AMSCO 400 Small Steam Sterilizers, AMSCO 400 Medium Steam Sterilizers
K212424 · STERIS Corporation · FLE · Oct 28, 2021 · General Hospital
Device Facts
Record ID
K212424
Device Name
AMSCO 400 Small Steam Sterilizers, AMSCO 400 Medium Steam Sterilizers
Applicant
STERIS Corporation
Product Code
FLE · General Hospital
Decision Date
Oct 28, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.6880
Device Class
Class 2
Indications for Use
The Amsco 400 Small Steam Sterilizer models 16, 16C, 16CS, 16S, 20, 20C, 20CS, and 20S are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations: - Prevacuum (Models 16, 16C and 20, 20C) – are equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. - Steam Flush Pressure-Pulse (SFPP) (Models 16CS, 16S and 20CS, 20S) are equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. The AMSCO 400 Medium Steam Sterilizers (Models 39SL and 39CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration: - Prevacuum equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. The AMSCO 400 Medium Steam Sterilizers (Models 36H, 48H, 36SL, 48SL, 60SL, 36CH, 48CH, 60CH, 36CSL, 48CSL and 60CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration: - Prevacuum - equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
Device Story
Steam sterilizer for healthcare facilities; processes heat/moisture-stable materials. Inputs: steam, pressure, temperature, time. Operation: utilizes Prevacuum, Gravity, Steam Flush Pressure-Pulse (SFPP), Liquid, Leak Test, and DART (Bowie-Dick) cycles. Outputs: sterilized instruments/fabric packs. Used in clinical settings; operated by healthcare staff. Device replaces obsolete control chip; otherwise identical to predicate. Output confirms sterilization completion; ensures patient safety by reducing infection risk.
Clinical Evidence
Bench testing only. Software validation performed per IEC 62304; all safety and functional tests passed.
Technological Characteristics
Steam sterilizer; chamber sizes range from 16"x16"x26" to 26"x37.5"x60". Features Prevacuum, Gravity, SFPP, Liquid, Leak Test, and DART cycles. Software-controlled via embedded firmware. Sterilization via steam heat.
Indications for Use
Indicated for sterilization of heat and moisture-stable materials in healthcare facilities. Liquid cycles are for non-patient contact use only.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".
October 28, 2021
STERIS Corporation % Anthoy Piotrkowski Director, Regulatory Affairs STERIS Corporation 5960 Heislev Rd Mentor, OH 44060
Re: K212424
Trade/Device Name: AMSCO 400 Small Steam Sterilizers, AMSCO 400 Medium Steam Sterilizers Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE Dated: September 24, 2021 Received: September 27, 2021
## Dear Anthoy Piotrkowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212424/S001 Device Name
Amsco 400 Small Steam Sterilizer
Indications for Use (Describe)
The Amsco 400 Small Steam Sterilizer models 16, 16C, 16CS, 16S, 20, 20CS, and 20S are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:
- Prevacuum (Models 16, 16C and 20, 20C) – are equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
- Steam Flush Pressure-Pulse (SFPP) (Models 16CS, 16S and 20CS, 20S) are equipped with SFPP, . Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
The Amsco 400 Small Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4-1):
Table 4-1.
Amsco 400 Small Prevacuum Steam Sterilizer Models 16, 16, 20, 20C factory-validated sterlization cycles and cycle values
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load |
|-------------------------------|--------------------------|-------------------|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prevac | 270°F<br>(132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to Table 4-3 for recommended<br>quantities. |
| Prevac | 270°F<br>(132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each and Fabric Packs. Refer to Table<br>4-3 for recommended quantities. |
| Prevac | 270°F<br>(132°C) | 4 minutes | 5 minutes | Single Fabric Pack |
| Prevac | 275°F<br>(135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| Prevac<br>(Immediate<br>Use) | 270°F<br>(132°C) | 4 minutes | 1 minutes | Unwrapped instrument tray with a single instrument or<br>multiple unwrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| Gravity | 250°F<br>(121°C) | 30<br>minutes | 15 minutes | Fabric Packs. Refer to Table 4-3 for recommended<br>quantities. |
| Gravity | 270°F<br>(132°C) | 15<br>minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| Gravity | 250°F<br>(121°C) | 30<br>minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| Gravity | 270°F<br>(132°C) | 25<br>minutes | 15 minutes | Fabric Packs. Refer to Table 4-3 for recommended<br>quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F<br>(132°C) | 3 minutes | 1 minutes | Unwrapped instrument tray with a single instrument or<br>multiple unwrapped instrument trays, non-porous load,<br>maximum weight 25 lbs (11.3 kg) each. Refer to<br>Table 4-3 for recommended quantities. |
{3}------------------------------------------------
| Gravity<br>(Immediate<br>Use) | 270°F<br>(132°C) | 10<br>minutes | 1 minutes | Unwrapped instrument tray with a single instrument or<br>multiple unwrapped instrument trays, non-porous load,<br>maximum weight 25 lbs (11.3 kg) each. Refer to<br>Table 4-3 for recommended quantities. |
|-------------------------------|------------------|----------------|-----------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Liquid* | 250°F<br>(121°C) | 45<br>minutes | N/A | Validated with 3 1000ml bottles. Refer to Table 4-4 for<br>recommended full load quantities. |
| DART<br>Warm-Up | 270°F<br>(132°C) | 3 minutes | 1 minute | N/A |
| DART | 270°F<br>(132°C) | 3 ½<br>minutes | 1 minute | Bowie-Dick Test Pack, DART Test Pack |
| Leak Test | N/A | N/A | N/A | N/A |
*The liquid cycle is for non-patient contact use only.
