← Product Code [NHA](/productcode/NHA) · K212664

# Sherlock (K212664)

_Open Implants, LLC · NHA · Nov 30, 2021 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212664

## Device Facts

- **Applicant:** Open Implants, LLC
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Nov 30, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Sherlock abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations. All digitally designed CAD/CAM customizations for Sherlock abutments are to be sent to an Open Implants-validated milling center for manufacture. Sherlock abutments are compatible with the implant systems listed in the Compatibility Table:

## Device Story

Sherlock is a dental implant abutment system comprising titanium blanks and multi-unit (straight, 17° angled, 30° angled) abutments with corresponding screws. Used in dental clinics, the device provides support for prosthetic restorations on specific OEM endosseous dental implants. Clinicians select the appropriate abutment based on the implant system; CAD/CAM customizations are digitally designed and sent to an Open Implants-validated milling center for fabrication. The abutments interface with the implant to provide functional and esthetic rehabilitation for edentulous patients. The system benefits patients by enabling customized prosthetic support across various implant platforms.

## Clinical Evidence

No clinical data included. Substantial equivalence supported by bench testing, including static compression fatigue testing (ISO 14801) on worst-case constructs and reverse engineering of OEM implant interfaces to confirm compatibility.

## Technological Characteristics

Materials: Ti-6Al-4V alloy (ASTM F136). Design: Internal connection abutments (Titanium blanks, multi-unit straight/angled). Dimensions: Platform diameters 3.0-6.5 mm. Sterilization: Non-sterile, end-user sterilized. Connectivity: CAD/CAM workflow via validated milling centers.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- Sherlock ([K193335](/device/K193335.md))

## Reference Devices

- Freadoll ® Patient Specific Dental Abutments ([K191609](/device/K191609.md))
- Encode® Patient Specific Dental Abutments ([K101608](/device/K101608.md))
- Straumann® CARES® Titanium Alloy (TAN) Abutment ([K150899](/device/K150899.md))
- Zimmer® Patient Specific Abutment, Internal Hex, Titanium ([K143505](/device/K143505.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Open Implants, LLC % Chris Brown Manager Aclivi, LLC 3250 Brackley Drive Ann Arbor, Michigan 48105

Re: K212664

Trade/Device Name: Sherlock Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 29, 2021 Received: November 1, 2021

Dear Chris Brown:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Device Name Sherlock

Indications for Use (Describe)

Sherlock abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations.

All digitally designed CAD/CAM customizations for Sherlock abutments are to be sent to an Open Implants-validated milling center for manufacture.

Sherlock abutments are compatible with the implant systems listed in the Compatibility Table:

| Compatible Implant Systems | Implant Body Diameter (mm) | Implant Platform Diameter (mm) |
|----------------------------|----------------------------|--------------------------------|
| Biomet 3i Certain          | 3.25                       | 3.4                            |
|                            | 4.0                        | 4.1                            |
|                            | 5.0                        | 5.0                            |
|                            | 6.0                        | 6.0                            |
| Straumann Bone Level       | 3.3                        | 3.3 (NC)                       |
|                            | 4.1, 4.8                   | 4.1, 4.8 (RC)                  |
| Zimmer TSV                 | 3.7, 4.1                   | 3.5                            |
|                            | 4.7                        | 4.5                            |
|                            | 6.0                        | 5.7                            |

#### Compatibility Table

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------|
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary Open Implants, LLC Sherlock K212664 November 30, 2021

#### ADMINISTRATIVE INFORMATION

| Manufacturer Name | Open Implants, LLC<br>800 West Cummings Ave, Suite 4900<br>Woburn, MA 01801<br>Telephone: +1 781-587-3242<br>Fax: n/a |
|-------------------|-----------------------------------------------------------------------------------------------------------------------|
| Official Contact  | Gregg Gellman, CEO                                                                                                    |
| Email:            | ggellman@openimplants.com                                                                                             |

