The Ulspira TS Nitric Oxide Therapy System is intended for use by healthcare professionals for the delivery of nitric oxide (NO) and the monitoring of inspired NO, NO2 and O2 concentrations for a patient undergoing inhaled Nitric Oxide (iNO) therapy. The Ulspira TS must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Refer to this material prior to use. The Ulspira TS primary NO therapy system delivers NO gas in the 0-80 ppm range while in the Constant Rate or flow sensing modes. This includes: • Continuous integrated monitoring for inspired NO, NO2 and a comprehensive alarm system. • A touch-screen user interface with a waveform display of the ventilation device flow as measured in the inspiratory limb. • Cylinder handling facilitated by manual or an automatic cylinder switch which is reactive to the detected gas supply state of NO cylinders, and a low O2 pressure alarm when using an oxygen cylinder. • An automated emergency dosing activated by certain high-risk alarms, which impact patient dosing, to avoid sudden cessation of therapy. • Compatibility with a wide inspiratory flow rate range of 0.25-120 1/min, utilizing an automatically detected low or high flow sensor. • An internal battery which provides at least two hours of uninterrupted therapy and a 12V DC inlet for additional external battery access. The integrated Ulspira TS pneumatic backup NO therapy system provides backup NO delivery capability that is intended to deliver a continuous flow of NO mixed with O2, for iNO therapy which allows continuous treatment during transit within hospitals.
Device Story
System delivers physician-prescribed nitric oxide (NO) gas into respiratory device circuits; monitors inspired NO, NO2, and O2 concentrations. Inputs include gas supply from cylinders and inspiratory flow data from ventilators. Operates via primary delivery system (Constant Rate or flow-sensing modes) and integrated pneumatic backup system. Features touch-screen UI with waveform display, automated cylinder switching, and emergency dosing algorithm triggered by high-risk alarms to prevent therapy cessation. Used by healthcare professionals in NICU and transport settings. Output includes controlled NO delivery and real-time gas monitoring/alarm data. Benefits include continuous, regulated iNO therapy and safety features to maintain dosing during system failures or transport.
Clinical Evidence
Bench testing only. Validation included system-level testing, regression testing, comparative waveform testing, software verification (code review, static analysis, unit/integration tests), biocompatibility (ISO 10993 series, ISO 18562 series), reprocessing validation, and human factors/usability engineering. Compliance with standards including ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, ISO 80601-2-12, and ISO 80601-2-55 confirmed performance requirements.
Technological Characteristics
Class II device; 21 CFR 868.5165. Materials meet ISO 10993 and ISO 18562 standards for breathing gas pathways. Features primary delivery system and pneumatic backup. Connectivity includes 12V DC inlet. Gas connectors per CGA V-5:2008 and CGA V-1:2013. Software-controlled with touch-screen UI, automated emergency dosing, and flow-sensing modes. Battery-powered (min 2 hours).
Indications for Use
Indicated for term and near-term (>34 weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. Used in NICU and during neonatal transport.
Regulatory Classification
Identification
The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.
Special Controls
The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”
The special control developed by the agency is a guidance document, entitled Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. This guidance document identifies the risks associated with these types of devices and contains information that will help manufacturers address those risks. This document is available on FDA's website at http://www.fda.gov/cdrh/ode/1157.pdf.
AeroNOx 2.0 Nitric Oxide Titration & Monitoring System (K193481)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). The logo includes the Department of Health & Human Services symbol on the left. To the right of the symbol is the FDA acronym in a blue box, followed by the words "U.S. Food & Drug Administration" in blue text.
June 30, 2023
Airgas Therapeutics Steve Miller V.P. Regulatory and Compliance 12800 West Little York Road Houston, TX 77041
Re: K212409
Trade/Device Name: ULSPIRA TS Nitric Oxide Therapy System Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric oxide administration apparatus Regulatory Class: Class II Product Code: MRN, MRO, MRP, MRO, CCL Dated: January 30, 2023 Received: January 31, 2023
Dear Steve Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee -S
James J. Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212409
Device Name Ulspira TS Nitric Oxide Therapy System
## Indications for Use (Describe)
The Ulspira TS Nitric Oxide Therapy System is intended for use by healthcare professionals for the delivery of nitric oxide (NO) and the monitoring of inspired NO, NO2 and O2 concentrations for a patient undergoing inhaled Nitric Oxide (iNO) therapy.
