K212472 · Triworks Group Srl · GEI · Jul 7, 2022 · General, Plastic Surgery
Device Facts
Record ID
K212472
Device Name
AgeJet
Applicant
Triworks Group Srl
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 7, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AgeJet system is an electrosurgical device used in dermatologic and general surgical for electro coagulation and hemostasis.
Device Story
AgeJet is an electrosurgical device for dermatologic and general surgical procedures; performs electrocoagulation and hemostasis. System consists of a generator, footswitch, handpiece/cable assembly, handpiece holder with thermal sensor, and disposable nozzles. Operates by delivering plasma energy to tissue surface; energy transfer generates electrothermal reaction for coagulation. Nitrogen gas (99.5% purity) required for plasma generation. Used in hospitals and clinics by healthcare professionals. Handpiece includes integral key to prevent use beyond operational life. Output controlled via footswitch. Provides consistent thermal damage profiles in tissue, comparable to predicate. Benefits include controlled tissue coagulation for surgical/dermatological applications.
Clinical Evidence
Bench testing only. In vitro testing performed on liver, kidney, muscle, and skin tissue samples. Evaluated thermal damage profiles at various power levels (1J, 2J, 4J) and distances. Results confirmed consistent thermal damage profiles (depths up to 350 microns) comparable to the predicate device. Biocompatibility evaluated per ISO 10993-1:2018. Electrical safety verified per EN 60601-1, EN 60601-1-2, and EN 60601-2-2.
Technological Characteristics
Electrosurgical device; 2.45GHz RF frequency; nitrogen gas-based plasma. Power: 110-240 VAC. Max power 260W. Output energy 0.5-41J. Repetition rate 1-3 Hz. Pulse width 4-15ms. Thermal effect depth up to 350 microns. Includes handpiece with thermal sensor and disposable nozzles with usage-tracking keys. Biocompatibility per ISO 10993-1:2018. Electrical safety per EN 60601 series.
Indications for Use
Indicated for dermatological and general surgical conditions requiring electrocoagulation and hemostasis.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
July 7, 2022
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Triworks Group SRL % Matthew Brulport President NEOconcepts LLC 2773 Alum Crossing Dr. Lewis Center, Ohio 43035
Re: K212472
Trade/Device Name: AgeJet Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: June 17, 2022 Received: June 21, 2022
Dear Matthew Brulport:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical
{1}------------------------------------------------
device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and -CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212472
Device Name AgeJet System
Indications for Use (Describe)
The Agelet system is an electrosurgical device and is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for TRIWORKS, a company that promotes "The Intelligent Beauty". The logo features a stylized trident symbol in gray, positioned above the company name. The name "TRIWORKS" is written in a sans-serif font, with a small green rectangle to the left and a red rectangle to the right, resembling the colors of the Italian flag.
# K212472. 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92(c).
| Submitter's Name: | Lorenzo Berchicci |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Company: | Triworks Group srl |
| Submitter's Address: | via Leone Belpulsi, 3 - 86100 Campobasso (CB) Operational<br>Office: via Don Giuseppe Mucciardi, 5 86020 Campochiaro<br>(CB) - Italy. |
| Phone: | +39 0874 1896435 |
| Fax: | +39 0874 1896435 |
| Email: | Lorenzo Berchicci 1.berchicci@triworks.it |
| Date Prepared: | July 4th 2022 |
| Device Trade Name: | AgeJet |
Device Common name: Electrosurgical Device
# Device Classification Information:
| Regulation<br>Number | Device<br>Classification<br>name | Device<br>Class | Product<br>Code | Classification<br>Panel | Type |
|----------------------|-------------------------------------------------------------------------|-----------------|-----------------|------------------------------|------------------------|
| 21 CFR<br>878.4400 | Electrosurgical<br>cutting and<br>coagulation device<br>and accessories | Class 2 | GEI | General &<br>Plastic Surgery | Traditional<br>510 (k) |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for TRIWORKS, a company that appears to be in the beauty industry. The logo features a stylized trident symbol in gray above the word "TRIWORKS" in black. Below "TRIWORKS" is a green and red line, and below that is the tagline "The Intelligent Beauty" in gray.
