← Product Code [DQK](/productcode/DQK) · K212345

# AcQMap High Resolution Imaging and Mapping System (K212345)

_Acutus Medical, Inc. · DQK · Aug 27, 2021 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212345

## Device Facts

- **Applicant:** Acutus Medical, Inc.
- **Product Code:** [DQK](/productcode/DQK.md)
- **Decision Date:** Aug 27, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1425
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed. When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either charge density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone. AND When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart. OR When used with conventional electrophysiology catheters, the AcQMap System provides information about the electrical activity of the heart and about catheter location during the procedure.

## Device Story

AcQMap System is a computer-based diagnostic recording system for use in EP labs. It processes inputs from ultrasound imaging, ECG/EGM recording, and impedance-based electrode localization. The system reconstructs 3D cardiac chamber anatomy and generates dynamic charge density or voltage maps to identify arrhythmia sources. It also tracks the position of AcQMap and conventional EP catheters. Physicians use the system to visualize electrical activity and guide clinical decision-making during procedures. The system includes a console, workstation, and auxiliary interface box. This modification adds cables and adapters to enable connectivity with various RF ablation generators, facilitating integration into existing clinical workflows.

## Clinical Evidence

No new clinical testing performed. Clinical evidence for the original system (K170948) is incorporated by reference. The DDRAMATIC-SVT study (prospective, non-randomized, open-label, 84 patients) demonstrated safety and effectiveness for the original system.

## Technological Characteristics

Programmable diagnostic computer and ultrasonic pulsed echo imaging system. Components: Console, workstation, auxiliary interface box, various cables/adapters. Sensing: Ultrasound (10 MHz), impedance-based localization (15-50 kHz), ECG/EGM (0.05-500 Hz). Power: 100-230 VAC. Safety: IEC 60601-1, Type CF/BF defibrillator-protected. Connectivity: Wired interface to catheters and RF generators. Software: Diagnostic imaging and mapping algorithms.

## Regulatory Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

## Predicate Devices

- AcQMap High Resolution Imaging and Mapping System, Model 900100 ([K201015](/device/K201015.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 27, 2021

Acutus Medical, Inc. Sindhu Sridhar Regulatory Affairs Manager 2210 Faraday Ave, Suite 100 Carlsbad, California 92008

Re: K212345

Trade/Device Name: AcQMap® High Resolution Imaging and Mapping System, Model 900100 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, IYO, ITX, Dated: July 26, 2021 Received: July 28, 2021

Dear Sindhu Sridhar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh Deoras Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212345

#### Device Name

AcQMap® High Resolution Imaging and Mapping System, Model 900100

#### Indications for Use (Describe)

The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed.

When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either charge density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.

#### AND

When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.

OR

When used with conventional electrophysiology catheters, the AcQMap System provides information about the electrical activity of the heart and about catheter location during the procedure.

Type of Use (Select one or both, as applicable)

| <span style="font-size:12px"></span> <span style="font-size:12px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:12px"></span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

# 510(K) NOTIFICATION K212345

# GENERAL INFORMATION [807.92(a)(1)]

# Date Prepared: 26 July 2021

| Applicant:                   | Contact Person:                  |
|------------------------------|----------------------------------|
| Acutus Medical, Inc.         | Sindhu Sridhar                   |
| 2210 Faraday Ave., Suite 100 | Regulatory Affairs Manager       |
| Carlsbad, CA 92008           | Acutus Medical, Inc.             |
| USA                          | 2210 Faraday Ave., Suite 100     |
| Phone: 1-442-232-6080        | Carlsbad, CA 92008               |
| Fax: 1-442-232-6081          | USA                              |
|                              | Phone: 1-442-999-7630            |
|                              | Email: sindhu.sridhar@acutus.com |

