← Product Code [MYN](/productcode/MYN) · K212519

# Overjet Caries Assist (K212519)

_Overjet, Inc. · MYN · May 10, 2022 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212519

## Device Facts

- **Applicant:** Overjet, Inc.
- **Product Code:** [MYN](/productcode/MYN.md)
- **Decision Date:** May 10, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2070
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device

## Indications for Use

The Overjet Caries Assist (OCA) is a radiological, automated, concurrent read, computer-assisted detection software intended to aid in the detection and segmentation of caries on bitewing radiographs. The device provides additional information for the dentist to use in their diagnosis of a tooth surface suspected of being carious. The device is not intended as a replacement for a complete dentist's review or that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment.

## Device Story

Overjet Caries Assist (OCA) is a cloud-native, software-only CAD device for dental clinics. It processes bitewing radiographs (input) to detect and segment carious lesions (output). The system operates in three layers: Network (data transmission), Decision (ML processing), and Presentation (dashboard). The Decision layer uses an ensemble of 3 U-Net based models to classify images, assign tooth numbers, and segment lesions. The dentist reviews the annotated images on a web-based dashboard, where they can filter, hide, or edit the AI-generated segmentation masks. The device serves as a concurrent reader to assist dentists in diagnostic decision-making; it does not replace clinical judgment or comprehensive review of patient history and in vivo assessment. It aims to improve caries detection sensitivity, potentially leading to earlier or more accurate treatment planning.

## Clinical Evidence

Clinical reader study with 13 US licensed dentists evaluating 352 radiographs. Primary endpoint: AFROC AUC improvement. Results: Overall reader sensitivity improved from 57.9% to 76.2% (assisted); specificity decreased from 99.3% to 98.4%. AFROC AUC increased by 0.057 (p < 0.001). Standalone testing on 7,129 surfaces showed 72.0% sensitivity and 98.1% specificity. Mean Dice scores for lesion segmentation were 0.69 (primary) and 0.75 (secondary).

## Technological Characteristics

Software-only, cloud-native application. Operates on web browsers (Edge, Chrome, Firefox). Uses ensemble of 3 U-Net based deep learning models for image classification, tooth segmentation, and caries detection. Compatible with standard dental radiograph formats (jpg, png, eop, jif, dicom). Requires minimum 500x500 resolution. Risk management per ISO 14971.

## Regulatory Identification

Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

## Predicate Devices

- Carestream Logicon Caries Detector ([P980025](/device/P980025.md))

## Reference Devices

- Overjet Dental Assist ([K210187](/device/K210187.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

May 10, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Overjet, Inc. % Adam Odeh Regulatory Contact 560 Harrison Ave, Unit 403 BOSTON MA 02118

Re: K212519

Trade/Device Name: Overjet Caries Assist Regulation Number: 21 CFR 892.2070 Regulation Name: Medical Image Analyzer Regulatory Class: Class II Product Code: MYN Dated: April 6, 2022 Received: April 7, 2022

Dear Mr. Adam Odeh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT 8B: Division of Radiological Imaging and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212519

Device Name Overjet Caries Assist

#### Indications for Use (Describe)

The Overjet Caries Assist (OCA) is a radiological, automated, concurrent read, computer-assisted detection software intended to aid in the detection and segmentation of caries on bitewing radiographs. The device provides additional information for the dentist to use in their diagnosis of a tooth surface suspected of being carious. The device is not intended as a replacement for a complete dentist's review or that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment.

Type of Use (Select one or both, as applicable)

| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;"><input type="checkbox"/> Over The Counter Use (21 CFR 801)</span> |
|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|

|X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 4: 510(k) Summary (K212519)

This summary of 510(k) information is being submitted in accordance with the requirements of 21CFR Part 807.92

- 1. Date Prepared: October 19, 2021
- 2. Applicant

Overjet, Inc. 560 Harrison Ave Unit 403 Boston, MA 02118 Contact Person: Adam N. Odeh Email: adam.odeh@overjet.ai

- 3. Trade Name Overjet Caries Assist
- Common Name 4. Medical Imaging Analyzer
- 5. Classification 21 CFR 892.2070. Product code MYN. Class 2. Radiology
- Device Description 6.

