← Product Code [GEX](/productcode/GEX) · K212573

# PicoLO Premium (K212573)

_Laseroptek Co., Ltd. · GEX · Dec 8, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212573

## Device Facts

- **Applicant:** Laseroptek Co., Ltd.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Dec 8, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

PicoLO Premium is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery. <Tattoo removal mode> 1064nm The 1064nm wavelength of the PicoLO Premium is indicated for tattoo removal for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple 532nm The 532mm wavelength of the PicoLO Premium is indicated for tattoo removal for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange. <PH mode> The PicoLO Premium is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. The DIA FX 1064 handpiece (1064mm) is also indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V. PicoLO Premium is also indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV.

## Device Story

PicoLO Premium is a multi-wavelength, pulsed Q-switched Nd:YAG laser system for dermatology/plastic surgery. Device delivers 1064nm and 532nm laser energy via articulated arm and handpieces to epidermis/dermis. System includes main body, color touch screen, and foot switch. Operated by clinicians in clinical settings. Embedded processor controls laser parameters. Thermal energy delivery targets tattoos, pigmented lesions, acne scars, and wrinkles. Benefits include non-invasive aesthetic and surgical treatment of skin conditions. Device is a modification of the predicate PicoLO system, featuring updated and additional handpieces.

## Clinical Evidence

Bench testing only. No clinical data presented. Device safety and performance verified through testing against standards: AAMI/ANSI ES60601-1, IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), IEC 60601-2-22 (laser safety), IEC 60825-1 (laser classification), ISO 14971 (risk management), and ISO 10993 (biocompatibility).

## Technological Characteristics

Q-switched Nd:YAG laser; wavelengths 1064nm/532nm; pulse duration 450ps (1064nm)/380ps (532nm); max energy 500mJ (1064nm)/350mJ (532nm); repetition rate 1-10Hz; beam delivery via articulated arm; cooling via internal water-to-air heat exchanger; TFT LCD touch screen interface; electrical 220-230VAC; compliant with 21 CFR 1040.10/1040.11.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- PicoLO Nd:YAG Picosecond Laser System ([K203491](/device/K203491.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 8, 2021

Laseroptek Co., Ltd. % Do Kim CEO BT Solutions, Inc. Unit 904, Eonju-ro 86-gil 5, Gangnam-gu Seoul, Seoul 06210 Korea. South

Re: K212573

Trade/Device Name: PicoLO Premium Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: August 13, 2021 Received: August 16, 2021

Dear Do Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K212573

Device Name PicoLO Premium

Indications for Use (Describe)

PicoLO Premium is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.

<Tattoo removal mode>

1064nm

The 1064nm wavelength of the PicoLO Premium is indicated for tattoo removal for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple

532nm

The 532mm wavelength of the PicoLO Premium is indicated for tattoo removal for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange.

<PH mode>

The PicoLO Premium is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. The DIA FX 1064 handpiece (1064mm) is also indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V. PicoLO Premium is also indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV.

Type of Use (Select one or both, as applicable)

| <div> <span></span>Research Use (21 CFR 201.3 Subject to Change)    </div>                   |
|----------------------------------------------------------------------------------------------|
| <div> <span></span><i>In Vitro</i> Diagnostic Use (21 CFR 809.3 Subject to Change)    </div> |

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 5. 510(k) Summary

## 1. General Information

| Applicant/Submitter: | Laseroptek Co., Ltd.                                                                                                                                                                                         |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:             | #114, #116, #117, #203, #204 Hyundai I Valley 31<br>Galmachi-Ro, 244beon-gil, Jungwon-Gu<br>Seongnam-Si, Gyeonggi-do, 13212<br>Rep. of Korea (South Korea)<br>Tel) +82.31.8023.5150<br>Fax) +82.31.8023.5151 |
| Contact Person:      | Do-Hyun Kim, BT Solutions, Inc.                                                                                                                                                                              |
| Address:             | 904, Eonju-ro 86-gil 5,<br>Gangnam-gu, Seoul, 06210, Republic of Korea<br>Tel) +82.2.538.9140<br>Email) ceo@btsolutions.co.kr                                                                                |
| Preparation Date:    | November 4, 2021                                                                                                                                                                                             |