## Device Description
The AgeJet system is an electro-surgical device intended for use in dermatological applications in hospital and clinics.
The AgeJet system comprises of the following.
## AgeJet Generator
The wheeled unit powered by120 VAC (available 110-240 VAC according to the market) with standard wall socket.
## Footswitch
This is a single pedal type for activation of the output.
# Handpiece/Cable Assembly
The handpiece and the cable assembled can be easily plugged and unplugged from the AgeJet Generator as reported in the User Manual. The handpiece has an integral key that is used by the generator to ensure it is not used beyond its operational life.
## Handpiece holder
The handpiece holder with the thermal sensor is connected to the generator according to the User Manual.
## Procedure Pack
Comprising one or more disposable nozzles that is connected to the handpiece, and which have an integral key that is used by the generator to ensure it is not used beyond its operational life.
# Nitrogen Gas Cylinder & Pressure regulator
These are not supplied as a part of the AqeJet system. Only medical grade nitrogen (99.5% purity) should be used. The Nitrogen tank is specified as 534mm high by 106mm diameter, a capacity of 560 litres at 2200psiG.
# Cylinder Holder Plate Kit
This kit includes screws, plate, cable holder shaft and cable tray. Please refer to the User Manual for the installation procedures.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image features the logo for TRIWORKS, a brand associated with beauty products. Above the brand name is a stylized trident symbol in gray. Below the brand name, there is a green rectangle on the left and a red rectangle on the right, possibly representing the colors of the Italian flag. The tagline "The Intelligent Beauty" is written in a simple font below the rectangles.
## Intended Use
The AgeJet system is an electrosurgical device used in dermatologic and general surgical for electro coagulation and hemostasis.
Plasma energy is delivered to the tissue and energy is rapidly transferred to the tissue surface. As the plasma energy passes through the tissue it generates an electrothermal reaction which is capable of coagulating tissue.
#### Indications for Use
The AgeJet system is an electrosurgical device that can be used in the treatment of dermatological and general surgical conditions through electro coagulation and hemostasis.
#### Predicate device(s)
The AgeJet system is substantially equivalent to the PlaDuo system K201735.The AgeJet system is predicated against Pladuo as both devices are electrosurgical devices intended for the treatment of dermatological and general surgical conditions through electro coagulation and hemostasis.
The key technological characteristics of the subject device and predicate device are summarized in the following table;
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for TRIWORKS. Above the company name is a gray trident-like symbol. Below the company name is the tagline "The Intelligent Beauty", with a green and red line above the tagline.
| Characteristics | Proposed device<br>AgeJet Device<br>K212472 | PlaDuo System<br>K201735 | Degree of equivalence |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Manufacturer name &<br>address. | Triworks SRL<br>Via S. Rocco, 58 Bis<br>86037<br>Palata CB<br>Italy | SheNB Co Ltd<br>Seongsui-ro,<br>148 Seongdong-gu<br>Seoul<br>South Korea | N/A |
| Device Trade Name | AgeJet | PlaDuo | N/A |
| Mode of Operation | Nitrogen Gas | Nitrogen Gas | Identical |
| Intended use | An electrosurgical device used in<br>dermatologic and general surgical<br>for electro coagulation and<br>hemostasis. | An electrosurgical device used in<br>dermatologic and general surgical<br>for electro coagulation and<br>hemostasis. | Identical |
| Principle of Operation | Plasma energy is delivered to the<br>skin and the energy is rapidly<br>transferred to the skin surface.<br>As the plasma energy passes<br>through the tissue it generates<br>an electrothermal reaction which<br>is capable of coagulating tissue. | Plasma energy is delivered to the<br>skin and the energy is rapidly<br>transferred to the skin surface. As<br>the plasma energy passes through<br>the tissue it generates an<br>electrothermal reaction which is<br>capable of coagulating tissue. | Identical |
