← Product Code [LZA](/productcode/LZA) · K212578

# Nitrile disposable examination gloves (K212578)

_Xingyu Medical Tech Co., Ltd. · LZA · Dec 2, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212578

## Device Facts

- **Applicant:** Xingyu Medical Tech Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Dec 2, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Nitrile disposable examination gloves are non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Nitrile disposable examination gloves serve as a physical barrier to prevent cross-contamination between patients and examiners. The device is a non-sterile, single-use glove manufactured from nitrile rubber. It is designed to be watertight and possess tensile properties that allow it to conform to the hand, facilitating necessary medical procedures. The gloves are intended for use in medical settings by healthcare professionals or other examiners. The device is used by donning it on the hand or finger prior to patient contact. By providing a protective barrier, the gloves reduce the risk of pathogen transmission, thereby protecting both the patient and the examiner during clinical examinations or procedures.

## Clinical Evidence

No clinical data was required or provided. Substantial equivalence is supported by bench testing, including physical dimensions, tensile strength, elongation, water leak (pinhole) testing, residual powder testing, and biocompatibility (skin irritation, sensitization, and acute systemic toxicity). All tests met acceptance criteria per ASTM D6319-19, ASTM D5151-19, ASTM D6124-06, ISO 10993-10, and ISO 10993-11.

## Technological Characteristics

Material: Nitrile rubber. Standards: ASTM D6319-19 (physical properties/dimensions), ASTM D5151-19 (watertightness), ASTM D6124-06 (residual powder), ISO 10993-10:2010 (irritation/sensitization), ISO 10993-11:2017 (acute systemic toxicity). Form factor: Disposable, non-sterile, ambidextrous glove. Shelf life: 3 years.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Xingyu Powder Free Nitrile Patient Examination Gloves, Blue Color ([K171616](/device/K171616.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 2, 2021

Xingyu Medical Tech Co., Ltd. Eva Li Consultant Shanghai Sungo Management Consulting Company Limited Room 1309, Dongfang Building, 1500# Century Ave Shanghai, Shanghai 200122 China

Re: K212578

Trade/Device Name: Nitrile disposable examination gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: October 24, 2021 Received: November 3, 2021

Dear Eva Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

542 of the Act); 21 CFR 1000-1050.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Ligun Zr

For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212578

#### Device Name Nitrile disposable examination gloves

Indications for Use (Describe)

Nitrile disposable examination gloves are non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K212578(Page1 / 5)

NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE,

CHINA

# 510 (K) Summary

The assigned 510(k) number is: K212578

Premarket Notification [510(k)] Summary

### 1.0 Submitter:

| Submitter's name :    | XINGYU MEDICAL TECH CO.,LTD.                           |
|-----------------------|--------------------------------------------------------|
| Submitter's Address : | NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, |
|                       | WEIFANG CITY, SHANDONG PROVINCE, CHINA                 |
| Phone number :        | 0086-18263618867                                       |
| Contact person:       | Cathrine Luan                                          |
| Email:                | cathrine@xingyugloves.com                              |

Submission Correspondent Primary contact: Ms. Eva Li Shanghai SUNGO Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: eatereva@hotmail.com

Secondary contact: Mr. Raymond Luo Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

Date of Preparation: 2021-11-23

#### 2.0 Name of the Device

Proprietary/Trade name: Nitrile disposable examination gloves Model(s):

| Model        | Color       | Size        |
|--------------|-------------|-------------|
| D5000        | Blue        | S, M, L, XL |
| D6000        | Black       | S, M, L, XL |
| D7000        | White       | S, M, L, XL |
| D8000        | Purple      | S, M, L, XL |
| Common Name: | Exam gloves |             |

| Common Name:           | Exam gloves               |
|------------------------|---------------------------|
| Classification Name:   | Patient examination glove |
| Device Classification: | Class I                   |
| Regulation Number:     | 21 CFR 880.6250           |
| Panel:                 | General Hospital          |
| Product Code:          | LZA                       |

{4}------------------------------------------------

NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE,

CHINA

# 3.0 Predicate Device

| Device Name:   | Xingyu Powder Free Nitrile Patient Examination Gloves, Blue Color |
|----------------|-------------------------------------------------------------------|
| Company name:  | Shandong Xingyu Gloves Co.,Ltd.                                   |
| 510(K) Number: | K171616                                                           |

# 4.0 Device Description:

# How the device functions:

Nitrile films form a barrier to prevent contamination between patient and examiner. Scientific concepts that form the basis for the device:

The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

Physical and performance characteristics such as design, materials and physical properties: The Nitrile glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151.