The Amsco 400 Small Steam Flush Pressure-Pulse (SFPP) Sterilizer is equipped with the following factoryprogrammed sterilization cycles and cycle values (Table 4-2):
## Table 4-2.
Amsco 400 Small Steam Flush Pressure-Pulse (SFPP)Sterilizer Models 16S, 16CS, 20S, factory-validated sterilization cycles and cycle values
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load |
|------------------------------|--------------------------|-------------------|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SFPP | 270°F<br>(132°C) | 4<br>minutes | 30 minutes | Double wrapped instrument trays, maximum<br>weight 25 lbs (11.3 kg) each and Fabric Packs.<br>Refer to Table 4-3 for recommended quantities. |
| SFPP | 270°F<br>(132°C) | 4<br>minutes | 20 minutes | Fabric Packs. Refer to Table 4-3 for<br>recommended quantities. |
| SFPP | 275°F<br>(135°C) | 3<br>minutes | 30 minutes | Double wrapped instrument trays, maximum<br>weight 25 lbs (11.3 kg) each. Refer to Table 4-3<br>for recommended quantities. |
| SFPP | 270°F<br>(132°C) | 4<br>minutes | 5 minutes | Single Fabric Pack |
| Prevac | 270°F<br>(132°C) | 4<br>minutes | 20 minutes | Fabric Packs. Refer to Table 4-3 for<br>recommended quantities. |
| Prevac | 270°F<br>(132°C) | 4<br>minutes | 30 minutes | Double wrapped instrument trays, maximum<br>weight 25 lbs (11.3 kg) each and Fabric Packs.<br>Refer to Table 4-3 for recommended quantities. |
| Prevac | 270°F<br>(132°C) | 4<br>minutes | 5 minutes | Single Fabric Pack |
| Prevac | 275°F<br>(135°C) | 3<br>minutes | 30 minutes | Double wrapped instrument trays, maximum<br>weight 25 lbs (11.3 kg) each. Refer to Table 4-3<br>for recommended quantities. |
| Prevac<br>(Immediate<br>Use) | 270°F<br>(132°C) | 4<br>minutes | 1 minutes | Unwrapped instrument tray with a single<br>instrument or multiple unwrapped instrument<br>trays, maximum weight 25 lbs (11.3 kg) each.<br>Refer to Table 4-3 for recommended quantities. |
| Gravity | 250°F<br>(121°C) | 30<br>minutes | 15 minutes | Fabric Packs. Refer to Table 4-3 for<br>recommended quantities. |
| Gravity | 270°F<br>(132°C) | 15<br>minutes | 30 minutes | Double wrapped instrument trays, maximum<br>weight 25 lbs (11.3 kg) each. Refer to Table 4-3<br>for recommended quantities. |
| Gravity | 250°F<br>(121°C) | 30<br>minutes | 30 minutes | Double wrapped instrument trays, maximum<br>weight 25 lbs (11.3 kg) each. Refer to Table 4-3<br>for recommended quantities. |
| Gravity | 270°F<br>(132°C) | 25<br>minutes | 15 minutes | Fabric Packs. Refer to Table 4-3 for<br>recommended quantities. |
{4}------------------------------------------------
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load |
|-------------------------------|--------------------------|-------------------|-----------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Gravity<br>(Immediate<br>Use) | 270°F<br>(132°C) | 3<br>minutes | 1 minutes | Unwrapped instrument tray with a single<br>instrument or multiple unwrapped instrument<br>trays, non-porous load, maximum weight 25 lbs<br>(11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F<br>(132°C) | 10<br>minutes | 1 minutes | Unwrapped instrument tray with a single<br>instrument or multiple unwrapped instrument<br>trays, non-porous load, maximum weight 25 lbs<br>(11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| Liquid* | 250°F<br>(121°C) | 45<br>minutes | N/A | Validated with 3 1000ml bottles. Refer to Table<br>4-4 for recommended full load quantities. |
| DART<br>Warm-Up | 270°F<br>(132°C) | 3<br>minutes | 1 minute | N/A |
| DART | 270°F<br>(132°C) | 3 ½<br>minutes | 1 minute | Bowie-Dick Test Pack, DART Test Pack |
| Leak Test | N/A | N/A | N/A | N/A |
*The liquid cycle is for non-patient contact use only.
