The Intri24 Sheath is indicated to provide a conduit for the insertion of endovascular devices into the vasculature while minimizing blood loss associated with such insertions.
Device Story
Intri24 Sheath is a single-use, sterile introducer sheath for peripheral vasculature. Device consists of a short single-lumen catheter with a hydrophilic coating, a proximal user-actuated hemostasis valve, and a stopcock with a flush port. A radiopaque marker at the distal tip aids fluoroscopic visualization. An included dilator, compatible with 0.035" guidewires, features a tapered leading edge to facilitate vessel insertion. The hemostasis valve, operated by spring-loaded push buttons, allows for the insertion or withdrawal of endovascular devices while maintaining hemostasis. Used in clinical settings by physicians to provide a conduit for endovascular procedures; benefits patient by facilitating device access and reducing blood loss.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including visual/dimensional inspection, lubricity/durability, guidewire compatibility, insertion/retraction forces, leak testing, flowrate, vacuum testing, sheath burst, and mechanical integrity (tensile, compression, torque). Biocompatibility testing conducted per ISO 10993-1:2018, including cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, complement activation, and thromboresistance.
Technological Characteristics
Single-lumen catheter with hydrophilic coating; includes dilator (0.035" guidewire compatible) and spring-loaded push-button hemostasis valve. Radiopaque distal marker. Materials meet ISO 10993-1:2018 biocompatibility standards. Sterile, single-use device.
Indications for Use
Indicated for patients requiring endovascular device insertion into the peripheral vasculature; provides conduit while minimizing blood loss.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
W.L. Gore DrySeal Sheath with Hydrophilic Coating (K121234)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
April 1, 2022
Inari Medical, Inc. Kit Cariquitan Vice President, Regulatory Affairs 9 Parker. Suite 100 Irvine, California 92618
Re: K212392
Trade/Device Name: Intri24 Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: February 25, 2022 Received: February 28, 2022
#### Dear Kit Cariquitan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Finn Donaldson Assistant Director (Acting) DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212392
Device Name Intri24 Sheath
Indications for Use (Describe)
The Intri24 Sheath is indicated to provide a conduit for the insertion of endovascular devices into the vasculature while minimizing blood loss associated with such insertions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| Date prepared | July 29, 2021 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Inari Medical, Inc.<br>6001 Oak Canyon, Suite 100<br>Irvine, CA 92618<br>877-923-4747 |
| Contact person | Kit Cariquitan<br>Vice President, Regulatory Affairs & Quality Assurance |
| Name of Device | Intri24 Sheath |
| Common name | Introducer Sheath |
| Regulation name | Catheter Introducer |
| Classification<br>number | 21 CFR 870.1340 |
| Product code | DYB |
| Regulatory class | II |
| Predicate device | W.L. Gore DrySeal Sheath with Hydrophilic Coating (K121234)<br>This product has not been subject to a design-related recall. |
| Reference device | ClotTriever Thrombectomy System (K180329) |
| Description | The Intri24 Sheath is a single-use, sterile medical device for use in the<br>peripheral vasculature. The Intri24 Sheath is an introducer sheath<br>consisting of a short single lumen catheter with a hydrophilic coating,<br>proximal hemostasis valve, and stopcock with flush port. A radiopaque<br>marker is positioned near the distal tip of the sheath to aid with<br>fluoroscopic visualization. The Intri24 Dilator is compatible with a<br>0.035" guidewire and has a tapered leading edge which aids insertion<br>and positioning of the Intri24 Sheath. |
| Indications for Use | The Intri24 Sheath is indicated to provide a conduit for the insertion of<br>endovascular devices into the vasculature while minimizing blood loss<br>associated with such insertions. |
| Comparison of<br>Technological<br>Characteristics with<br>the Predicate Device | The indications for use and principles of operation of the Intri24 Sheath<br>and the predicate device are substantially the same. Both devices act<br>as conduit for the insertion and removal of endovascular devices. The<br>design and materials of construction of both devices is also<br>substantially the same. The shafts of both devices contain a hydrophilic<br>coating to reduce the insertion force through skin and tissue. Both<br>sheaths also include an appropriately sized dilator (0.035 in guidewire<br>compatible) with a tapered tip to aid in dilation of the target vessel when<br>inserting the sheath.<br><br>The Intri24 Sheath contains a user-actuated hemostasis valve identical<br>to that used on the ClotTriever catheter and sheath (K180329). The<br>hemostasis valve on the Intri24 Sheath allows for inserting or<br>withdrawing endovascular devices through the sheath. The user<br>actuated hemostasis valve is operated by spring-loaded push buttons.<br>The W.L. Gore DrySeal hemostasis valve contains an inflatable ring<br>which closes when saline is injected and opens when saline is<br>withdrawn from the ring. |
| Summary of<br>substantial<br>equivalence | The Intri24 Sheath's design, materials, mechanism of action, and<br>performance specifications are substantially equivalent to the predicate<br>device.<br><br>Non-Clinical Testing<br><br>In accordance with the Design Failure Modes and Effects Analysis,<br>verification and validation testing were identified to support the<br>substantial equivalence of the Intri24 Sheath. This testing demonstrated<br>compliance with relevant product specifications. These tests included:<br>Visual and Dimensional Inspection Dye Staining Lubricity and Durability Guidewire Compatibility Insertion Force Locking Cap Force Kink Radius and Dilator Retraction Force Leak Testing Simulated Use Flowrate Testing Vacuum Testing Sheath Burst Push-Button Force Tensile, Compression, and Torque Testing Placement Resistance Testing |
| | |
| Pouch Seal Visual Inspection and Dye Penetration<br>● | |
| Pouch, Peel, Seal Strength<br>● | |
| Sterilization Validation<br>● | |
| Biocompatibility Testing | |
| The following biocompatibility tests were conducted in accordance<br>with ISO 10993-1:2018 and successfully passed: | |
| L-929 MEM Elution | |
| Kligman Maximization<br>● | |
| Intracutaneous Injection<br>● | |
| Systemic Injection<br>● | |
| Pyrogen<br>● | |
| Blood Hemolysis<br>● | |
| Complement Activation<br>● | |
| Thromboresistance<br>● | |
| Platelet and Leukocyte Count<br>● | |
| Partial Thromboplastin Time<br>● | |
| Animal and clinical testing were not required for the determination of<br>substantial equivalence. | |
| Test results demonstrated that all acceptance criteria were met;<br>therefore, the device conforms to established product specifications. | |
| Conclusion | |
| testing demonstrated that the Intri24<br>Non-clinical<br>Sheath is<br>substantially equivalent to the W.L. Gore DrySeal Sheath with<br>Hydrophilic Coating with respect to design, materials, function, and<br>intended use. | |
| | |
{4}------------------------------------------------
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.