#### DEVICE NAME AND CLASSIFICATION

| Trade/Proprietary Name: | Sherlock                                                                                     |
|-------------------------|----------------------------------------------------------------------------------------------|
| Common Name:            | Abutment, Implant, Dental, Endosseous                                                        |
| Regulation Name:        | Endosseous dental implant abutment                                                           |
| Regulation Number:      | 21 CFR 872.3630                                                                              |
| Device Class:           | Class II                                                                                     |
| Product Code:           | NHA                                                                                          |
| Review Panel:           | Dental Products Panel                                                                        |
| Reviewing Branch:       | Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)<br>Dental Devices (DHT1B) |

#### PREDICATE DEVICE INFORMATION

The devices within this submission are substantially equivalent in indications, intended use and design principles to the following predicate and reference devices:

| 510(k)  | Predicate Device Name                        | Company Name       |
|---------|----------------------------------------------|--------------------|
| K193335 | Sherlock                                     | Open Implants, LLC |
| 510(k)  | Reference Device Name                        | Company Name       |
| K191609 | Freadoll ® Patient Specific Dental Abutments | Biomet 3i, Inc.    |

| 510(k)  | Reference Device Name                                     | Company Name       |
|---------|-----------------------------------------------------------|--------------------|
| K101608 | Encode® Patient Specific Dental Abutments                 | Biomet 3i, Inc     |
| K150899 | Straumann® CARES® Titanium Alloy (TAN) Abutment           | Straumann USA, LLC |
| K143505 | Zimmer® Patient Specific Abutment, Internal Hex, Titanium | Zimmer Dental, Inc |

{4}------------------------------------------------

#### INDICATIONS FOR USE

Sherlock abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations.

All digitally designed CAD/CAM customizations for Sherlock abutments are to be sent to an Open Implants-validated milling center for manufacture.

Sherlock abutments are compatible with the implant systems listed in the Compatibility Table:

| Compatible Implant Systems | Implant Body Diameter (mm) | Implant Platform Diameter (mm) |
|----------------------------|----------------------------|--------------------------------|
| Biomet 3i Certain          | 3.25                       | 3.4                            |
| Biomet 3i Certain          | 4.0                        | 4.1                            |
| Biomet 3i Certain          | 5.0                        | 5.0                            |
| Biomet 3i Certain          | 6.0                        | 6.0                            |
| Straumann Bone Level       | 3.3                        | 3.3 (NC)                       |
| Straumann Bone Level       | 4.1, 4.8                   | 4.1, 4.8 (RC)                  |
| Zimmer TSV                 | 3.7, 4.1                   | 3.5                            |
| Zimmer TSV                 | 4.7                        | 4.5                            |
| Zimmer TSV                 | 6.0                        | 5.7                            |

### Compatibility Table

### DEVICE DESCRIPTION

Sherlock is a dental implant abutment system that includes four (4) abutment designs compatible with three (3) OEM implant systems. The Subject device abutment platform diameters range from 3.0 mm to 6.5 mm, and the corresponding compatible implant body diameters also range from 3.0 mm to 6.5 mm. The subject device includes the following abutment designs: Titanium blank, multi-unit angled 17°, and multiunit angled 30°. The system also includes corresponding abutment screws.

The following table shows the Subject device abutments for each of the Compatible implant platforms.

| Sherlock Platform Diameter/<br>Compatible Implant System | Subject Device Abutment Designs |                     |                       |                       |                 |
|----------------------------------------------------------|---------------------------------|---------------------|-----------------------|-----------------------|-----------------|
|                                                          | Titanium Blank                  | Multi-unit Straight | Multi-Unit Angled 17° | Multi-Unit Angled 30° | Titanium Screws |
| Biomet 3i Certain                                        |                                 |                     |                       |                       |                 |
| 3.4 mm                                                   | X                               | X                   | X                     | X                     | X               |
| 4.1 mm                                                   | X                               | X                   | X                     | X                     | X               |
| 5.0 mm                                                   | X                               | X                   | X                     | X                     | X               |
| 6.0 mm                                                   | X                               | X                   | X                     | X                     | X               |
| Straumann Bone Level                                     |                                 |                     |                       |                       |                 |
| 3.3 mm (NC)                                              | X                               | X                   | X                     | X                     | X               |
| 4.1 mm (RC)                                              | X                               | X                   | X                     | X                     | X               |
| 4.8 mm (RC)                                              | X                               | X                   | X                     | X                     | X               |
| Zimmer TSV                                               |                                 |                     |                       |                       |                 |
| 3.5 mm                                                   | X                               | X                   | X                     | X                     | X               |
| 4.5 mm                                                   | X                               | X                   | X                     | X                     | X               |
| 5.7 mm                                                   | X                               | X                   |                       |                       | X               |

All abutments and screws are manufactured from Ti-6Al-4V alloy conforming to ASTM F136 and are provided nonsterile to the end user. All digitally designed customized Titanium Blank abutments are to an Open Implants-validated milling center for manufacture.