| Frequency | 2.45Ghz | 2.45Ghz | Identical |
| Modulation | Pulsed Carrier wave | Pulsed Carrier Wave | Identical |
| Max Power (w) overall | 260W<br>(260-160 = 100W) | 100W | Identical |
| Cooling compressor<br>power (W) | 160W | Not Applicable | PlaDuo does not have patient<br>cooling system |
| Maximum Magnetron<br>output | 900W<br>(AgeJet uses only 11% of total at<br>all energy levels) | Not known | - |
| Output Energy (joule) | 0.5-41 | 4 | Identical |
| Characteristics | Proposed device<br>AgeJet Device<br>K212472 | PlaDuo System<br>K201735 | Degree of equivalence |
| Repetition rate | 1-3 Hz | 1-3Hz | Identical |
| Pulse Width (single<br>pulse) | 4 to 15ms | 5 to 15ms | Equivalent |
| Depth of Thermal<br>effect | Up to 350 microns | Up to 400 microns | Equivalent |
| Gas requirement | Medical grade Nitrogen 99.9% | Medical grade Nitrogen 99.9% | Identical |
| Electrical supply<br>Voltage | AC 110 - 240 Volts<br>50/60 Hz | AC 100 - 230 Volts<br>50/60 Hz | Equivalent |
| Mains connection | Detachable Power Cord to IEC<br>type appliance inlet<br>Mains Switch controlled ON/OFF<br>Operation | Detachable Power Cord to IEC<br>type appliance inlet<br>Mains Switch controlled ON/OFF<br>Operation | Identical |
| Device classification | Class II | Class II | Identical |
| Regulation number | 21 CFR 878.4400 | 21 CFR 878.4400 | Identical |
| Rx/OTC | RX | RX | Identical |
| Device classification<br>name | Electrosurgical cutting and<br>coagulation device and<br>accessories | Electrosurgical cutting and<br>coagulation device and<br>accessories | Identical |
| Device product code,<br>classification name | GEI | GEI | Identical |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for TRIWORKS, a company that is branded as "The Intelligent Beauty". The logo features a stylized trident above the company name. The trident is gray, and the company name is in black, with a green bar above the "TRI" and a red bar above the "WORKS".
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for TRIWORKS. The logo features a stylized trident symbol in gray above the word "TRIWORKS" in blue. Below "TRIWORKS" are two small rectangles, one green and one red, side by side. Underneath the rectangles is the tagline "The Intelligent Beauty" in gray.
#### Similarities and Differences between the subject and predicate device: Key Similarities
The AgeJet system is substantially equivalent to the PlaDuo System (K201735), as both are electrosurgical devices used in dermatologic and general surgical for electro coagulation and hemostasis.
Both systems exact their effect on the tissue through the same mechanism. Plasma energy is delivered to the skin and the energy is rapidly transferred to the skin surface. As the plasma energy passes through the tissue it generates an electrothermal reaction which is capable of coagulating tissue.
The AgeJet system and PlaDuo have the same principle of operation, mode of action and equivalent energy outputs to a maximum of 4Joule.
Both devices have similar incremental precision settings of 0.25J and 0.1 Joule respectively.
The AgeJet, and PlaDuo systems have identical treatment pulse repetition rates 1.0 to 3.0 Hz.
The pulse of UHF required to generate a given energy level or single Pulse width are equivalent.
The RF output frequency of the proposed device is identical to the predicate device 2.45Ghz.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for TRIWORKS, a company that specializes in beauty products. The logo features a stylized trident symbol in gray, with the word "TRIWORKS" written below it in a sans-serif font. Underneath the word "TRIWORKS" are two horizontal bars, one green and one red, which may represent the Italian flag. The tagline "The Intelligent Beauty" is written in a smaller font below the bars.
#### Differences
There are no significant differences in terms of the key device parameters that raise questions in terms of safety and efficacy.
Where there are differences between the AgeJet system and the predicate device, these occur in terms of the devices services which do not alter efficacy parameters of the device and where necessary have been tested and conform to relevant standards.
