K212625 · Shanghai Microport EP Medtech Co., Ltd. · DRC · Mar 23, 2022 · Cardiovascular
Device Facts
Record ID
K212625
Device Name
PathBuilder Transseptal Needle
Applicant
Shanghai Microport EP Medtech Co., Ltd.
Product Code
DRC · Cardiovascular
Decision Date
Mar 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1390
Device Class
Class 2
Indications for Use
The PathBuilder™ Transseptal Needle is used to puncture the interatrial septum during a transseptal catheterization procedure to gain left heart access.
Device Story
The PathBuilder™ Transseptal Needle is a manual surgical instrument used by clinicians during transseptal catheterization. It consists of a needle with a beveled distal tip, a curved section for positioning, a main body, a pointer flange for orientation, a hub for aspiration or fluid injection, and a straight stylet. The device is inserted into a catheter introducer/dilator to facilitate puncture of the interatrial septum. By providing access to the left heart, it enables subsequent diagnostic or therapeutic procedures. The device is single-use and sterile.
Clinical Evidence
No clinical data was used to support this submission. Substantial equivalence is based on bench testing, including biocompatibility (ISO 10993-1), mechanical performance (puncture force, joint forces, dimensional verification), and sterilization validation (ISO 11135).
Technological Characteristics
Manual transseptal needle; components include hub, pointer flange, curved needle, and stylet. Materials comply with ISO 10993-1 biocompatibility standards. Functional performance complies with ISO 11070:2014. Sterilization via Ethylene Oxide (ISO 11135). Single-use device.
Indications for Use
Indicated for patients undergoing transseptal catheterization procedures requiring puncture of the interatrial septum to gain left heart access.
Regulatory Classification
Identification
A trocar is a sharp-pointed instrument used with a cannula for piercing a vessel or chamber to facilitate insertion of the cannula.
Special Controls
*Classification.* Class II (special controls). Except for trocars that are reprocessed for multiple use, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.
March 23, 2022
Shanghai MicroPort EP MedTech Co., Ltd. Tian Xia RA Engineer Building 23&28, Lane 588, Tianxiong Rd. Shanghai China
Re: K212625
Trade/Device Name: PathBuilder™ Transseptal Needle Regulation Number: 21 CFR 870.1390 Regulation Name: Trocar Regulatory Class: Class II Product Code: DRC Dated: February 15, 2022 Received: February 22, 2022
Dear Tian Xia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212625
Device Name PathBuilder Transseptal Needle
Indications for Use (Describe)
The PathBuilder Transseptal Needle is used to puncture the interatrial septum during a transseptal catheterization procedure to gain left heart access.
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the EverPace logo. The logo consists of a stylized "E" and "O" connected together, with the "E" in blue and the "O" in red. To the right of the symbol is the word "EverPace" in blue font, with a trademark symbol in the upper right corner.
## 510(k) # K212625 510(k) Summary
### Submitter Information
- Shanqhai MicroPort EP MedTech Co., Ltd. A. Company Name: B. Company Address: Building 23&28, Lane 588, Tianxiong Rd., 201318 Shanghai, PEOPLE'S REPUBLIC OF CHINA C. Company Phone: +86 21 38954600*3613 D. Company Facsimile: +86 21 20903925 E. Email: txia@everpace.com F. Contact Person: Xia, Tian G. Summary Prepared on: February 15, 2022
### ldentification of Proposed Device
- PathBuilder™ Transseptal Needle A. Trade Name:
- B. Common Name: Transseptal Needle
- C. Classification Name: Trocar
- 21 CFR 870.1390 D. Regulation Number
- E. Product Code: DRC
- F. Device Class: Class II
- G. Review Panel Cardiovascular
- H. 510(k) Number K212625
## Device Description
PathBuilder™ Transseptal Needle is consisted of thin section, curved section, main body, pointer flange, hub and stylet. The distal end of PathBuilder™ Transseptal Needle is curved for easy positioning within the heart when used together with the catheter introducer. The outer diameter of needle stepped down within the curved section for better fitness to the dilator which fixes the maximum length of the needle into the dilator. The distal tip of the needle is beveled to facilitate the transseptal process. The proximal end of PathBuilder™ Transseptal Needle is with a pointer flange which indicates the curve orientation of the needle. And a hub is equipped for aspiration, fluid injection/infusion etc. The stylet is straight. The proximal end of the stylet is curved to lock on the hub when inserted into the needle lumen. The stylet is used to help insert the needle into the dilator.
### Intended Use Statement
The PathBuilder™ Transseptal Needle is used to puncture the interatrial septum during a transseptal catheterization procedure to gain left heart access.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the EverPace logo. The logo consists of a stylized "E" and "O" connected together, with the "E" in blue and the "O" in red. To the right of the symbol is the word "EverPace" in blue, with a small "TM" symbol in the upper right corner.