The following table lists STERIS's recommended loads by sterilizer size:
| Sterilizer Size | Wrapped Instrument Trays | Fabric Packs |
|-----------------|--------------------------|--------------|
| 16" x 16" x 26" | 2 | 4 |
| 20" x 20" x 38" | 3 | 6 |
The following table is a guideline for liquid cycle processing:
Table 4-4 Amsco 400 Small Steam Sterilizer Liquid* Cycle Guideline
| Sterilizer Size | Number of Containers for full load | Number of Containers for validation | Volume of Liquid In One Container | Minimum Recommended Sterilize Time at 250°F (121°C) |
|-----------------|------------------------------------|-------------------------------------|-----------------------------------|-----------------------------------------------------|
| 16" x 16" x 26" | 15 | 3 | 1000 ml | 45 minutes |
| 20" x 20" x 38" | 32 | 3 | 1000 ml | 45 minutes |
*The liquid cycle is for non-patient contact use only.
The Amsco 400 Small Steam Sterilizer is offered in the following chamber sizes:
- 16" x 16" x 26" (Models 16, 16C, 16CS, and 16S) ●
- 20" x 20" x 38" (Models 20, 20C, 20CS, and 20S) .
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
X
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect
{5}------------------------------------------------
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
PSC Publishing Services (301) 443-6740 EF
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212424/S001 Device Name
AMSCO 400 Medium Steam Sterilizers (Models 39SL and 39CSL)
Indications for Use (Describe)
The AMSCO 400 Medium Steam Sterilizers (Models 39SL and 39CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration:
- Prevacuum equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. ●
The AMSCO 400 Medium Prevacuum Steam Sterilizers are equipped with the following factoryprogrammed sterilization cycles and cycle values (Table 4.5):
## Table 4.5.
AMSCO 400 Medium Prevacuum Steam Sterilizer factory-validated sterilization cycles and cycle values
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load |
|-------------------------------|--------------------------|-------------------|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prevac | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to Table 4.6 for recommended<br>quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each and Fabric Packs. Refer to Table<br>4.6 for recommended quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack |
| Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4.6 for recommended<br>quantities. |
| Prevac<br>(Immediate<br>Use) | 270°F (132°C) | 4 minutes | 1 minutes | Unwrapped instrument tray with a single instrument or<br>multiple unwrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each. Refer to Table A-2 for<br>recommended quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 15 minutes | Fabric Packs. Refer to Table 4.6 for recommended<br>quantities. |
| Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4.6 for recommended<br>quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4.6 for recommended<br>quantities. |
| Gravity | 270°F (132°C) | 25 minutes | 15 minutes | Fabric Packs. Refer to Table 4.6 for recommended<br>quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 3 minutes | 1 minutes | Unwrapped instrument tray with a single instrument or<br>multiple unwrapped instrument trays, non-porous load,<br>maximum weight 25 lbs (11.3 kg) each. Refer to Table<br>4.6 for recommended quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 10 minutes | 1 minutes | Unwrapped instrument tray with a single instrument or<br>multiple unwrapped instrument trays, non-porous load,<br>maximum weight 25 lbs (11.3 kg) each. Refer to Table<br>4.6 for recommended quantities. |
| Liquid* | 250°F (121°C) | 45 minutes | N/A | Validated with 3 1000ml bottles. Refer to Table 4.7 for<br>recommended full load quantities. |
| DART<br>Warm-Up | 270°F (132°C) | 3 minutes | 1 minute | N/A |
| Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Recommended Load |
| DART | 270°F (132°C) | 3 ½ minutes | 1 minute | Bowie-Dick Test Pack, DART Test Pack |
| Leak Test | N/A | N/A | N/A | N/A |
{7}------------------------------------------------
*The liquid cycle is for non-patient contact use only.
The following table lists STERIS' full loads by sterilizer size:
## Table 4.6
### AMSCO 400 Medium Steam Sterilizer full load per sterilizer size
| Sterilizer Size | Wrapped Instrument Trays | Fabric Packs |
|-----------------|--------------------------|--------------|
| 26" x 26" x 39" | 4 | 12 |
The following table is a guideline for liquid cycle processing:
Table 4.7 AMSCO 400 Medium Steam Sterilizer Liquid Cycle Guideline
| Sterilizer Size | Number of Containers for full load | Number of Containers for validation | Volume of Liquid In One Container | Minimum Recommended Sterilize Time at 250°F (121°C) |
|-----------------|------------------------------------|-------------------------------------|-----------------------------------|-----------------------------------------------------|
| 26" x 26" x 39" | 80 | 3 | 1000 ml | 45 minutes |
*The liquid cycle is for non-patient contact use only.
