K212402 · Mivi Neuroscience, Inc. · QJP · Dec 16, 2021 · Cardiovascular
Device Facts
Record ID
K212402
Device Name
MIVI Q Distal Access Catheter
Applicant
Mivi Neuroscience, Inc.
Product Code
QJP · Cardiovascular
Decision Date
Dec 16, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The MIVI Q Distal Access Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.
Device Story
Single-lumen, variable stiffness catheter; used to facilitate insertion/guidance of microcatheters into peripheral, coronary, or neurovascular vessels. Input: manual manipulation by physician under fluoroscopic guidance. Operation: catheter introduced via 8F guide catheter/6F guide sheath; advanced over guidewire/microcatheter; stainless-steel control wire and docking station facilitate movement; hydrophilic coating reduces friction. Output: physical placement of microcatheter at target vessel site. Used in clinical settings (e.g., cath lab) by physicians. Benefits: enables access to distal vasculature for diagnostic/therapeutic interventions.
Clinical Evidence
Bench testing only. Testing included dimensional verification, simulated use (system introduction), liquid leakage under pressure, and tensile strength. Biocompatibility testing included cytotoxicity, irritation/intracutaneous reactivity, sensitization, acute systemic toxicity, pyrogenicity, hemolysis, unactivated partial thromboplastin time, and complement activation. All tests passed.
Indicated for patients requiring endovascular insertion and guidance of microcatheters into peripheral, coronary, or neurovascular vessels during diagnostic or therapeutic procedures for arterial disease or damage.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date December 16, 2021. The month is spelled out, followed by the day and year. The text is in a simple, sans-serif font and is left-aligned.
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below that.
MIVI Neuroscience, Inc. Janel Hurtado Director of Regulatory 6545 City West Parkway Eden Prairie, Minnesota 55344
Re: K212402
Trade/Device Name: MIVI Q Distal Access Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, DQY Dated: November 11, 2021 Received: November 17, 2021
Dear Janel Hurtado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212402
Device Name MIVI Q Distal Access Catheter
#### Indications for Use (Describe)
The MIVI Q Distal Access Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.
**Type of Use (Select one or both, as applicable)**
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) SUMMARY K212402
[as required by 21 CFR \$807.92(c)]
Date Prepared: December 14, 2021
| | 510(k) Submitter | Contact |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| | MIVI Neuroscience, Inc.<br>6545 City West Parkway<br>Eden Prairie, MN 55344<br>(952) 944-3834 | Janel Hurtado<br>Regulatory Affairs Director<br>Email: jhurtado@mivineuro.com<br>(952) 944-3834 |
| General Information | | |
| Trade Name | MIVI Q Distal Access Catheter | |
| Common Name | Distal access catheter | |
| Classification<br>Information | Device Classification: Class II<br>Common Name: Percutaneous catheter<br>Regulation Number: 21 CFR 870.1250<br>Product Code: QJP; Panel: Neurology<br>Product Code: DQY; Panel: Cardiovascular | |
| Predicate Device | MIVI Q Distal Access Catheter (K192558; April 6, 2020) | |
## Device Description
The MIVI Q Distal Access Catheter (Q Catheter) is a single-lumen, variable stiffness catheter with radiopaque markers on the distal and proximal end of the catheter portion for angiographic visualization. The catheter shaft has a hydrophilic coating to reduce friction during use. The proximal portion of the catheter is a stainless-steel control (push) wire. The Q Catheter may be introduced via an 8F guide catheter/6F guide sheath and over a guidewire/microcatheter into the arterial vasculature until the desired vessel is reached. The Q Docking Station may be used to facilitate insertion and extraction of the Q Catheter through a hemostasis valve attached to the 8F guide catheter/6F guide sheath. The pin vise may be used to advance the catheter.
### Intended Use / Indications for Use
The MIVI Q Distal Access Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.
### Substantial Equivalence Comparison
The MIVI Q Distal Access Catheter (pre-loaded with Docking Station) has identical indications, principle of operation, and similar design as the currently marketed predicate catheters.
