Dark Blue                                                                                                                                                                                                                                                  | Similar                |
| Materials                        | Nitrile                                                                                                                                                                                                                                                    | Nitrile                                                                                                                                                                                                                                                    | Same                   |
| Prescription vs.<br>OTC          | OTC                                                                                                                                                                                                                                                        | OTC                                                                                                                                                                                                                                                        | Same                   |
| Contact<br>Durations             | Limited ≤ 24 hours                                                                                                                                                                                                                                         | Limited ≤ 24 hours                                                                                                                                                                                                                                         | Same                   |
| Sterile vs. Non-<br>Sterile      | Non-Sterile                                                                                                                                                                                                                                                | Non-Sterile                                                                                                                                                                                                                                                | Same                   |
| Disposable vs.<br>Non-Disposable | Disposable                                                                                                                                                                                                                                                 | Disposable                                                                                                                                                                                                                                                 | Same                   |
| Single Use vs.<br>Reusable       | Single Use                                                                                                                                                                                                                                                 | Single Use                                                                                                                                                                                                                                                 | Same                   |
| Dimensions-<br>Width             | Complies with: ASTM D6319-19<br>70mm min                                                                                                                                                                                                                   | Complies with: ASTM D6319-10<br>70mm min                                                                                                                                                                                                                   | Same                   |
| Dimensions-                      | Complies with: ASTM D6319-19                                                                                                                                                                                                                               | Complies with: ASTM D6319-10                                                                                                                                                                                                                               | Same                   |

{7}------------------------------------------------

K212735(Page 4 / 7)

XINGYU MEDICAL TECH CO.,LTD.

NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE,

CHINA

| Thickness                                                                                                     | Palm - 0.05mm min. Finger –<br>0.05mm min                                                                                                                                                 |                           | Palm - 0.05mm min. Finger –<br>0.05mm min                                                                                                                                                 |                                        |                            |                      |         |
|---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|----------------------------|----------------------|---------|
| Physical<br>Properties                                                                                        | Complies with: ASTM D6319-19<br>minimum:<br>Tensile Strength:<br>Before Aging ≥14 MPa, min.<br>After Aging ≥14 MPa, min.<br>Elongation: Before Aging 500%, min.<br>After Aging 400%, min. |                           | Complies with: ASTM D6319-10<br>minimum:<br>Tensile Strength:<br>Before Aging ≥14 MPa, min.<br>After Aging ≥14 MPa, min.<br>Elongation: Before Aging 500%, min.<br>After Aging 400%, min. |                                        | Same                       |                      |         |
| Freedom from<br>holes                                                                                         | Complies with: ASTM D6319-19<br>and ASTM D5151-19 G-1,<br>AQL 1.5                                                                                                                         |                           | Complies with: ASTM D6319-10<br>and ASTM D5151-06 G-1,<br>AQL 1.5                                                                                                                         |                                        | Same                       |                      |         |
| Powder or<br>Powder Free                                                                                      | Powder Free                                                                                                                                                                               |                           | Powder Free                                                                                                                                                                               |                                        | Same                       |                      |         |
| Residual<br>Powder                                                                                            | Complies with ASTM D6319-19                                                                                                                                                               |                           | Complies with ASTM D6319-10                                                                                                                                                               |                                        | Same                       |                      |         |
| Biocompatibility                                                                                              | Complies with AAMI/ANSI/ISO<br>10993-10: Not a skin irritant Not<br>a skin sensitizer AAMI/ANSI/ ISO<br>10993-11: Non-Toxic                                                               |                           | Complies with AAMI/ANSI/ISO<br>10993-10: Not a skin irritant<br>Not a skin sensitizer<br>AAMI/ANSI/ ISO 10993-11: Non-Toxic                                                               |                                        | Same                       |                      |         |
| Chemotherapy<br>Drugs Tested<br>with Minimum<br>Breakthrough<br>Detection Time<br>as tested per<br>ASTM D6978 | Chemotherapy Drug                                                                                                                                                                         | Concentration             | Breakthrough<br>time                                                                                                                                                                      | Chemotherapy<br>Drug                   | Concentration              | Breakthrough<br>time | Simliar |
|                                                                                                               | Carmustine (BCNU)                                                                                                                                                                         | 3.3mg/ml (3,300ppm)       | 47.9<br>(68.2,47.9,69.3)                                                                                                                                                                  | Carmustine<br>(BCNU)                   | 3.3 mg/ml<br>(3,300 ppm)   | 12.4 minutes         |         |
|                                                                                                               | Cyclophosphamide<br>(Cytoxan)                                                                                                                                                             | 20.0mg/ml (20,000ppm)     | > 240 min                                                                                                                                                                                 | Cisplatin                              | 1.0 mg/ml<br>(1,000 ppm)   | >240 minutes         |         |
|                                                                                                               | Doxorubicin HCl                                                                                                                                                                           | 2.0 mg/ml (2,000ppm)      | > 240 min                                                                                                                                                                                 | Cyclophosphamide<br>(Cytoxan)          | 20.0 mg/ml<br>(20,000 ppm) | >240 minutes         |         |
|                                                                                                               | Etoposide                                                                                                                                                                                 | 20.0 mg/ml<br>(20,000ppm) | > 240 min                                                                                                                                                                                 | Dacarbazine<br>(DTIC)                  | 10.0 mg/ml<br>(10,000 ppm) | >240 minutes         |         |
|                                                                                                               | Fluorouracil                                                                                                                                                                              | 50.0 mg/ml<br>(50,000ppm) | > 240 min                                                                                                                                                                                 | Doxorubicin<br>Hydrochloride           | 2.0 mg/ml<br>(2,000 ppm)   | >240 minutes         |         |
|                                                                                                               | Paclitaxel                                                                                                                                                                                | 6.0 mg/ml (6,000ppm)      | > 240 min                                                                                                                                                                                 | Etoposide<br>(Toposar)                 | 20.0 mg/ml<br>(20,000 ppm) | >240 minutes         |         |
|                                                                                                               | Thio Tepa                                                                                                                                                                                 | 10.0 mg/ml<br>(10,000ppm) | > 240 min                                                                                                                                                                                 | Fluorouracil                           | 50.0 mg/ml<br>(50,000 ppm) | >240 minutes         |         |
|                                                                                                               | Bleomycin<br>Sulfate(Blenoxane)                                                                                                                                                           | 15mg/ml (15,000ppm)       | > 240 min                                                                                                                                                                                 | Methotrexate                           | 25 mg/ml<br>(25,000 ppm)   | >240 minutes         |         |
|                                                                                                               | Carboplatin                                                                                                                                                                               | 10 mg/ml (10,000ppm)      | > 240 min                                                                                                                                                                                 | Mitomycin C                            | 0.5 mg/ml (500<br>ppm)     | >240 minutes         |         |
|                                                                                                               | Cisplatin                                                                                                                                                                                 | 1.0 mg/ml<br>(1,000ppm)   | > 240 min                                                                                                                                                                                 | Paclitaxel (Taxol)                     | 6.0 mg/ml<br>(6,000 ppm)   | >240 minutes         |         |
|                                                                                                               | Cytarabine                                                                                                                                                                                | 100mg/ml(100,000ppm)      | > 240 min                                                                                                                                                                                 | Thio Tepa                              | 10.0 mg/ml<br>(10,000 ppm) | 27.4 minutes         |         |
|                                                                                                               | Dacarbazine                                                                                                                                                                               | 10.0mg/ml (10,000ppm)     | > 240 min                                                                                                                                                                                 | Vincristine Sulfate<br>(Oncovin)       | 1.0 mg/ml<br>(1,000 ppm)   | >240 minutes         |         |
|                                                                                                               | Idarubicin HCL                                                                                                                                                                            | 1mg/ml (1,000ppm)         | > 240 min                                                                                                                                                                                 | Do Not Use with Carmustine or Thiotepa |                            |                      |         |
|                                                                                                               | Ifosfamide                                                                                                                                                                                | 50mg/ml (50,00ppm)        | > 240 min                                                                                                                                                                                 |                                        |                            |                      |         |
|                                                                                                               | Mechlorethamine<br>HCL                                                                                                                                                                    | 1mg/ml (1,000ppm)         | > 240 min                                                                                                                                                                                 |                                        |                            |                      |         |
|                                                                                                               | MESNA                                                                                                                                                                                     | 100mg/ml<br>(100,000ppm)  | > 240 min                                                                                                                                                                                 |                                        |                            |                      |         |
|                                                                                                               | Methotrexate                                                                                                                                                                              | 25mg/ml (25,000ppm)       | > 240 min                                                                                                                                                                                 |                                        |                            |                      |         |
|                                                                                                               | Mitomycin C                                                                                                                                                                               | 0.5 mg/ml (500ppm)        | > 240 min                                                                                                                                                                                 |                                        |                            |                      |         |
|                                                                                                               | Mitoxantrone                                                                                                                                                                              | 2mg/ml(2,000ppm)          | > 240 min                                                                                                                                                                                 |                                        |                            |                      |         |
|                                                                                                               | Trisenox                                                                                                                                                                                  | 1 mg/ml(1,000ppm)         | > 240 min                                                                                                                                                                                 |                                        |                            |                      |         |
|                                                                                                               | Vincristine Sulfate                                                                                                                                                                       | 1 mg/ml (1.000ppm)        | > 240 min                                                                                                                                                                                 |                                        |                            |                      |         |

