# Yadu Medical Face Mask (Level 1) (K212330)

_Henan Yadu Industrial Co., Ltd. · FXX · Mar 14, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212330

## Device Facts

- **Applicant:** Henan Yadu Industrial Co., Ltd.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Mar 14, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

Y ADU Surgical Masks (Model E (ear loops, sterile), Model T (tie on straps, sterile/non-sterile/) are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. It is intended to be used in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided sterile and non sterile.

## Device Story

YADU Surgical Masks are single-use, three-layer, flat-pleated masks; intended for use by healthcare personnel and patients in clinical settings for infection control. Device consists of inner/outer spun-bond polypropylene layers and a middle melt-blown polypropylene filter layer; includes a polyvinyl chloride-coated iron wire nose piece for fit. Variants include ear loops (spandex/polyester) or tie-on straps (spun-bond polypropylene). Masks are provided sterile (EO) or non-sterile. Device acts as a physical barrier to reduce exposure to blood, body fluids, and particulates. Healthcare providers wear the mask to protect themselves and patients from cross-contamination. Benefits include reduced risk of pathogen transfer and fluid exposure during medical procedures.

## Clinical Evidence

Bench testing only. No clinical data. Performance verified against ASTM F2100-18, ASTM F2101-19 (BFE ≥95%), ASTM F2299-03 (particle filtration ≥95%), EN 14683-2019 (differential pressure <5.0 mm H2O/cm²), and ASTM F1862-17 (synthetic blood penetration at 80 mmHg). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

## Technological Characteristics

Three-layer flat-pleated mask. Materials: Spun-bond polypropylene (outer/inner), melt-blown polypropylene (filter), spandex/polyester or polypropylene (straps), PVC-coated iron wire (nose piece). ASTM F2100 Level 1. Sterilization: EO (ISO 11135-1). Dimensions: 17.5 x 9.5 cm. No software or electronic components.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Mask ([K212120](/device/K212120.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 14, 2022

HeNan YADU Industrial Co., Ltd. % Charles Shen Director Manton Business and Technology Services 37 Winding Ridge Oakland, New Jersey 07436

Re: K212330

Trade/Device Name: YADU Surgical Masks (Model E (ear loops, sterile/non-sterile), Model T (tie on straps, sterile/non-sterile)) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: February 9, 2022 Received: February 17, 2022

Dear Charles Shen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)

regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212330

Device Name

Y ADU Surgical Masks (Model E (ear loops, sterile), Model T (tie on straps, sterile/non-sterile/)

Indications for Use (Describe)

Y ADU Surgical Masks (Model E (ear loops, sterile), Model T (tie on straps, sterile/non-sterile/) are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. It is intended to be used in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided sterile and non sterile.

Type of Use (Select one or both, as applicable):

|  | Registration Use (Part 21 CFR 201.2 Subpart D) |
|--|------------------------------------------------|
|  | Over-The-Counter Use (21 CFR 201.2 Subpart C)  |

_ | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary-K212330:

This summary of 510k safety and effectiveness information is being submitted in accordance with the requirements of 21CFR 807.92

#### 1.0 Submitter Information

HeNan YADU Industrial Co., Ltd. No.234, West Jianpu Road 453400 Changyuan, Henan Province, CHINA Tel: (086) 373-2157057 Submitter's FDA Registration Number:3016452400

### Submission Correspondent

Charles Shen Manton Business and Technology Services 37 Winding Ridge Oakland, NJ 07436 Tel: 608-217-9358 Email: cyshen@aol.com

Date of Summary: March 13, 2022

#### 2.0 Device Information

| Proprietary Name: | YADU Surgical Masks                                                                   |
|-------------------|---------------------------------------------------------------------------------------|
| Common Name:      | Surgical Mask, Medical Face Mask                                                      |
| Models:           | Model E (ear loop, sterile/non-sterile), Model T (tie on straps, sterile/non-sterile) |

#### 3.0 Device Classification

| Classification Name: | Surgical Apparel |
|----------------------|------------------|
|----------------------|------------------|

Classification Regulation: 21 CFR 878.4040 Class: Class II Product Code: FXX

#### 4.0 Predicate Device Information:

| Manufacturer: | Hubei Xinxin Non-woven Co., Ltd. |
|---------------|----------------------------------|
| Product Name: | Surgical Mask                    |
| 510(K)#:      | K212120                          |

{4}------------------------------------------------

#### 5.0 Device description:

The YADU Surgical Masks are single use, three-layer, flat-pleated style with ear loops and nose piece. They have three layers structure: inner layer, and outer layer. The inner and outer layers are made of spun-bond polypropylene (SPP), and the middle layer is made of melt blown polypropylene filter.

Y ADU Surgical Masks have two variants: The model with ear loop and the model with tie-on bands. The mask is held in place over the users' mouth and nose by two elastic ear loops or tie-on bands welded to the side of the mask. The ear loops are made from spandex and polyester elastic, while tie-on bands are made from SPP straps. Neither elastic ear loops nor tie-on bands are made with natural rubber latex.

Y ADU Surgical Masks also have nose piece that allows the user to fit the mask around his/her nose. The nose piece is made of polyvinyl chloride coated iron wire.