#### 2. Device Name and Code

| Device Trade Name:   | PicoLO Premium                                                                         |
|----------------------|----------------------------------------------------------------------------------------|
| Common Name:         | Nd:YAG Laser System                                                                    |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Product Code:        | GEX                                                                                    |
| Regulation Number:   | 878.4810                                                                               |
| Classification:      | Class II                                                                               |
| Review Panel:        | General & Plastic Surgery (ODE)                                                        |

## 3. Predicate Device

PicoLO Premium is substantially equivalent to the following devices

| Applicant            | Device Name                              | 510(k) Number |
|----------------------|------------------------------------------|---------------|
| Laseroptek Co., Ltd. | PicoLO Nd:YAG Picosecond<br>Laser System | K203491       |

#### Table 5.1 Predicate device

{4}------------------------------------------------

#### 4. Device Description

PicoLO Premium is a multi-wavelength, pulsed laser system designed for the treatment of benign pigmented lesions. A key feature of the device is its ability to produce multiple laser wavelengths (1064 nm and 532 nm). The PicoLO Premium consists of a set of Q-switched Nd:YAG lasers, controlled by an embedded processor, to be used in dermatology. The laser system uses focusing optics to deliver a pattern of thermal energy to the epidermis and dermis. This system consists of main body, color touch screen, articulated arm, hand piece and foot switch.

#### 5. Indications / Intended Use

PicoLO Premium is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.

<Tattoo removal mode>

1064nm

The 1064nm wavelength of the PicoLO Premium is indicated for tattoo removal for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple.

532nm

The 532nm wavelength of the PicoLO Premium is indicated for tattoo removal for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange.

<PH mode> The PicoLO Premium is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV.

The DIA FX 1064 handpiece (1064nm) is also indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V.

PicoLO Premium is also indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV.

#### 6. Technical Characteristics in Comparison

#### 6.1. Technical Characteristics in Comparison to Predicate Device

PicoLO Premium is substantially equivalent to the following legally marketed predicate device.

|                                | Proposed Device     | Proposed Device     |
|--------------------------------|---------------------|---------------------|
| 510(K) Number                  | K203491             |                     |
| Product Code                   | GEX                 | GEX                 |
| Classification /<br>Regulation | Class II/878.4810   | Class II/878.4810   |
| Manufacturer                   | Laseroptek Co.,Ltd. | Laseroptek Co.,Ltd. |