The design parameters for the CAD/CAM Titanium Blank custom abutment are: Minimum wall thickness – 0.41 to 1.6 mm (varies by implant line); Minimum post height for single-unit restoration - 4.0 mm;

{5}------------------------------------------------

Maximum Correction Angle - 30°; Minimum gingival height - 0.5 mm to 0.8 mm (varies by implant line); Maximum gingival height - 5 mm.

### PERFORMANCE DATA

Non-clinical data submitted to demonstrate substantial equivalence included: biocompatibility according to ANSI/AAMI ST72; biocompatibility testing of the K193335 Predicate device to ISO 10993-5 in support of the Subject device; reverse engineering of OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility; and static compression fatigue testing according to ISO 14801; sterilization validations of the K193335 Predicate device in support of the Subject device. For each compatible OEM implant line, worst-case constructs were subjected to static compression fatigue testing. No animal or clinical data is included in this premarket notification.

### EQUIVALENCE TO MARKETED DEVICE

Overall, the Subject device is substantially equivalent in indications and design principles to the Predicate device listed above. Provided at the end of this summary are tables comparing the Indications for Use Statements and the technological characteristics of the Subject, Predicate device, and Reference devices.

The Indications for Use Statement (IFUS) of the Subject device is substantially equivalent to that of the K193335 Predicate device. Differences in the list of compatible implant systems do not affect the intended use as an endosseous dental implant abutment for support of a prosthesis to restore chewing function.

Similarly, the differences between the Subject device IFUS and that of each of the Reference devices are related to the specific device names and design features, validated milling centers, and the compatible implant lines. None of these minor differences impact substantial equivalence because all IFUS express equivalent intended use to facilitate dental prosthetic restorations, and the indications are expressed equivalently using different specific wording.

Subject device abutments are substantially equivalent in intended use and abutment designs to the sponsor's K193335 Predicate device. All are intended for use with endosseous dental implants in the maxilla and mandible to provide functional and esthetic rehabilitation of the edentulous maxilla and mandible. Reference devices introduce additional restorative interfaces and restorative platform and implant diameters with similar intended uses.

The Subject device and K193335 Predicate devices are for single-unit or multi-unit restorations, have internal implant interface connections, and are made of Ti-6Al-4V ELI alloy (abutment screws). The Subject device abutment designs are substantially equivalent to that of the K193335 Predicate device designs: Titanium Blank; Straight or Angled 17° and Angled 30° Multi-Unit abutments.

The Subject device includes abutment designs for implant restorative platforms ranging from 3.3 mm. The K19335 Predicate included implant platform sizes of 3.5 mm to 3.9 mm. Substantial equivalence of the smaller and larger implant restorative platform diameters of the Subject device are supported by the Reference devices K101608, K150899, and K143505 and through performance testing of the Subject device.

The Subject device is to be sterilized by the end-user, using the same methods as previously validated for the sponsor's K19335 Predicate device.

Minor differences in the designs, dimensions, sizes, or compatible OEM implant lines between the Predicate device, and the Reference devices do not affect substantial equivalence. Additional implant/abutment interfaces and dimensions are supported by Reference devices. The Subject, Predicate and Reference devices encompass the same range of physical dimensions. These minor differences are related to the compatible OEM implant designs and are mitigated by mechanical performance testing.

{6}------------------------------------------------

### CONCLUSION

Overall, the Indications for Use statements for the Subject and Predicate devices are substantially equivalent differing only in the list of compatible implant system systems.

Overall, the Technological Characteristics, mode of operation and materials of the Subject device are substantially equivalent to that of the Predicate device with additional compatible implant systems supported by Reference devices.