The AMSCO 400 Medium Steam Sterilizer is offered in the following chamber sizes:
- 26" x 26" x 39" (Models 39SL and 39CSL) ●
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
BF SC Publishing Services (301) 443-6740
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212424/S001 Device Name
AMSCO 400 Medium Steam Sterilizers (Models 36H, 48H, 60H, 36SL, 48SL, 60CH, 36CSL, 48CSL and 60CSL)
Indications for Use (Describe)
The AMSCO 400 Medium Steam Sterilizers (Models 36H, 48H, 36SL, 48SL, 60SL, 36CH, 48CH, 60CH, 36CSL, 48CSL and 60CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration:
- Prevacuum - equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles
The Amsco 400 Small Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4.8):
## Table 4.8.
Amsco 400 Medium Prevacuum Steam Sterilizer factory-validated sterilization cycles and cycle values
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load |
|-----------------|--------------------------|-------------------|------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Prevac | 270°F<br>(132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to Table 4.9 for recommended<br>quantities. |
| Prevac | 270°F<br>(132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each and Fabric Packs. Refer to Table 4.9<br>for recommended quantities. |
| Prevac | 270°F<br>(132°C) | 4 minutes | 5 minutes | Single Fabric Pack |
| Prevac | 275°F<br>(135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4.9 for recommended<br>quantities. |
| Gravity | 250°F<br>(121°C) | 30<br>minutes | 15 minutes | Fabric Packs. Refer to Table 4.9 for recommended<br>quantities. |
| Gravity | 270°F<br>(132°C) | 15<br>minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4.9 for recommended<br>quantities. |
| Gravity | 250°F<br>(121°C) | 30<br>minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4.9 for recommended<br>quantities. |
| Gravity | 270°F<br>(132°C) | 25<br>minutes | 15 minutes | Fabric Packs. Refer to Table 4.9 for recommended<br>quantities. |
| Liquid* | 250°F<br>(121°C) | 45<br>minutes | N/A | Validated with 3 1000ml bottles. Refer to Table 4.10 for<br>recommended full load quantities. |
| DART<br>Warm-Up | 270°F<br>(132°C) | 3 minutes | 1 minute | N/A |
| DART | 270°F<br>(132°C) | 3 ½<br>minutes | 1 minute | Bowie-Dick Test Pack, DART Test Pack |
| Leak Test | N/A | N/A | N/A | N/A |
* The liquid cycle is for non-patient contact use only.
The following table lists STERIS's recommended loads by sterilizer size:
| Table 4.9 | | | | | | |
|-------------------------------------------------------------------------|--|--|--|--|--|--|
| Amsco 400 Medium Steam Sterilizer recommended loads per sterilizer size | | | | | | |
| Sterilizer Size | Wrapped Instrument Trays | Fabric Packs |
|-------------------|--------------------------|--------------|
| 26" x 37.5" x 36" | 9 | 18 |
{9}------------------------------------------------
| 26" x 37.5" x 48" | 12 | 30 |
|-------------------|----|----|
| 26" x 37.5" x 60" | 15 | 36 |
The following table is a guideline for liquid cycle processing:
## Table 4.10 AMSCO 400 Medium Steam Sterilizer Liquid Cycle Guideline
| Sterilizer Size | Number of Containers for full load | Number of Containers for validation | Volume of Liquid In One Container | Minimum Recommended Sterilize Time at 250°F (121°C) |
|-------------------|------------------------------------|-------------------------------------|-----------------------------------|-----------------------------------------------------|
| 26" x 37.5" x 36" | 112 | 3 | 1000 ml | 45 minutes |
| 26" x 37.5" x 48" | 154 | 3 | 1000 ml | 45 minutes |
| 26" x 37.5" x 60" | 196 | 3 | 1000 ml | 45 minutes |
*The liquid cycle is for non-patient contact use only.
The AMSCO 400 Medium Steam Sterilizer is offered in the following chamber sizes:
- 26" x 37.5" x 36" (Models 36H, 36CH, 36CH, 36CSL, 36CSL) ●
- 26" x 37.5" x 48" (Models 48H, 48CH, 48CH, 48CSL) ●
- 26" x 37.5" x 60" (Models 60H, 48CH, 48CH, 48CSL) ●
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
вн PSC Publishing Services (301) 443-6740
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for STERIS. The logo consists of the word "STERIS" in large, bold, sans-serif font, with the registered trademark symbol to the right of the word. Below the word is a graphic of several horizontal, wavy lines in blue, resembling water waves.
