K212650 · Gyrus Acmi, Inc. · ERL · Jan 19, 2022 · Ear, Nose, Throat
Device Facts
Record ID
K212650
Device Name
Celeris, Disposable Sinus Debrider
Applicant
Gyrus Acmi, Inc.
Product Code
ERL · Ear, Nose, Throat
Decision Date
Jan 19, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Celeris System (reusable power pack and Disposable Sinus Debrider) is intended for cutting, debriding, and removal of thin bone and soft tissue in general ENT and Sinus/Rhinology procedures and applications would include: · FESS (Functional Endoscopic Sinus Surgery) – Including Endoscopic approaches for: Polypectomy, Sphenoidectomy, Maxillary Antrostomy, Uncinectomy, Frontal Sinusotomy · Turbinate Reduction / Turbinoplasty, including sub-mucosal resection.
Device Story
Celeris System comprises a reusable power pack and a single-use Disposable Sinus Debrider (DSD) handpiece with integrated motor and blade. Used in ENT/sinus surgery (e.g., FESS, turbinoplasty) by surgeons. Input: electrical power from wall outlet; user input via handle trigger or footswitch. Operation: motor oscillates inner blade to cut/debride tissue; suction removes tissue/fluid. Bipolar models connect to external electrosurgical generator for RF coagulation. Output: mechanical tissue resection and RF-based hemostasis. Benefits: efficient tissue removal and coagulation in endoscopic procedures. Surgeon visualizes site endoscopically; device facilitates precise tissue management.
Clinical Evidence
No clinical data. Evidence based on non-clinical performance testing (electrical, mechanical, functional, stability) and preclinical bench tissue (ex vivo) studies using bovine tissue. Bench testing evaluated thermal margin, thermal impact, and coagulation performance. Biocompatibility testing performed per ISO 10993-1.
Technological Characteristics
System includes reusable power pack and disposable handpiece. Materials: plastics and stainless steel (patient-contacting). Principle: electric motor-driven oscillating blade with suction; optional bipolar RF coagulation. Dimensions: 2mm and 4mm blade diameters. Connectivity: universal power cord; external electrosurgical generator for RF. Sterilization: ETO (disposable handpiece). Software: controller board for motor speed/torque regulation.
Indications for Use
Indicated for patients undergoing general ENT and Sinus/Rhinology procedures, including FESS (polypectomy, sphenoidectomy, maxillary antrostomy, uncinectomy, frontal sinusotomy) and turbinate reduction/turbinoplasty. No specific age or gender contraindications provided.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
January 19, 2022
Gyrus ACMI, Inc. Dolan Mills Program Manager, Regulatory Affairs 800 West Park Drive Westborough, Massachusetts 01581
Re: K212650
Trade/Device Name: Celeris, Disposable Sinus Debrider Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, nose, and throat electric or pneumatic surgical drill Regulatory Class: Class II Product Code: ERL Dated: December 17, 2021 Received: December 20, 2021
Dear Dolan Mills:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shu-Chen Peng Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212650
#### Device Name
Celeris Disposable Sinus Debrider Model numbers: DSDPP100, DSD4000BA, DSD4000SA, DSD2000BA, DSD40MMSC
#### Indications for Use (Describe)
The Celeris System (reusable power pack and Disposable Sinus Debrider) is intended for cutting, debriding, and removal of thin bone and soft tissue in general ENT and Sinus/Rhinology procedures and applications would include:
· FESS (Functional Endoscopic Sinus Surgery) – Including Endoscopic approaches for: Polypectomy, Sphenoidectomy, Maxillary Antrostomy, Uncinectomy, Frontal Sinusotomy
· Turbinate Reduction / Turbinoplasty, including sub-mucosal resection.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K212650 510(k) Summary
# General Information
| Manufacturer: | Gyrus ACMI, Inc., an Olympus company<br>9600 Louisiana Ave. North<br>Brooklyn Park, MN 55445 USA<br>Phone: 1-763-416-3000 |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 3011050570 |
| 510(k) Submitter: | Gyrus ACMI, Inc.<br>800 West Park Dr. |
| Establishment Registration Number: | Westborough, MA 01581<br>3003790304 |
| Contact Person: | Dolan Mills<br>Program Manager, Regulatory Affairs |
| Date Prepared: | Jan 14, 2022 |
| Device Description | |
| Classification Name: | ENT Electric or Pneumatic Surgical Drill |
| Regulatory Class | Class 2 |
R Regulation Number Review Panel Product Code
Project Name:
Trade Name(s):
Generic/Common Name: Model Numbers:
21 CFR 874.4250 Ear, Nose, & Throat Panel ERL
Gyrus ACMI DSD - Desoto
Celeris, Disposable Sinus Debrider
Electrical Surgical Drill / Shaver DSDPP100, DSD4000SA, DSD4000BA, DSD2000BA, DSD40MMSC
{4}------------------------------------------------
## Predicate Device
Gyrus ACMI Inc. Diego® Elite:
K123429
### Device Description and Technological Characteristics
The Gyrus ACMI Disposable Sinus Debrider (DSD) Celeris System is similar in form and function to the Gyrus ACMI blades and handpieces found within the predicate Diego Elite System cleared under K123429, and its predicate PK Diego cleared under K034004.
