K212434 · MIPM Mammendorfer Institut für Physik und Medizin GmbH · KOI · May 5, 2022 · Anesthesiology
Device Facts
Record ID
K212434
Device Name
Neuromuscular Transmission Monitor TOF3D
Applicant
MIPM Mammendorfer Institut für Physik und Medizin GmbH
Product Code
KOI · Anesthesiology
Decision Date
May 5, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2775
Device Class
Class 2
Indications for Use
The TOF3D is used to objectively monitor the level of neuromuscular transmission by measuring muscle contraction following stimulation. The TOF3D can also be used as a peripheral nerve stimulator (without the objective measuring function) for subjective monitoring. The device is intended for use for adolescents greater than 18 through 21 years of age, and adults.
Device Story
Neuromuscular Transmission Monitor TOF3D monitors neuromuscular transmission (NMT) levels during surgery or in intensive care units. Device stimulates peripheral nerves (ulnar, posterior tibial, or facial) and measures resulting muscle contraction via acceleromyography (AMG). Output displayed on integrated LCD; aids medical staff in maintaining neuromuscular block levels and assessing recovery. Used by qualified medical staff in clinical settings. Provides objective monitoring via AMG or subjective monitoring via nerve stimulation. Benefits include precise titration of neuromuscular blocking agents and improved patient safety during recovery.
Clinical Evidence
No clinical or animal testing performed. Substantial equivalence demonstrated via bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-10), EMC (IEC 60601-1-2), software verification/validation, and biocompatibility (ISO 10993-1).
Technological Characteristics
Acceleromyography (AMG) sensor; constant current electrical stimulator (0-60 mA, max 300V); monophasic pulses (200/300 μs). Includes surface temperature sensor. Standards: IEC 60601-1, IEC 60601-1-6, IEC 60601-2-10, IEC 60601-1-2, ISO 10993-1. LCD display. Biocompatible skin-contact materials.
Indications for Use
Indicated for adolescents >18 through 21 years and adults requiring objective or subjective monitoring of neuromuscular transmission during surgery or in intensive care units.
Regulatory Classification
Identification
An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a circular seal with an abstract design. To the right of the seal, the text "FDA U.S. FOOD & DRUG ADMINISTRATION" is displayed in a bold, sans-serif font. The word "FDA" is in a blue square, and the rest of the text is in a smaller font size.
May 5, 2022
MIPM Mammendorfer Institut für Physik und Medizin GmbH Lukas Spögler Regulatory Affairs Oskar-von-Miller Str. 6 Mammendorf, Bavaria 82291 Germany
Re: K212434
Trade/Device Name: Neuromuscular Transmission Monitor TOF3D Regulation Number: 21 CFR 868.2775 Regulation Name: Electrical Peripheral Nerve Stimulator Regulatory Class: Class II Product Code: KOI Dated: April 6, 2022 Received: April 11, 2022
Dear Lukas Spögler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Todd Courtnev Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212434
Device Name
Neuromuscular Transmission Monitor TOF3D
### Indications for Use (Describe)
The TOF3D is used to objectively monitor the level of neuromuscular transmission by measuring muscle contraction following stimulation. The TOF3D can also be used as a peripheral nerve stimulator (without the objective measuring function) for subjective monitoring. The device is intended for use for adolescents greater than 18 through 21 years of age, and adults.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo featuring a stylized depiction of Leonardo da Vinci's Vitruvian Man, enclosed within a circle and a square. Encircling the top portion of the image is the text "HOMINUM SALUTI MEDICINAM EXERCENS", written in a serif font. Below the image is the acronym "MIPM", written in a bold, sans-serif font.
# 5. 510(k) Summary
The following information is provided in accordance with 21 CFR 807.92 for the Permarket 510(k) Summary:
#### 5.1 Submitter Information
| Company: | Jennifer Rosenheimer<br>Managing Director |
|------------------------|-----------------------------------------------------------|
| | MIPM Mammendorfer Institut für Physik und Medizin<br>GmbH |
| | Oskar-von-Miller Str. 6 |
| | Mammendorf, Bavaria 82291, Germany |
| | Telephone: +49 8145 / 92 09 0 |
| | Fax: +49 8145 / 92 09 33 |
| | quality@mipm.com |
| Contact: | Lukas Spögler |
| | Regulatory Affairs |
| | MIPM Mammendorfer Institut für Physik und Medizin<br>GmbH |
| | Oskar-von-Miller Str. 6 |
| | Mammendorf, Bavaria 82291, Germany |
| | Telephone: +49 8145 / 92 09 0 |
| | Fax: +49 8145 / 92 09 33 |
| | regulatory@mipm.com |
| Date Summary Prepared: | April 27, 2022 |
#### 5.2 Name of the Device
| Trade Name: | Neuromuscular Transmission Monitor TOF3D |
|----------------------|------------------------------------------|
| Common Name: | Neuromuscular Transmission Monitor |
| Classification Name: | Electrical peripheral nerve stimulator |
| Review Panel: | Anesthesiology (AN) |
| Regulation: | 868.2775 |
| Class: | Class II |
| Product Code: | KOI |
#### 5.3 Equivalence Claimed to Predicate Device
The Neuromuscular Transmission Monitor TOF3D is equivalent to the TOF-WATCH SX (K992598), manufactured by ORGANON TEKNIKA CORP..