# DEVICE INFORMATION [807.92(a)(2)]

| Trade Name:          | AcQMap® High Resolution Imaging and Mapping System, Model 900100             |
|----------------------|------------------------------------------------------------------------------|
| Generic/Common Name: | Programable diagnostic computer and<br>Ultrasonic pulsed echo imaging system |
| Classification:      | Class II / 21 CFR § 870.1425 and<br>Class II / 21 CFR § 892.1560             |
| Product Code(s):     | DQK, IYO, ITX                                                                |

# PREDICATE DEVICES [807.92(a)(3)]

| Predicate Device                                                      | Manufacturer         | FDA 510(k) |
|-----------------------------------------------------------------------|----------------------|------------|
| AcQMap High Resolution<br>Imaging and Mapping System,<br>Model 900100 | Acutus Medical, Inc. | K201015    |

{4}------------------------------------------------

## DEVICE DESCRIPTION [807.92(a)(4)]

The AcQMap High Resolution Imaging and Mapping System, Model 900100 operates outside of the sterile field and consists of the AcQMap Console, the AcQMap Workstation and the AcQMap Auxiliary Interface Box.

The AcQMap High Resolution Imaging and Mapping System, Model 900100 ("AcQMap System Model, 900100") is a diagnostic recording system. This computer-based system is intended for use in the Electrophysiology (EP) Lab, and it is capable of imaging, navigation and mapping of the atrial chambers of the heart.

The AcQMap System hardware consists of three functional subsystems:

- . Ultrasound imaging,
- . ECG and EGM recording; and
- Impedance based electrode Localization. .

The AcQMap System, Model 900100 is used in conjunction with the associated AcQMap 3D Imaging and Mapping Catheter models 900003 and 900009 (cleared under K201341). The AcQMap System provides:

- . 3-D cardiac chamber reconstruction - Contact and non-contact (ultrasound),
- . Three-dimensional position of the AcQMap Catheter and conventional electrophysiology catheters,
- . Cardiac electrical activity as waveform traces,
- . Contact LAT and voltage amplitude maps
- . Remapping of the chamber at any time during the procedure; and
- Dynamic, three-dimensional, charge density maps overlaid on the cardiac chamber reconstruction . to show chamber-wide electrical activation.

The AcQMap System, Model 900100 is intended to create a surface reconstruction of the cardiac chamber as well as an electrical map of the substrate. The surface reconstruction and electrical map are then used by physicians to identify the source(s) of the arrhythmia.

Additionally, the AcQMap System allows physicians to perform traditional contact mapping activities, including establishing a coordinate system, localizing conventional electrophysiology catheters relative to one another within the coordinate system, recording contact electrograms, and initiating a procedure without the AcQMap Catheter present. Based on the information captured in the contact electrograms, the physician may decide to treat an arrythmia without deploying the AcQMap Catheter.

The modifications to the AcQMap System, Model 900100 includes addition of new accessories (cables and adapters) to facilitate connectivity between the cleared AcQMap High Resolution Imaging and Mapping System and compatible RF Ablation Generators.

{5}------------------------------------------------

# INDICATIONS FOR USE [807.92(a)(5)]

The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed.

When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either charge density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.

# AND

When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.

OR

When used with conventional electrophysiology catheters, the AcQMap System provides information about the electrical activity of the heart and about catheter location during the procedure.

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(A)(6)[

Tables 1 and 2 provides a comparison of the modified AcQMap System classification for use against the predicate device. Table 3 provides a comparison of the technological characteristics for the modified AcQMap System against the predicate device.