Overall, the data included in this premarket notification demonstrate substantial equivalence of Subject device to the sponsor's Predicate device.

The basis for the belief that the Subject device is substantially equivalent to the Predicate device and is summarized in the following comparison tables.

{7}------------------------------------------------

| Subject Device                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                               | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Reference Device                                                                                                                                                                                                                                                                                          | Reference Device                                                                                                                                                                                                                                                                                        |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                               |  |  |  |  |  |  |  |  |  |  |  |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|
| Sherlock<br>Open Implants, LLC                                                                                                                                                                                                                                                                                                                                                                 | Sherlock<br>Open Implants, LLC<br>K193335                                                                                                                                                                                                                                                                                                                                                      | Encode® Patient Specific Dental<br>Abutments<br>Biomet 3i, Inc<br>K101608<br><br>BIOMET 3i Dental Abutments are<br>intended for use as accessories to<br>endosseous dental implant to<br>support a prosthetic device in a<br>partially or completely edentulous<br>patient. A dental abutment is<br>intended for use to support single<br>and multiple tooth prosthesis, in<br>the mandible or maxilla. The<br>prosthesis can be screw retained or<br>cement retained to the abutment.<br>Restorative Components:<br><br>· Temporary Healing Abutments<br>are intended for use to shape and<br>maintain the soft tissue opening<br>during healing.<br><br>· Castable restorative components<br>are intended for use as accessories<br>to endosseous dental implants to<br>aid in the fabrication of dental<br>prosthetics.<br><br>· Screw components are intended<br>for use as accessories to<br>endosseous dental implants for<br>retention of screw retained<br>abutments to the dental implant. | Straumann® CARES® Titanium Alloy<br>(TAN) Abutments<br>Straumann USA, LLC<br>K150899<br><br>The Straumann CARES® TAN<br>abutments are indicated for single<br>tooth replacement and multiple<br>tooth restorations. The prosthetic<br>restoration can be cemented or<br>directly veneered/screw-retained. | Zimmer® Patient Specific Abutment,<br>Internal Hex, Titanium<br>Zimmer Dental, Inc<br>K143505<br><br>The Zimmer® Patient Specific<br>Abutment is used as a terminal or<br>intermediate abutment for a cemented<br>prosthesis. The abutment can be used<br>for a single or multiple-unit<br>restoration. |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                               |  |  |  |  |  |  |  |  |  |  |  |
| Sherlock abutments are intended<br>to be used in conjunction with<br>endosseous dental implants in<br>the maxillary or mandibular arch<br>to provide support for single-unit<br>or multi-unit prosthetic<br>restorations.<br><br>All digitally designed CAD/CAM<br>customizations for Sherlock<br>abutments are to be sent to an<br>Open Implants-validated milling<br>center for manufacture. | Sherlock abutments are intended<br>to be used in conjunction with<br>endosseous dental implants in<br>the maxillary or mandibular arch<br>to provide support for single-unit<br>or multi-unit prosthetic<br>restorations.<br><br>All digitally designed CAD/CAM<br>customizations for Sherlock<br>abutments are to be sent to an<br>Open Implants-validated milling<br>center for manufacture. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                         |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                               |  |  |  |  |  |  |  |  |  |  |  |
| Sherlock abutments are<br>compatible with the implant<br>systems listed in the<br>Compatibility Table:                                                                                                                                                                                                                                                                                         | Sherlock abutments are<br>compatible with the implant<br>systems listed in the<br>Compatibility Table:                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                         |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |                                                                                                                                                               |  |  |  |  |  |  |  |  |  |  |  |
| Compatibility Table   Compatible<br>Implant<br>Systems Implant<br>Body<br>Diameter<br>(mm) Implant<br>Platform<br>Diameter<br>(mm)   Biomet 3i 3.25 3.4   Certain 4.0 4.1    5.0 5.0    6.0 6.0   Straumann 3.3 3.3 (NC)   Bone Level 4.1, 4.8 4.1, 4.8 (RC)    3.7, 4.1 3.5   Zimmer TSV 4.7 4.5    6.0 5.7                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                         |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  | Compatible<br>Implant<br>Systems Implant<br>Body<br>Diameter<br>(mm) Implant<br>Platform<br>Diameter<br>(mm)   NobelActive® 3.5 3.5 (NP)    4.3, 5.0 3.9 (RP) |  |  |  |  |  |  |  |  |  |  |  |