#### 5.4 Device Description
The Neuromuscular Transmission Monitor TOF3D is capable to monitor the level of neuromuscular transmission (NMT) during surgery of in the intensive care unit by stimulating different nerves and measuring the response of the respective muscles to the stimulation. The different locations for monitoring are for instance the ulnar nerve/adductor pollicis muscle, the posterior tibial nerve/flexor halluces brevis muscle and the facial nerve/orbicularis oculi muscle.
The Neuromuscular Transmission Monitor TOF3D uses acceleromyography (AMG) measurement for recording of evoked muscle responses. The results are shown on the LCD display of the device and shall aid qualified medical staff to maintain the proper level of neuromuscular block and to determine the level of recovery from neuromuscular block.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo with a blue border. Inside the border is a drawing of a human figure in the style of Leonardo da Vinci's Vitruvian Man. The figure is surrounded by a circle and a square. Above the figure, the text "HOMINUM SALUTI MEDICINAM EXERCENS" is written in a circular pattern. Below the figure, the letters "MIPM" are written in blue.
#### ર.ર Indications for Use
The TOF3D is used to objectively monitor the level of neuromuscular transmission by measuring muscle contraction following stimulation. The TOF3D can also be used as a peripheral nerve stimulator (without the objective measuring function) for subjective monitoring. The device is intended for use for adolescents greater than 18 through 21 years of age, and adults.
#### 5.6 Substantial Equivalence Discussion
| Features | TOF3D<br>(subject device) | TOF-WATCH SX<br>(predicate device) |
|-----------------------------------------------|------------------------------|------------------------------------|
| Stimulation patterns | | |
| TOF | Yes | Yes |
| PTC | Yes | Yes |
| 1 Hz | Yes | Yes |
| 0.1 Hz | Yes | Yes |
| DBS | Yes | Yes |
| TET | Yes | Yes |
| Nerve location | No | Yes |
| Maximum simulation voltage | 300V (60mA, 5kΩ) | 300V (60mA, 5kΩ) |
| Stimulation current range constant<br>current | 0 - 60 mA | 0 - 60 mA |
| Stimulation pulse width | Monophasic, 200 μs or 300 μs | Monophasic, 200 μs or 300 μs |
| Surface temperature sensor | Yes | Yes |
The Neuromuscular Transmission Monitor TOF3D shares the same indications for use, device operation, overall technical and functional capabilities with its predicate device TOF-WATCH SX (K992598). Both devices rely on the use of acceleromyograph to measure muscle response to electrical stimulation to objectively monitor the level of neuromuscular transmission. The additional functionality of the TOF-WATCH SX (nerve location device - localize nerves for loco-regional anesthesia) is based on the same technical principle used for the nerve stimulation functionality, therefore additional risks in clinical application are not to be expected.
The Neuromuscular Transmission Monitor TOF3D is substantially equivalent to the predicate device TOF-WATCH SX (K992598).
#### 5.7 Performance Data
### Electrical safety and electromagnetic compatibility (EMC)
The Neuromuscular Transmission Monitor TOF3D complies with the IEC 60601-1, IEC 60601-1-6 and IEC 60601-2-10 standards for safety and performance and the IEC 60601-1-2 standard for EMC-
### Software Verification and Validation Testing
The software for this device was considered as a "moderate2 level of concern. Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices".
### Biocompatibility testing
The biocompatibility evaluation for the Neuromuscular Transmission Monitor TOF3D was conducted in accordance with ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". Based on the results of the biocompatibility evaluation, it can be concluded that the virological safety risks of the TOF3D sensors with accessories are acceptable for use as medical device surface skin contact pars.
### Animal and Clinical Studies
No animal or clinical testing was required to demonstrate the substantial equivalence of this device to its predicate.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for an organization called MIPM. The logo features a stylized drawing of a human figure inside of a square. The human figure is in the style of Leonardo da Vinci's Vitruvian Man. The text around the figure reads "HOMINUM SALUTI MEDICINAM EXERCENS".
#### Conclusion 5.8
The subject device has the same indications for use, technological characteristics, and design and operating principles as the predicate device. The identified differences do not raise different questions of safety and effectiveness. The performance of the device is supported by non-clinical testing and risk management activities. The Neuromuscular Transmission Monitor TOF3D is substantially equivalent to the TOF-WATCH SX, cleared under K992598.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.