← Product Code [LZA](/productcode/LZA) · K212661

# Disposable Nitrile Examination Glove (K212661)

_Inner Mongolia Cureguard Medical Technology Co., Ltd. · LZA · Nov 19, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212661

## Device Facts

- **Applicant:** Inner Mongolia Cureguard Medical Technology Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Nov 19, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

## Device Story

Disposable nitrile examination glove; non-sterile; ambidextrous; white color. Worn by clinicians during medical examinations to provide a barrier against contamination. Device functions as a physical barrier; no electronic or mechanical components. Manufactured to meet ASTM standards for physical dimensions, tensile strength, and watertightness. Used in clinical settings; single-use only.

## Clinical Evidence

Bench testing only. No clinical data. Testing included physical dimensions (ASTM D6319), watertightness (ASTM D5151), residual powder (ASTM D6124), tensile strength/elongation (ASTM D412), and biocompatibility (ISO 10993-5 cytotoxicity, ISO 10993-10 irritation and sensitization). All tests passed.

## Technological Characteristics

Material: Nitrile. Design: Ambidextrous, powder-free, non-sterile. Standards: ASTM D6124 (residual powder), ASTM D5151 (watertightness), ASTM D6319 (physical dimensions/specifications), ASTM D412 (tensile properties). Biocompatibility: ISO 10993-5, ISO 10993-10.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color ([K171422](/device/K171422.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

November 19, 2021

Inner Mongolia Cureguard Medical Technology Co., Ltd. Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161, East Lujiazui Rd., Pudong Shanghai, Shanghai 200120 China

Re: K212661

Trade/Device Name: Disposable Nitrile Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: July 26, 2021 Received: August 23, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K212661

Device Name Disposable Nitrile Examination Glove

Indications for Use (Describe)

The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K212661

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

## 1.0 Submitter's Information

Name: Inner Mongolia Cureguard Medical Technology Co.,Ltd. Address: Room 326, Management Committee of New Industrial Park, Tumote youqi, Baotou, Inner Mongolia Autonomous Region 014100, China. Phone Number: +86-13485097856 Contact: Guo Hua Date of Preparation: Jul.26,2021

#### Designated Submission Correspondent

Mr. Boyle Wanq Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn

#### 2.0 Device Information

Trade name: Disposable Nitrile Examination Glove Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove XS,S, M. L. XL Model(s):

#### 3.0 Classification

Production code: LZA Requlation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

#### 4.0 Predicate Device Information

Ever Global (Vietnam) Enterprise Corp Manufacturer: Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422

#### 5.0 Indication for Use

{4}------------------------------------------------

The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

### 6.0 Device Description

The subject device is powder free nitrile patient examination gloves. The subject device is white color. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The subject device is non-sterile.

| Item                        | Subject Device<br>(K212661)                                                                                                                                                                                     | Predicate Device<br>(K171422)                                                                                                                                                                                                         |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code                | LZA                                                                                                                                                                                                             | LZA                                                                                                                                                                                                                                   |
| Regulation No.              | 21CFR880.6250                                                                                                                                                                                                   | 21CFR880.6250                                                                                                                                                                                                                         |
| Class                       | I                                                                                                                                                                                                               | I                                                                                                                                                                                                                                     |
| Intended Use                | The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner. | The Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. |
| Material                    | Nitrile                                                                                                                                                                                                         | Nitrile                                                                                                                                                                                                                               |
| Powdered or<br>Powered free | Powdered free                                                                                                                                                                                                   | Powdered free                                                                                                                                                                                                                         |
| Design Feature              | Ambidextrous                                                                                                                                                                                                    | Ambidextrous                                                                                                                                                                                                                          |
| Colorant                    | White                                                                                                                                                                                                           | White/ Blue/ Black/ Pink                                                                                                                                                                                                              |
| Labeling Information        | Single-use indication, powder free, device color, device name, glove size and quantity,Non-Sterile                                                                                                              | Single-use indication, powder free, device color, device name, glove size and quantity, Non-Sterile                                                                                                                                   |
| Dimensions(mm)              | Length:<br>XS/S: $\ge$ 220;<br>M/L/XL: $\ge$ 230;<br>Width:<br>XS: 70 $\pm$ 10;<br>S: 80 $\pm$ 10;<br>M: 95 $\pm$ 10;                                                                                           | Length:<br>XS/S: $\ge$ 220; M: $\ge$ 235;<br>L/XL: $\ge$ 245<br>Width:<br>XS: 75 $\pm$ 5;<br>S: 85 $\pm$ 5;<br>M: 95 $\pm$ 5;                                                                                                         |