The Ulspira TS must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Refer to this material prior to use.
The Ulspira TS primary NO therapy system delivers NO gas in the 0-80 ppm range while in the Constant Rate or flow sensing modes. This includes:
• Continuous integrated monitoring for inspired NO, NO2 and a comprehensive alarm system.
· A touch-screen user interface with a waveform display of the ventilation device flow as measured in the inspiratory limb.
· Cylinder handling facilitated by manual or an automatic cylinder switch which is reactive to the detected gas supply state of NO cylinders, and a low O2 pressure alarm when using an oxygen cylinder.
• An automated emergency dosing activated by certain high-risk alarms, which impact patient dosing, to avoid sudden cessation of therapy.
· Compatibility with a wide inspiratory flow rate range of 0.25-120 1/min, utilizing an automatically detected low or high flow sensor.
• An internal battery which provides at least two hours of uninterrupted therapy and a 12V DC inlet for additional external battery access.
The integrated Ulspira TS pneumatic backup NO therapy system provides backup NO delivery capability that is intended to deliver a continuous flow of NO mixed with O2, for iNO therapy which allows continuous treatment during transit within hospitals.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
The following summary is provided in accordance with 21 CFR 807.92:
# A. DATE
May 31, 2023
# B. SUBMITTER / SPONSOR
Airgas Therapeutics 12800 West Little York Road Houston, TX 77041
- Contact Person: Steve Miller Airgas Therapeutics V.P. Regulatory and Compliance Phone (713) 896-2280
# C. DEVICE
| Trade Name of Device: | ULSPIRA TS Nitric Oxide Therapy System |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Nitric Oxide Administration Apparatus (Primary)<br>Nitric Oxide Administration Apparatus (Backup)<br>Nitric Oxide Analyzer<br>Nitrogen Dioxide Analyzer<br>Oxygen Gas Analyzer |
| Classification: | Class II- 21 CFR 868.5165 |
| Classification Name: | Nitric Oxide Administration Apparatus |
{4}------------------------------------------------
MRN (Primary), MRO, MRP, MRQ, CCL Product Codes:
- D. PREDICATE DEVICE K200389, INOmax DSIR Plus
- E. REFERENCE DEVICE(S) K171696, NOxBOXi Nitric Oxide Delivery System K193481, AeroNOx 2.0 Nitric Oxide Titration & Monitoring System
### F. DEVICE DESCRIPTION
Ulspira TS Nitric Oxide Therapy System delivers physician-prescribed NO therapy gas and monitors inspired Nitric Oxide, Nitrogen Dioxide, and Oxygen gas in combination with a respiratory device.
The main device functionalities of the Ulspira TS Nitric Oxide Therapy System include:
- A primary delivery system to administer NO gas into a respiratory device circuit.
- Monitoring of NO, NO2, and O2 gas concentrations close to the patient interface. ।
- System includes a user interface that contains all controls used to set the NO delivery and monitoring parameters. All set i parameters as well as other information are shown on the user interface screen.
- -The system will produce visual and audible alarms if vital parameters vary beyond preset or default limits.
- -The system includes an integrated pneumatic back-up system for manual hand bagging in order to the rapy in the event of a failure of the primary delivery system and during manual ventilation.
The Ulspira TS system consists of the base unit, the mobile cart and bedside rail holder, and various components and accessories, including the gas regulators and patient kits for use with validated respiratory devices.
### G. INDICATIONS FOR USE
The Ulspira TS Nitric Oxide Therapy System is intended for the delivery of nitric oxide (NO) and the monitoring of inspired NO, NO2 and O2 concentrations for a patient undergoing inhaled Nitric Oxide (iNO) therapy.
The Uspira TS must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardic evidence of pulmonary hypertension. The primary targeted clinical
{5}------------------------------------------------
setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Refer to this material prior to use.
