Medical Surgical Mask

K212299 · Wepon Medical Technology Co., Ltd. · FXX · Jan 4, 2022 · General, Plastic Surgery

Device Facts

Record IDK212299
Device NameMedical Surgical Mask
ApplicantWepon Medical Technology Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateJan 4, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Story

Single-use, flat-pleated surgical mask; provides barrier protection against microorganisms, body fluids, and particulates. Constructed of polypropylene non-woven outer/inner layers and meltblown polypropylene filter layer; includes adjustable nose clip (polypropylene/iron) and spandex ear loops. Worn by healthcare personnel and patients in clinical settings to reduce infection risk. Device functions as physical barrier; performance validated against ASTM F2100-19 Level 2 and Level 3 standards for bacterial/particulate filtration, breathability (differential pressure), fluid resistance, and flammability. Non-sterile, disposable.

Clinical Evidence

No clinical data. Bench testing only: Bacterial Filtration Efficiency (ASTM F2101), Differential Pressure (EN14683), Sub-micron Particulate Filtration Efficiency (ASTM F2299), Synthetic Blood Penetration (ASTM F1862), and Flammability (16 CFR Part 1610). Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation).

Technological Characteristics

Flat-pleated mask; 3-layer construction (polypropylene non-woven/meltblown/non-woven). Nose clip: polypropylene/iron. Ear loops: spandex. ASTM F2100-19 Level 2 and 3 compliant. Biocompatibility: ISO 10993-5, ISO 10993-10. Non-sterile.