# K212424/S001 510(k) Summary For AMSCO 400 Small Steam Sterilizers
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Contact: Tony Piotrkowski Director, Regulatory Affairs
Telephone: (440) 392-7437 Email:
Summary Date: August 3, 2021
{11}------------------------------------------------
### Device Name 1.
| Trade Name: | AMSCO 400 Small Steam Sterilizer (cleared as<br>AMSCO Chimeron Small Steam Sterilizer) |
|------------------------|----------------------------------------------------------------------------------------|
| Device Class: | Class II |
| Common/usual Name: | Steam Sterilizer |
| Classification Name: | Sterilizer, Steam |
| Classification Number: | 21 CFR 880.6880 |
| Product Code: | FLE |
### 2. Predicate Devices
K111223 - AMSCO Chimeron Small Steam Sterilizer (renamed AMSCO 400 before commercialization)
The proposed device is identical to the predicate except for the modification to replace an obsolete control chip.
### 3. Description of Devices
The AMSCO 400 Steam Sterilizers are the same as the predicate devices except for the specific modifications described in this submission. There are no differences in the indications for use or technology (including features and principles of operation) between the proposed and predicate devices. Therefore, the differences between the proposed and predicate devices are limited to the described modifications, and these proposed changes raise no new concerns of safety and effectiveness when compared to the predicate device.
### 4. Intended Use / Indications for Use
The Amsco 400 Small Steam Sterilizer models 16, 16CS, 16CS, 16S, 20, 20C, 20CS, and 20S are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:
- Prevacuum (Models 16, 16C and 20, 20C) are equipped with Prevacuum, ● Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
- . Steam Flush Pressure-Pulse (SFPP) (Models 16CS, 16S and 20CS, 20S) - are equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
{12}------------------------------------------------
The Amsco 400 Small Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4-1):
# Table 4-1.
Amsco 400 Small Prevacuum Steam Sterilizer Models 16, 16, 20, 20C factory-validated sterilization cycles and cycle values
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load |
|-------------------------------|--------------------------|-------------------|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prevac | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to Table 4-3 for recommended<br>quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each and Fabric Packs. Refer to Table 4-3<br>for recommended quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack |
| Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4-3 for recommended<br>quantities. |
| Prevac<br>(Immediate<br>Use) | 270°F (132°C) | 4 minutes | 1 minutes | Unwrapped instrument tray with a single instrument or<br>multiple unwrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 15 minutes | Fabric Packs. Refer to Table 4-3 for recommended<br>quantities. |
| Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4-3 for recommended<br>quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25<br>lbs (11.3 kg) each. Refer to Table 4-3 for recommended<br>quantities. |
| Gravity | 270°F (132°C) | 25 minutes | 15 minutes | Fabric Packs. Refer to Table 4-3 for recommended<br>quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 3 minutes | 1 minutes | Unwrapped instrument tray with a single instrument or<br>multiple unwrapped instrument trays, non-porous load,<br>maximum weight 25 lbs (11.3 kg) each. Refer to Table<br>4-3 for recommended quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 10 minutes | 1 minutes | Unwrapped instrument tray with a single instrument or<br>multiple unwrapped instrument trays, non-porous load,<br>maximum weight 25 lbs (11.3 kg) each. Refer to Table<br>4-3 for recommended quantities. |
| Liquid* | 250°F (121°C) | 45 minutes | N/A | Validated with 3 1000ml bottles. Refer to Table 4-4 for<br>recommended full load quantities. |
| DART<br>Warm-Up | 270°F (132°C) | 3 minutes | 1 minute | N/A |
| DART | 270°F (132°C) | 3 ½<br>minutes | 1 minute | Bowie-Dick Test Pack, DART Test Pack |
| Leak Test | N/A | N/A | N/A | N/A |
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load |
| SFPP | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each and Fabric Packs. Refer to<br>Table 4-3 for recommended quantities. |
| SFPP | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to Table 4-3 for recommended<br>quantities. |
| SFPP | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| SFPP | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack |
| Prevac | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to Table 4-3 for recommended<br>quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each and Fabric Packs. Refer to<br>Table 4-3 for recommended quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack |
| Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| Prevac<br>(Immediate<br>Use) | 270°F (132°C) | 4 minutes | 1 minutes | Unwrapped instrument tray with a single instrument<br>or multiple unwrapped instrument trays, maximum<br>weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| Gravity | 250°F (121°C) | 30<br>minutes | 15 minutes | Fabric Packs. Refer to Table 4-3 for recommended<br>quantities. |
| Gravity | 270°F (132°C) | 15<br>minutes | 30 minutes | Double wrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| Gravity | 250°F (121°C) | 30<br>minutes | 30 minutes | Double wrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each. Refer to Table 4-3 for<br>recommended quantities. |
| Gravity | 270°F (132°C) | 25<br>minutes | 15 minutes | Fabric Packs. Refer to Table 4-3 for recommended<br>quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 3 minutes | 1 minutes | Unwrapped instrument tray with a single instrument<br>or multiple unwrapped instrument trays, non-porous<br>load, maximum weight 25 lbs (11.3 kg) each. Refer<br>to Table 4-3 for recommended quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 10<br>minutes | 1 minutes | Unwrapped instrument tray with a single instrument<br>or multiple unwrapped instrument trays, non-porous<br>load, maximum weight 25 lbs (11.3 kg) each. Refer<br>to Table 4-3 for recommended quantities. |
| Liquid* | 250°F (121°C) | 45<br>minutes | N/A | Validated with 3 1000ml bottles. Refer to Table 4-4<br>for recommended full load quantities. |
| Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Recommended Load |
| DART<br>Warm-Up | 270°F (132°C) | 3 minutes | 1 minute | N/A |
| DART | 270°F (132°C) | 3 ½ minutes | 1 minute | Bowie-Dick Test Pack, DART Test Pack |
| Leak Test | N/A | N/A | N/A | N/A |
*The liquid cycle is for non-patient contact use only.