Overjet Caries Assist (OCA) is a radiological automated concurrent read computer-assisted detection (CAD) software intended to aid in the detection and segmentation of caries on bitewing radiographs. The device provides additional information for the clinician to use in their diagnosis of a tooth surface suspected of being carious. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.

OCA is a software-only device which operates in three layers - a Network Layer, a Presentation Layer, and a Decision Layer (as shown in the data flow diagram below). Images are pulled in from a clinic/dental office, and the Machine Learning model creates predictions in the Decision Layer and results are pushed to the dashboard, which are in the Presentation Layer.

The Machine Learning System within the Decision Layer processes bitewing radiographs and annotates suspected carious lesions. It is comprised of four modules:

- Image Classifier The model evaluates the incoming radiograph and predicts the ● image type between Bitewing and Periapical Radiograph. This classification is

{4}------------------------------------------------

used to support the data flow of the incoming radiograph. As part of the classification of the image type any non-radiographs are classified as "junk" and not processed. These include patient charting information, or other non-bitewing or periapical radiographs. OCA shares classifier and Tooth Number modules with the Overjet Dental Assist product cleared under K210187.

- . Tooth Number Assignment module - This module analyzes the processed image and determines what tooth numbers are present and provides a pixel wise segmentation mask for each tooth number.
- Caries module - This module outputs a pixel wise segmentation mask of all carious lesions using an ensemble of 3 U-Net based models. The shape and location of every carious lesion is contained in this mask as the carious lesions' predictions.
- Post Processing The overlap of tooth masks from the Tooth Number . Assignment Module and carious lesions from the Caries Module is used to assign specific carious lesions to a specific tooth. The Image Post Processor module annotates the original radiograph with the carious lesions' predictions.
- 7. Indications for Use

The Overjet Caries Assist (OCA) is a radiological, automated, concurrent read, computer-assisted detection software intended to aid in the detection and segmentation of caries on bitewing radiographs. The device provides additional information for the dentist to use in their diagnosis of a tooth surface suspected of being carious. The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment.

- 8. Intended Patient Population:
The intended patient population of the device is patients that have permanent dentition, and are at least 18 years of age.

- 9. Warning and Limitations:
	- The safety and effectiveness of the system has not been established in patients with primary or mixed dentition.
	- The device should only be used by trained dentists.
	- Overjet Caries Assist system assists only in potential caries detection, not interpretation or diagnosis. It should not be relied on as the sole decision-making tool for diagnosis or treatment.
	- Only images from a supported dental radiograph system as defined in the manual can be ●

{5}------------------------------------------------

used.

- . The Overjet Dental Assist system is not intended for images smaller than 500 x 500 resolution. Overjet cannot guarantee the accuracy of results when Overjet Caries Assist is used on images of lower resolution.
- The product is not 100% sensitive, and some caries will not be detected. This can delay ● necessary treatment.
- Gross decay detection is not supported by Overjet Caries Assist, as the algorithm has been trained to detect caries within the tooth structure, rather than to detect a lack of tooth structure as is often observed with severe decay.
- Endodontic access may be mistaken by the product as caries, due to the similarities on the radiographs.
- Overiet Caries Assist cannot detect carious lesions that are not visible to a dentist on bitewing radiographs (e.g., obstructed by radiopaque restorations).
- . The product has the potential for false positive or false negative outputs. This could result in unnecessary treatment on rare occasions. Final clinical determination is the responsibility of the treating clinician with the assessment of the actual patient's dentition and multiple clinical indicators of treatment unrelated to the prediction.
- The dentist should use all appropriate clinical information to render a final clinical . opinion, with radiographic interpretation being one component of the determination process.
- The dentist is responsible for reviewing the segmentation accuracy prior to making ● diagnostic decisions, and for manually adjusting segmentation when deemed necessary.