#### 7.0 Technological Characteristic Comparison Table

{5}------------------------------------------------

|                        |                 | L: 105±10;                                                                                                                                                                       | L: 105±5;                                                               |
|------------------------|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
|                        |                 | XL: 115±10                                                                                                                                                                       | XL: 115±5                                                               |
| Thickness(mm)          |                 | Finger: ≥0.05;                                                                                                                                                                   | Finger: ≥0.05;                                                          |
|                        |                 | Palm: ≥0.05                                                                                                                                                                      | Palm: ≥0.05                                                             |
| Physical<br>Properties | Before<br>Aging | Tensile<br>Strength<br>14MPa, min<br>Ultimate<br>Elongation<br>500% min                                                                                                          | Tensile<br>Strength<br>14MPa, min<br>Ultimate<br>Elongation<br>500% min |
|                        | After<br>Aging  | Tensile<br>Strength<br>14MPa, min                                                                                                                                                | Tensile<br>Strength<br>14MPa, min                                       |
|                        |                 | Ultimate<br>Elongation<br>400%min                                                                                                                                                | Ultimate<br>Elongation<br>400%min                                       |
|                        |                 | Freedom from<br>Holes                                                                                                                                                            |                                                                         |
| Powder Content         |                 | Meet the requirements of<br>ASTM D6124 <2.0mg                                                                                                                                    | Meet the requirements of<br>ASTM D6124                                  |
| Biocompatibility       |                 | ISO 10993-10;<br>Under the conditions of the<br>study, not an irritant or a<br>sensitizer<br>ISO 10993-5<br>Under conditions of the<br>study, device extract is not<br>cytotoxic | Comply with<br>ISO10993-10<br>/                                         |

Analysis: The physical dimensions are little different with that of the predicate, but they all meet the requirements of ASTM D6319.

# 8.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

{6}------------------------------------------------

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

| Test<br>Methodology | Purpose                                             | Acceptance Criteria                                                                                                                 | Results                                                                                                                                                                                                                                      |
|---------------------|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM D6319          | Physical<br>Dimensions<br>Test                      | Length(mm):<br>XS/S: $\ge$ 220;<br>M/L/XL: $\ge$ 230;<br>Width:<br>XS: 70±10;<br>S: 80±10;<br>M: 95±10;<br>L: 105±10;<br>XL: 115±10 | Length(mm):<br>>230<br>Width(mm):<br>XS: 73-76;<br>S: 80-83<br>M: 95-97<br>L: 110-114<br>XL: 118-120<br>Pass                                                                                                                                 |
|                     |                                                     | Thickness (mm) :<br>Finger: $\ge$ 0.05<br>Palm: $\ge$ 0.05                                                                          | XS:<br>Finger: 0.07-0.11<br>Palm: 0.08-0.10<br>S:<br>Finger: 0.08-0.10<br>Palm: 0.08-0.11<br>M:<br>Finger: 0.08-0.11<br>Palm: 0.08-0.11<br>L:<br>Finger: 0.08-0.12<br>Palm: 0.09-0.12<br>XL:<br>Finger: 0.08-0.12<br>Palm: 0.08-0.11<br>Pass |
| ASTM D5151          | Watertightness<br>Test for<br>Detection of<br>Holes | Meet the requirements of ASTM D5151 AQL 2.5                                                                                         | XS:0/125 leaks<br>S:0/125 leaks<br>M:0/125 leaks<br>L: 1/125 leaks<br>XL: 1/125 leaks<br>Pass                                                                                                                                                |
| ASTM D6124          | Powder<br>Content                                   | Meet the requirements of ASTM D6124 < 2.0mg                                                                                         | XS:0.04mg<br>S:0.06mg                                                                                                                                                                                                                        |

Table 2: Summary of Non-clinical Testing Table

{7}------------------------------------------------

|              |                        |                 |                        |        | M:0.08mg                                                                                           |
|--------------|------------------------|-----------------|------------------------|--------|----------------------------------------------------------------------------------------------------|
|              |                        |                 |                        |        | L:0.08mg                                                                                           |
|              |                        |                 |                        |        | XL:0.09mg                                                                                          |
|              |                        |                 |                        |        | Pass                                                                                               |
| ASTM D412    | Physical<br>properties | Before<br>Aging | Tensile<br>Strength    | ≥14MPa | XS:15.5-17.9<br>S:16.1-18.1<br>M: 15.4-17.9<br>L:15.3-17.7<br>XL:15.4-17.7<br>Pass                 |
|              |                        |                 |                        |        |                                                                                                    |
|              |                        |                 | Ultimate<br>Elongation | ≥500%  | XS:524-565<br>S:520-567<br>M: 528-567<br>L:521-566<br>XL:525-567<br>Pass                           |
|              |                        |                 |                        |        |                                                                                                    |
|              |                        |                 |                        |        |                                                                                                    |
|              |                        | After<br>Aging  | Tensile<br>Strength    | ≥14MPa | XS:15.3-17.4<br>S:15.4-17.8<br>M:15.5-17.8<br>L:15.5-17.5<br>XL:15.9-17.8<br>Pass                  |
|              |                        |                 |                        |        |                                                                                                    |
|              |                        |                 | Ultimate<br>Elongation | ≥400%  | XS:524-568<br>S:528-563<br>M:527-570<br>L:534-563<br>XL:530-569<br>Pass                            |
|              |                        |                 |                        |        |                                                                                                    |
|              |                        |                 |                        |        |                                                                                                    |
| ISO 10993-5  | Cytotoxicity           | Non-cytotoxic   |                        |        | Under conditions<br>of the study, did<br>not show potential<br>toxicity to L-929<br>cells.<br>Pass |
| ISO 10993-10 | Irritation             | Non-irritating  |                        |        | Under the<br>conditions of the<br>study, not an<br>irritant.<br>Pass                               |

{8}------------------------------------------------

| ISO 10993-10 | Sensitization | Non-sensitizing | Under conditions<br>of the study, not a<br>sensitizer.<br>Pass |
|--------------|---------------|-----------------|----------------------------------------------------------------|
|--------------|---------------|-----------------|----------------------------------------------------------------|

#### 9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

#### 10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device, Disposable Nitrile Examination Glove, is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K212661](https://fda-staging.innolitics.com/device/K212661)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com