The Ulspira TS primary NO therapy system delivers NO gas in the Constant Rate or flow sensing modes. This includes:
- Continuous integrated monitoring for inspired NO, NO2 and a comprehensive alarm system. .
- A touch-screen user interface with a waveform display of the ventilation device flow as measured in the inspiratory limb. .
- Cylinder handling facilitated by manual or an automatic cylinder switch which is reactive to the detected gas supply state of NO . cylinders, and a low O2 pressure alarm when using an oxygen cylinder.
- An automated emergency dosing algorithm activated by certain high-risk alarms, which impact patid sudden . cessation of therapy.
- Compatibility with a wide inspiratory flow rate range of 0.25-120 |/min, utilizing an automatically detected low or high flow . sensor.
- An internal battery which provides at least two hours of uninterrupted therapy and a 12V DC inlet for additional external battery access.
The integrated Ulspira TS pneumatic backup NO therapy system provides backup NO delivery capability that is intended to deliver a continuous flow of NO mixed with O2, for iNO therapy which allows continuous treatment within hospitals.
{6}------------------------------------------------
# H. TECHNOLOGICAL CHARACTERISTICS - COMPARISON TO PREDICATE
| Device name | INOmax DSIR Plus | Ulspira TS | Comparison |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison of general attributes, indication, patient population, operating environment, etc. | | | |
| Product code(s) | -<br>- MRN<br>- MRP<br>- MRQ | -<br>- MRN, MRO<br>- MRP<br>- MRQ<br>- CCL | Similar; the additional product codes (MRO – Apparatus Nitric Oxide Back-up<br>delivery, CCL – Analyzer O2) are also part of the INOmax DSIR Plus system. |
| Regulation number | - 21 CFR 868.5165 | - 21 CFR 868.5165 | Identical |
| Regulation<br>Description | - Nitric oxide administration apparatus | - Nitric oxide administration apparatus | Identical |
| Device name | INOmax DSIR Plus | Ulspira TS | Comparison |
| Indications for Use | The INOmax DSIR Plus delivery system delivers<br>INOMAX (nitric oxide for inhalation) therapy gas into<br>the inspiratory limb of the patient breathing circuit in<br>a way that provides a constant concentration of nitric<br>oxide (NO), as set by the user, to the patient<br>throughout the inspired breath. It uses a specially<br>designed injector module, which enables tracking of<br>the ventilator waveforms and the delivery of a<br>synchronized and proportional dose of NO. It may be<br>used with most ventilators. | The Ulspira TS Nitric Oxide Therapy System is intended for use<br>by healthcare professionals for the delivery of nitric oxide (NO)<br>and the monitoring of inspired NO, NO2 and O2 concentrations<br>for a patient undergoing inhaled Nitric Oxide (iNO) therapy. | Similar<br>Both devices deliver controlled concentrations of NO into the inspiratory limb<br>of a respiratory device circuit. Both devices contain gas monitors with alarms<br>for NO, O2 and NO2. |
| | | The Ulspira TS must only be used in accordance with the<br>indications, contraindications, warnings and precautions<br>described in the nitric oxide drug packaging inserts and<br>labeling and is indicated for use in term and near-term<br>(>34weeks gestation) neonates with hypoxic respiratory failure<br>associated with clinical or echocardiographic evidence of<br>pulmonary hypertension. The primary targeted clinical setting<br>is the Neonatal Intensive Care Unit (NICU) and secondary<br>targeted clinical setting is the transport of neonates. Refer to<br>this material prior to use. | Ulspira TS system facilitates cylinder switching.<br>Ulspira TS system has an emergency dosing algorithm to avoid sudden<br>cessation of therapy, activated by certain high-risk alarms. |
