Powder Free Nitrile Examination Gloves, Non-sterile,Tested For Use with Chemotherapy Drugs and Opioid Fentanyl Citrate (Blue)
K212613 · Koon Seng Sdn Bhd · LZA · Dec 21, 2021 · General Hospital
Device Facts
Record ID
K212613
Device Name
Powder Free Nitrile Examination Gloves, Non-sterile,Tested For Use with Chemotherapy Drugs and Opioid Fentanyl Citrate (Blue)
Applicant
Koon Seng Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Dec 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A Nitrile powder free examination glove is a disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between examiner and patient. This specialty glove has also been tested for use with Chemotherapy drugs and the Opioid Fentanyl citrate.
Device Story
Nitrile synthetic, blue, non-sterile, disposable, single-use examination glove. Used by healthcare professionals to prevent cross-contamination between examiner and patient. Tested for permeation resistance against chemotherapy drugs and Fentanyl citrate per ASTM D6978-05 (2019). Provides barrier protection; users must consult drug labeling/SDS to determine appropriate protection levels. Warning: Do not use with Carmustine or Thiotepa due to low breakthrough times.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM D6978-05 (2019) for chemotherapy/opioid permeation, ASTM D412/D573 for physical properties, ASTM D5151 for water-leak (AQL 1.5), and ASTM D6124 for residual powder. Biocompatibility confirmed via ISO 10993-10 (irritation/sensitization) and ISO 10993-11 (systemic toxicity).
Technological Characteristics
Nitrile synthetic material; blue color; non-sterile; powder-free. Dimensions and physical properties meet ASTM D6319-10(2019). Permeation resistance tested per ASTM D6978. Biocompatibility per ISO 10993-10 and ISO 10993-11. Standalone device.
Indications for Use
Indicated for use as a disposable medical examination glove worn on the examiner's hand or finger to prevent cross-contamination between examiner and patient. Tested for use with specific chemotherapy drugs and the opioid Fentanyl citrate.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Blue Colored, Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (K192954)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
December 21, 2021
Koon Seng Sdn Bhd % Wava Truscott President Truscott MedSci Associates, LLC. 180 Burkemeade Ct. Roswell, Georgia 30075
Re: K212613
Trade/Device Name: Powder Free Nitrile Examination Gloves, Non-sterile,Tested For Use with Chemotherapy Drugs and Opioid Fentanyl Citrate (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ, QPO Dated: November 15, 2021 Received: November 22, 2021
Dear Wava Truscott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known)
## K212613
#### Device Name
Trade Name: Powder Free Nitrile Examination Gloves, Non-sterile, Tested for Use with Chemotherapy Drugs and the Opiod Fentanyl Citrate (Blue)
Indications for Use (Describe) A Nitrile powder free examination glove is a disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between examiner and patient. This specially glove has also been tested with Chemotherapy drugs and the Opioid Fentanyl citrate.
| | | | Chemotherapy Drugs and the opioid Fentanyl citrate were tested for breakthrough detection times: |
|--|--|--|--------------------------------------------------------------------------------------------------|
|--|--|--|--------------------------------------------------------------------------------------------------|
| TEST<br>CHEMOTHERAPY DRUG | CONCENTRATION<br>TESTED | MINIMUM BREAKTHROUGH<br>DETECTION TIME (MINUTES) |
|----------------------------|-------------------------|--------------------------------------------------|
| Carmustine (BCNU) | 3.3 mg/ml | 13.0 |
| Cisplatin | 1.0 mg/ml | No Breakthrough Up To 240 minutes |
| Cyclophosphamide (Cyloxan) | 20.0 mg/ml | No Breakthrough Up To 240 minutes |
| Cytarabine | 100.0 mg/ml | No Breakthrough Up To 240 minutes |
| Dacarbazine (DTIC) | 10.0 mg/ml | No Breakthrough Up To 240 minutes |
| Doxorubicin Hydrochloride | 2.0 mg/ml | No Breakthrough Up To 240 minutes |
| Etoposide (Toposar) | 20.0 mg/ml | No Breakthrough Up To 240 minutes |
| Fluorouracil | 50.0 mg/ml | No Breakthrough Up To 240 minutes |
| Ifosfamide | 50.0 mg/ml | No Breakthrough Up To 240 minutes |
| Methotrexate | 25.0 mg/ml | No Breakthrough Up To 240 minutes |
| Mitomycin C | 0.5 mg/ml | No Breakthrough Up To 240 minutes |
| Mitoxantrone | 2.0 mg/ml | No Breakthrough Up To 240 minutes |
| Pacitaxel (Taxol) | 6.0 mg/ml | No Breakthrough Up To 240 minutes |
| Thiotepa | 10.0 mg/ml | 37.9 |
| Vincristine Sulfate | 1.0 mg/ml | No Breakthrough Up To 240 minutes |
Important: Carmustine and Thiolepa have extremely low minimal breakthrough times of 13.0 minutes and 37.9 minutes respectively. Warning: Do not use with Carmusline or Thiotepa.