The Celeris DSD handpiece is a single-use disposable debrider handle and blade permanently combined with a built-in motor that plugs into a reusable power pack that contains a power supply for motor control. The power pack is also part of the subject system. The power supply is able to auto-adapt to any voltage via the universal power cord connection. Motor control requires software on a controller board inside the power pack.
The handpiece incorporates a rotatable blade and attached RF cable for bipolar models. The blades are either standard, bipolar, or malleable. The bipolar function requires a separate electrosurgical generator (not part of this system), connected via a cable. The cutting performance is equivalent to the current Gyrus ACMI predicate models.
A trigger style button on the DSD handle, or analog footswitch (sold separately) connected to the power pack, activates the blade oscillation and a standard electrosurgical unit footswitch (sold separately) powers the binolar effect for bipolar blades. The electrosurgical generator and footswitch alone will control the RF energy delivery.
A nosecone on the DSD handpiece allows the blade cutting window to rotate left or right. The blades are offered in 2mm and 4mm variants, either straight or malleable, and standard or bipolar. The malleable blade angle is flexible, and the design allows the blade to be bent with the provided bending fixture up to 60° at the user's discretion.
A standard suction tube (sold separately) is attached to the DSD handpiece proximal suction port and a clip attaches the tubing to the power cable. The handpiece power cable plugs directly into the power pack. The non-sterile power pack plugs into a standard power outlet and only provides power to the connected DSD handpiece. When the handpiece is activated power is sent to the motor which oscillates a gear which in turn oscillates the inner blade. For bipolar models, the energy lead from the cable is connected directly to the blade and energy is provided by the separate electrosurgical generator footswitch. For bipolar models, insulation around the cutting window limits energy delivery to the surgical site.
{5}------------------------------------------------
### Intended Use / Indications
The Celeris System (reusable power pack and Disposable Sinus Debrider) is intended for cutting, coagulation, debriding, and removal of thin bone and soft tissue in general ENT and Sinus/Rhinology procedures. Specific procedures and applications would include:
- FESS (Functional Endoscopic Sinus Surgery) - Including Endoscopic approaches for: Polypectomy, Ethmoidectomy, Sphenoidectomy, Maxillary Antrostomy, Uncinectomy, Frontal Sinusotomy
- . Turbinate Reduction / Turbinoplasty, including sub-mucosal resection.
### Compliance to Standards
The following standards were used during the design and testing of the DSD:
IEC 60601-1: 2005 IEC 60601-1-2: 2014 IEC 60601-2-2: 2017 ISO 10993-1: 2018 ISO 10993-5: 2009 ISO 10993-7: 2008 ISO 10993-10: 2010 ISO 11135: 2014 ISO 11607-1: 2019 ISO 11737-2: 2019 ISO 15223-1: 2016 IEC 62304: 2015 ASTM F88/F88M - 2016 ASTM F1886: 2016 ASTM D4169-14: 2016
Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971. Design verification sample sizes and tests were identified and performed as a result of risk analysis assessment.
### Summary of Performance Testing
The following non-clinical, preclinical tests and usability studies were conducted:
{6}------------------------------------------------
# Non-Clinical / Preclinical Performance
Evidence of safety and effectiveness was obtained from two primary areas:
- 1) non-clinical (electrical, mechanical, functional, stability) performance testing
- 2) preclinical (bench tissue, simulated use) evaluations and testing
The bipolar device thermal performance has been tested per FDA guidance document, "FDA Guidance for Industry and FDA Staff, Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery', Mar 2020.
The power pack has been tested per FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", issued on May 11, 2005.
Electrical safety and EMC compatibility: Basic safety and performance testing was performed in accordance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2.