### ldentification of Predicate Device
- A. Product Name: BRK™ Transseptal Needle
- B. Manufacturer: St. Jude Medical
- C. 510(k) Number: K122587
- D. Regulation Number 21 CFR 870.1390
- E. Product Code: DRC
- F. Device Class: Class II
### Non-Clinical Performance Testing
Non-clinical performance testing was completed for the PathBuilder™ Transseptal Needle to support its substantial equivalence to the predicate device. The test results demonstrated that the proposed device complies with the following standards and guidance:
- (1) Biocompatibility Verification: The biological safety of the transseptal needle was verified as per the requirements of ISO 10993-1:2018 Biological evaluation of medical devices- Part 1:Evaluation and testing within a risk management process and FDA's modified ISO guidelines in accordance with the FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
- Cytocompatiblity a)
- b) Sensitization
- Irritation C)
- Acute Systemic Toxicity ರು)
- Hemocompatibility e)
- f) Complement Activation
- Thrombogenicity ರು)
- (2) Bench Validation: Validation testing of the transseptal needle was performed to validate the design of the device. This testing included visual inspection, dimensional verification, packaging testing (visual inspection, sealing strength, dye penetration) and mechanical failure/performance (Joint forces between each parts). Additionally, the compatibly of the needle and dilator was validated with compatibility, puncture force, radio detectability, luer fitting. All samples were sterilized and aged prior to testing. All tested samples passed bench testing.
- (3) Cleaning and Sterilization Validation: Validation testing was performed to demonstrate that the transseptal needle could be sterilized in accordance with ISO 11135:2014 Sterilization of health care products-
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the EverPace logo. The logo consists of a stylized "E" and "O" connected together, followed by the word "EverPace" in a sans-serif font. The "E" and part of the "O" are in blue, while the rest of the "O" is in red. There is a trademark symbol next to the word EverPace.
Ethylene Oxide: Requirements for development, validation and routine control of a sterilization process for medical device, AAMI TIR 28:2009 Product Adoption and Process Equivalency for Ethylene Oxide sterilization and the FDA guidance document.
- (4) Shelf Life Validation: Validation testing was performed to demonstrate the shelf life of the transseptal needle is three years.
### Clinical Tests Conclusion
No clinical study was used to support this submission.
| Description | Proposed Device | Predicate Device (K122587) | Remark |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Product Code | DRC | DRC | SE |
| Regulation<br>No. | 21 CFR 870.1390 | 21 CFR 870.1390 | SE |
| Class | Class II | Class II | SE |
| Intended Use | The PathBuilderTM Transseptal<br>Needle is used to puncture the<br>interatrial septum during a<br>transseptal catheterization<br>procedure to gain left heart<br>access. | The BRKTM Transseptal Needle<br>is used to puncture the<br>interatrial septum during a<br>transseptal catheterization<br>procedure to gain left heart<br>access. | SE |
| Indications<br>for Use | The PathBuilderTM Transseptal<br>Needle is used to puncture the<br>interatrial septum during a<br>transseptal catheterization<br>procedure to gain left heart<br>access. | The BRKTM Transseptal Needle<br>is used to puncture the<br>interatrial septum during a<br>transseptal catheterization<br>procedure to gain left heart<br>access. | SE |
| | Hub | Hub | SE |
| Configuration | Pointer flange | Pointer flange | SE |
| | Needle tip | Needle tip | SE |
| Dimensional<br>Comparison | Effective length | Effective length | SE |
| | Diameter | Diameter | SE |
| | Effective length of stylet | Effective length of stylet | SE |
| | Outer diameter of stylet | Outer diameter of stylet | SE |
| | Curve shape | Curve shape | SE |
| Functional<br>performance | Comply with ISO 11070: 2014 | Comply with ISO 11070: 2014 | SE |
| Biological<br>characteristics | Comply with ISO 10993-1 | Comply with ISO 10993-1 | Similar |
| Intended | Professional healthcare | Professional healthcare | SE |
### Comparison to Predicate Device、
Shanghai MicroPort EP MedTech Co., Ltd.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for EverPace. The logo consists of a stylized "E" and "O" in blue and red, respectively, followed by the word "EverPace" in blue. The "TM" symbol is located to the right of the word "EverPace".
| healthcare<br>environment | | environment | |
|---------------------------|---------------|---------------|----|
| Sterilization<br>method | EO Sterilized | EO Sterilized | SE |
| Single use | Yes | Yes | SE |
| Shelf life | 3 years | 3 years | SE |
### Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
The proposed and predicate devices share the same intended use and fundamental scientific technology, including principles of operation and mechanism of action. Design and technological differences between the proposed and predicate devices do not raise any new concerns of safety and effectiveness. The results of verification and validation testing demonstrate that the PathBuilder™ Transseptal Needle is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.