| TEST OPIOID DRUG | CONCENTRATION<br>TESTED | MINIMUM BREAKTHROUGH<br>DETECTION TIME (MINUTES) |
|-------------------------------|-------------------------|--------------------------------------------------|
| Fentanyl Citrate (Injectable) | 100.0 mg/2ml | No Breakthrough up to 240 minutes |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo featuring two golden hands overlapping each other, set against a black background. The logo is encircled by a golden ring, and the text "Delivering Quality Excellence in Gloves" is written around the ring. The text is in a smaller font size and follows the curve of the circle.
QUALITY, TRANSPARENCY, PARTNER
#### OON SENG SDN. BHD = (107051-H)
(A member of the KS Group) Corporate Sales Office: 1321, 13th Floor, BlockA, Lift Lobby &, Damancara Intan No.1,
Jin. SS 20/27, 47400 Petaling Jaya, Selangor. Factory 1 : PTD 3274, Batu 41/2, Jalan Bakri, 84000 Muar, Johor, Factory 2 : PTD 16058, Kaw. Perindustrian Tangkak, 84900 Tangkak, Johor.
Email : koonseng@koonseng.com Website : www.koonseng.com Tel : 603- 7733 1388
Tel : 606-9868446 , 606-986 9257 Fax: 606-986 6908 Tel: 606-979 1778 , 606-978 2780 Fax: 606-979 1779
Image /page/3/Picture/6 description: The image contains several certification logos. There is an SMG logo, a CE logo, an ISO 22000 registered logo, and an ISO 13485 registered logo. These logos likely indicate that the product or company associated with the image meets certain standards for quality, safety, or environmental management.
# K212613: 510(k) Summary
| 1.0 | Submitter | | | |
|-----|------------------------------------------------------|---|--------------------------------------------------------------------------------------------------------------------------------------|--|
| | Name | : | Koon Seng, Sdn. Bhd. | |
| | Address | : | PTD 16058, Kawasan Perindustrian Tangkak,<br>84900 Tangkak, Johor, Malaysia | |
| | Contact | : | Yeo Pai Shuang | |
| | Phone No. | : | +603-7733 1388 | |
| | Fax No. | : | +603-7733 6499 | |
| | Date of Preparation | : | June 29, 2021 | |
| 2.0 | Name of the Device | | | |
| | Common Name | : | Nitrile Powder Free Examination Glove | |
| | Trade Name | : | Powder Free Nitrile Examination Gloves, Non-sterile,Tested for Use<br>with Chemotherapy Drugs and the Opioid Fentanyl Citrate (Blue) | |
| | Classification Name | : | Patient Examination Glove Specialty (21CFR 880.6250) | |
| | Product Codes | : | LZA, LZC, OPJ, QDO | |
| | Device Class | : | Class I | |
| 3.0 | Predicate: Identification of Legally Marketed Device | | | |
| | Predicate 510(k) No. | : | K192954 | |
| | Trade Name | : | Blue Colored, Powder Free Nitrile Examination Gloves Tested for Use with<br>Chemotherapy Drugs and Fentanyl Citrate | |
| | Owner | : | Comfort Rubber Gloves Industries Sdn. Bhd. | |
| | Product Codes | : | LZA, LZC, QDO | |
| | Device Class | : | Class I | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Koon Seng. Above the name is the text "Quality - Transparency - Partnership". Below the name is the text "Since 1987".