# 5.0 Device Indication for use:

Nitrile disposable examination gloves are non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

| Features &<br>Description | Predicate Device                                                                                                                                                                                                                             | Proposed Device                                                                                                                                                                                                                     | Result of<br>Comparison |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Company                   | Shandong Xingyu Gloves<br>Co.,Ltd                                                                                                                                                                                                            | XINGYU MEDICAL TECH<br>CO.,LTD.                                                                                                                                                                                                     | –                       |
| 510(K) Number             | K171616                                                                                                                                                                                                                                      | K212578                                                                                                                                                                                                                             | –                       |
| Product name              | Xingyu Powder Free Nitrile<br>Patient Examination Gloves,<br>Blue                                                                                                                                                                            | Nitrile disposable<br>examination gloves                                                                                                                                                                                            | –                       |
| Product Code              | LZA                                                                                                                                                                                                                                          | LZA                                                                                                                                                                                                                                 | Same                    |
| Size                      | Small/ Medium/ Large/X large                                                                                                                                                                                                                 | Small/Medium/Large/X large                                                                                                                                                                                                          | Same                    |
| Indications for Use       | Powder Free Nitrile Patient<br>Examination Gloves, Blue<br>Color is a disposable device<br>intended for medical<br>purposes that is worn on the<br>examiner's hand or finger to<br>prevent contamination<br>between patient and<br>examiner. | Nitrile disposable<br>examination gloves are non-<br>sterile disposable device<br>intended for medical<br>purposes that is worn on the<br>examiner's hand or finger to<br>prevent contamination<br>between patient and<br>examiner. | Same                    |

# 6.0 Summary of the Technological Characteristics of the Device:

{5}------------------------------------------------

# K212578(Page3 / 5)

NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE,

CHINA

| Device Description<br>and Specifications                 | Meets ASTM D6319-10<br>(Reapproved 2015)                                                                                                                 | Meets ASTM D6319-19                                                                                                                                                                                  | same        |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Dimensions<br>Length<br>ILS-2 AQL4.0                     | 232 mm min for all sizes                                                                                                                                 | 230 mm min for all size                                                                                                                                                                              | Similar*1   |
| Dimensions --<br>Width<br>IL S-2 AQL4.0                  | S: 76-88 mm<br>M: 89-102 mm<br>L: 108-119mm<br>XL: 115-128 mm                                                                                            | S: 83-89 mm<br>M: 88-100 mm<br>L: 103-111 mm<br>XL: 114-119 mm                                                                                                                                       | Similar*1   |
| Dimensions --<br>Thickness IL S-2<br>AQL4.0              | Thickness (mm) min. Finger<br>0.08<br>Palm 0.08                                                                                                          | Thickness (mm) min. Finger<br>0.09<br>Palm 0.05                                                                                                                                                      | Similar*1   |
| Physical Properties<br>IL S- 2 AQL4.0                    | Aging<br>Elongation (%)<br>Tensile<br>Strength<br>(MPa)                                                                                                  | Aging<br>Elongation (%)<br>Tensile<br>Strength<br>(MPa)                                                                                                                                              | Similar*2   |
|                                                          | Before After<br>530- 460-<br>600 580<br>18-25 17-22                                                                                                      | Before After<br>≥500% ≥400%<br>≥14 ≥14                                                                                                                                                               |             |
| Freedom from<br>Pinholes<br>Inspection Level I<br>AQL2.5 | Meets<br>• 21 CFR 800.20<br>• ASTM D6319-10 (Reapproved<br>2015) Tested in accordance<br>with ASTM D5151<br>(Reapproved 2015)<br>with acceptable results | Meets<br>•ASTM D 6319-19 Tested in<br>accordance with ASTM<br>D5151-19 with acceptable<br>results                                                                                                    | same        |
| Residual Powder                                          | Meets ASTM D 6124-06<br>(Reaffirmation 2011)<br>Results generated values<br>below 2mg of residual<br>powder                                              | Meets ASTM D 6124-06 (2017)<br>Results generated values<br>below 2mg of residual<br>powder                                                                                                           | same        |
| Materials used to<br>fabricate the devices               | Nitrile                                                                                                                                                  | Nitrile                                                                                                                                                                                              | same        |
| Shelf life                                               | NA                                                                                                                                                       | 3 years                                                                                                                                                                                              | Different*1 |
| Compare data<br>supporting Device<br>Performance         | Meets ASTM D5151-06<br>(Reapproved 2015) ASTM<br>D6319-10 (Reapproved 2015)<br>ASTM D6124-06<br>(Reaffirmation 2011)                                     | Meets ASTM D5151-<br>19 ASTM D6319-19<br>ASTM D6124-06(2017)                                                                                                                                         | same        |
| Single Patient Use                                       | Single Patient Use                                                                                                                                       | Single Patient Use                                                                                                                                                                                   | same        |
| Biocompatibility                                         | Under the conditions of this<br>study, the test article was a<br>non-irritant or non-sensitizer.<br>Meets ISO 10993-10:2010                              | Under the conditions of this<br>study, the test article was a non-<br>irritant, non- sensitizer and does<br>not pose acute systemic toxicity.<br>Meets ISO 10993-10:2010.<br>Meets ISO 10993-11:2017 | Similar*3   |

{6}------------------------------------------------

K212578(Page4 / 5)

NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE,

CHINA

# Similar*1 analysis :

The Dimensions (length, width and fingers thickness) of the proposed deice is not same with the predicate device, however the Dimensions of the subject device all meet the criteria of the ASTM D6319-19. The test results show the differences do not affect compliance with the associated standards for the proposed device.