# 10. Predicate Device

Device - Logicon Caries Detector Manufacturer - Carestream Dental PMA - P980025 (down-classified to Class II under 85 FR 3548, Jan. 22, 2020)

{6}------------------------------------------------

# 9. Substantial Equivalence

| Device                          | Carestream<br>Logicon<br>Caries Detector                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Overjet Caries<br>Assist<br>(proposed)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k                            | P980025                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | K212519                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Regulation No.<br>/ Description | CFR 892.2070<br>Medical image analyzer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | CFR 892.2070<br>Medical image analyzer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Indications                     | The Logicon Caries Detector is a<br>software device that is an aid in the<br>diagnosis of caries that have<br>penetrated into the dentin, on<br>un-restored proximal surfaces of<br>secondary dentition through the<br>statistical analysis of digital<br>intra-oral radiographic imagery. The<br>device provides additional<br>information for the clinician to use<br>in his/her diagnosis of a tooth<br>surface suspected of being carious.<br>It is designed to work in conjunction<br>with an existing Carestream dental<br>RVG digital X-ray radiographic<br>system with dental imaging<br>software (dis) for Windows XP or<br>higher. | The Overjet Caries Assist (OCA) is a<br>radiological, automated, concurrent<br>read, computer-assisted detection<br>software intended to aid in the<br>detection and segmentation of caries<br>on bitewing radiographs. The device<br>provides additional information for<br>the dentist to use in their diagnosis of<br>a tooth surface suspected of being<br>carious. The device is not intended as<br>a replacement for a complete dentist's<br>review or their clinical judgment that<br>takes into account other relevant<br>information from the image, patient<br>history, and actual in vivo clinical<br>assessment. |
| End User                        | Dentist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Dentist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Patient Population              | Patients requiring dental services,<br>all sexes, no age restriction                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Patients requiring dental services,<br>all sexes, at least 18 years of age,<br>and with permanent dentition.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Platform                        | Windows PC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Web - Edge, Chrome, Firefox                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| OS                              | Microsoft Window 7, 8, 10                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Any                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| User Interface                  | Mouse, Keyboard                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Mouse, Keyboard, Trackpad                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Image Input<br>Sources          | Images can be scanned, loaded<br>from connected Carestream image<br>solutions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Images imported from the<br>radiographic device, or from the<br>practice management system, from<br>Carestream or Schick sensors                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Image format                    | Carestream                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | jpg, png, eop, jif, dicom                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Processing<br>Architectu<br>re  | The software provides graphical<br>representation of the density change<br>in a tooth, by looking for a pattern<br>of density dips starting at the tooth<br>surface, penetrating the enamel and<br>going into the dentin. Enamel is<br>represented by 10 green lines and<br>dentin by 5 blue lines. If a pattern<br>suggestive of caries exists, the dips<br>are highlighted with red dots to<br>warn the dentist.                                                                                                                                                                                                                           | Three layers:<br>- The Network layer works with the<br>practice PACS or EMR to transmit<br>the image and meta-data to Overjet.<br>- The decision layer processes the<br>image to ensure it is the correct data<br>type, and then annotates it via the<br>algorithm<br>- The presentation layer displays the<br>annotated image in a non-diagnostic<br>viewer. The dentist can filter,<br>display, hide, create and edit the<br>annotations presented.                                                                                                                                                                        |
| Data Source                     | Bitewing radiographs acquired<br>from Carestream dental RVG<br>digital X-ray radiographic<br>system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Digital files of Bitewing radiographs<br>whose longer edge is greater than 500<br>pixel resolution                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Output                          | • Outline of suspected region<br>• Tooth Density<br>• Lesion (caries) probability                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Caries detection and segmentation<br>on radiograph resulting in outline of<br>suspected caries                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Performance<br>Testing          | Increase in dentist's sensitivity of<br>approximately 20%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Increase in dentist's sensitivity of<br>greater than 15%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Level of Concern                | Moderate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Moderate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |

{7}------------------------------------------------

Overjet Caries Assist is determined to be substantially equivalent to the Carestream Logicon Caries Detector approved in P980025 and later down-classified to Class II under 85 FR 3548, Jan. 22, 2020. Both systems are software intended to support dental professionals in their diagnosis and treatment planning for caries.