The Surgical Masks provide Level-1 protection per ASTM F2100 definition. They have blue color and are sold in both sterile and non-sterile versions and are intended to be single use, disposable devices.

| Sterilization | Model               |
|---------------|---------------------|
| Non-sterile   | Tie-on (Model T)    |
|               | Ear loops (Model E) |
| EO Sterile    | Tie-on (Model T)    |
|               | Ear loops (Model E) |

#### 6.0 Indications for Use:

Y ADU Surgical Masks is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. It is intended to be used in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided sterile and non sterile.

#### 7.1 Comparison to Predicate Devices

Y ADU Surgical Masks, manufactured by "HeNan YADU Industrial Co., Ltd." are compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.

- (1) K212120, "Surgical Mask" Model XX009 Ear loop Level 1 and Model XX008 Tie-on Level 1, manufactured by "Hubei Xinxin Non-woven Co., Ltd."

{5}------------------------------------------------

The following table shows similarities and differences of use, design, and material between our device and the predicate devices.

| Description           | Subject Device                                                                                                                                                                                                                                                                                                                                                                                      | Predicate Device (K212120)                                                                                                                                                                                                                                                                                                                                                                         |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication<br>for Use | YADU Surgical Masks is<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids and<br>particulate material. It is intended<br>for use in infection control<br>practices to reduce the potential<br>exposure to blood and body fluids.<br>This is a single use, disposable<br>device, provided sterile and non<br>sterile. | The surgical masks are intended<br>to be worn to protect both the<br>patient and healthcare personnel<br>from transfer of<br>microorganisms, body fluids,<br>and particulate material. These<br>masks are intended for use in<br>infection control practices to<br>reduce the potential exposure to<br>blood and body fluids. This is a<br>single-use, disposable device,<br>provided non-sterile. |
| Layers                | Three                                                                                                                                                                                                                                                                                                                                                                                               | Three                                                                                                                                                                                                                                                                                                                                                                                              |
| Materials             | Outer layer is made of 100%<br>spun- bond polypropylene.<br>Middle player is made of<br>100%<br>meltblown<br>polypropylene filter media.<br>Inner layer is made of 100%<br>spun- bond polypropylene.<br>Ear-loops are made of polyester<br>and spandex.<br>Ties are made from Spunbond<br>Polypropylene<br>The nose piece is iron and plastics                                                      | Outer layer is made of 100%<br>spun- bond polypropylene.<br>Middle player is made of<br>100% meltblown<br>polypropylene filter media.<br>Inner layer is made of 100%<br>spun- bond polypropylene.<br>Ear-loops are made of<br>spandex<br>Ties are made from Spunbond<br>Polypropylene<br>The nose piece is iron and<br>plastics                                                                    |
| Dimensions            | 17.5 x 9.5cm                                                                                                                                                                                                                                                                                                                                                                                        | 17.5 x 9.5cm                                                                                                                                                                                                                                                                                                                                                                                       |
| Mask style            | Flat Pleated                                                                                                                                                                                                                                                                                                                                                                                        | Flat Pleated                                                                                                                                                                                                                                                                                                                                                                                       |
| Design                | Ear Loop and Tie-on Bands                                                                                                                                                                                                                                                                                                                                                                           | Ear Loop (XX009) and Tie-on<br>(XX008)                                                                                                                                                                                                                                                                                                                                                             |
| Color                 | Blue                                                                                                                                                                                                                                                                                                                                                                                                | Blue                                                                                                                                                                                                                                                                                                                                                                                               |
| Sterility             | Sterile and Non-sterile                                                                                                                                                                                                                                                                                                                                                                             | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                        |
| Shelf life            | Sterile: 3 years<br>Non-sterile: None stated                                                                                                                                                                                                                                                                                                                                                        | Sterile: N/A<br>Non-sterile: 2 years                                                                                                                                                                                                                                                                                                                                                               |
| Single Use            | Yes                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                |
| ASTM<br>F2100 Level   | Level 1                                                                                                                                                                                                                                                                                                                                                                                             | Level 1                                                                                                                                                                                                                                                                                                                                                                                            |

Table 1: Comparison of Intended Use, Design, and Material

{6}------------------------------------------------

The differences between the subject device and predicate device do not raise different questions of safety and effectiveness.

#### Non-Clinical Study Summary 8.1

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

- . Sterilization Validation: ISO-11135-1: 2014
- ASTM F2100: 2018 Standard Specification for Performance of Materials Used in Medical Face Masks
- ASTM F2101-19 Standard Test Method for Evaluating the Bacterial Filtration ● Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
- ASTM F1862-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
- ASTM F2299-03 (Reapproved 2017) Standard Test Method for Determining the ● Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
- EN 14683-2019 Medical Face Masks Requirements and Test Methods ●
- In Vitro Cytotoxicity Test: ISO 10993-5: 2009 ●
- Skin Irritation Test: ISO 10993-10: 2010 ●
- Skin Sensitization Test: ISO 10993-10: 2010 ●

