Disposable Medical Nitrile Examination Gloves (Non-sterile)

K212600 · Suqian Linglian Medical Technology Co., Ltd. · LZA · Nov 18, 2021 · General Hospital

Device Facts

Record IDK212600
Device NameDisposable Medical Nitrile Examination Gloves (Non-sterile)
ApplicantSuqian Linglian Medical Technology Co., Ltd.
Product CodeLZA · General Hospital
Decision DateNov 18, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Disposable Medical Nitrile Examination Gloves (Non Sterile) are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Story

Disposable nitrile examination gloves; blue; powder-free; non-sterile. Worn by clinicians or patients on hands to provide barrier protection against contamination. Single-use. Manufactured from nitrile compound. Performance verified against ASTM D6319-19 standards for physical dimensions, tensile strength, elongation, and freedom from holes. Biocompatibility assessed per ISO 10993 standards. No clinical data required.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM D6319-19 (physical dimensions, tensile strength, elongation, residual powder, freedom from holes) and ISO 10993 (biocompatibility: skin irritation, sensitization, and in vitro cytotoxicity).

Technological Characteristics

Nitrile compound; blue; finger-textured; non-sterile; single-use. Dimensions and physical properties comply with ASTM D6319-19. Biocompatibility per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner. Suitable for over-the-counter use.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. November 18, 2021 Sugian Linglian Medical Technology Co., Ltd % Johnson Liu Consultant CNMED Consultant 31 Archer St Upper MT Gravatt, OLD 4122 Australia Re: K212600 Trade/Device Name: Disposable Medical Nitrile Examination Gloves (Non-sterile) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: August 6, 2021 Received: August 17, 2021 Dear Johnson Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) #### K212600 Device Name Disposable Medical Nitrile Examination Gloves (Non Sterile) Indications for Use (Describe) Disposable Medical Nitrile Examination Gloves (Non Sterile) are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Summary (K212600) #### Disposal Medical Nitrile Examination Gloves (Non Sterile) Preparation Date: Aug 6th, 2021 #### 1. SUBMITTER Company Name: SUQIAN LINGLIAN MEDICAL TECHNOLOGY CO.. LTD Company Address: No 1 Rd Wanpi Industrial Park, Xinshuguan Road, Shuyang County, Suqian Jiangsu, CN 223699 Contact Person: Johnson Liu Telephone Number: +614-0158-9995 Email: Johnson(@cnmed.com.au #### 2. NAME OF THE DEVICE Trade Name / Proprietary Name: Disposable Medical Nitrile Examination Gloves (Non Sterile) Device Name: Disposable Medical Nitrile Gloves Device Classification Name: Patient Examination Gloves Device Class: Class I Device Classification Number: 21 CFR 880.6250 Product Code: LZA ### 3. IDENTIFICATION OF THE LEGALLY MARKETED DEVICE Predicate Device: K203191 Applicant: Nathan Trading Co., Ltd Device Name: LYDUS NITRILE EXAMINATION GLOVES, POWDER FREE Device Classification Name: Patient Examination Gloves Device Classification Number: 21 CFR 880.6250 Device Class: Class I Product Code: LZA Review Panel: General Hospital ### 4. DEVICE DESCRIPTION The subject device in this 510(k) Notification is Disposable Medical Nitrile Examination Gloves (Non sterile). The subject device is a patient examination glove made from nitrile compound, blue color, powder free and non-sterile (Per 21 CFR 880.6250, Class I) for single use only. The device meets the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application. ## 5. INDICATIONS FOR USE OF THE DEVICE Disposable Medical Nitrile Examination Gloves (Non Sterile) are disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. {4}------------------------------------------------ # 6. TECHNOLOGICAL CHARACTERISTIC COMPARISON