#### 4.0 Description of Subject Device
The K212613 gloves are produced as nitrile synthetic gloves. They are manufactured without natural rubber latex. The nitrile gloves are blue in color, disposable, single-use only, non-sterile, medical examination gloves. K212613 meets all the requirements of ASTM D6978-10(2019) Standard Specifications for Examination Gloves for Medical Applications. They have been tested for use with chemotherapy drugs and the opioid Fentanyl citrate, both of which were evaluated using ASTM D6378-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
#### 5.0 Indications for Use
A Nitrile powder free examination glove is a disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between examiner and patient. This specialty glove has also been tested for use with Chemotherapy drugs and the Opioid Fentanyl citrate.
| Test Chemotherapy Drug | Concentration | Minimum Breakthrough Detection Time<br>(minutes) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|--------------------------------------------------|
| Carmustine (BCNU) | 3.3 mg/mL | 13 min. |
| Cisplatin | 1.0 mg/mL | >240 min. |
| Cyclophosphamide (Cytoxan) | 20.0 mg/mL | >240 min. |
| Cytarabine | 100 mg/mL | >240 min. |
| Dacarbazine | 10.0 mg/mL | >240 min. |
| Doxorubicin | 2.0 mg/mL | >240 min. |
| Etoposide | 20.0 mg/mL | >240 min. |
| Fluorouracil | 50.0 mg/mL | >240 min. |
| Ifosfamide | 50.0 mg/mL | >240 min. |
| Methotrexate | 25.0 mg/mL | >240 min. |
| Mitomycin C | 0.5 mg/mL | >240 min. |
| Mitoxantrone | 2.0 mg/mL | >240 min. |
| Paclitaxel | 6.0 mg/mL | >240 min. |
| ThioTepa | 10.0 mg/mL | 37.9 min |
| Vincristine Sulfate | 1.0 mg/mL | >240 min. |
| Important: Carmustine and ThioTepa have extremely low minimum breakthrough times of<br>13.0 minutes and 37.9 minutes respectively. | | |
| Warning: Do not use with Carmustine or ThioTepa. | | |
| Test Opioid Drug | Concentration | Minimum Breakthrough Detection Time<br>(minutes) |
| Fentanyl Citrate Injection | 100mcg/2mL | >240 min. |
| Please Note: Glove used for protection against possible Chemotherapy Drug exposure<br>should be selected specifically for the type of drugs used. Users should review drug labeling<br>or material safety data sheets for the drugs being used to determine an adequate level of<br>protection. | | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Koon Seng. The logo includes the words "Quality • Transparency • Partnership" at the top. Below that is the name "KOON SENG" in large, bold letters. Finally, the words "Since 1987" are at the bottom of the logo.
# Technological Characteristic Comparison:
Shown below is a comparison of The Powder Free Nitrile Examination Gloves, Non-sterile, Tested for Use with Chemotherapy Drugs and Opioid Fentanyl Citrate (Blue) and Blue Colored, Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.
## Glove Comparison Table":
| Subject | Standards | Koon Seng Subject<br>Device<br>510(k): K212613 | Predicate Device:<br>510(k): K192954 | Comparison K212613<br>Subject glove to<br>Predicate K192954 |
|-----------------------------------------------------|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturers | NA | Koon Seng Sdn. Bhd. | Comfort Rubber Gloves<br>Industries Sdn. Bhd. | Different |
| Trade Name | NA | Powder Free Nitrile<br>Examination Gloves, Non-<br>Sterile, Tested for Use with<br>Chemotherapy Drugs and the<br>Opiod F<br>entanyl Citrate (Blue) | Blue Colored, Powder Free<br>Nitrile Examination Glove<br>Tested for Use with<br>Chemotherapy Drugs and<br>Fentanyl Citrate | Different |
| Common<br>Name | ASTM D6319 -<br>10(2019) | Nitrile Powder free Patient<br>Examination Glove | Nitrile Patient<br>Examination Glove | Similar |
| Device Class | 21 CFR<br>880.6250 | Class 1 | Class 1 | Same |
| Base Material | NA | Nitrile | Nitrile | Same |
| Color | NA | Blue | Blue | Same |