Table 5.2 Comparison table between Predicate device and Proposed device

{5}------------------------------------------------

### PicoLO Premium 510(k) Summary

| Device Name                               | PicoLO Nd:YAG Picosecond<br>Laser System                                                                                                                                                                          | PicoLO Premium                                                                                                                                                                                               | Pulse Duration<br>(Pulse Width) | 450ps (1064nm), (Accuracy ±20%)<br>380ps (532nm), (Accuracy ±20%)                   | 450ps (1064nm), (Accuracy ±20%)<br>380ps (532nm), (Accuracy ±20%)                                                 |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Clearance Date                            | April 7, 2021                                                                                                                                                                                                     | N/A                                                                                                                                                                                                          | Pulse Energy<br>(max)           | Tattoo Mode<br>500mJ (1064nm), (Accuracy<br>±20%)<br>350mJ (532nm), (Accuracy ±20%) | Tattoo Mode<br>500mJ (1064nm), (Accuracy<br>±20%)<br>350mJ (532nm), (Accuracy ±20%)                               |
| Intended Use /<br>Indications for<br>Use: | PicoLO laser system is intended<br>for use in surgical and aesthetic<br>applications in the medical<br>specialties of dermatology and<br>general and plastic surgery.                                             | PicoLO Premium is intended for<br>use in surgical and aesthetic<br>applications in the medical<br>specialties of dermatology and<br>general and plastic surgery.                                             |                                 | PH Mode*                                                                            | PH Mode*                                                                                                          |
|                                           | <Tattoo Mode>                                                                                                                                                                                                     | <Tattoo removal mode>                                                                                                                                                                                        |                                 | 400mJ (1064nm), (Accuracy<br>±20%)<br>200mJ (532nm), (Accuracy ±20%)                | 400mJ (1064nm), (Accuracy<br>±20%)<br>200mJ (532nm), (Accuracy ±20%)                                              |
|                                           | 1064nm                                                                                                                                                                                                            | 1064nm                                                                                                                                                                                                       | Peak Power<br>(Gigawatts)       | 1.1                                                                                 | 1.1                                                                                                               |
|                                           | The 1064nm wavelength of the<br>PicoLO laser system is indicated<br>for tattoo removal for all skin types<br>(Fitzpatrick I-VI) to treat the<br>following tattoo colors: black,<br>brown, green, blue and purple. | The 1064nm wavelength of the<br>PicoLO Premium is indicated for<br>tattoo removal for all skin types<br>(Fitzpatrick I-VI) to treat the<br>following tattoo colors: black,<br>brown, green, blue and purple. | Aiming Beam                     | Laser diode, 635nm/ <5mW                                                            | Laser diode, 635nm/ <5mW                                                                                          |
|                                           | 532nm                                                                                                                                                                                                             | 532nm                                                                                                                                                                                                        | Repetition Rate<br>(Hz)         | Single, M3, M5, 1~10Hz<br>(Accuracy: ± 20%)                                         | Single, M3, M5, 1~10Hz<br>(Accuracy: ± 20%)                                                                       |
|                                           | The 532nm wavelength of the<br>PicoLO laser system is indicated<br>for tattoo removal for Fitzpatrick<br>skin types I-III to treat the<br>following tattoo colors: red, yellow<br>and orange.                     | The 532nm wavelength of the<br>PicoLO Premium is indicated for<br>tattoo removal for Fitzpatrick skin<br>types I-III to treat the following<br>tattoo colors: red, yellow and<br>orange.                     | Spot size (mm)                  | 1064 (10 mm)<br>532 (7 mm)<br>Collimator (20 mm)<br>Zoom (2~7 mm)                   | DIA FX 1064 (10mm)<br>DIA FX 1064 S (5mm)<br>DIA FX 532 (7mm)<br>S20 (20mm)<br>Collimator (10mm)<br>Zoom (2~7 mm) |
|                                           | <PH Mode>                                                                                                                                                                                                         | <PH mode>                                                                                                                                                                                                    | Laser Type                      | Q-switched Nd:YAG Laser                                                             | Q-switched Nd:YAG Laser                                                                                           |
|                                           | The PicoLO laser system is also<br>indicated for benign pigmented<br>lesions removal for Fitzpatrick<br>Skin Types I-IV.                                                                                          | The PicoLO Premium is also<br>indicated for benign pigmented<br>lesions removal for Fitzpatrick<br>Skin Types I-IV.                                                                                          | Activation                      | Via foot-switch                                                                     | Via foot-switch                                                                                                   |
|                                           | The 1064 handpiece (1064nm) is<br>also indicated for the treatment of<br>acne scars in Fitzpatrick Skin<br>Types II-V.                                                                                            | The DIA FX 1064 handpiece<br>(1064nm) is also indicated for the<br>treatment of acne scars in<br>Fitzpatrick Skin Types II-V.                                                                                | Display                         | TFT LCD Touch screen                                                                | TFT LCD Touch screen                                                                                              |
|                                           | The PicoLO laser system is also<br>indicated for treatment of wrinkles<br>in Fitzpatrick Skin Types I-IV.                                                                                                         | PicoLO Premium is also indicated<br>for treatment of wrinkles in<br>Fitzpatrick Skin Types I-IV.                                                                                                             | Cooling System                  | Internal water to air heat exchanger                                                | Internal water to air heat exchanger                                                                              |
| Wavelength                                | 1064/532 nm (Accuracy ±20%)                                                                                                                                                                                       | 1064/532 nm (Accuracy ±20%)                                                                                                                                                                                  | Electrical Power                | 220-230VAC, 50/60Hz                                                                 | 220-230VAC, 50/60Hz                                                                                               |
| Beam Delivery<br>System                   | Articulated Arm with Handpiece                                                                                                                                                                                    | Articulated Arm with Handpiece                                                                                                                                                                               |                                 |                                                                                     |                                                                                                                   |
| System<br>Dimensions(mm)                  | 350(W) x 1080(L) x 970(H)                                                                                                                                                                                         | 350(W) x 1080(L) x 970(H)                                                                                                                                                                                    |                                 |                                                                                     |                                                                                                                   |
| System Weight<br>(kg)                     | 110 kg                                                                                                                                                                                                            | 110 kg                                                                                                                                                                                                       |                                 |                                                                                     |                                                                                                                   |