## 7. Summary of Non-Clinical Testing

Physical performance qualities of the proposed device were evaluated per ASTM D6319- 19, Standard Specification for Nitrile Examination Gloves for Medical Application. Permeation testing was conducted to support the addition of the labeling claim: Tested for use with 

{8}------------------------------------------------

NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE, CHINA

chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05(2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs

To summarize, the performance testing of the subject device were conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

ASTM D 6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application ASTM D 6124-06 (2017) Standard Test Method for Residual Powder on Medical Gloves

ASTM D 5151-19 Standard Test Method for Detection of Holes in Medical Gloves

ASTM D 6978-05(2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs

| Test<br>standard<br>followed                                                         | Test<br>conducted                                    | Acceptance criteria                |                                      |       |        | Conclusion |      |
|--------------------------------------------------------------------------------------|------------------------------------------------------|------------------------------------|--------------------------------------|-------|--------|------------|------|
|                                                                                      |                                                      | S                                  | M                                    | L     | XL     |            |      |
| ASTM D6319-19                                                                        | Physical<br>Dimensions                               | Length (mm)                        | ≥220                                 |       | ≥230   | pass       |      |
|                                                                                      |                                                      | Width (mm) ±10                     | 80±10                                | 95±10 | 110±10 | 120±10     | pass |
|                                                                                      |                                                      | Thickness<br>Finger<br>Palm        | ≥0.05 mm                             |       |        |            | pass |
| ASTM D6319-19                                                                        | Physical<br>properties                               | Before Aging                       | Tensile ≥14Mpa; Elongation ≥<br>500% |       |        |            | pass |
|                                                                                      |                                                      | After Aging                        | Tensile ≥14Mpa; Elongation ≥<br>400% |       |        |            | pass |
|                                                                                      |                                                      |                                    |                                      |       |        |            |      |
| ASTM D6319-19<br>Test method in<br>accordance with<br>ASTM D5151-19                  | Watertight<br>ness Test<br>for detection<br>of holes | Batch size=35000, sample 125<br>≤7 |                                      |       |        | pass       |      |
|                                                                                      |                                                      |                                    |                                      |       |        |            |      |
|                                                                                      |                                                      |                                    |                                      |       |        |            |      |
| ASTM standard<br>D 6319-19<br>Test method<br>in<br>accordance with<br>D6124-06(2017) | Powder<br>Residual                                   | ≤2 (mg/glove)                      |                                      |       |        | pass       |      |
|                                                                                      |                                                      |                                    |                                      |       |        |            |      |
|                                                                                      |                                                      |                                    |                                      |       |        |            |      |