Indications for Use

Indicated for use by patients and healthcare personnel to protect against transfer of microorganisms, body fluids, and particulate material; intended for infection control to reduce exposure to blood and body fluids.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 4, 2022 Wepon Medical Technology Co., Ltd. Yongxian Sun Manager Representative West side of North Baizhang Road, Chengdong Street Wenling, Zhejiang 317500 China Re: K212299 Trade/Device Name: Medical Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 21, 2021 Received: December 2, 2022 Dear Yongxian Sun: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212299 Device Name Medical Surgical Mask ### Indications for Use (Describe) The Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------|-----------------------------------------------------------| | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. {3}------------------------------------------------ # 510(k) Summary K212299 ### I Submitter - Device submitter: Wepon Medical Technology Co., Ltd. West side of North Baizhang Road, Chengdong Street, Wenling City, Zhejiang Province, China - Contact person: Yongxian Sun Management Representative Phone: +86 13586268021 E-mail: syx@wepon.cc # II Proposed Device | 510(k) Number: | K212299 | |--------------------|-----------------------| | Trade/Device Name: | Medical Surgical Mask | | Regulation Number: | 21 CFR 878.4040 | | Regulation Name: | Surgical Apparel | | Regulatory Class: | Class II | | Product code: | FXX | | Review Panel: | General Hospital | # III Predicate Devices | 510(k) Number: | K210030 | |--------------------|---------------------------------------------| | Trade/Device Name: | Medical Surgical Mask | | Regulation Number: | 21 CFR 878.4040 | | Regulation Name: | Surgical Apparel | | Classification: | Class II | | Product Code: | FXX | | Manufacturer: | Tianjin Aoshang Outdoor Equipment Co., Ltd. | #### IV Device description The Medical Surgical Mask is single use, flat-pleated masks that are provided in blue. The Medical Surgical Mask is available in two types, which are Level 3 based on ASTM F2100-19. The inner and outer layers of the mask are made of polypropylene non-woven fabric, the middle layer is made of polypropylene meltblown fabric, and the nose clip is made of polypropylene and iron wire. Users can adjust the nose clip according to the shape of the nose, and fix the mask on the bridge of the nose to prevent the mask from falling off. The ear loops are made of spandex. The ear loops are held in place over the users' mouth and nose by two ear loops welded to the mask. {4}------------------------------------------------ This is a single use, disposable device(s), provided non-sterile. # V Indication for use The Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. # VI Non-Clinical Test Conclusion The Medical Surgical Mask was tested in accordance with the tests recommended in the FDA guidance document, Guidance for Industry and FDA Staff Surgical Masks - Premarket Notification [510(k)] Submission issued March of 2004. Based upon the guidance document the following testing has been performed. | Performance Testing | | | | | |-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|--------------------------------------------|--------------------------------------------|--------| | Test Methodology | Purpose | Acceptance Criteria for<br>Level 2 Barrier | Acceptance Criteria for<br>Level 3 Barrier | Result | | Bacterial Filtration<br>Efficiency ASTM<br>F2101 | Measure bacterial<br>filtration efficiency | ≥98% | ≥98% | Passed | | Differential Pressure<br>(mm H2O/cm2)<br>EN14683:2019 Annex C | Determine<br>breathability of a<br>mask | <6.0 H2O/cm² | <6.0 H2O/cm² | Passed | | Sub-micron Particulate<br>Filtration Efficiency<br>ASTM F2299/<br>F2299M-03 | Measure initial<br>particle<br>filtration efficiency | ≥98% | ≥98% | Passed | | Resistance to<br>Penetration by Synthetic<br>Blood ASTM F1862/F<br>1862M-2017 | Evaluate the<br>resistance to<br>penetration by<br>impact of small<br>volume of synthetic<br>blood | 120 mmHg | 160 mmHg | Passed | | Flame spread 16CFR<br>Part 1610-2008 | Response of<br>materials to heat and<br>flame | Class I | Class I | Passed | # Biocompatibility Testing {5}------------------------------------------------ The biocompatibility evaluation for the Medical Surgical Mask was conducted in accordance with ISO 10993-1:2018 Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing within a Risk Management Process, as recognized by FDA. The Medical Surgical Mask is classified as a surface contacting device. Specific biocompatibility tests were selected under the guidance of ISO 10993-1:2018 Annex A. | Biocompatibility Evaluation | | | | | |-----------------------------|-------------------|--------------|-----------------------|--------| | | Biological Effect | Standard | Results | | | 1 | Cytotoxicity | ISO 10993-5 | Not cytotoxic | Passed | | 2 | Sensitization | ISO 10993-10 | Non sensitizing | Passed | | 3 | Irritation | ISO 10993-10 | Negligibly irritating | Passed | # VII Clinical Test Conclusion No clinical study is included in this submission. # VIII Summary of Technological characteristics Table 1 Comparison of Medical Surgical Mask | Item | Subject device | Predicate device<br>(K210030) | Discussion | | |-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | Product name | Medical Surgical Mask | Medical Surgical Mask | NA | | | Product Code | FXX | FXX | identical | | | Intended use | The Medical Surgical Mask is<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids<br>and particulate material. These<br>surgical masks are intended<br>for use in infection control<br>practices to reduce the<br>potential exposure to blood<br>and body fluids. This is a<br>single use, disposable<br>device(s), provided<br>non-sterile. | The Medical Surgical Masks<br>are intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These surgical masks<br>are intended for use in<br>infection control practices to<br>reduce the potential exposure<br>to blood and body fluids. This<br>is a single use, disposable<br>device(s), provided non-sterile. | identical | | | Mask style | Flat-pleated | Flat-pleated | identical | | | ASTM F2100 Level | Level 2 and Level 3 | Level 2 and Level 3 | identical | | | Design feature | Level 2: Ear loop | Level 2: Ear loop | identical | | | | | Level 3: Ear loop | Level 3: Ear loop and Tie-on | | | Color | | Blue | Blue | identical | | Dimension | | 17.5cmx9.5cm | 17.5cmx9.5cm | identical | | Sterility | | Non-Sterile | Non-Sterile | identical | | Use | | Single Use, Disposable | Single Use, Disposable | identical | | Particulate<br>filtration<br>efficiency | | Level 2 and Level 3: average 99.96% | Level 2 mask: average 99.71%<br>Level 3 mask: average 99.93% | different | | Bacterial filtration<br>efficiency | | Level 2 and Level 3: average<br>99.9% | Level 2 mask: average 99.7%<br>Level 3 mask: average 99.9% | different | | Differential pressure | | Level 2 and Level 3: average<br>$2.8 mmH_2O/cm_2$ | Level 2 mask: average<br>$2.8mmH_2O/cm_2$ | different | | | | | Level 3 mask: average 4.0 | | | | | | $mmH_2O/cm_2$ | | | | | | EN 14683 | | | | | | | | | Flammability | | Class 1 | Class 1 | identical | | Fluid resistance | | Level 2: Pass at 120mmHg<br>Level 3: Pass at 160mmHg | Level 2: Pass at 120mmHg<br>Level 3: Pass at 160mmHg | identical | | | | | | | | Label/Labeling | | Complied with 21 CFR part<br>801 | Complied with 21 CFR part<br>801 | identical | | | | | | | | Patient Contacting Material | Outer facing<br>layer | polypropylene non-woven<br>fabric | Spunbond Polypropylene | Different | | | Middle layer | polypropylene meltblown<br>fabric | Meltblown Polypropylene<br>Filter | | | | Inner facing<br>layer | polypropylene non-woven<br>fabric | Spunbond Polypropylene | | | | Nose clip | polypropylene and iron wire | PE and Iron | | | | Ear loops | spandex | Spandex | | | | Biocompatibility | | ISO 10993-5 and ISO<br>10993-10;<br>Under the conditions of the<br>study, the proposed device<br>extract was determined to be<br>non-cytotoxic, non-sensitizing,<br>and non-irritating. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | {6}------------------------------------------------ Different - Particulate filtration efficiency The test result for particulate filtration efficiency for the proposed device is different from the two predicate devices. However, the test result for the proposed device can meet the requirements of level 2 mask and Level 3 mask. Thus, this difference will not affect the safety and effectiveness between the {7}------------------------------------------------ proposed device and the two predicate devices. #### Different - Bacterial filtration efficiency The test result for bacterial filtration efficiency for the proposed device is different from the two predicate devices. However, the test result for the proposed device can meet the requirements of level 2 mask and Level 3 mask. Thus, this difference will not affect the safety and effectiveness between the proposed device and the two predicate devices. #### Different - Differential pressure The test result for differential pressure for the proposed device is different from the two predicate devices. However, the test result for the proposed device can meet the requirements of level 3 mask based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness between the proposed device and the two predicate devices. #### Different - Patient Contacting Material The patient contacting material for the proposed device is different from the two predicate devices. However, biocompatibility test has been performed on the proposed device and the results does not show any adverse effect. Thus, this difference will not affectiveness between the proposed device and the two predicate devices. #### IX Conclusion The proposed device has the same the intended use as the predicate device. It presents similar technological characteristics as the predicate device including the performance parameters and biocompatibility. The non-clinical testing demonstrates that the device is as effective and performs as well as the legally marketed device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...