# Comparison of Indications for Use Statement

{8}------------------------------------------------

| Comparison                                 | Subject Device                                                                   | Predicate Device                                                                                             | Reference Device                                                                                    | Reference Device                                                                 | Reference Device                                                                                                  |
|--------------------------------------------|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
|                                            | Sherlock                                                                         | Sherlock                                                                                                     | Encode®@ Patient Specific Dental Abutments                                                          | Straumann® CARES® Titanium Alloy (TAN) Abutments                                 | Zimmer® Patient Specific Abutment, Internal Hex,                                                                  |
|                                            | Open Implants                                                                    | Open Implants, LLC                                                                                           | Biomet 3i, Inc                                                                                      | Straumann USA, LLC                                                               | Titanium                                                                                                          |
|                                            |                                                                                  | K193335                                                                                                      | K101608                                                                                             | K150899                                                                          | Zimmer Dental, Inc                                                                                                |
|                                            |                                                                                  |                                                                                                              |                                                                                                     |                                                                                  | K143505                                                                                                           |
| Intended Use                               | Functional and esthetic rehabilitation of<br>the edentulous maxilla and mandible | Functional and esthetic rehabilitation of the<br>edentulous maxilla and mandible                             | Functional and esthetic rehabilitation of the<br>edentulous maxilla and mandible                    | Functional and esthetic rehabilitation of the edentulous<br>maxilla and mandible | Functional and esthetic rehabilitation of the<br>edentulous maxilla and mandible                                  |
| Reason for<br>Predicate/Reference          | Not Applicable                                                                   | Abutment configurations listed, material,<br>sterilization, biocompatibility, CAD/CAM<br>abutment technology | Implant/abutment interface, prosthetic<br>diameters, CAD/CAM abutment technology                    | Implant/abutment interface, prosthetic diameters, CAD/CAM<br>abutment technology | Implant/abutment interface, prosthetic diameters,<br>CAD/CAM abutment technology                                  |
|                                            | Titanium Blank                                                                   | Titanium Blank                                                                                               | Abutment types, not specified, but consistent<br>with Titanium Blank or Ti-Base/Hybrid<br>abutment. | Titanium Blank / Stock                                                           | Titanium Blank / Stock                                                                                            |
|                                            | Minimum wall thickness - 0.41 - 0.65 mm<br>(varies by implant line);             | Minimum wall thickness - 0.42 mm;                                                                            | Minimum wall thickness - not specified;                                                             | Minimum wall thickness - 0.4 mm;                                                 | Minimum wall thickness - not specified;                                                                           |
|                                            | Minimum post height for single-unit<br>restoration - 4.0 mm;                     | Minimum post height for single-unit<br>restoration - 4.0 mm;                                                 | Minimum post height for single unit<br>restoration - not specified                                  | Minimum post height for single unit restoration - not specified                  | Minimum post height for single unit restoration - not<br>specified                                                |
|                                            | Maximum Correction Angle - 30°;                                                  | Maximum Correction Angle - 30°;                                                                              | Maximum correction angle - 30°;                                                                     | Maximum correction angle - NC - 0°, RC - 30°;                                    | Maximum correction angle - 30°;                                                                                   |
|                                            | Gingival height - 0.5 to 5 mm;                                                   | Gingival Height - 0.5 mm to 5 mm;                                                                            | Gingival Height - 0.25 minimum (3.4 mm<br>internal connection)                                      | Maximum gingival height - no limit;                                              | Maximum gingival height - no limit;                                                                               |
|                                            | Gingival Diameter - 3.0 - 11.9 mm (varies<br>by implant line)                    | Gingival Diameter – 3.4 – 11.9 mm                                                                            | Gingival Diameter - 3.8 - 16 mm                                                                     | Gingival Diameter - not specified                                                | Gingival Diameter:<br>3.5mm Platform: 3.5- 9.0 mm<br>4.5mm Platform: 4.5- 10.0 mm<br>5.7mm Platform: 5.7- 12.0 mm |
|                                            | Abutment height - 6.4 - 15 mm (varies by<br>implant line)                        | Abutment height - 7.3 - 15 mm                                                                                | Abutment height - 4.75 - 15 mm                                                                      | Abutment Height- not specified                                                   | Abutment Height- not specified                                                                                    |