| Product Codes | 21 CFR<br>880.6250 | LZA, LZC, OPJ, QDO | LZA, LZC, QDO | Similar OPJ is new: denotes<br>tested with Chemotherapy<br>drugs to further define the<br>LZC specialty designation |
| Sizes offered | ASTM D3767 | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large | Same |
| Dimensions: | ASTM D6319 -<br>10(2019) | Meets | Meets | Same |
| Sterile v.<br>Non-Sterile | NA | Non-Sterile | Non-Sterile | Same |
| Prescription or<br>OTC | NA | OTC | OTC | Same |
| Single Use-<br>Disposable | NA | Single use-Disposable | Single use-Disposable | Same |
| Thickness:<br>Finger & palm | ASTM D6319 -<br>10(2019) | Meets | Meets | Same |
| Before aging<br>Physical<br>Properties | ASTM D412<br>ASTM D6319 -<br>10(2019) | Meets | Meets | Same |
| After aging<br>Physical<br>Properties | ASTM D412<br>ASTM D6319 | Meets | Meets | Same |
| Freedom from<br>Holes | ASTM D5151<br>ASTM D6319 | Meets AQL 1.5 | Meets AQL 1.5 | Same |
| Subject | Standards | Koon Seng Subject<br>Device<br>510(k): K212613 | Predicate Device:<br>510(k): K192954 | Comparison K212613<br>Subject glove to<br>Predicate K192954 |
| Powder-Free | ASTM D6124<br>< 2mg/glove | Meets | Meets | Same |
| Intended Use<br>(part of<br>Indications for<br>Use) | | A Nitrile powder free<br>examination glove is a<br>disposable device intended<br>for medical purposes, worn<br>on the examiner's hand or<br>finger to prevent<br>contamination between<br>examiner and patient. This<br>specialty glove has also been<br>tested for use with<br>Chemotherapy drugs and the<br>Opioid Fentanyl citrate. | A patient examination<br>glove is a disposable<br>device intended for<br>medical purpose that is<br>worn on the examiner's<br>hand to prevent<br>contamination between<br>patient and examiner. | Similar |
| Indications for<br>Use<br>Chemotherapy | ASTM D6978 | Chemotherapy drugs:<br>All >240min Breakthrough<br>Times except Carmustine &<br>ThioTEPA<br>• Cisplatin, 1mg/ml<br>• Cyclophosphamide<br>(Cytoxan), 20mg/ml<br>• Cytarabine, 100mg/ml<br>• Dacarbazine, 10mg/ml<br>• Doxorubicin HCL, 2mg/ml<br>• Etoposide, 20mg/ml<br>• Fluorouracil, 50mg/ml<br>• Ifosfamide, 50mg/ml<br>• Methotrexate, 25mg/ml<br>• Mitomycin C, 0.5mg/ml<br>• Mitoxantrone, 2mg/ml<br>• Paclitaxel, 6mg/ml<br>• Vincristine Sulfate,mg/ml<br><br>• Carmustine (BCNU), 3.3<br>mg/ml:<br>Breakthrough: 13 min.<br><br>• ThioTepa, 10mg/ml<br>Breakthrough: 37.9 min. | Chemotherapy drugs:All<br>>240min Breakthrough<br>Times except Carmustine &<br>ThioTEPA<br>•Cisplatin 1mg/ml<br>• Cyclophosphamide<br>(Cytoxan) 20mg/ml<br>• Dacarbazine (DTIC) 10<br>mg/ml<br>• Doxorubicin HCL, 2mg/ml<br>• Hydrochloride 2mg/ml<br>• Etoposide (Toposar)<br>20mg/ml<br>• FluorouracilS0mg/ml<br>• Paclitaxel (Taxol)<br>6mg/ml<br><br>• Carmustine (BCNU) 3.3<br>mg/ml:<br>Breakthrough: 18.2 min.<br><br>• ThioTepa (THT) 10mg/ml:<br>Breakthrough: 57.3 min. | Similar |
| Subject | Standards | Koon Seng Subject<br>Device<br>510(k): K212613 | Predicate Device:<br>510(k): K192954 | Comparison K212613<br>Subject glove to<br>Predicate K192954 |
| Cautions &<br>Warnings | | Important: Carmustine and<br>ThioTEPA have extremely low<br>minimal breakthrough times of<br>13.0 minutes and<br>37.9 minutes respectively.<br><b>Warning:</b> Do not use with<br>Carmustine or ThioTepa | Please note that the<br>following drugs have<br>extremely low permeation<br>times: Carmustine (BCNU):<br>18.2minutes and Thiotepa:<br>57.3 minutes. <b>Warning:</b><br>Do not use with<br>Carmustine. | Similar: except Subject<br>device added ThioTepa to<br>the " <b>Warning:</b> Do not use<br>with..." listing. Technically,<br>it is not required to do so<br>as the cut-off to require<br>the warning is below 30<br>minutes. |