Mechanical and Functional: Verification and comparison bench studies were conducted to evaluate the mechanical and functional performance as compared to the predicate.
Stability: Representative samples were subjected to accelerated aging to confirm that the device maintains functionality and continues to meet specifications over time. The results of the accelerated age testing demonstrate that the device will be stable for the stated shelf-life. In addition, real time age testing will confirm the results of the accelerated age testing. Representative samples were also subjected to distribution testing.
Preclinical: Evidence obtained from preclinical bench tissue (ex vivo) and simulated use studies demonstrate that the DSD performs substantially equivalent to the predicate devices in relevant aspects associated with usability, cutting, coagulation, and removal of tissue.
Bench tissue - evaluated ex vivo using bovine tissue:
- Thermal margin -
- -Thermal impact
- Visual comparison of coagulation -
- Microscopic measurement of thermal margin and impact -
Software: Device software development and validation were completed per the level of concern and guidance.
Biocompatibility testing: The disposable handpieces are classified in accordance with ISO 10993-1, as an External Communication Device, Tissue/Bone/Dentin, for limited exposure (<24 hours.). ISO 10993-1 and FDA guidelines recommend that these
{7}------------------------------------------------
devices have supporting data for cytotoxicity, sensitization, and acute systemic toxicity. Full GLP biocompatibility testing to ISO 10993-1 is on file for these devices.
No animal or clinical testing was conducted. The use of the device type has been documented in published literature and indicates safe and effective use for the target procedures and expected patient populations.
Performance testing demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate devices.
# Material
The composition of the DSD is similar to the predicate devices. The handle and power pack is not patient-contacting. Patient contacting components are limited to the shaver blade portion up to the nosecone.
Patient contacting materials are primarily plastics and stainless steel.
# Sterilization and Shelf Life Discussion
The disposable handpieces are provided sterile, single-use. The devices are sterilized with ETO, using a validated cycle to provide a sterility assurance level of 10-9.
The Shelf Life period for the device was determined via testing and through an analysis of the shelf-life stability of the materials used in the design of the device, as well as an analysis of the packaging materials and processes used with other Gyrus ACMI devices. Shelf-life studies are on file to support the labeled shelf life.
The reusable power pack is provided non-sterile.
# Substantial Equivalence
The DSD bipolar models deliver high frequency electrosurgical energy to coagulate tissue in a similar manner as the predicate RF devices.
The DSD cuts and removes tissue in the same manner as the predicate device. The DSD inner blade oscillates at a high rate of speed and as tissue is pulled into the cutting window via suction it is resected and removed. The oscillating portion of the DSD blade is driven by an electric motor inside the handle portion of the device that incorporates a gear transmission to provide the desired range of speeds and torque. The speed is limited by the motor within the handle. The hollow inner blade is attached to a suction supply to facilitate removal of tissue and fluid from the surgical site.
{8}------------------------------------------------
The power pack provides power to the DSD handpiece motor for blade rotation only. The power pack provides power in a similar manner as the predicate console, but focuses on power delivery only.
| Predicate comparison table to outline differences and similarities between the | | |
|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| subject and predicate devices | | |
| Description | Subject Device DSD | Predicate Device (K123429) |
| Intended Use | Cutting, coagulation, debriding, and removal of thin bone and soft tissue in general ENT and Sinus/Rhinology procedures. | Cutting, coagulation, drilling, debriding, and removal of bone and soft and hard tissue in general ENT and Sinus/Rhinology, Nasopharyngeal / Laryngology, and Head & Neck procedures. |
| Prescription/over-the-counter use | Rx Only | Rx Only |
| Size(s) | OD: 2mm, 4mm<br>Working Length:<br>3.9 to 4.3 in. | OD of similar blades: 2 to 4 mm<br>Working Length of similar blades:<br>4.3 to 5 in |
| Method of dissection | Conventional rotating cutting blade combined with RF energy hemostasis for specific models | Conventional rotating cutting blade combined with RF energy hemostasis for specific models |
# Conclusion
The performance of the DSD was compared against performance requirements and the predicate systems listed above. Performance requirements were based on the predicate systems. Testing demonstrated that the performance requirements were met, and that the subject device exhibited comparable performance characteristics to the predicate. Any differences have been validated and demonstrate that the technological features do not raise new questions of safety and efficacy.
In summary, the Gyrus ACMI Cleris DSD is substantially equivalent to the predicate devices and does not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.