{13}------------------------------------------------
The Amsco 400 Small Steam Flush Pressure-Pulse (SFPP) Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4-2):
# Table 4-2.
Amsco 400 Small Steam Flush Pressure-Pulse (SFPP)Sterilizer Models 16S, 16CS, 20S, 20CS factory-validated sterilization cycles and cycle values
{14}------------------------------------------------
*The liquid cycle is for non-patient contact use only.
The following table lists STERIS's recommended loads by sterilizer size:
| Table 4-3 | |
|-----------|--|
|-----------|--|
Amsco 400 Small Steam Sterilizer recommended loads per sterilizer size
| Sterilizer Size | Wrapped Instrument Trays | Fabric Packs |
|-----------------|--------------------------|--------------|
| 16" x 16" x 26" | 2 | 4 |
| 20" x 20" x 38" | 3 | 6 |
The following table is a guideline for liquid cycle processing:
| Sterilizer Size | Number of Containers<br>for full load | Number of Containers<br>for validation | Volume of Liquid<br>In One Container | Minimum Recommended<br>Sterilize Time at 250°F<br>(121°C) |
|-----------------|---------------------------------------|----------------------------------------|--------------------------------------|-----------------------------------------------------------|
| 16" x 16" x 26" | 15 | 3 | 1000 ml | 45 minutes |
| 20" x 20" x 38" | 32 | 3 | 1000 ml | 45 minutes |
Table 4-4 Amsco 400 Small Steam Sterilizer Liquid* Cycle Guideline
*The liquid cycle is for non-patient contact use only.
The Amsco 400 Small Steam Sterilizer is offered in the following chamber sizes:
- 16" x 16" x 26" (Models 16, 16C, 16CS, and 16S) ●
- 20" x 20" x 38" (Models 20, 20C, 20CS, and 20S) ●
#### 5. Summary of Nonclinical Tests
The modified sterilizers have the same indications for use, intended use and same technological characteristics as the predicate devices. Performance testing to demonstrate substantial equivalence to the predicate, using the same methods as were used to validate the original design, is summarized below.
| Test | Result | Conclusion |
|-----------------------------------|--------------------------------------|------------|
| Software validation per IEC 62304 | All safety and functional tests pass | PASS |
{15}------------------------------------------------
### 6. Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the AMSCO 400 Small Steam Sterilizer is as safe, as effective and performs as well or better than the legally marketed predicate device K111223, Class II (21 CFR 880.6880), product code FLE.
{16}------------------------------------------------
# K212424/S001 510(k) Summary For AMSCO 400 Medium 26" x 26" Steam Sterilizers
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Contact: Tony Piotrkowski Director, Regulatory Affairs
Telephone: (440) 392-7437 tony_piotrkowski@steris.com Email:
Summary Date: August 3, 2021
{17}------------------------------------------------
### Device Name 6.
| Trade Name: | AMSCO 400 Medium26" x 26" Steam Sterilizer |
|------------------------|--------------------------------------------|
| Device Class: | Class II |
| Common/usual Name: | Steam Sterilizer |
| Classification Name: | Sterilizer, Steam |
| Classification Number: | 21 CFR 880.6880 |
| Product Code: | FLE |
### 7. Predicate Devices
K112403 - Chimeron Medium 26" x 26" Steam Sterilizer (renamed AMSCO 400 before commercialization)
The proposed device is identical to the predicate except for the modification to replace an obsolete control chip.
### 8. Description of Devices
The AMSCO 400 Steam Sterilizers are the same as the predicate devices except for the specific modifications described in this submission. There are no differences in the indications for use or technology (including features and principles of operation) between the proposed and predicate devices. Therefore, the differences between the proposed and predicate devices are limited to the described modifications, and these proposed changes raise no new concerns of safety and effectiveness when compared to the predicate device.
### 9. Intended Use / Indications for Use
The AMSCO 400 Medium Steam Sterilizers (Models 39SL and 39CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration:
- Prevacuum equipped with Prevacuum, Gravity, Liquid, Leak Test and DART ● (Bowie-Dick) cycles.
