Lilliput APD System

K212522 · Fresenius Medical Care Renal Therapies Group, LLC · FKX · Apr 15, 2022 · Gastroenterology, Urology

Device Facts

Record IDK212522
Device NameLilliput APD System
ApplicantFresenius Medical Care Renal Therapies Group, LLC
Product CodeFKX · Gastroenterology, Urology
Decision DateApr 15, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Lilliput APD System is intended for Automated Peritoneal Dialysis (APD) therapy for the treatment of end-stage renal disease (ESRD) in clinical and home settings. The supported peritoneal dialysis therapy types include: Continuous Cycling Peritoneal Dialysis (CCPD), Intermittent Peritoneal Dialysis (IPD), Peritoneal Dialysis Plus™ Therapy (PD+), Tidal Peritoneal Dialysis (TPD), Nocturnal Intermittent Peritoneal Dialysis (NIPD) and adapted Automated Peritoneal Dialysis (aAPD).

Device Story

Lilliput APD System performs automated peritoneal dialysis (APD) for end-stage renal disease (ESRD) patients. System comprises electromechanical cycler, sterile single-use disposable set, USB key, and Kinexus Gateway. Cycler uses peristaltic pump to move dialysate between solution bags, patient, and drain; includes integrated warmer to heat solution to 37°C. User interacts via touchscreen and front panel keys to program treatment parameters (infusion volume, dwell time). USB key stores/transfers prescription data; Kinexus Gateway enables communication with peripheral devices and remote data storage. Used in home or clinical settings by patients or clinicians. Device monitors fluid volumes and pressures to ensure therapy delivery. Benefits include automated, controlled delivery of dialysis therapy, reducing manual intervention and improving treatment consistency.

Clinical Evidence

No clinical data. Evidence based on bench testing, including simulated treatments, functional design verification, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-1), and human factors validation.

Technological Characteristics

Electromechanical cycler with PC/ABS housing and foil heater. Disposable set: silicone, PC, PA12, PVC, PP, HDPE, EVA. Sterilization: 100% ethylene oxide (SAL 10^-6). Connectivity: USB and Kinexus Gateway. Essential performance: flow rates 100-300 mL/min, volume accuracy ±5% or 15 mL, temperature 37°C ±2°C.

Indications for Use

Indicated for adult chronic peritoneal dialysis in home and clinical settings.

Regulatory Classification

Identification

(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.