| Table 1: Comparison with Currently Marketed Predicate | | | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| Feature | Subject Device<br>MIVI Q Distal Access Catheter | Predicate Device<br>MIVI Q Distal Access Catheter | |
| 510(k) Holder &<br>Manufacturer | MIVI Neuroscience, Inc. | MIVI Neuroscience, Inc. | |
| 510(k)# | K212402 | K192558 | |
| Table 1: Comparison with Currently Marketed Predicate | | | |
| Feature | Subject Device<br>MIVI Q Distal Access Catheter | Predicate Device<br>MIVI Q Distal Access Catheter | |
| Classification<br>Information | Percutaneous catheter<br>21 CFR 870.1250 (Class II)<br>ProCode: QJP, DQY | | |
| Indications / Intended Use / Principle of Operation | | | |
| Indications for Use | The MIVI Q Distal Access<br>Catheter is indicated for use with<br>compatible guide catheters in<br>facilitating the insertion and<br>guidance of microcatheters into a<br>selected blood vessel in the<br>peripheral, coronary and neuro<br>vascular systems. | The Q Catheter is indicated for use<br>with compatible guide catheters in<br>facilitating the insertion and<br>guidance of microcatheters into a<br>selected blood vessel in the<br>peripheral, coronary and neuro<br>vascular systems. | |
| Principle of<br>Operation | Used to endovascularly insert and guide microcatheters under fluoroscopy<br>during diagnostic and/or therapeutic procedures for patients with arterial<br>disease or damage. | | |
| Design Information | | | |
| Configuration | Distal | Single-lumen, variable stiffness, braided (5F & 6F) / coiled (3F & 4F) | |
| | Proximal | Control (push) wire through a<br>Docking Station assembly | |
| | Tip | Straight | |
| | Radiopaque<br>Markers | Two (2) - Distal and Proximal | |
| | Catheter<br>Coating | Hydrophilic | |
| | Catheter Sizes | 3F, 4F, 5F, 6F | |
| | | | |
| | 3F | 154 | 148 |
| | 4F | 141 | 135 |
| Total Length<br>(cm) | 5F | 136 | 130 |
| | 6F | 136 | 130 |
| | | | |
| | | | |
| Extensible Length | | | |
| | 3F | 40-42 | |
| | 4F | 27-29 | |
| Extensible length (cm) with:<br>90 cm Guide | 5F | 22-24 | |
| | 6F | 22-24 | |
| | | | |
| | | | |
| | 3F | 35-37 | |
| | 4F | 22-24 | |
| 95 cm Guide | 5F | 17-19 | |
| | 6F | 17-19 | |
| | | | |
| | | | |
| Control Wire<br>Length | 108 | 104 | |
| Table 1: Comparison with Currently Marketed Predicate | | | |
| Feature | Subject Device<br>MIVI Q Distal Access Catheter | Predicate Device<br>MIVI Q Distal Access Catheter | |
| Pin Vise Shape | Looped shape (Pin Vise not removable) | | |
| Guidewire<br>Compatibility<br>(cm) | 3F | 0.018" | |
| | 4F | 0.035" | |
| | 5F | | |
| | 6F | | |
| Catheter<br>Biomaterials | Patient contacting materials are identical between the subject and currently<br>marketed predicate and accessories. | | |
| Accessories / Packaging / Sterilization / Shelf Life | | | |
| Provided Accessories | 9F Rotating Hemostasis Valve Y-Connector | None | |
| Required Accessories<br>(not provided) | 8F guide catheter / 6F guide sheath 90-95 cm in length | | |
| Package<br>Configuration | Catheter inserted in a plastic tube, mounted on an insert card, and sealed in<br>a pouch. Sealed pouch packaged in carton along with Instructions for Use. | | |
| Sterile & Non-<br>pyrogenic | Yes | | |
| Sterilization Method | | Ethylene Oxide (EO) | |
| Shelf Life | | 3 years | |
{4}------------------------------------------------
# P a g e | 2
{5}------------------------------------------------
# Page | 3
The technological differences between the modified MIVI Q Distal Access Catheter and the currently marketed predicate do not raise new questions of safety and effectiveness.
## Non-Clinical Performance Testing
## Bench Testing
Bench testing was completed to assess the performance of the subject device, as shown in the following table.
| Test | Test Summary | Result |
|----------------------------------------|--------------------------------------------------|--------|
| Dimensional<br>Verification | The device must meet dimensional specifications. | Pass |
| System Introduction (Simulated<br>Use) | The device must function as intended. | Pass |
| Liquid Leakage under Pressure | The device must hold a hydrostatic pressure. | Pass |
| Tensile Strength | Tensile strength pull force minimum must be met. | Pass |
{6}------------------------------------------------
### Page | 4
### Biocompatibility
Biocompatibility testing was completed to assess the subject device, as shown in the following table.
| Test | Test Summary | Result |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------|
| Cytotoxicity | Extracted with MEM with 10% FBS at 37 °C for 24<br>hours and administered on mouse fibroblast L929 cells. | PASS |
| Irritation or Intracutaneous<br>Reactivity | Extracted with 0.9% NaCl and cottonseed oil (CSO) at<br>37 °C for 72 hours and intracutaneously injected in<br>rabbits. | PASS |
| Sensitization | Extracts: 0.9% NaCl, CSO for 72 hours at 37 °C. | PASS |
| Acute Systemic Toxicity | Extracted with 0.9% NaCl and CSO at 37 °C for 72<br>hours and injected in mice. | PASS |
| Pyrogenicity (Material-<br>mediated) | Extracted with 0.9% NaCl at 37 °C for 72 hours and<br>injected intravenously in rabbits. | PASS |
| Hemolysis | Extracted from PBS at 37 °C for 24 hours, administered<br>on rabbit blood; or directly contacting rabbit blood in<br>vitro. | PASS |
| Unactivated Partial<br>Thromboplastin Time | Human plasma was exposed to the test article at a ratio of<br>6 cm2/ml for 15 minutes at 37 °C. | PASS |
| Complement Activation | Human plasma was exposed to the test article at a<br>ratio of 6 cm2/mL for 90 mins at 37 °C, in triplicate. | PASS |
### Sterilization
The sterilization method is unchanged. Confirmatory testing on the final packaging configuration was completed and met specification.
### Shelf Life and Expiration Dating
The labeled shelf life of the modified device is the same as that of the predicate, 3 years from the date of sterilization. The bench testing described above was performed to support the shelflife of the modified device.
### Conclusion
The modified MIVI Q Distal Access Catheter has identical indications, principle of operation, and similar design as the currently marketed MIVI Q Catheter predicate device. The technological differences between the modified Q Catheter and the currently marketed predicate do not raise new questions of safety and effectiveness. Based on the predicate comparison, risk assessment, and device testing results, the modified MIVI Q Distal Access Catheter has been shown to be appropriate for its intended use and is considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.