{9}------------------------------------------------

### K212735(Page 6 / 7)

# XINGYU MEDICAL TECH CO.,LTD.

NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE,

CHINA

| ASTM D 6978-<br>05(2019) | Chemotherapy<br>Drugs Tested<br>with Minimum<br>Breakthrough<br>Detection<br>Time | Chemotherapy Drug                  | Concentration             | Breakthrough<br>time     |
|--------------------------|-----------------------------------------------------------------------------------|------------------------------------|---------------------------|--------------------------|
|                          |                                                                                   | Carmustine (BCNU)                  | 3.3mg/ml (3,300ppm)       | 47.9<br>(68.2,47.9,69.3) |
|                          |                                                                                   | Cyclophosphamide<br>(Cytoxan)      | 20.0mg/ml (20,000ppm)     | > 240 min                |
|                          |                                                                                   | Doxorubicin HCl                    | 2.0 mg/ml (2,000ppm)      | > 240 min                |
|                          |                                                                                   | Etoposide                          | 20.0 mg/ml<br>(20,000ppm) | > 240 min                |
|                          |                                                                                   | Fluorouracil                       | 50.0 mg/ml<br>(50,000ppm) | > 240 min                |
|                          |                                                                                   | Paclitaxel                         | 6.0 mg/ml (6,000ppm)      | > 240 min                |
|                          |                                                                                   | ThioTepa                           | 10.0 mg/ml<br>(10,000ppm) | > 240 min                |
|                          |                                                                                   | Bleomycin<br>Sulfate(Blenoxane)    | 15mg/ml (15,000ppm)       | > 240 min                |
|                          |                                                                                   | Carboplatin                        | 10 mg/ml (10,000ppm)      | > 240 min                |
|                          |                                                                                   | Cisplatin                          | 1.0 mg/ml<br>(1,000ppm)   | > 240 min                |
|                          |                                                                                   | Cytarabine                         | 100mg/ml(100,000ppm)      | > 240 min                |
|                          |                                                                                   | Dacarbazine                        | 10.0mg/ml (10,000ppm)     | > 240 min                |
|                          |                                                                                   | Idarubicin HCl                     | 1mg/ml (1,000ppm)         | > 240 min                |
|                          |                                                                                   | Ifosfamide                         | 50mg/ml (50,00ppm)        | > 240 min                |
|                          |                                                                                   | Mechlorethamine<br>HCl             | 1mg/ml (1,000ppm)         | > 240 min                |
|                          |                                                                                   | MESNA                              | 100mg/ml<br>(100,000ppm)  | > 240 min                |
|                          |                                                                                   | Methotrexate                       | 25mg/ml (25,000ppm)       | > 240 min                |
|                          |                                                                                   | Mitomycin C                        | 0.5 mg/ml (500ppm)        | > 240 min                |
|                          |                                                                                   | Mitoxantrone                       | 2mg/ml( 2,000ppm)         | > 240 min                |
|                          |                                                                                   | Trisenox                           | 1 mg/ml(1,000ppm)         | > 240 min                |
|                          |                                                                                   | Vincristine Sulfate                | 1 mg/ml (1,000ppm)        | > 240 min                |
|                          |                                                                                   | Do Not Use with Carmustine (BCNU). |                           |                          |

### 8. Biocompatibility

The following tests were performed to evaluate the biocompatibility of the Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs)

- ISO 10993-10: Primary Skin Irritation
- ISO 10993-10: Dermal Sensitization
- · ISO 10993-11: Acute Systemic Toxicity

| Test conducted   | Standard                                                                 | Acceptance<br>criteria          | Result                                                                                                  |
|------------------|--------------------------------------------------------------------------|---------------------------------|---------------------------------------------------------------------------------------------------------|
| Biocompatibility | Skin Irritation Test Extraction<br>Method ISO 10993-10: 2010             | non-irritation                  | Passes<br>Under the conditions of<br>the study, the subject<br>device is non-<br>irritation             |
|                  | Skin sensitization the guinea<br>pig maximization ISO 10993-<br>10: 2010 | non- sensitization              | Passes<br>Under the conditions of<br>the study, the subject<br>device is non-sensitization              |
|                  | Acute Systemic Toxicity<br>Test ISO 10993-11: 2017                       | non- acute systemic<br>toxicity | Passes<br>Under the conditions of<br>the study, the subject<br>device is non-acute<br>systemic toxicity |

{10}------------------------------------------------

## 9. Conclusion

Based on the nonclinical tests data, it can be concluded that the Nitrile disposable examination gloves (Tested for use with Chemotherapy Drugs) is as safe, as effective, and performs as well as the predicate device, Medline Nitrile Powder-Free Dark Blue Examination Gloves (Tested for use with Chemotherapy Drugs) by Medline Industries, Inc. K200960.

---

**Source:** [https://fda-staging.innolitics.com/device/K212735](https://fda-staging.innolitics.com/device/K212735)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).