| | The INOmax DSIR Plus provides continuous integrated<br>monitoring of inspired O2, NO2, and NO, and a<br>comprehensive alarm system.<br><br>The INOmax DSIR Plus incorporates a battery that<br>provides up to 6 hours of uninterrupted NO delivery in<br>the absence of an external power source.<br><br>The INOmax DSIR Plus includes a backup NO delivery<br>capability that provides a fixed flow of 250 mL/min of<br>NO which along with user supplied 10 L/min of oxygen<br>provides 20 ppm in the gas flow to a patients<br>breathing circuit. It may also use the INOblender for<br>backup. | The Ulspira TS primary NO therapy system delivers NO gas in<br>the 0-80 ppm range while in the Constant Rate or flow sensing<br>modes. This includes:<br>• Continuous integrated monitoring for inspired NO, NO2 and<br>O2 and a comprehensive alarm system.<br>• A touch-screen user interface with a waveform display of the<br>ventilation device flow as measured in the inspiratory limb. | Ulspira TS system has a broader inspiratory flow rate range utilizing two flow<br>sensors, high and low.<br>Indications for both devices follow the respective drug labeling for nitric oxide<br>(currently neonates). |
| | | • Cylinder handling facilitated by manual or an automatic<br>cylinder switch which is reactive to the detected gas supply<br>state of NO cylinders, and a low O2 pressure alarm when using<br>an oxygen cylinder. | |
| | The target patient population is controlled by the drug<br>labeling for INOMAX and is currently neonates. The<br>primary targeted clinical setting is the Neonatal<br>Intensive Care Unit (NICU) and secondary targeted<br>clinical setting is the transport of neonates.<br>(Ref. K200389) | • An automated emergency dosing algorithm activated by<br>certain high-risk alarms, which impact patient dosing, to avoid<br>sudden cessation of therapy. | |
| | | • Compatibility with a wide inspiratory flow rate range of 0.25-<br>120 l/min, utilizing an automatically detected low or high flow<br>sensor. | |
| | | • An internal battery which provides at least two hours of<br>uninterrupted therapy and a 12V DC inlet for additional<br>external battery access. | |
| | | The integrated Ulspira TS pneumatic backup NO therapy<br>system provides backup NO delivery capability that is intended<br>to deliver a continuous flow of NO mixed with O₂, for iNO<br>therapy which allows continuous treatment during transit<br>within hospitals. | |
| Device name | INOmax DSIR Plus | Ulspira TS | Comparison |
| Physical<br>dimensions and<br>weight (excl.<br>carrier/cart) | Weight: 5.3 kg<br>Width: 350 mm<br>Depth: 160 mm<br>Height: 220 mm | Weight: 7.0 kg<br>Width: 320 mm<br>Depth: 150 mm<br>Height: 300 mm | Different but substantially equivalent |
| NO gas connectors | CGA 626 | CGA 626 | Identical |
| NO Injection<br>location | NO injection connected between ventilator and<br>humidifier. | NO injection connected between ventilator and humidifier. | Identical |
| Battery backup | Yes | Yes | Identical |
| Battery backup<br>time | 6h | 2h | Different, however both devices meet the US Food and Drug Administration<br>special controls guidance. |
| Power Supply -<br>Main | Voltage: 100-240V, 50 - 60Hz | Voltage: 100-240V, 50 - 60Hz | Identical |
| Automated Pre-Use<br>check | Yes | Yes | Identical |
| Alarms | | | |
| NO Delivery/Flow<br>Sensor Alarms | Yes | Yes | Substantially equivalent<br>Both devices include alarms for failures in injection/delivery and for stop in<br>delivery/no treatment of patient. The Ulspira TS system also includes specific<br>alarms in relation to the respiratory device/ventilator flow. |
| Power<br>Supply/Battery<br>Alarms | Yes | Yes | Substantially equivalent<br>Both devices include alarm(s) for low battery power, additionally the Ulspira TS<br>system also includes technical alarms in relation to failures in different parts of<br>the power supply system. |
| NO, NO2, O2<br>Monitoring Alarms | Yes | Yes | Substantially equivalent |
| Device name | INOmax DSIR Plus | Ulspira TS | Comparison |
| Primary NO administration system | | | |
| NO administration<br>principle | NO delivery into the inspiratory limb of a ventilation<br>device's patient circuit. | NO delivery into the inspiratory limb of a ventilation device's<br>patient circuit. | Identical…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.