ASTM F2100 Testing for Level 1

|  | Table 2 - Sterile masks, ear loops, 3 years real time aged |
|--|------------------------------------------------------------|
|--|------------------------------------------------------------|

| Name of the test                                                      | Purpose                                                      | Acceptance<br>Criteria | Results    |            |            |
|-----------------------------------------------------------------------|--------------------------------------------------------------|------------------------|------------|------------|------------|
|                                                                       |                                                              |                        | Lot 1      | Lot 2      | Lot 3      |
| ASTM F2101-19<br>Bacterial Filtration<br>Efficiency (BFE)             | To evaluate the<br>bacterial efficiency<br>of the mask       | $\geq$ 95%             | 95.1-97.4% | 95.1-98.4% | 95.1-98.4% |
| ASTM F2299– 03<br>(Reapproved 2017)<br>Penetration by<br>Particulates | To evaluate particle<br>filtration efficiency<br>of the mask | $\geq$ 95%             | 95.1-97.9% | 95.1-99.6% | 95.1-97.4% |

{7}------------------------------------------------

| EN 14683-2019<br>Differential<br>Pressure | To evaluate<br>differential<br>pressure of the<br>device | < 5.0 mm<br>H2O/cm²                                                 | 3.7-4.2                                                           | 3.7-4.1                                | 3.7-4.1    |            |            |
|-------------------------------------------|----------------------------------------------------------|---------------------------------------------------------------------|-------------------------------------------------------------------|----------------------------------------|------------|------------|------------|
|                                           |                                                          | ASTM F1862-17<br>Resistance to<br>penetration by<br>Synthetic Blood | To evaluate<br>resistance to<br>penetration by<br>Synthetic Blood | No<br>penetration<br>under 80 mm<br>Hg | 32/32 pass | 32/32 pass | 32/32 pass |

### Table 3 - Sterile masks, tie on straps, 3 years real time aged*

| Name of the test | Purpose      | Acceptance Criteria | Results   |           |           |
|------------------|--------------|---------------------|-----------|-----------|-----------|
| 16 CFR part 1610 | Flammability | Class 1             | 32/32 IBE | 32/32 IBE | 32/32 IBE |

### Table 4 - Non-sterile masks, ear loops, unaged

| Name of the test                                                     | Purpose                                                           | Acceptance Criteria                    | Results           |                   |                  |
|----------------------------------------------------------------------|-------------------------------------------------------------------|----------------------------------------|-------------------|-------------------|------------------|
|                                                                      |                                                                   |                                        | Lot 1             | Lot 2             | Lot 3            |
| ASTM F2101-19<br>Bacterial Filtration<br>Efficiency (BFE)            | To evaluate the<br>bacterial efficiency<br>of the mask            | ≥ 95%                                  | 99.8-99.9%        | 99.2-99.9%        | 99.8-99.9%       |
| ASTM F2299–03<br>(Reapproved 2017)<br>Penetration by<br>Particulates | To evaluate particle<br>filtration efficiency<br>of the mask      | ≥ 95%                                  | 99.5 – 99.9<br>%  | 99.5 – 99.9<br>%  | 99.5 – 99.9<br>% |
| EN 14683-2019<br>Differential<br>Pressure                            | To evaluate<br>differential<br>pressure of the<br>device          | < 5.0 mm<br>H2O/cm2                    | 3.1 – 3.6<br>N=32 | 3.1 – 3.5<br>N=32 | 3.0 – 3.<br>N=32 |
| ASTM F1862-17<br>Resistance to<br>penetration by<br>Synthetic Blood  | To evaluate<br>resistance to<br>penetration by<br>Synthetic Blood | No<br>penetration<br>under 80 mm<br>Hg | 32/32 pass        | 32/32 pass        | 32/32 pass       |

### Table 5 - Non-sterile mask, tie on straps, unaged*

| Name of the test | Purpose      | Acceptance<br>Criteria | Results   |           |           |
|------------------|--------------|------------------------|-----------|-----------|-----------|
|                  |              |                        | Lot 1     | Lot 2     | Lot 3     |
| 16 CFR part 1610 | Flammability | Class 1                | 32/32 IBE | 32/32 IBE | 32/32 IBE |

* Testing on masks with ear loops is representative of masks with tie on straps for all tests but flammability. Therefore, flammability was conducted on masks with tie on straps.

{8}------------------------------------------------

# Table 6 – Biocompatibility

| Identification of test | Purpose                             | Acceptance criteria | Results Sterile Mask | Results Non-Sterile Mask |
|------------------------|-------------------------------------|---------------------|----------------------|--------------------------|
| ISO 10993-5            | Assess the cytotoxicity of the mask | No Cytotoxicity     | No Cytotoxicity      | No Cytotoxicity          |
| ISO 10993-10           | Assess sensitization of the mask    | No Sensitization    | No Sensitization     | No Sensitization         |
| ISO 10993-10           | Assess irritation of the mask       | No Irritation       | No Irritation        | No Irritation            |

#### 9.0 Clinical Study Summary

Clinical data is not needed for YADU Surgical Masks

#### 10.0 Conclusion

The conclusion drawn from the non-clinical tests demonstrates that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K212330](https://fda-staging.innolitics.com/device/K212330)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).