Special Controls

*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue. April 15, 2022 Fresenius Medical Care Renal Therapies Group, LLC Denise M. Oppermann Senior Director, RA Devices 920 Winter Street Waltham, MA 02451 Re: K212522 Trade/Device Name: Lilliput™ APD System Regulation Number: 21 CFR\$ 876.5630 Regulation Name: Peritoneal dialysis system and accessories Regulatory Class: II Product Code: FKX Dated: March 11, 2022 Received: March 14, 2022 Dear Denise M. Oppermann: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Glenn B. Bell, Ph.D. Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212522 Device Name Lilliput™ APD System Indications for Use (Describe) The Lilliput™ APD System is indicated for adult chronic peritoneal dialysis in home and clinical settings Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left side of the logo is a blue symbol that consists of three chevron-like shapes stacked on top of each other. To the right of the symbol, the text "FRESENIUS MEDICAL CARE" is written in a bold, blue font. The words "MEDICAL CARE" are stacked below the word "FRESENIUS". # 5. 510(K) SUMMARY This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92. # 5.1. Submitter's Information | Name: | Fresenius Medical Care Renal Therapies Group, LLC | |-------------------|---------------------------------------------------| | Address: | 920 Winter Street<br>Waltham, MA 02451-1457 | | Phone: | (781) 996-9103 | | Fax: | (781) 699-9635 | | Contact Person: | Denise Oppermann, Senior Director | | Preparation Date: | 10 August 2021 | # 5.2. Device Name | Trade Name: | Lilliput™ APD System | |--------------------|--------------------------------------------| | Common Name: | System, Peritoneal, Automatic Delivery | | Regulation Name: | Peritoneal Dialysis System and Accessories | | Regulatory Class: | Class II per 21 CFR § 876.5630 | | Product Code: | FKX | | Product Code Name: | System, Peritoneal, Automatic Delivery | | FDA Review Panel: | Gastroenterology/Urology | # 5.3. Legally Marketed Predicate Device The legally marketed predicate devices are the Liberty PDx Cycler cleared under K141145 and the Fresenius Liberty Select Cycler cleared under K181108. # Device Description 5.4. # 5.4.1. Device Identification The Lilliput APD System is available in one (1) configuration that includes an electromechanical cycler (hereinafter referred to as "Cycler"), disposable set (hereinafter referred to as "Disposable Set''), USB key, and Kinexus Gateway. The Disposable Sets are available in three (3) configurations: - . Lilliput Disposable Set, One Patient Connector (Low feature) - Lilliput Disposable Set, One Patient Connector with Extended Drain and Patient Lines (Medium Feature) - . Lilliput Disposable Set, Two Patient Connectors with Extended Drain and Patient Lines (High Feature) {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left is a blue icon that looks like three chevrons stacked on top of each other, pointing downwards. To the right of the icon is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on top and "MEDICAL CARE" on the bottom. K212522 #### 5.4.2. Device Characteristics The Cycler is a software-controlled electromechanical device. The Disposable Sets are single-use Class II devices designed to operate with the Cycler to perform Automated Peritoneal Dialysis (APD) therapy for the treatment of end-stage renal disease (ESRD). The Disposable Sets are provided sterile and non-pyrogenic and are sterilized using ethylene oxide (EO). #### 5.4.3. Environment of Use The Lilliput APD System is prescribed for use in both professional and home healthcare settings #### Brief Written Description of the Device 5.4.4. The Lilliput APD System includes a Cycler, Disposable Set, USB Key, Kinexus Gateway, and optional accessories that are designed for use with the Cycler or Disposable Set. The Lilliput APD system components are described in Table 1. | Component | Description | |------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Cycler | Software-controlled electromechanical device | | User Interface | Touchscreen display and front panel keys which enable<br>user interaction with the Cycler | | Disposable Set | Composed of a cartridge and flexible tubing lines used for<br>exchange of fluid between the dialysate bag, patient, and<br>drain. The cartridge contains the functional components of<br>the peristaltic pump that interface with the Cycler to<br>enable fluid exchange. | | Cartridge Loader | Motorized loading mechanism that enables the interface<br>between the Cycler and Disposable Set by moving the<br>cartridge into place when it is inserted into the Cycler | | Pump System | Composed of a peristaltic pump and pump motor<br>· Peristaltic pump: Formed by the mechanical coupling of<br>the pump drive (with rotating shaft) in the Cycler and<br>the pump head in the cartridge<br>· Pump motor: Electromechanical motor that rotates the<br>Cycler pump drive | | Warmer | Heats dialysate solution prior to solution flow to the<br>patient | | USB Key | Memory device used by both patients and clinical staff to<br>store and transfer prescription and treatment data to and<br>from the cycler | | Kinexus Gateway | Allows communication between the Lilliput APD System,<br>off-the-shelf (OTS) peripheral devices, and remote data<br>storage services | #### Table 1: Lilliput APD System Components Description {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Fresenius Medical Care. On the left is a blue symbol