| Table 1. Comparison of Classification with the Predicate Device                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                       |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics                                                                   | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Rationale for Substantial Equivalence                                                                                                                                                                                                                                                 |
|                                                                                   | AcQMap® High Resolution Imaging and Mapping System, Model 900100                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | AcQMap® High Resolution Imaging and Mapping System, Model 900100 (K201015)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                       |
| 510(k) Number                                                                     | K212345                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | K201015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | --                                                                                                                                                                                                                                                                                    |
| Classification/<br>Regulation<br>Number/<br>Regulation Name/<br>Product Code      | Class II/ 21 CFR § 870.1425/ Programmable diagnostic computer/ DQK<br>Class II/ 21 CFR § 892.1560/ Ultrasonic pulsed echo imaging system/ IYO, ITX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Class II/ 21 CFR § 870.1425/ Programmable diagnostic computer/ DQK<br>Class II/ 21 CFR § 892.1560/ Ultrasonic pulsed echo imaging system/ IYO, ITX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                                                                                             |
| Table 2. Comparison of Indications for Use with the Predicate Device              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                       |
| Characteristics                                                                   | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Rationale for Substantial<br>Equivalence                                                                                                                                                                                                                                              |
|                                                                                   | AcQMap® High Resolution Imaging and<br>Mapping System, Model 900100                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | AcQMap® High Resolution Imaging and<br>Mapping System, Model 900100 (K201015)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                       |
| 510(k) Number                                                                     | K212345                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | K201015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Identical                                                                                                                                                                                                                                                                             |
| Indications for<br>Use                                                            | The AcQMap System is intended for use in<br>patients for whom electrophysiology procedures<br>have been prescribed.<br><br>When used with the AcQMap Catheters, the<br>AcQMap System is intended to be used to<br>reconstruct the selected chamber from<br>ultrasound data for purposes of visualizing the<br>chamber anatomy and displaying electrical<br>impulses as either charge density-based or<br>voltage-based maps of complex arrhythmias that<br>may be difficult to identify using conventional<br>mapping systems alone.<br>AND<br>When used with the specified Patient Electrodes,<br>the AcQMap System is intended to display the<br>position of AcQMap Catheters and conventional<br>electrophysiology (EP) catheters in the heart.<br>OR<br>When used with conventional electrophysiology<br>catheters, the AcQMap System provides<br>information about the electrical activity of the<br>heart and about catheter location during the<br>procedure. | The AcQMap System is intended for use in<br>patients for whom electrophysiology procedures<br>have been prescribed.<br><br>When used with the AcQMap Catheters, the<br>AcQMap System is intended to be used to<br>reconstruct the selected chamber from<br>ultrasound data for purposes of visualizing the<br>chamber anatomy and displaying electrical<br>impulses as either charge density-based or<br>voltage-based maps of complex arrhythmias that<br>may be difficult to identify using conventional<br>mapping systems alone.<br>AND<br>When used with the specified Patient Electrodes,<br>the AcQMap System is intended to display the<br>position of AcQMap Catheters and conventional<br>electrophysiology (EP) catheters in the heart.<br>OR<br>When used with conventional electrophysiology<br>catheters, the AcQMap System provides<br>information about the electrical activity of the<br>heart and about catheter location during the<br>procedure. |                                                                                                                                                                                                                                                                                       |
| Table 3: Comparison of Technological Characteristics Against the Predicate Device |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                       |
| Characteristics                                                                   | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Rationale for Substantial<br>Equivalence                                                                                                                                                                                                                                              |
| Patient Anatomy                                                                   | Intracardiac Structures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Intracardiac Structures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Identical                                                                                                                                                                                                                                                                             |
| Testing to Support<br>Substantial<br>Equivalence                                  | Software V/V Electromagnetic and Electrical Safety Verification Testing, Accuracy Testing, and Animal Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Software V/V Electromagnetic and Electrical Safety Verification Testing, Accuracy Testing, and Animal Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                                             |
| System Safety<br>Standards                                                        | IEC 60601-1:2005 /A1:2012 IEC 60601-1-2:2014 IEC 60601-1-6:2010/A1:2013 IEC 60601-2-25:2015 IEC 60601-2-37:2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | IEC 60601-1:2005 /A1:2012 IEC 60601-1-2:2014 IEC 62366-1: 2015 IEC 60601-2-25:2015 IEC 60601-2-37:2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Different. IEC 60601-1-6:<br>2010/ A1: 2013 has been<br>withdrawn and replaced<br>with IEC 62366-1: 2015.                                                                                                                                                                             |
| Physical Characteristics                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                       |
| System<br>Components<br>&<br>Accessories                                          | Console Workstation Workstation Cable Auxiliary Interface Box ECG Input Cable Ampere Ablation Catheter Adapter Cable Ampere RF Generator Adapter Cable ECG Output Cable Ablation Reference Cable Ablation Electrogram Cable ECG w/Snaps Cable ECG POST Cable 2mm Pin Jumper Set Patient Electrode Kit Ampere generator adapter SmartAblate generator adapter Maestro generator adapter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Console Workstation Workstation Cable Auxiliary Interface Box ECG Input Cable Ampere Ablation Catheter Adapter Cable Ampere RF Generator Adapter Cable ECG Output Cable Ablation Reference Cable Ablation Electrogram Cable ECG w/Snaps Cable ECG POST Cable 2mm Pin Jumper Set Patient Electrode Kit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | All system components are<br>identical, except the new<br>accessories. Verification<br>testing on the new<br>accessories demonstrates<br>that the AcQMap System<br>and the new accessories<br>perform as intended. There<br>are no different questions of<br>safety or effectiveness. |
| Table 3: Comparison of Technological Characteristics Against the Predicate Device |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                       |
| Characteristics                                                                   | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Rationale for Substantial<br>Equivalence                                                                                                                                                                                                                                              |
|                                                                                   | AcQMap® High Resolution Imaging and Mapping System, Model 900100                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | AcQMap® High Resolution Imaging and Mapping System, Model 900100 (K201015)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                       |
|                                                                                   | Adapter Cable- Short - Ablation Adapter Adapter Cable- Long - Ablation Adapter SmartAblate Adapter Cable - Catheter Maestro Adapter Cable - Catheter Maestro Adapter Cable - Generator                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                       |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