### Similar*2 analysis :

Physical Properties of the proposed deice is not same with the predicate device, however the Physical Properties meet the criteria of the ASTM D6319-19. The test results show the differences do not affect compliance with the associated standards for the proposed device.

#### Similar*3 analysis :

Compare the predicate device, except the Skin Irritation Test and Skin sensitization test, the proposed device conduct the Acute Systemic Toxicity Test meets ISO 10993-11: 2017 additional and show it non-acute systemic toxicity. The biocompatibility test results show the differences do not affect compliance with the associated standards for the proposed device.

#### Different*1 analysis :

The proposed device determined with a 3 year shelf life. This shelf life determination was conducted under experiment per ASTM D7161-16. The observed differences do not affect compliance with the associated standards for the proposed device.

# 7.0 Assessment of Non-Clinical Performance Data:

The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate conformance with the applicable standards. The non-clinical performance testing completed for the XINGYU MEDICAL TECH CO.,LTD. nitrile disposable examination gloves demonstrated that the subject device met the acceptance criteria or specification for the applicable test methodology or standard as shown below.

- . Dimension per ASTM D6319-19
- . Tensile strength (Before aging/ After aging) and Elongated rate (Before aging/ After aging) per ASTM D6319-19
- Water leak test on pinhole per ASTM D6319-19/D6124-06 (2017)
- Powder Residual tests per ASTM D6319-19
- . Biocompatibility tests per ISO 10993-10: 2010 and ISO 10993-11: 2017

Performance of the nitrile disposable examination gloves are tested with the following technological characteristics as compared to ASTM or equivalent standards.

| Test<br>methodology | Purpose                | Acceptance criteria |                                  |       |        | Result |      |
|---------------------|------------------------|---------------------|----------------------------------|-------|--------|--------|------|
|                     |                        | S                   | M                                | L     | XL     |        |      |
| ASTM D6319-19       | Physical<br>Dimensions | Length ( mm )       | ≥220                             | ≥230  |        | pass   |      |
|                     |                        | Width ( mm ) ±10    | 80±10                            | 95±10 | 110±10 | 120±10 | pass |
|                     |                        | Thickness Finger    | ≥0.05 mm                         |       |        |        | pass |
|                     |                        | Palm                | ≥0.05 mm                         |       |        |        | pass |
| ASTM D6319-19       | Physical<br>properties | Before Aging        | Tensile≥14Mpa; Elongation ≥ 500% |       |        |        | pass |
|                     |                        | After Aging         | Tensile≥14Mpa; Elongation ≥ 400% |       |        |        | pass |

{7}------------------------------------------------

# K212578(Page5 / 5)

NO.2189 YAOQIAN ROAD, GAOMI ECONOMIC DEVELOPMENT ZONE, WEIFANG CITY, SHANDONG PROVINCE,

|  | CHINA |
|--|-------|
|--|-------|

| ASTM D6319-19<br>Test method in<br>accordance with<br>ASTM D5151-19               | Watertight<br>ness Test<br>for<br>detection<br>of holes | Batch size=35000, sample 125<br>$≤$ 7 | pass |
|-----------------------------------------------------------------------------------|---------------------------------------------------------|---------------------------------------|------|
| ASTM standard<br>D 6319-19<br>Test method in<br>accordance with<br>D6124-06(2017) | Powder<br>Residual                                      | $≤$ 2 (mg/glove)                      | pass |

| Purpose          | Standard                                                                 | Acceptance<br>criteria          | Result                                                                                                  |
|------------------|--------------------------------------------------------------------------|---------------------------------|---------------------------------------------------------------------------------------------------------|
| Biocompatibility | Skin Irritation Test Extraction<br>Method ISO 10993-10: 2010             | non-irritation                  | Passes<br>Under the conditions of<br>the study, the subject<br>device is non-<br>irritation             |
| Biocompatibility | Skin sensitization the guinea<br>pig maximization ISO 10993-<br>10: 2010 | non- sensitization              | Passes<br>Under the conditions of<br>the study, the subject<br>device is non-sensitization              |
| Biocompatibility | Acute Systemic Toxicity<br>Test ISO 10993-11: 2017                       | non- acute systemic<br>toxicity | Passes<br>Under the conditions of<br>the study, the subject<br>device is non-acute<br>systemic toxicity |

# 8.0 Assessment of Clinical Performance Data:

Clinical data is not needed to demonstrate the performance or safety of the subject glove as compared to the predicate device or applicable standards.

# 9.0 Conclusion:

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(k) submission K212578, the Nitrile disposable examination gloves, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K171616.

---

**Source:** [https://fda-staging.innolitics.com/device/K212578](https://fda-staging.innolitics.com/device/K212578)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).