that looks like three chevrons stacked on top of each other, pointing downwards. To the right of the symbol is the text "FRESENIUS MEDICAL CARE", with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line. The text is also in blue. K212522 #### Table 1: Lilliput APD System Components Description | Component | Description | |----------------------|-------------------------------------------------------------------------------------------------------------------------| | Optional Accessories | Optional accessories designed for use with the Cycler or<br>Disposable Set (e.g., cart, travel case, solution bag case) | The Cycler is a software-controlled electromechanical device designed for use in APD therapy for the treatment of ESRD. The software controls the functions of the Cycler during peritoneal dialysis (PD) treatment. The Cycler is designed as a table-top unit for use with a single-use, dedicated disposable set. The Lilliput APD System may be prescribed for either clinical or home treatment settings. Treatment settings, such as the amount of solution to be infused and the length of time the solution remains in the peritoneal cavity, are programmed into the Cycler. During treatment, the Cycler heats the peritoneal dialysis solution prior to patient infusion, measures and delivers a predetermined amount of fluid to the patient, and monitors the drained volume. The Lilliput Disposables Sets are used as accessories to the Cycler. The Disposables sets are composed of a cartridge and, depending on the configuration, 7 or 8 fluid lines. The fluid lines consist of: - 1 drain line (4 feet or 20 feet, depending on the configuration) with a female Luerlock connector. - 1 patient connection line (with 1 or 2 stay safe® patient connectors, depending on the configuration) - 2 warming pouch lines (1 warmer return line and 1 warmer inlet line) ● - 2 or 3 dialysate bag lines (with Safe-Lock® connectors), depending on the ● configuration. - . 1 last bag inlet line When used in conjunction with the Cycler, the Disposable Sets provide a defined, sterile fluid path with secure connections (stay safe, Safe-Lock. and Luer lock) to alternately draw the prescribed dialysate solution to and from the patient's peritoneal cavity. #### 5.4.5. Materials of Use #### 5.4.5.1. Cycler The materials of use for the Cycler are provided in Table 2. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the Fresenius Medical Care logo. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right, also in blue. The symbol is made up of three downward-pointing chevrons stacked on top of each other. | Component | Material | |------------------|--------------------------------------------| | Cycler Housing | Plastic (PC/ABS Cycoloy C6600) | | Warmer | Foil heating plates | | Cartridge Loader | Aluminum/Plastic (PC/ABS<br>Cycoloy C6600) | #### Lilliput Cycler Materials Table 2: # 5.4.5.2. Disposable Sets The Disposable Sets are classified as externally communicating, blood path indirect, prolonged contact (>24 hours to 30 days) duration, Class II (Category B) devices in accordance with FDA guidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (04 September 2020). With the exception of the cartridge and warming pouch components, the 3 Disposable Sets are constructed from materials identical to those of the predicate device. Materials used in the manufacture of the Disposable Sets are detailed in Table 3. | Component | Materials | |-----------------------|-------------------------------------------------------------------------------------| | Cartridge | Silicone<br>Polycarbonate (PC)<br>Polyamide 12 (PA12) | | Tubing and Connectors | Polyvinyl chloride (PVC)<br>Acrylic polymer<br>Polypropylene (PP)<br>Silicone<br>PC | | Warming Pouch | PVC<br>PC | | Clamps | High-density polyethylene (HDPE)<br>PP | | Connector Caps | Ethylene-vinyl acetate (EVA)<br>PP<br>Polyethylene (PE) | | Solvents | Cyclohexanone | Table 3: Disposable Set Materials of Use {7}------------------------------------------------ Image /page/7/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left is a blue symbol that looks like three downward pointing arrows stacked on top of each other. To the right of the symbol is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on top of "MEDICAL CARE". #### 5.4.6. Key Performance Specifications/Characteristic The Lilliput essential performance requirements are provided in Table 4. #### Lilliput Essential Performance Table 4: | Feature | Specification | |--------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------| | Dialyzing solution flow rate during inflow | 150-300 mL/min | | Dialyzing solution flow rate during outflow | 100-250 mL/min | | Dialyzing solution volume balancing<br>(inflow volume)<br>(FILL) | Volume accuracy is the larger of (at 2X standard<br>deviation):<br>$\leq$ 15 mL<br>or<br>$\pm$ 5% | | Dialyzing solution volume balancing<br>(outflow volume)<br>(DRAIN) | Volume accuracy is the larger of (at 2X standard<br>deviation):<br>$\leq$ 15 mL<br>or<br>$\pm$ 5% | | Volume Balancing<br>(Symmetric Volume Accuracy) | Symmetric volume accuracy is the larger of (at 2X<br>standard deviation):<br>$\leq$ 12 mL<br>or<br>±2% | | PD dwell time | Dependent on prescription. Dwell time accuracy ±1<br>minute for the calculated dwell time | | Dialyzing solution temperature | 37°C+2°C | # ર.