| Table 3: Comparison of Technological Characteristics Against the Predicate Device (Continued) |                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                          |
|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| Characteristics                                                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Rationale for Substantial<br>Equivalence |
| Visual/Mapping<br>Characteristics                                                             | 3-D cardiac chamber reconstructions – Contact and non-contact (ultrasound); Three-dimensional position of the AcQMap Catheter and conventional electrophysiology catheters; Cardiac electrical activity as waveform traces; Contact LAT and voltage amplitude maps; Remapping of the chamber at any time during the procedure; Dynamic, three-dimensional, Charge Density maps overlaid on the cardiac chamber reconstruction to show chamber-wide electrical activation. | 3-D cardiac chamber reconstructions - Contact and non-contact (ultrasound); Three-dimensional position of the AcQMap Catheter and conventional electrophysiology catheters; Cardiac electrical activity as waveform traces; Contact LAT and voltage amplitude maps; Remapping of the chamber at any time during the procedure; Dynamic, three-dimensional, Charge Density maps overlaid on the cardiac chamber reconstruction to show chamber-wide electrical activation. | Identical                                |
| Visualization<br>Device/Catheter                                                              | AcQMap Catheter (electrodes & transducers) or Conventional electrophysiology catheters                                                                                                                                                                                                                                                                                                                                                                                    | AcQMap Catheter (electrodes & transducers) or Conventional electrophysiology catheters                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                |
| Physical Characteristics - Console/Amplifier Comparison                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                          |

{9}------------------------------------------------

| Table 3: Comparison of Technological Characteristics Against the Predicate Device (Continued) |                                               |                                               |                                          |
|-----------------------------------------------------------------------------------------------|-----------------------------------------------|-----------------------------------------------|------------------------------------------|
| Characteristics                                                                               | Subject Device                                | Predicate Device                              | Rationale for Substantial<br>Equivalence |
| Dimensions                                                                                    | 99 cm L x 58 cm W x 76 cm D                   | 99 cm L x 58 cm W x 76 cm D                   | Identical                                |
| Weight Maximum                                                                                | 80 kg                                         | 80 kg                                         |                                          |
| Power Requirement                                                                             | 100-127 VAC, 50/60 Hz, 220-230 VAC, 50 Hz     | 100-127 VAC, 50/60 Hz, 220-230 VAC, 50 Hz     |                                          |
| Input Current                                                                                 | 4.6 A                                         | 4.6 A                                         |                                          |
| Fuse protection                                                                               | 250 V, 6.3A, two high breaking capacity fuses | 250 V, 6.3A, two high breaking capacity fuses |                                          |