The AMSCO 400 Medium Prevacuum Steam Sterilizers are equipped with the following factory-programmed sterilization cycles and cycle values (Table 4.5):
{18}------------------------------------------------
# Table 4.5.
AMSCO 400 Medium Prevacuum Steam Sterilizer factory-validated sterilization cycles and cycle values
| AMSCO 400 Medium Prevacuum Steam Sterilizer factory-validated sterilization cycles and cycle values | | | | |
|-----------------------------------------------------------------------------------------------------|--------------------------|-------------------|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load |
| Prevac | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to Table 4.6 for recommended<br>quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each and Fabric Packs. Refer to<br>Table 4.6 for recommended quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack |
| Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each. Refer to Table 4.6 for<br>recommended quantities. |
| Prevac<br>(Immediate<br>Use) | 270°F (132°C) | 4 minutes | 1 minutes | Unwrapped instrument tray with a single instrument<br>or multiple unwrapped instrument trays, maximum<br>weight 25 lbs (11.3 kg) each. Refer to Table A-2 for<br>recommended quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 15 minutes | Fabric Packs. Refer to Table 4.6 for recommended<br>quantities. |
| Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each. Refer to Table 4.6 for<br>recommended quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Double wrapped instrument trays, maximum weight<br>25 lbs (11.3 kg) each. Refer to Table 4.6 for<br>recommended quantities. |
| Gravity | 270°F (132°C) | 25 minutes | 15 minutes | Fabric Packs. Refer to Table 4.6 for recommended<br>quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 3 minutes | 1 minutes | Unwrapped instrument tray with a single instrument<br>or multiple unwrapped instrument trays, non-porous<br>load, maximum weight 25 lbs (11.3 kg) each. Refer to<br>Table 4.6 for recommended quantities. |
| Gravity<br>(Immediate<br>Use) | 270°F (132°C) | 10 minutes | 1 minutes | Unwrapped instrument tray with a single instrument<br>or multiple unwrapped instrument trays, non-porous<br>load, maximum weight 25 lbs (11.3 kg) each. Refer to<br>Table 4.6 for recommended quantities. |
| Liquid* | 250°F (121°C) | 45 minutes | N/A | Validated with 3 1000ml bottles. Refer to Table 4.7 for<br>recommended full load quantities. |
| DART<br>Warm-Up | 270°F (132°C) | 3 minutes | 1 minute | N/A |
| DART | 270°F (132°C) | 3 ½ minutes | 1 minute | Bowie-Dick Test Pack, DART Test Pack |
| Leak Test | N/A | N/A | N/A | N/A |
*The liquid cycle is for non-patient contact use only.
The following table lists STERIS' full loads by sterilizer size:
| Table 4.6 | | |
|-----------------------------------------------------------------|--------------------------|--------------|
| AMSCO 400 Medium Steam Sterilizer full load per sterilizer size | | |
| Sterilizer Size | Wrapped Instrument Trays | Fabric Packs |
| 26" x 26" x 39" | 4 | 12 |
{19}------------------------------------------------
The following table is a guideline for liquid cycle processing:
# Table 4.7 AMSCO 400 Medium Steam Sterilizer Liquid Cycle Guideline
| Sterilizer Size | Number of<br>Containers<br>for full<br>load | Number of<br>Containers<br>for validation | Volume of<br>Liquid In One<br>Container | Minimum<br>Recommended<br>Sterilize Time at<br>250°F (121°C) |
|-----------------|---------------------------------------------|-------------------------------------------|-----------------------------------------|--------------------------------------------------------------|
| 26" x 26" x 39" | 80 | 3 | 1000 ml | 45 minutes |
*The liquid cycle is for non-patient contact use only.
The AMSCO 400 Medium Steam Sterilizer is offered in the following chamber sizes:
- 26' x 26' x 39" (Models 39SL and 39CSL) ●
### 10. Summary of Nonclinical Tests
The modified sterilizers have the same indications for use, intended use and same technological characteristics as the predicate devices. Performance testing to demonstrate substantial equivalence to the predicate, using the same methods as were used to validate the original design, is summarized below.
| Test | Result | Conclusion |
|-----------------------------------|--------------------------------------|------------|
| Software validation per IEC 62304 | All safety and functional tests pass | PASS |
### 6. Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the AMSCO 400 Medium 26" x 26" Steam Sterilizer is as safe, as effective and performs as well or better than the legally marketed predicate device K112403, Class II (21 CFR 880.6880), product code FLE.