ડ. Intended Use The Lilliput APD System is intended for Automated Peritoneal Dialysis (APD) therapy for the treatment of end-stage renal disease (ESRD) in clinical and home settings. The supported peritoneal dialysis therapy types include: Continuous Cycling Peritoneal Dialysis (CCPD), Intermittent Peritoneal Dialysis (IPD), Peritoneal Dialysis Plus™ Therapy (PD+), Tidal Peritoneal Dialysis (TPD), Nocturnal Intermittent Peritoneal Dialysis (NIPD) and adapted Automated Peritoneal Dialysis (aAPD). # 5.6. Indications for Use The Lilliput™ APD System is indicated for adult chronic peritoneal dialysis in home and clinical settings. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for Fresenius Medical Care. On the left is a blue symbol that looks like three chevrons stacked on top of each other, pointing downwards. To the right of the symbol is the text "FRESENIUS MEDICAL CARE", with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line, both in a bold, blue font. # 5.7. Comparison of Technological Characteristics with the Predicate Device The following technological characteristics of the Lilliput APD System are substantially equivalent to the primary predicate device Liberty PDx Cycler (K141145) and secondary predicate Liberty Select Cycler (K181108): - Intended Use - Indications for Use - Fundamental Scientific Technology/Operating Principle - . Technological Characteristics - . Essential Performance Requirements - Sterilization Method, Packaging, and Sterility Label Claims (Disposable Set) ● # 5.8. Sterilization Testing (Disposable Sets) The Disposable Sets are sterilized by exposure to 100% ethylene oxide (EO). The sterility assurance level (SAL) is 10-6. Sterility and non-pyrogenicity are claimed for the fluid pathway of the disposable set. # EO Residual Testing 5.8.1. Residual testing for EO and ethylene chlorohydrin (ECh) was performed in accordance with AAMI/ANSI/ISO 10993-7:2008/(R) 2012 Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals. Acceptable results (i.e., < 4.6 mg/device for EO and ECh) were obtained for the subject disposable sets. # 5.8.2. Bacterial Endotoxin (Pvrogenicity) Testing The Disposable Sets were tested for bacterial endotoxin (pyrogenicity) with Limulus Amebocyte Lysate (LAL) in accordance with ANSI/AAMI/ST72:2019 Bacterial Endotoxins – Test Methodologies, Routine Monitoring, and Alternatives to Batch Testing. It was determined the Disposable Sets are non-pyrogenic (< 20 EU/device). # 5.8.3. Sterile Barrier Testing Sterility of the Lilliput Disposable Sets' fluid pathway is maintained by the device itself. This includes vented caps with connectors (vented tortuous path or hydrophobic membrane/filter), tubing, and the cartridge portion of the Disposable Set. The vented caps were qualified as part of the sterile barrier by microbial challenge tests in accordance with ANSI/AAMI/ISO 11607-1. The tubing and other components were qualified as part of the sterile barrier through leak testing and the structural integrity test adapted from ISO 8638. Testing was performed on samples after aging and distribution simulation. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left is a blue symbol that looks like three chevrons stacked on top of each other, pointing downwards. To the right of the symbol is the text "FRESENIUS MEDICAL CARE", with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line, both in blue. # 5.9. Performance Data # 5.9.1. Cycler Performance Testing Summary Testing conducted to support the determination of substantial equivalence for the Cycler is provided in Table 5. | Test Conducted | Test Method Description | |----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Simulated Treatments | Simulated peritoneal dialysis treatments were performed in a simulated patient use environment. | | Essential Performance | Simulated peritoneal dialysis treatments were performed in a simulated patient use environment. System essential performance was verified during simulated peritoneal dialysis treatments | | System and Cycler Subsystem and Functional Design Verification | System functional requirements including user interface, solution management, pressure sensors, pumping system, and heating were verified according to pre-determined performance specifications. | | Operational Environment | System operational environment performance was verified under simulated operational environment conditions. | | Safety System | System safety requirements including alarms, solution flow stoppage, and sensor functionality were verified according to pre-determined performance specifications | | Shipping and Packaging | Shipping and packaging verification according to ASTM D4169-16 and ASTM D7386-16 | Table 5: Cycler Performance Testing Summary # 5.9.2. Lilliput Disposable Set Performance Testing Testing conducted to support the determination of substantial equivalence for the Disposable Set is provided in Table 6. | Test Conducted | Test Method Description | |---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Performance Test: Load to close<br>the tubing clamp | The maximum load force (lbf) to close the Lilliput clamp was<br>determined using a calibrated electromechanical force gauge<br>(Instron Machine) | | Performance Test: Load to push the<br>stay•safe PIN connector | The maximum load force (lbf) required for the PIN (Trigger<br>plunger) from the stay safe PIN connector to insert into a<br>catheter connector was determined using a calibrated<br>electromechanical force gauge (Instron Machine) | Table 6: Lilliput Disposable Set Performance Testing Summary {10}------------------------------------------------ Image /page/10/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left is a blue symbol that looks like three chevrons stacked on top of each other, pointing downwards. To the right of the symbol is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line. | Test Conducted | Test Method Description | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Occlusion test | The ability of the patient, solution, and drain line clamps to<br>occlude the patient, solution lines, and drain line was determined<br>by subjecting to a positive pressure of 1 bar at 15°C or 39°C for<br>10 min and observing for leakage of air bubbles. | | Compression test | The maximum load force (lbf) that the cartridge ports must<br>withstand was determined using a calibrated electromechanical<br>force gauge (Instron Machine) | | Bond Strength | A calibrated electromechanical force gauge (Instron Machine)<br>was used to perform a pull-off test for each bonded engagement | | Shipping Test | A simulated shipping and distribution test was conducted per<br>ASTM D4169-16 Standard Practice for Performance Testing of<br>Shipping Containers and Systems, Distribution Cycle 6,<br>Assurance Level II to ensure that the product's structural<br>integrity is maintained during shipping | | Heater Bag Seal | The ability of the heater bag (warming pouch) material to<br>withstand pressure was determined by subjecting it to positive<br>pressure of 1 bar at 39°C for 10 min and observing for leaks or<br>pressure decay | | Structural Integrity | The ability of the disposable sets to withstand 1.5X the labeled<br>maximum positive and negative pressures was determined | #### Lilliput Disposable Set Performance Testing Summary Table 6: # 5.9.3. Biocompatibility Testing Biocompatibility testing was conducted in accordance with ISO 10993-1:2018 and FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices –Part 1: Evaluation and testing within a risk management process" (04 September 2020). The following endpoints were evaluated to support the biological safety of the Lilliput Disposable Sets. - Semi-quantitative Leachable Chemical Evaluation, 1.5% Dextrose, 4.25% Dextrose and 7.5% Icodextrin (Volatiles, Semi-Volatiles, Non-Volatiles, Metals, and Elements) - Cytotoxicity ● - Sensitization ● - Irritation ● - Material-Mediated Pyrogenicity - Hemocompatibility - Systemic Toxicity ● {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons, stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line. - Genotoxicity A Toxicological Risk Assessment was also performed. # Human Factors Validation Testing 5.9.4. The Lilliout APD System was validated for safe and effective use in accordance with FDA guidance document Applying Human Factors and Usability Engineering to Medical Devices (03 February 2016). # 5.9.5. Electrical Safety and Electromagnetic Compatibility (EMC) Electrical Safety testing was performed in accordance with IEC 60601-1:2005. Electromagnetic Compatibility testing was performed in accordance with IEC 60601-1-2:2014. Electromagnetic Compatibility information within this submission is provided in accordance with FDA guidance document Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices (11 July 2016). # 5.9.6. Software Verification and Validation Testing Unit, software, regression (system verification), and validation testing were performed to demonstrate the effectiveness of the software and to confirm operation of the machine. Software verification information within this submission is provided in accordance with the following FDA guidance documents: - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (11 May 2005) - Guidance for Off-The-Shelf Software Use in Medical Devices (27 September 2019) ● - Content of Premarket Submissions for Management of Cybersecurity in Medical ● device (02 October 2014) - . Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software (14 January 2005) - . Design Considerations and Pre-Market Submission recommendations for Interoperable Medical Devices (06 September 2017) # 5.9.7. Animal Studies No animal studies were conducted. # 5.9.8. Clinical Studies No clinical studies were conducted. # 5.10. Conclusion The information provided in this submission, including design verification, risk management, electrical safety, electromagnetic compatibility (EMC), biocompatibility, and usability testing, demonstrates the Lilliput APD System functions as intended and supports the determination of {12}------------------------------------------------ Image /page/12/Picture/0 description: The image contains the logo for Fresenius Medical Care. The logo consists of a blue geometric shape on the left, resembling three downward-pointing chevrons stacked on top of each other. To the right of the geometric shape is the text "FRESENIUS" in bold, blue letters, with "MEDICAL CARE" in a smaller, bold, blue font below it. The background of the image is white. substantial equivalence to the predicate devices. Test results demonstrate that the differences between the proposed and the predicate devices do not raise any new concerns with regard to safety or effectiveness. The Indications for Use, technological characteristics, design, and performance requirements of the Lilliput APD System are substantially equivalent to those of the predicate devices. FMCRTG concludes that within the meaning of the Medical Device Amendments Act of 1976, the Lilliput APD System is safe and effective for the intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...