#### Electrical safety and safety standards
To demonstrate safety and effectiveness of the AgeJet system and to demonstrate substantial equivalence to the predicate devices, Triworks Group srl has completed several non-clinical performance tests. The AgeJet meets established requirements for overall design, electrical safety, software validation and usability studies confirming that the design outputs meet design input requirements and established specifications.
The AgeJet system successfully passed testing per internal verification/validation requirements and national/international standards illustrated below: EN 60601-1:2006 EN 60601-1-2:2015
EN 60601-2-2:2018
## Other Non-Clinical Performance testing
To demonstrate safety and effectiveness and substantial equivalence the AgeJet system has undergone several non-clinical performance tests in line with recognized standards in terms of general requirements, biocompatibility, and software.
The following non-clinical performance data is provided in support of the substantial equivalence determination.
#### Biocompatibility
Materials used in the AgeJet system have been evaluated in respect to material composition, use and patient/user contact according to ISO 10993-1: 2018.
#### Software verification and validation testing
Software has been classified in accordance with EN 62304:2015 and Guidance for the Content of Premarket Submissions for Software Contained in Medical
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for TRIWORKS. The logo features a stylized trident symbol in gray, above the word "TRIWORKS" in a sans-serif font. Below the word, there are two horizontal bars in green and red, resembling the colors of the Italian flag. Underneath the bars, the tagline "The Intelligent Beauty" is written in a smaller, lighter gray font.
Devices Guidance for Industry and FDA Staff May 2005. Verification and Validation in accordance with the software risk concern provides evidence that all design specifications were met.
#### Performance testing
Performance tests were conducted to assess the performance characteristics of AgeJet and ensure its functionality and stability when used in accordance with the manufacturer indications. The device was tested to verify that the output power value falls within the manufacturers specified range, thus ensuring the efficiency of the device. The results show that AgeJet works as expected, according to the manufacturers specifications.
In vitro testing of the AgeJET nitrogen plasma device was performed on four different types of tissue as per Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery (DOC-FDA-2014-D-0217): liver, kidney, muscle, and skin. Treatment settings included the three power levels, low (1 J), medium (2 J) and maximum (4 J), at the minimum clinically relevant distance (5 mm) from the tissue surface and at temperature close to physiological 37°C. All test settings were triplicated and collected biopsy samples analyzed by H&E histology method. The testing demonstrated that the treatments produced clearly detectable thermal damage profiles in treated tissues with depths ranging from a few tens of microns up to a few hundred microns for a sinqle pulse treatment and several stacked pulses, respectively. The study demonstrated the ability of the AgeJet device to achieve consistent thermal damage profiles in line with the target treatment and comparable to the profiles produced by the predicate device. Thus, it can be concluded that treatment by such device at the appropriate testing settings will possess a desirable clinical treatment effect.
#### Statement of Substantial Equivalence:
513(i) of the FD&C Act (21 U.S.C. 360c(i) states that for substantial equivalence a proposed device is required to have the same intended use and equivalent technological characteristics as the predicate device. Where there are differences in technological characteristics, these can be negated by appropriate clinical or non-clinical performance testing demonstrating that the proposed device is as safe and effective as the predicate device, and that the proposed device does not
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for TRIWORKS, a company that specializes in beauty products. The logo features a stylized trident above the word "TRIWORKS" in a sans-serif font. Below the word "TRIWORKS" are two horizontal bars, one green and one red, which may represent the Italian flag. The tagline "The Intelligent Beauty" is located below the bars.
raise any different questions of safety and effectiveness than the predicate device for the same intended use.
Triworks Group srl has demonstrated that the AgeJet device has the same intended use as the predicate device, employs equivalent technological characteristics, and has similar thermal effects in tissue. Where there are minor differences, these occur only in the basic services of the device. The sponsor demonstrates that the device does not pose any additional questions regarding safety and efficacy relative to the predicate.
Therefore, the AgeJet system, as designed, and manufactured, has been demonstrated to be substantially equivalent to the referenced predicate, the PlaDuo System K201735.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.