|                                            | Multi-Unit Straight                                                              | Multi-Unit Straight                                                                                          |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | PD = 3.3 - 4.8 mm                                                                | PD = 3.5, 3.9 mm                                                                                             |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | GH = 1, 1.5, 2, 2.5, 3, 3.5, 4, 5 mm                                             | GH = 1.5, 2, 3, 5 mm                                                                                         |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | GD = 4.8 mm                                                                      | GD = 4.8 mm                                                                                                  |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | PH = 4.4 mm (includes insert)                                                    | PH = 4.4 mm (includes insert)                                                                                |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | AH = 7.4 - 12.5 mm (includes insert)                                             | AH = 7.5 -11 mm (includes insert)                                                                            |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | Multi-Unit 17°                                                                   | Multi-Unit 17°                                                                                               |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | PD = 3.3 - 4.8 mm                                                                | PD = 3.5, 3.9 mm                                                                                             |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | GH = 2, 2.5, 3, 3.5, 4 mm                                                        | GH = 3.5, 5 mm                                                                                               |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | GD = 4.8 mm                                                                      | GD = 4.8 mm                                                                                                  |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | PH = 4.4 mm (includes insert)                                                    | PH = 4.4 mm (includes insert)                                                                                |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | AH = 7.9 - 12.1 mm (includes insert)<br>Multi-Unit 30°                           | AH = 7.3 - 10.8 mm (includes insert)<br>Multi-Unit 30°                                                       |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | PD = 3.3 - 4.8 mm                                                                | PD = 3.5, 3.9 mm                                                                                             |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | GH = 3, 3.5, 4, 4.5, 5 mm                                                        | GH = 3.5, 5 mm                                                                                               |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | GD = 4.8 mm                                                                      | GD = 4.8 mm                                                                                                  |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | PH = 4.4 mm (includes insert)                                                    | PH = 4.4 mm (includes insert)                                                                                |                                                                                                     |                                                                                  |                                                                                                                   |
|                                            | AH = 8.3 - 12.5 mm (includes insert)                                             | AH = 7.7 - 11.2 mm (includes insert)                                                                         |                                                                                                     |                                                                                  |                                                                                                                   |
| Abutment/Implant<br>Platform Diameter (mm) | 3.3 - 6.0                                                                        | 3.5, 3.9                                                                                                     | 3.4, 4.1, 5.0, 6.0                                                                                  | 3.3, 4.1, 4.8                                                                    | 3.5, 4.5, 5.7 mm                                                                                                  |
| Material<br>(Abutment and Screw)           | Ti-6AL-4V Alloy                                                                  | Ti-6AL-4V Alloy                                                                                              | Titanium Alloy or Biocompatible Zirconia TZP                                                        | Titanium Alloy (Ti-6AL-7NB)                                                      | Titanium 6Al-4V Alloy                                                                                             |
| Abutment/Implant<br>Interface              | Internal Connection                                                              | Internal Connection                                                                                          | Internal Connection and External Connection                                                         | Internal Connection                                                              | Internal Connection                                                                                               |
| Prosthesis Attachment                      | Cement-retained<br>Screw-retained                                                | Cement-retained<br>Screw-retained                                                                            | Cement-retained<br>Screw-retained                                                                   | Cement-retained<br>Screw-retained                                                | Cement-retained                                                                                                   |
| Restoration                                | Single-unit<br>Multi-unit                                                        | Single-unit<br>Multi-unit                                                                                    | Single-unit<br>Multi-unit                                                                           | Single-unit                                                                      | Single-unit<br>Multi-unit                                                                                         |

Comparison of Technological Characteristics

PD = Abutment/Implant Platform (Restorative) Diameter
GH = Gingival Height
GD = Gingival Diameter
PH = Post Height

AH = Abutment Height

---

**Source:** [https://fda-staging.innolitics.com/device/K212664](https://fda-staging.innolitics.com/device/K212664)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com