Both software systems automatically annotate suspected carious lesions for the dentist to review. The Logicon software displays suspected carious lesions as dips in annotated lines within the radiograph, green lines for the enamel and blue lines for the dentin. The Overjet Caries Assist presents suspected carious lesions as segmented polygons outlining the prediction. Both systems allow users to visualize the radiograph with the annotations, add their own annotations, and use the information as part of their diagnostic process.

Other similarities include both systems have no direct contact with the patient, both systems evaluate oral cavity radiographs, both systems utilize standard image types, and both systems connect to practice management systems. Logicon compatibility is limited to Carestream products

Some differences between the systems include the location of the software, the user interface, and the availability of additional features. A primary difference is the Carestream 

{8}------------------------------------------------

Logicon is a local software while Overjet Caries Assist is a cloud native application. While Logicon and Overjet Caries Assist have different user interfaces, both are accessed by computer and are intended for dental professionals to review annotations on dental radiographs. Overjet does not believe that the differences raise a concern of substantial equivalence and these differences do not interfere with the ability of the Overjet software to achieve its intended use.

## 10. Performance Testing

Overjet has conducted performance testing according to FDA's "Guidance for Industry and Food and Drug Administration Staff Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data - Premarket Notification [510(k)] Submissions Document" issued on: July 3, 2012, as part of the development process of the caries model. Performance testing included standalone testing and a clinical reader evaluation. All testing demonstrated that the Overjet Caries Assist software met prespecified requirements.

#### Standalone Testing

Standalone performance of the Overjet AI algorithm for the 352 images is shown in the tables below. Sensitivity and specificity were summarized based on surfaces, and 95% CIs were provided based on treating the subject as the basis of a cluster. A total of 7,129 surfaces were available and included in the analysis.

#### Sensitivity

Overall standalone sensitivity was 72.0% (62.9%, 81.1%). When broken down further. sensitivity was 74.4% (64.4%, 84.4%) for primary caries, and 62.5% (46.6%, 78.4%) for secondary caries.

#### Specificity

Overall standalone specificity was 98.1% (97.7%, 98.5%).

#### Subgroup Analyses:

Subgroup analyses were also performed for age, gender, and sensor, as well as by primary vs secondary classification, and associated restoration (for secondary caries).

{9}------------------------------------------------

|                         |                                          | Ground Truth                                     |                                            |                                    |
|-------------------------|------------------------------------------|--------------------------------------------------|--------------------------------------------|------------------------------------|
| Assessment              | Overjet<br>AI                            | Caries<br>Present                                | No Caries<br>Present                       | Total (%)                          |
| Observed Counts         | Caries Present<br>No Caries<br>Total (%) | 175 (2.5%)<br>68 (1.0%)<br>243 (3.4%)            | 131 (1.8%)<br>6755 (94.8%)<br>6886 (96.6%) | 306 (4.3%)<br>6823 (95.7%)<br>7129 |
| Diagnostic<br>Statistic | Measure<br>Sensitivity<br>Specificity    | Estimate<br>72.0% (175/243)<br>98.1% (6755/6886) | 95% CI<br>62.9%, 81.1%<br>97.7%, 98.5%     |                                    |

#### Standalone Performance of Overjet AI Algorithm based on Surfaces

· Confidence interval adjusted for multiple images and caries per subject based on clustered binary data methodology.

#### Lesion Segmentation:

Dice coefficient analysis was performed to compare pixel-level metrics of each carious lesion with the lesion tracing provided by ground truthers. Dice scores were calculated only for true positives.

For 66 images containing primary caries, the mean Dice score was 0.69 (0.66, 0.72) with a standard deviation of 0.122). For 30 images containing secondary caries, the mean Dice score was 0.75 (0.71, 0.79) with a standard deviation of 0.112.

Thus, it is unlikely that the automatic lesion segmentation presents any new risks, and any such potential risks are mitigated by the fact that Overjet Caries Assist allows the dentist to modify the margins of any segmented lesions.

## Clinical Evaluation - Reader Improvement

Overjet evaluated the Overjet Caries Assist in a multi reader fully crossed reader improvement study. 13 US licensed dentists were asked to evaluate 352 radiographs (104 containing caries / 248 without caries). Ground truth was established by the consensus labels of three US licensed dentists, and non-consensus labels were adjudicated by a Dental Radiologist. Half of the data set contained unannotated images, and the second half contained radiographs that had been processed through the OCA model. The radiographs were presented to the reader in alternating groups.

