This Digital Mobile Diagnostic X-Ray System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
Device Story
Mobile digital X-ray system for diagnostic imaging in clinical settings. Operates via battery or mains power; features motorized movement controlled by capacitive handlebar sensors. System includes X-ray generator (20-50 kW), rotating anode X-ray tube, and manual collimator. Input: X-ray radiation; Output: digital radiographic images. Operated by qualified doctors or technicians. Advanced configuration includes secondary head-assembly touch screen and 4-step telescopic arm. Image acquisition managed by Canon CXDI Control Software NE on integrated touch screen console. Ethernet/Wi-Fi connectivity supports digital detector communication. Benefits include mobile diagnostic capability for patients in various positions (prone, supine, standing, sitting).
Clinical Evidence
Bench testing only. No clinical data required as all digital X-ray receptor panels have previous FDA clearance. System performance verified through compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, 60601-2-28, 60601-1-6, and 62304 standards.
Indicated for adult and pediatric patients for diagnostic radiographic imaging of the skull, spine, chest, abdomen, and extremities. Not for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are displayed side-by-side.
November 3, 2021
Stephanix % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
Re: K212515
Trade/Device Name: MOVIX DReamy Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL, MQB Dated: August 5, 2021 Received: August 10, 2021
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
, for
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212515
Device Name MOVIX DReamy
Indications for Use (Describe)
This Digital Mobile Diagnostic X-Ray System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary: 510(k) Number K212515
Image /page/3/Picture/1 description: The image shows the logo for Stephanix Medical Imaging Solutions. The logo features a blue square with a stylized "X" inside, followed by the word "STEPHANIX" in large, bold, blue letters. Below the company name, the words "MEDICAL IMAGING SOLUTIONS" are written in smaller, lighter blue letters.
Stephanix 10 Rue Jean Moulin La Ricamarie Auvergne-Rhone-Alpes, FR 42150 (Tel): +33 4 77 47 81 60 Registration Number: 3006972752 Date Prepared: October 22, 2021 Contact: Sandie Perret Quality Manager
- 1) Identification of the Device: Trade/Device Name: MOVIX DReamy Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 2) Equivalent legally marketed device: K192011 Trade/Device Name: PhoeniX Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 3) Reference devices: We employ these cleared devices (K131106, K133693, K170332 and K171270 with software update cleared in K190368) without modification: Trade/Device Name: Canon Digital X-ray Panels CANON CXDI-401C Wireless CANON CXDI-701C Wireless CANON CXDI-801C Wireless CANON CXDI-710C Wireless CANON CXDI-810C Wireless CANON CXDI-410C Wireless Regulation Number: 21 CFR 892. 1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB
- 4) Indications for Use: This Digital Mobile Diagnostic X-Ray System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts.
{4}------------------------------------------------
Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
- 5) Description of the Device: These are modified versions of the predicate mobile digital diagnostic x-ray systems. They feature motorized movement and full battery operation. Various Canon digital X-ray panels are supplied with the system. (See list above.) The Mobile X-Ray Unit MOVIX DReamy has a Basic configuration or Advanced configuration. Advanced configurations include:
- Second screen on head assembly.
- Smooth movements of head assembly.
- Telescopic arm in four steps instead three steps for basic configuration.
The Mobile X-Ray unit MOVIX DReamy is provided with touch screen to operate as a control console. The Digital Imaging System is composed by image receptors and application for image acquisition (control console & image processing controller). The Image acquisition software "CANON CXDI Control Software NE" runs on MOVIX DReamy and it is displayed on touch screen. It is the user interface and compatible digital detectors are listed above. All have FDA Clearance. The Advanced configuration has a second Touch Screen Monitor located on head-assembly.