FOR THE PROPOSED AND PREDICATE DEVICES | | DEVICE PERFORMANCE | | REMARKS | |----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | CHARACTERISTICS | SUBJECT | PREDICATE | | | 510(k) Number | Pending | K203191 | - | | | Disposable Medical Nitrile | LYDUS Nitrile Examination | | | Device Name | Examination Gloves (Non Sterile) | Gloves, Powder Free | - | | Product Code | LZA | LZA | Same | | Indications for Use | Disposable Medical Nitrile<br>Examination Gloves (Non Sterile) are<br>disposable devices intended for<br>medical purpose that are worn on the<br>examiner's hand to prevent<br>contamination between patient and<br>examiner. The device is for over the-<br>counter use. | LYDUS Nitrile Examination<br>Gloves, Powder Free is a<br>disposable device<br>intended for medical purposes<br>that is worn on the examiner's<br>hand or fingers to prevent<br>contamination between patient<br>and examiner. | Same | | Materials of Use<br>(ASTM<br>D6910/D6910M-19) | Nitrile compound | Nitrile compound | Same | | Color | Blue | Blue | Same | | Texture | Finger Textured | Finger Textured | Same | | Size<br>(ASTM D6319-19) | Small, Medium, Large, | Small, Medium, Large,<br>Extra Large | Same | | Sterilization | Non-sterile | Non-sterile | Same | | Usage | Single use | Single use | Same | | Dimensions<br>(ASTM D6319-19) | Length Min. 230 min Width Min<br>95+/-10 mm (for medium size) | Length Min. 230 min Width Min<br>95+/-10 mm (for medium size) | Same | | Physical Properties<br>(ASTM D6319-19) | Before Aging<br>Tensile Strength<br>Min 14 Mpa Ultimate Elongation<br>Min 500%<br>After Aging<br>Tensile Strength<br>Min 14 Mpa Ultimate Elongation<br>Min 400% | Before Aging<br>Tensile Strength<br>Min 14 Mpa Ultimate<br>Elongation Min 500%<br>After Aging<br>Tensile Strength<br>Min 14 Mpa Ultimate<br>Elongation Min 400% | Same | | Thickness<br>(ASTM D6319-19) | Palm min 0.05 mm<br>Finger min 0.05 mm | Palm min 0.05 mm<br>Finger min 0.05 mm | Same | | Powder Free<br>(ASTM D6319-19) | <2 mg/glove | <2 mg/glove | Same | | Freedom from Holes (Water<br>Tight -1000 ml) – ASTM<br>D6319-19 (Cross Reference<br>D5151) AQL 2.5 | Passed | Passed | Same | | Biocompatibility -<br>SKIN SENSITIZATION -<br>ISO 10993-10: 2010 (E) | Under the conditions of study, the<br>test article did not show significant<br>evidence of causing skin<br>sensitization in the guinea pig. | Under the conditions of study<br>not a sensitizer | Same | | Biocompatibility -<br>SKIN IRRITATION - ISO<br>10993-10: 2010 (E) | Under the conditions of<br>the study the sample did not induce<br>skin irritation. | Under the conditions of<br>the study not an irritant | Same | | Biocompatibility - IN<br>VITRO CYTOTOXICITY -<br>ISO<br>10993-5: 2009(E) | Under the conditions of the<br>cytotoxicity study, mild<br>Systemic cytotoxicity (Grade 2)<br>observed. It complied with the<br>criteria in ISO 10993-5:2009. | Exhibit cytotoxic<br>reactivity at 100%<br>extract concentration<br>(Grade 4 with neat extract).<br>Non-cytotoxic<br>reactivity at 50%, 25%,<br>12.5% and 6.25%<br>extract concentration. | Similar | | Manufacturer(s) | SUQIAN LINGLIAN<br>MEDICAL TECHNOLOGY<br>CO., LTD | Nathan Trading Co., Ltd | ------ | {5}------------------------------------------------ There are no significant differences between the two products and are identical in terms of intended use, materials, design, manufacturing methods. Both devices meet the ASTM standard D6319-19. {6}------------------------------------------------ # 7. NON-CLINICAL TESTING SUMMARY ## PERFORMANCE DATA | Test Method | Purpose | Acceptance Criteria | Result | |-----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ASMT D6319-19<br>Standard Specification<br>for Nitrile Examination<br>Gloves for Medical<br>Application - Physical<br>Dimensions Test | To determine the<br>width, length, and<br>thickness of the<br>gloves | Width:<br>80mm ±10mm (for small size)<br>95mm ±10mm (for medium size)<br>110 mm ±10mm (for large size)<br>Length:<br>220 mm (Minimum) (for small size)<br>230 mm (Minimum)(for medium, large<br>size)<br>Thickness:<br>Finger -0.05 mm (Minimum) (for small,<br>medium, large size)<br>Palm -0.05 mm (Minimum)(for small,<br>medium, large size) | Width<br>small size<br>82 to 89mm<br>medium size<br>91 to 96mm<br>large size<br>115 to 116mm<br>Length<br>small size<br>230mm (minimum)<br>medium size<br>240mm (minimum)<br>large size<br>238mm (minimum)<br>Thickness<br>Finger<br>small size<br>0.12mm (minimum)<br>medium size<br>0.13mm (minimum)<br>large size<br>0.12mm (minimum)<br>Palm<br>small size<br>0.08 mm (minimum)<br>medium size<br>0.08mm (minimum)<br>large size<br>0.09 mm (minimum)<br>Passed | | ASMT D6319-19<br>Standard Specification<br>for Nitrile Examination<br>Gloves for Medical<br>Application - Physical<br>Requirements Test | To determine the<br>tensile strength and<br>ultimate elongation<br>before and after<br>acceleration aging | Before Acceleration Aging:<br>Tensile Strength (MPa):<br>14 (Minimum)<br>Ultimate Elongation (%):<br>500 (Minimum)<br>After Acceleration Aging:<br>Tensile Strength (MPa):<br>14 (Minimum)<br>Ultimate Elongation (%):<br>400 (Minimum) | Before Acceleration Aging:<br>Tensile Strength (MPa):<br>20 (Minimum)<br>Ultimate Elongation (%):<br>521 (Minimum)<br>After Acceleration Aging:<br>Tensile Strength (MPa):<br>19 (Minimum)<br>Ultimate Elongation (%):<br>505 (Minimum)<br>Passed | | ASTM D6319-19<br>(ASTM D5151-11)<br>Standard Test Method<br>for Detection of Holes in<br>Medical Gloves | To determine the<br>holes in the gloves | (Mean)<br>AQL 2.5 | Passed | | ASMT D6319-19<br>(ASTM D6124-11)<br>Standard Test Method<br>for Residual Powder on<br>Medical Gloves | To determine the<br>residual powder in<br>the gloves | < 2.0 mg/glove | 0.33 mg/glove, Passed | | Test Method | Purpose | Acceptance Criteria | Result | | ISO 10993-10 Biological<br>evaluation of medical<br>devices — Part 10: Tests<br>for skin irritation and<br>skin sensitization | To determine the<br>potential of the material<br>under test to produce<br>skin irritation in rabbits | Under the condition of<br>study, testing articles<br>are not an irritant | Under the conditions of<br>study, the test article did<br>not show significant<br>evidence of causing skin<br>sensitization in the guinea<br>pig. | | ISO 10993-10 Biological<br>evaluation of medical<br>devices — Part 10: Tests<br>for skin irritation and skin<br>sensitization | To determine the skin<br>sensitization potential of<br>the material both in<br>terms of induction and<br>elicitation in guinea pigs. | Under the conditions of<br>the study, the testing<br>articles are not a<br>sensitizer. | Under the conditions of<br>the study the sample did<br>not induce skin irritation. | | ISO 10993-5 Biological<br>evaluation of medical<br>devices — Part 5: Tests<br>for in vitro cytotoxicity | To evaluate the in vitro<br>cytotoxic potential of the<br>test item (both inner and<br>outer surface) Extracts in<br>L-929 mouse fibroblasts<br>cells using elution<br>method | Under the conditions of<br>study, no more than<br>grade 2 cytotoxic<br>reaction | Mild (Grade 2)<br>cytotoxicity reaction<br>observed. | {7}------------------------------------------------ #### BIO-COMPATIBILITY DATA # 8.CLINICAL TESTING SUMMARY Not applicable - Clinical data is not needed for gloves cleared by the 510(k) process. ## 9. CONCLUSION The conclusions drawn from the non-clinical test demonstrate that the subject device in 510(k) submission, the Disposable Medical Nitrile Examination Gloves (Non Sterile) are as safe, as effective, and performs as well as or better than the legally marketed predicate device K203191- LYDUS Nitrile Examination Gloves, Powder Free.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...