← Product Code [ODP](/productcode/ODP) · K212715

# Vy Spine ClariVy Cervical IBF System (K212715)

_Vy Spine, LLC · ODP · Dec 20, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212715

## Device Facts

- **Applicant:** Vy Spine, LLC
- **Product Code:** [ODP](/productcode/ODP.md)
- **Decision Date:** Dec 20, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Vy Spine™ ClariVy™ Cervical IBF System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The device systems are designed for use with autogenous bone graft to facilitate fusion. One device may be used per intervertebral space. The implants are intended to be used with legally cleared supplemental spinal fixation cleared for the implanted level. The Vy Spine™ ClariVy™ Cervical IBF System is intended for use at one level in the cervical spine, from C3 to T1, for treatment of cervical degenerate disc disease (DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The Vy Spine™ ClariVy™ Cervical IBF System is to be used in patients who have six weeks of non-operative treatment.

## Device Story

Intervertebral body fusion spacer; inserted into anterior column of cervical spine; used to treat cervical degenerative disc disease. System includes PEEK Optima LT1 or PEEK Optima LT1-HA spacers with Tantalum markers; optional Titanium alloy bone screws for fixation. Used by surgeons in clinical settings; requires supplemental spinal fixation. Device maintains disc space height; facilitates fusion via autogenous bone graft. Benefits include stabilization of spinal segment and reduction of discogenic neck pain.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical mechanical testing performed on the predicate device (K173102) per ASTM F-2077, ASTM F-2267, and ASTM Draft F-04.25.02.02, including static/dynamic axial compression, torsion, shear, expulsion, and subsidence testing.

## Technological Characteristics

Spacer composed of PEEK Optima LT1 or PEEK Optima LT1-HA with Tantalum markers; Titanium alloy bone screws. Mechanical testing per ASTM F-2077, ASTM F-2267, and ASTM Draft F-04.25.02.02. No software or electrical components.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Reliance Cervical IBF ([K173102](/device/K173102.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 20, 2021

Vy Spine, LLC Jordan Hendrickson Operations Manager 2236 Capital Circle NE, Suite 103-1 Tallahassee. Florida 32308

Re: K212715

Trade/Device Name: Vy Spine™ ClariVy™ Cervical IBF System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP, OVE Dated: November 16, 2021 Received: November 18, 2021

Dear Mr. Hendrickson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212715

Device Name Vy SpineTM ClariVyTM Cervical IBF System

#### Indications for Use (Describe)

The Vy Spine™ ClariVy™ Cervical IBF System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The device systems are designed for use with autogenous bone graft to facilitate fusion. One device may be used per intervertebral space. The implants are intended to be used supplemental spinal fixation cleared for the implanted level.

The Vy Spine™ ClariVy™ Cervical IBF System is intended for use at one level in the cervical spine, from C3 to T1, for treatment of cervical degenerate disc disease (DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The Vy Spine™ ClariVy™ Cervical IBF System is to be used in patients who have six weeks of non-operative treatment.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

Vy Spine, LLC 2236 Capital Circle NE, Suite 103-1 Tallahassee, FL 32308 Telephone: 866-489-7746 Fax: 850-597-8571

#### Contact: Jordan Hendrickson Operations Manager

| 510(k) Number:                      | K212715                                |
|-------------------------------------|----------------------------------------|
| Common or Usual Name:               | Intervertebral Body Fusion Device      |
| Proposed Proprietary or Trade Name: | Vy Spine™ ClariVy™ Cervical IBF System |
| Classification Name:                | Intervertebral Body Fusion Device      |
| Regulation Number:                  | 21 CFR 888.3080                        |
| Product Code:                       | ODP, OVE                               |
| Regulatory Classification:          | Class II                               |

### Substantial Equivalence

The Vy Spine™ ClariVy™ Cervical IBF System is substantially equivalent to the primary predicate device Reliance Cervical IBF (K173102). The Vy Spine™ ClariVy™ Cervical IBF System is equivalent to these commercially available devices in terms of material, intended use, levels of attachment, size range, and use with supplemental fixation.

### Device Description

The Vy Spine™ ClariVy™ Cervical IBF System is comprised of implants components. The implant component, the Vy Spine™ ClariVy™ Cervical IBF device, is a spacer, which inserts between vertebral bodies in the anterior column of the cervical spine. The spacer may be made of PEEK Optima LT1 or PEEK Optima LT1-HA with Tantalum markers. The Vy Spine™ ClariVy™ Cervical IBF System may also include Titanium alloy bone screws to secure the device to the vertebral body.

### Intended Use/Indications for Use

The Vy Spine™ ClariVy™ Cervical IBF System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The device systems are designed for use with autogenous bone graft to facilitate fusion. One device may be used per intervertebral space. The implants are intended to be used with legally cleared supplemental spinal fixation cleared for the implanted level.

The Vy Spine™ ClariVy™ Cervical IBF System is intended for use at one level in the cervical spine, from C3 to T1, for treatment of cervical degenerate disc disease (DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic

7 December 2021

{4}------------------------------------------------

studies). The Vy Spine™ ClariVy™ Cervical IBF System is to be used in patients who have six weeks of non-operative treatment.

# Non-Clinical Testing

The Reliance Cervical IBF System (K173102) has undergone Non-Clinical Testing using various ASTM Standards at a third party facility. Mechanical testing was performed on the Reliance Cervical IBF System following ASTM F-2077, ASTM F-2267, and ASTM Draft F-04.25.02.02 standards. These mechanical tests included static axial compression, dynamic axial compression, static axial torsion, dynamic axial torsion, static compressive shear, dynamic compressive shear, expulsion, and subsidence. The subject Vy Spine™ ClariVy™ Cervical IBF System has the same material, design, sizes, indication of use & biocompatibility as the predicate devices.

# Technological Characteristics

The subject Vy Spine™ ClariVy™ Cervical IBF System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to have equivalent technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and function. This device does not contain software or electrical equipment.

---

**Source:** [https://fda-staging.innolitics.com/device/K212715](https://fda-staging.innolitics.com/device/K212715)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