In Session 1, the readers were asked to outline suspected caries, and to review predictions from the OCA model. Each reader was asked to provide a rating of 1-4 for their confidence in the label (1 low confidence, 4 high confidence). A 30-day washout period was utilized to limit recollection bias. Following the washout, the readers were presented the same data set but with alternate grouping. If a reader saw a radiograph in the unpredicted state in session 1, they were presented with the Overjet Caries Assist

{10}------------------------------------------------

predictions in session 2.

The results were compared against a consensus ground truth, and the sensitivity, specificity, and alternative free response receiver operating characteristic (AFROC) was evaluated to characterize the performance of the readers with and without viewing the model annotations.

## Unassisted vs Assisted Sensitivity:

Overall reader sensitivity improved from 57.9% (48.9%, 66.0%) to 76.2% (68.4%, 82.6%) unassisted vs assisted. For primary caries, reader sensitivity improved from 60.5% (49.3%, 69.2%) to 79.4% (71.0%, 85.9%). For secondary caries, reader sensitivity improved from 49.8% (39.5%, 60.6%) to 63.0% (52.0%, 74.8%).

## Unassisted vs Assisted Specificity:

Overall reader specificity decreased slightly from 99.3% (99.1%, 99.5%) to 98.4% (94.5%, 98.8%) unassisted vs assisted.

# Subgroup Analyses:

Subgroup analyses were performed for age, gender, and sensor, primary vs secondary classification, and associated restoration (for secondary caries).

## Unassisted vs Assisted Dice Scores:

As with standalone testing, Dice scores were calculated in comparison to ground truth for readers with and without Overjet Caries Assist.

For primary caries, mean Dice scores improved from 0.67 unassisted (standard deviation 0.012) to 0.69 assisted (SD 0.011). For secondary caries, mean Dice scores improved from 0.65 unassisted (SD 0.017) to 0.74 assisted (SD 0.017). These data suggest improved segmentation when using Overjet Caries Assist, though results were not statistically significant.

# AFROC Scores:

Readers provided confidence scores for any detected caries, which were used to calculate AUC for weighted AFROC scores. For the average of all readers, AUC increased from 0.593 (0.686, 0.743) to 0.649 (0.744, 0.820), for an increase in AUC of 0.057 (0.039, 0.098) unassisted to assisted. This increase was statistically significant, with an overall p-value less than 0.001.

## Summary

{11}------------------------------------------------

Increases in overall AFROC numbers clearly demonstrate improvement in caries detection by dentists when aided by Overjet Caries Assist (0.057 increase). This aligns with the observed increase in sensitivity for both primary and secondary caries (18.9% improvement for primary; 13.2% improvement for secondary caries). When considered alongside the observed decrease in overall specificity, which was less than 1%, it is clear that Overjet Caries Assist demonstrates a clear benefit for caries detection.

- 11. General Safety and Effectiveness Concerns
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of this device. Risk management was conducted according to ISO 14971 which ensured, via a risk analysis, the identification and mitigation of potential hazards. Any potential hazards were controlled via software development and design, verification, and validation testing. In addition, general and special controls of the FD&C Act established for Radiological Computer Assisted Detection and Diagnosis Software are in place to further mitigate any safety and or effectiveness risks.

- 12. Assessment of Non-Clinical Performance Data
Overjet Caries Assist has been verified and validated according to Overjet's design control processes. All supporting documentation has been included in this 510(k) Premarket Notification. Verification activity included unit, integration, and system level testing. Validation testing included performing a pivotal reader study to compare the clinical performance of dentists using CAD detections from the Overjet Caries Assist software when applied to dental radiographs to that of dentists not using Overjet Caries Assist.

# 13. Conclusion

Overjet Caries Assist is substantially equivalent to the predicate device, Carestream Logicon Caries Detector. Any differences do not raise any concerns about the safety or efficacy of the device.

---

**Source:** [https://fda-staging.innolitics.com/device/K212519](https://fda-staging.innolitics.com/device/K212519)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).