The MOVIX DReamy Mobile X-Ray Unit is provided with separate battery packs for X-Ray generation and motorized movements of the unit can operate connected to mains or in stand-alone mode, that is, operating without mains being present or unplugged from mains. The unit is connected to mains to charge the batteries; Rating: the input voltage range goes from 100 V~ to 240 V~, 1 phase, 1.1 kVA. New X-ray generator (model SHFM) is mounted on Battery Mobile X-Ray unit MOVIX DReamy This X-Ray generator for MOVIX DReamy Mobile X-Ray Unit has a radiogenic unit mounted on headassembly and comprising an X-ray tube with rotating anode and its circuit for high voltage. The electronic and associated software to control the X-ray generation are placed on mobile cart. The Battery Mobile X-Ray Unit MOVIX DReamy is provided with different output powers: 20 kW, 32 kW, 40 kW and 50 kW. There are available two X-ray tube inserts with rotating anode manufactured by CANON ELECTRON TUBES & DEVICES:
- XRR-3331 insert.
- E7886 insert.
The Manual Beam Limiting Device is from Ralco, model: R108 F. External interface (controls) and covers are provided by Sedecal. There are two versions of collimator assembly, Basic and Advanced.
Image /page/4/Picture/10 description: The image is a comparison of two medical devices, labeled "Basic" and "Advanced". The "Basic" device has a white and gray body with a black control panel that has two knobs and a port. The "Advanced" device has a similar white and gray body, but it also has a display screen and additional components.
{5}------------------------------------------------
| Basic | Advanced |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Knobs to adjust the collimator blades.<br>Collimator light push button.<br>Rail system to install external additional<br>filtration.<br>Measuring tape to measure the SID.<br>Double laser (optional)<br>Motorized filter (optional) | Knobs to adjust the collimator blades.<br>Collimator light push button.<br>Rail system to install external additional filtration.<br>Measuring tape to measure the SID.<br>Double laser (optional)<br>Motorized filter (optional)<br>Rear knobs to adjust the collimator blades (optional).<br>Additional options selectable with TFT:<br>Variable additional filtration.<br>Collimator blades aperture. |
The Battery Mobile X-Ray Unit MOVIX DReamy is driven by holding the handlebar with the hands. It is provided to internal sensor (gauges) that control the direction and speed of each wheel depending the force applied by the operator.
The handlebar is capacitive handlebar and when operator put his handlebar the movements are enabled. The height of handlebar is adaptable to improve the user commodity.
Four buttons are provided on the hand grips placed on Head-Assembly. They control the motion of each driving wheel and allow fine positioning adjustment of the patient (movements with low speed because this control is not intended for long displacement).
The Battery Mobile X-Ray unit MOVIX DReamy supports:
- Ethernet communication (with hospital and backup cable for digital detectors) ●
- Wireless Wi-Fi communications (with digital detectors and remote hand switch). The wireless . communication is supported by Access Point and one single internal antenna.
Options:
Barcode scanner for register patient information. Dosimetry for Dose Area Product (DAP).
Tube Arm specifications:
-Counterbalanced tube arm.
- Tube arm reach 1220 mm horizontally for both configurations (Advanced option with 4 steps and Basic option with 3 steps).
-Column rotation: ± 317 °.
- Maximum SID to floor is 2020 mm for both configurations.
- Minimum SID to floor is 530 mm for both configurations.
Mobile Unit:
- Weight: the heaviest configuration 520 kg.
-Size (L x W x H):
- Mobile on parking position: 1220 x 540 x 1290 mm. o
- Mobile fully extended (Telescopic column & Telescopic arm): 2290 x 1790 x 2230 mm. O
-Frontal anti-collision system.
- Motorized unit up to 5,5 km/h.
- Motor drive with two independent drive motors, one for each wheel (forward and reverse).
-Touch screen console:
- o Main control console: 19 inch.