{6}------------------------------------------------

# PicoLO Premium

#### 510(k) Summary

PH mode* is intended for 3 newly added indications, as below;

- 1) benign pigmented lesions removal for Fitzpatrick Skin Types I-IV,
- the treatment of acne scars in Fitzpatrick Skin Types II-V, and 2)
- 3) the treatment of wrinkles in Fitzpatrick Skin Types I-IV.

{7}------------------------------------------------

#### PicoLO Premium 510(k) Summary

#### 7. Performance Data

Non-clinical tests: Testing conducted on the PicoLO Premium shows that it refers to the relevant mandatory performance standards for laser products 21 CFR 1040.10 and 1040.11. Other performance, such as electromagnetic compliance, etc, were tested using following standards:

- PicoLO Premium is tested and evaluated according to AAMI/ANSI ES60601-1:2005 and -A1:2012. All the results presented in the submission demonstrate general requirements for basic safety and essential performance.
- -Effect to the device by electromagnetic disturbances were tested and evaluated according to the FDA-recognized consensus standard IEC 60601-1-2: 2014. All the results presented here demonstrated the requirements and tests for electromagnetic disturbances.
- -PicoLO Premium is tested and evaluated according to FDA-recognized consensus standard IEC 60601-1-6:2010/AMD1:2013. All the results presented here demonstrated the General requirements for safety - Collateral Standard: Usability.
- PicoLO Premium is tested and evaluated according to FDA-recognized consensus standard -IEC 60601-2-22: 2007 (Third Edition) + A1:2012. All the results presented here demonstrated the particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.
- -Safety of laser products is evaluated according IEC 60825-1: 2014. All the results presented here demonstrated the equipment classification and requirements.
- -Risk management was recorded according to the FDA-recognized consensus standard ISO 14971: 2012. All the results presented here demonstrated the application of risk management to medical devices.
- -Usability was documented according to the FDA-recognized consensus standard IEC 62366: 2008. All the results presented here demonstrated the application of usability engineering to medical devices.
- -Biocompatibility was tested and evaluated according to FDA-recognized consensus standard ISO 10993-5: 2009 and ISO 10993-10: 2010.

#### 8. Substantial Equivalence

The PicoLO Premium, subject of this submission, is modifications of the PicoLO Nd:YAG Picosecond Laser System cleared under K203491.

The modifications are:

- (1) changing the names of three handpieces, and
- (2) adding new two handpieces.

These modifications do not affect technological principles between the modified device and the predicate device.

Based upon the predicted overall performance characteristics for the PicoLO Premium, Laseroptek Co. Ltd. believes that no significant differences exist in usage of its underlying technological principles between PicoLO Premium and the predicate device.

{8}------------------------------------------------

#### 9. Conclusions

The technological characteristics of the subject device PicoLO Premium are comparable to the predicate device for comparable indications for use. Thus, subject device PicoLO Premium is concluded to be substantially equivalent to the predicate.

---

**Source:** [https://fda-staging.innolitics.com/device/K212573](https://fda-staging.innolitics.com/device/K212573)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