{20}------------------------------------------------
# K212424/S001 510(k) Summary For AMSCO 400 Medium Steam Sterilizers
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Contact: Tony Piotrkowski Director, Regulatory Affairs
Telephone: (440) 392-7437 Email: tony_piotrkowski@steris.com
Summary Date: August 3, 2021
{21}------------------------------------------------
### Device Name 11.
| Trade Name: | AMSCO 400 Medium Steam Sterilizer |
|------------------------|-----------------------------------|
| Device Class: | Class II |
| Common/usual Name: | Steam Sterilizer |
| Classification Name: | Sterilizer, Steam |
| Classification Number: | 21 CFR 880.6880 |
| Product Code: | FLE |
### 12. Predicate Devices
K112055 - AMSCO 400 Medium Steam Sterilizer (renamed AMSCO 400 before commercialization)
The proposed device is identical to the predicate except for the modification to replace an obsolete control chip.
### 13. Description of Devices
The AMSCO 400 Steam Sterilizers are the same as the predicate devices except for the specific modifications described in this submission. There are no differences in the indications for use or technology (including features and principles of operation) between the proposed and predicate devices. Therefore, the differences between the proposed and predicate devices are limited to the described modifications, and these proposed changes raise no new concerns of safety and effectiveness when compared to the predicate device.
### 14. Intended Use / Indications for Use
The AMSCO 400 Medium Steam Sterilizers (Models 36H, 48H, 60H, 36SL, 48SL, 60SL, 36CH, 48CH, 60CH, 36CSL, 48CSL and 60CSL) are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in the following configuration:
- Prevacuum equipped with Prevacuum, Gravity, Liquid, Leak Test and DART ● (Bowie-Dick) cycles
The Amsco 400 Small Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4.8):
{22}------------------------------------------------
## Table 4.8.
Amsco 400 Medium Prevacuum Steam Sterilizer factory-validated sterilization cycles and cvcle values
| Cycles | Sterilize<br>Temperature | Sterilize<br>Time | Dry Time | Recommended Load |
|-----------------|--------------------------|-------------------|------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Prevac | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to Table 4.9 for recommended quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4.9 for recommended quantities. |
| Prevac | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack |
| Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4.9 for recommended quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 15 minutes | Fabric Packs. Refer to Table 4.9 for recommended quantities. |
| Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4.9 for recommended quantities. |
| Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4.9 for recommended quantities. |
| Gravity | 270°F (132°C) | 25 minutes | 15 minutes | Fabric Packs. Refer to Table 4.9 for recommended quantities. |
| Liquid* | 250°F (121°C) | 45 minutes | N/A | Validated with 3 1000ml bottles. Refer to Table 4.10 for recommended full load quantities. |
| DART<br>Warm-Up | 270°F (132°C) | 3 minutes | 1 minute | N/A |
| DART | 270°F (132°C) | 3 ½ minutes | 1 minute | Bowie-Dick Test Pack, DART Test Pack |
| Leak Test | N/A | N/A | N/A | N/A |
*The liquid cycle is for non-patient contact use only.
The following table lists STERIS's recommended loads by sterilizer size:
# Table 4.9
Amsco 400 Medium Steam Sterilizer recommended loads per sterilizer size
| Sterilizer Size | Wrapped Instrument Trays | Fabric Packs |
|-------------------|--------------------------|--------------|
| 26" x 37.5" x 36" | 9 | 18 |
| 26" x 37.5" x 48" | 12 | 30 |
| 26" x 37.5" x 60" | 15 | 36 |
The following table is a guideline for liquid cycle processing:
{23}------------------------------------------------
## Table 4.10
## AMSCO 400 Medium Steam Sterilizer Liquid Cycle Guideline
| Sterilizer Size | Number of Containers for full load | Number of Containers for validation | Volume of Liquid In One Container | Minimum Recommended Sterilize Time at 250°F (121°C) |
|-------------------|------------------------------------|-------------------------------------|-----------------------------------|-----------------------------------------------------|
| 26" x 37.5" x 36" | 112 | 3 | 1000 ml | 45 minutes |
| 26" x 37.5" x 48" | 154 | 3 | 1000 ml | 45 minutes |
| 26" x 37.5" x 60" | 196 | 3 | 1000 ml | 45 minutes |
*The liquid cycle is for non-patient contact use only.
The AMSCO 400 Medium Steam Sterilizer is offered in the following chamber sizes:
- 26" x 37.5" x 36" (Models 36H, 36CH, 36SL, 36CSL)
- 26" x 37.5" x 48" (Models 48H, 48CH, 48CH, 48CSL) ●
- 26" x 37.5" x 60" (Models 60H, 48CH, 48CH, 48CSL)
#### 15. Summary of Nonclinical Tests
The modified sterilizers have the same indications for use, intended use and same technological characteristics as the predicate devices. Performance testing to demonstrate substantial equivalence to the predicate, using the same methods as were used to validate the original design, is summarized below.
| Test | Result | Conclusion |
|-----------------------------------|--------------------------------------|------------|
| Software validation per IEC 62304 | All safety and functional tests pass | PASS |
#### 6. Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the AMSCO 400 Medium Steam Sterilizer is as safe, as effective and performs as well or better than the legally marketed predicate device K112055, Class II (21 CFR 880.6880), product code FLE.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.