- Additionally, second touch screen of 8.4 inch for Advanced configuration O
{6}------------------------------------------------
## Substantial Equivalence Chart 6)
| Characteristic | Phoenix K192011 (Sedecal) | MOVIX DReamy |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use: | Intended for use by a qualified/trained<br>doctor or technician on both adult and<br>pediatric subjects for taking diagnostic<br>radiographic exposures of the skull, spinal<br>column, chest, abdomen, extremities, and<br>other body parts. Applications can be<br>performed with the patient sitting,<br>standing, or lying in the prone or supine<br>position. Not for mammography | This Digital Mobile Diagnostic X-Ray System is<br>Intended for use by a qualified/trained doctor<br>or technician on both adult and pediatric<br>subjects for taking diagnostic radiographic<br>exposures of the skull, spinal column, chest,<br>abdomen, extremities, and other body parts.<br>Applications can be performed with the patient<br>sitting, standing, or lying in the prone or supine<br>position. Not for mammography. (SAME,<br>device description added) |
| Configuration | Mobile System with digital x-ray panel<br>and image acquisition computer | SAME |
| X-ray<br>Generator(s) | kW rating: 20 kW, 32 kW, 40 kW and 50 kW.<br>kV range: from 40 kV to 150 kV in 1 kV steps.<br>mA range: from 10 mA to 630 mA / 640 mA /<br>650 mA. | SAME |
| Collimator | Ralco R108F | SAME |
| Photos | Image: Phoenix K192011 | Image: MOVIX DReamy |
| Digital X-ray<br>Panel<br>Supplied | CANON CXDI-401C Wireless<br>CANON CXDI-701C Wireless<br>CANON CXDI-801C Wireless<br>Plus:<br>CANON CXDI-710C Wireless<br>CANON CXDI-810C Wireless<br>CANON CXDI-410C Wireless | SAME |
| Characteristic | PhoeniX K192011 (Sedecal) | MOVIX DReamy |
| Software | Canon control software CXDI-NE | SAME |
| Panel<br>Interface | Ethernet or Wi-Fi wireless | SAME |
| Meets US<br>Performance<br>Standard | YES 21 CFR 1020.30 | SAME |
| Power<br>Source | Universal power supply, from 100 V~ to<br>240 V~. 1 phase, 1.1 kVA | SAME |
{7}------------------------------------------------
- 7) The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness. Safety and Effectiveness, comparison to predicate device. The results of bench testing indicate that the new devices are as safe and effective as the predicate devices. Proper system operation is fully verified upon installation. We verified that the modified combination of components worked properly and produced diagnostic quality images as good as our predicate generator/panel combination.
- 8) Summary of non-clinical testing: Systems covering all generator/panel combinations were assembled and tested and found to be operating properly. Firmware was validated according to the FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005. Because the system uses Wi-Fi and Ethernet, we observed the recommendations contained in the FDA Guidance Document: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014. The digital panel software employed was already reviewed by FDA in K190368. Because the units are indicated for the pediatric population (as well as the adult population) we reviewed and took into account the FDA guidance Pediatric Information for X-ray Imaging Device Premarket Notifications.
The MOVIX DReamy Units have been tested to be in compliance with the following International Standards:
- a) IEC 60601-1:2005+A1:2012 (Edition 3.1)
- b) IEC 60601-1-2:2014 (Edition 4.0)
- c) IEC 60601-1-3:2008+A1:2013 (Edition 2.1)
- d) IEC 60601-2-54:2009+A1:2015 (Edition 1.1)
- e) IEC 60601-2-28:2010 (Edition 2.0)
- f) IEC 60601-1-6:2010 + A1:2013 (Edition 3.1)
- g) IEC 62304:2006 + A1:2016 (Edition 1.1)
- 9) Summary of clinical testing: Clinical testing was not required to establish substantial equivalence because all digital x-ray receptor panels have had previous FDA clearance.
- 10) Conclusion: After analyzing bench and clinical tests, it is the conclusion of STEPHANIX that the new MOVIX DReamy Digital Diagnostic Mobile X-Ray Systems are as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.