| Indications for<br>Use: Fentanyl | ASTM D6978 | Fentanyl Citrate (Injectable)<br>100mg/2ml:Minimum<br>breakthrough time: >240 | Fentanyl Citrate 100mg/2ml:<br>Minimum breakthrough time:<br>>240 | Similar: Fentanyl Citrate<br>concentration & diluent<br>same. US DEA required<br>testing street injectable<br>term by test lab in test<br>results |
| Biocompatibility | ISO 10993-<br>11: Tests<br>for Systemic<br>Toxicity | Passes<br>Under the conditions of The<br>Systemic Toxicity test the<br>predicate glove showed no<br>adverse biological reaction. | (Used Accepted Alternative:<br>ISO 10993-5 (Cytotoxicity) | Similar: Used Different<br>alternative<br>biocompatibility<br>test method: Both<br>acceptable |
| | ISO 10993-5<br>Cytotoxicity | (Used Accepted Alternative:<br>ISO 10993-11 Acute Systemic<br>Toxicity) | Passed at: Exhibits severe<br>cytotoxicity reactivity at<br>100%, and 66% extract<br>concentrations and no<br>cytotoxicity reactivity at<br>44%, 30%, 20% and 15%<br>extract concentrations<br>under the condition of this<br>test. | Similar: Used Different<br>Accepted Alternative<br>Biocompatibility test |
| | ISO 10993-10<br>Skin<br>Irritation | Passes<br>Under conditions of this<br>study, subject glove is non-<br>irritating | Passes<br>Under conditions of the<br>study, the device is non-<br>irritating | Same |
| | ISO 10993-10<br>Dermal<br>Sensitization | Passes<br>Under conditions of this<br>study, subject glove is a<br>non-sensitizer | Passes<br>Under conditions of the<br>study, the device is non-<br>sensitization | Same |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
## Summary of Non-Clinical Performance Tests conducted
| Test Method | Purpose | Acceptance Criteria | Results |
|--------------------------------------|-------------------------------------|------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| ASTM D6978-2005<br>(2019) | Permeation by<br>Chemotherapy Drugs | Minimum detection time >240<br>minute | All Chemotherapy Drugs & the<br>Opioid Passed >240min. except<br>Carmustine & ThioTepa |
| ASTM D412-2016<br>ASTM D573-04(2019) | Physical properties | Tensile strength<br>(min=14Mpa); Elongation (min<br>400%) | Pass |
| ASTM D5151-19 | Water-leak test | AQL 1.5 (ISO 2859-1) | Pass |
| ASTM D6124-06(2017) | Residual powder test | <2mg/glove | Pass |
| ISO 10993-10:2010<br>(2016) | Dermal Irritation | Under conditions of this study,<br>the subject device is non-<br>irritating | Pass |
| ISO 10993-10:2010<br>(2016) | Skin Sensitization | Under conditions of this study,<br>the subject device is non-<br>sensitizing | Pass |
#### Non-clinical performance testing method full Titles:
- ASTM D412 Test Methods for Vulcanized Rubber and Thermoplastic Elastomers— Tension O
- O ASTM D573 Test Method for Rubber—Deterioration in an Air Oven x ASTM D3578 Specificationfor Rubber Examination Gloves
- ASTM D6319 Standard Specification for Nitrile Examination Gloves for Medical Application O
- ASTM D5151 Test Method for Detection of Holes in Medical Gloves O
- O ASTM D6124 Test Method for Residual Powder on Medical Gloves
- O ISO 2859 Sampling Procedures and Tables for Inspection by Attributes Test results show that under the conditions of the testing, there is no difference in physical attributes between the proposed device and the predicate device.
- O ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs
Biocompatibility Testing was performed utilizing: ISO 10993 Biological Evaluation of Medical Devices
- ISO 10993 Part 10: Tests for Irritation and Sensitization. Both Skin Irritation and Dermal O Magnuson/Kligman Sensitization performed.
- ISO 10993 Part 11: Tests for assessment of Systemic Toxicity O
#### No human clinical or animal performance testing was performed.
Conclusion: The conclusions drawn from comparing the physical attributes and the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as, or better than, the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.