| Table 3: Comparison of Technological Characteristics Against the Predicate Device (Continued) |                                                                                                                                                                                                                   |                                                                                                                                                                                                                   |                                       |
|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| Characteristics                                                                               | Subject Device                                                                                                                                                                                                    | Predicate Device                                                                                                                                                                                                  | Rationale for Substantial Equivalence |
|                                                                                               | AcQMap® High Resolution Imaging and Mapping System, Model 900100                                                                                                                                                  | AcQMap® High Resolution Imaging and Mapping System, Model 900100 (K201015)                                                                                                                                        |                                       |
| System Specifications                                                                         |                                                                                                                                                                                                                   |                                                                                                                                                                                                                   |                                       |
| Safety Information                                                                            | IEC 60601-1, Class I, Type Defibrillator Protected CF, continuous operation, no sterilization, equipment not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide | IEC 60601-1, Class I, Type Defibrillator Protected CF, continuous operation, no sterilization, equipment not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide | Identical                             |
| Ingress Protection                                                                            | The Console is rated IP20                                                                                                                                                                                         | The Console is rated IP20                                                                                                                                                                                         | Identical                             |
| Functional and Performance Characteristics                                                    |                                                                                                                                                                                                                   |                                                                                                                                                                                                                   |                                       |
| Ultrasound Output                                                                             | Frequency: 10 MHz+/-400 kHz<br>Maximum Voltage: 50V p-p                                                                                                                                                           | Frequency: 10 MHz+/-400 kHz<br>Maximum Voltage: 50V p-p                                                                                                                                                           | Identical                             |

K212345

{10}------------------------------------------------

|                                                                                               | Maximum Power: 1 W peak                                             | Maximum Power: 1 W peak                                                       |                                          |
|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------|-------------------------------------------------------------------------------|------------------------------------------|
| Ultrasound<br>Performance                                                                     | Single operating mode                                               | Single operating mode                                                         | Identical                                |
|                                                                                               | Thermal Index less than 1.0                                         | Thermal Index less than 1.0                                                   |                                          |
|                                                                                               | Mechanical Index less than 1.0                                      | Mechanical Index less than 1.0                                                |                                          |
| Localization<br>Output                                                                        | Frequency: Variable 15 kHz to 50 kHz                                | Frequency: Variable 15 kHz to 50 kHz                                          | Identical                                |
|                                                                                               | Maximum current: 1.2mA                                              | Maximum current: 1.2mA                                                        |                                          |
| ECG & EGM<br>Input                                                                            | Bandwidth: 0.05 Hz to 500 Hz                                        | Bandwidth: 0.05 Hz to 500 Hz                                                  | Identical                                |
|                                                                                               | Resolution: +/-1uV                                                  | Resolution: +/-1uV                                                            |                                          |
|                                                                                               | Timing Accuracy: +/-1.6 microsecond                                 | Timing Accuracy: +/-1.6 microsecond                                           |                                          |
| Table 3: Comparison of Technological Characteristics Against the Predicate Device (Continued) |                                                                     |                                                                               |                                          |
| Characteristics                                                                               | Subject Device                                                      | Predicate Device                                                              | Rationale for Substantial<br>Equivalence |
|                                                                                               | AcQMap® High Resolution Imaging and Mapping<br>System, Model 900100 | AcQMap® High Resolution Imaging and Mapping<br>System, Model 900100 (K201015) |                                          |
| Front Panel Connectio…

---

**Source:** [https://fda-staging.innolitics.com/device/K212345](https://fda-staging.innolitics.com/device/K212345)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
