K212288 · Kai Medtech, LLC · QJP · Jun 2, 2022 · Cardiovascular
Device Facts
Record ID
K212288
Device Name
Minjie Catheter System
Applicant
Kai Medtech, LLC
Product Code
QJP · Cardiovascular
Decision Date
Jun 2, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Minjie Catheter System is indicated for the introduction of interventional devices into the peripheral and neurovasculature.
Device Story
The Minjie Catheter System is a single-lumen, flexible, variable-stiffness composite catheter with a Nitinol structure, used to facilitate the introduction and selective placement of interventional devices into target blood vessels in the peripheral and neurovasculature. It features a radiopaque marker band for fluoroscopic visualization and a hydrophilic coating on the distal section to reduce friction. The device is operated by physicians in a clinical setting. During a procedure, the catheter is navigated through the vasculature; the physician monitors its position via fluoroscopy. The device serves as a conduit for other interventional tools, aiding in reaching target sites such as the M1 and M2 segments of the middle cerebral artery. By providing a stable, trackable pathway, it enables precise delivery of therapeutic devices, potentially improving procedural outcomes.
Clinical Evidence
No clinical testing was conducted. Safety and functionality were validated through non-clinical bench testing (including simulated use, mechanical integrity, and biocompatibility) and an animal study comparing the subject device to the predicate. Bench testing confirmed compliance with ISO 10555-1 and ISO 80369 standards.
{0}------------------------------------------------
June 2, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Kai Medtech, LLC Ricardo Olivo Senior Director, Quality and Reguatory Affairs 22651 Lambert Street, Suite 107 Lake Forest, California 92630
Re: K212288
Trade/Device Name: Minjie Catheter System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, DQY Dated: April 29, 2022 Received: May 3, 2022
Dear Ricardo Olivo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212288
Device Name Minjie Catheter System
Indications for Use (Describe)
The Minjie Catheter System is indicated for the introduction of interventional devices into the peripheral and neruovasculature.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K212288
This 510(k) Summary for the Minjie Catheter System is submitted in accordance with the requirements of 21 CFR Part 807.92 and following the recommendations outlined in FDA guidance, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]," issued on July 28, 2014.
#### Submitter
| Submitter: | Kai Medtech, LLC<br>22651 Lambert Street, Suite 107<br>Lake Forest, CA 92630 USA |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ricardo Olivo<br>Senior Director, Quality and Regulatory Affairs<br>+1 (949) 767-8960, Extension 109<br>ricardo.olivo@kaimtgroup.com |
| Date Prepared: | June 01, 2022 |
## Device Name and Details
| Trade Name of Device: | Minjie Catheter System |
|------------------------------------|----------------------------------------------------------|
| Common Name of<br>Device: | Distal Access Catheter |
| Device Classification: | Class II |
| Device Classification<br>Name: | Percutaneous Catheter |
| Regulation Number/<br>Description: | 21 CFR 870.1250, Percutaneous Catheter |
| Classification<br>Product Code: | Primary Product Code: QJP<br>Secondary Product Code: DQY |
| Review Panel: | Neurology |
{4}------------------------------------------------
| 510(k)<br>Number | Product<br>Code | Trade Name of Device | Device Manufacturer |
|--------------------------|-----------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------|
| <b>Predicate Device</b> | | | |
| K150107 | DQY | Arc™ Intracranial Support<br>Catheter and Arc™ Mini<br>Intracranial Support<br>Catheter | Micro Therapeutics, Inc.<br>d/b/a ev3 Neurovascular |
| <b>Reference Devices</b> | | | |
| K161152 | DQY | Navien Intracranial Support<br>Catheter | Micro Therapeutics, Inc.<br>d/b/a ev3 Neurovascular |
| K140080 | DQY | ENVOY® Distal Access (DA)<br>Guiding Catheter | Codman & Shurtleff, Inc. |
Table 1: Legally Marketed Predicate and Reference Devices
## Device Description
The Minjie Catheter System consists of the Minjie Catheter, Introducer Peel Away Tool and packaging hoop with Elbow Flush Luer. The Minjie Catheter is a single lumen. flexible, variable stiffness composite catheter with a Nitinol structure. A radiopaque marker band on the distal tip of the catheter is used for visualization under fluoroscopy. The distal section of the catheter is coated with a hydrophilic coating, to reduce friction during intravascular use. The Minjie Catheter System dimensions are included in the individual device label. The devices are supplied sterile and are intended for single use only.
## Indications for Use Statement
The Miniie Catheter System is indicated for the introduction of interventional devices into the peripheral and neurovasculature.
## Comparison of Technological Characteristics with the Predicate Device
A comparison of the technological characteristics of the subject device (Minjie Catheter System) and predicate device (Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter) demonstrates that the technological characteristics of the subject Minjie Catheter System are substantially equivalent to the technological characteristics of the predicate Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter previously cleared under K150107. Refer to Table 2 below for a comparison between the Minjie Catheter System and Arc™ Intracranial Support Catheter and Arc™ Mini Intracranial Support Catheter.
{5}------------------------------------------------
| | Predicate Device, ArcTM<br>Intracranial Support Catheter<br>and ArcTM Mini Intracranial<br>Support Catheter (K150107) | Subject Device, Minjie Catheter<br>System (K212288) | Comparison |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Regulation | DQY (21 CFR 870.1250) | QJP, DQY (21 CFR 870.1250) | Similar |
| Indications for Use<br>Statement | The ArcTM Intracranial<br>Support Catheter and ArcTM<br>Mini Intracranial Support<br>Catheter are indicated for<br>the introduction of<br>interventional<br>devices into the peripheral<br>and neurovasculature. | The Minjie Catheter System is<br>indicated for the introduction of<br>interventional devices into the<br>peripheral and neurovasculature. | Same |
| Function /<br>Principle of Operation | Facilitate introduction and<br>selective placement of<br>interventional devices into target<br>blood vessels in the peripheral<br>and neuro vasculature. | Facilitate introduction and<br>selective placement of<br>interventional devices into<br>target blood vessels in the<br>peripheral and neuro<br>vasculature | Similar |
| Catheter Shaft<br>Materials | PTFE lined polymeric catheter<br>with hydrophilic coating | PTFE lined polymeric catheter<br>with hydrophilic coating | Similar |
| Catheter Shaft<br>Support | Nitinol | Nitinol | Similar |
| Marker Band | Platinum | Platinum/Iridium | Testing<br>demonstrated<br>that the<br>difference does<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Usable Length | 132cm - 135cm | 115cm - 131cm | Similar |
| Coating Length | 40cm - 45cm | 60cm | Testing<br>demonstrated<br>that the<br>difference does<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Distal Inner Diameter<br>(ID) | 0.061" | 0.068" (1.73mm) | Testing<br>demonstrated<br>that the<br>difference does<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Distal Outer Diameter<br>(OD) (Max) | 0.071" | 0.081" (2.06mm) | Testing<br>demonstrated<br>that the |
| | | | |
| Proximal ID | 0.069" | 0.068" (1.73mm) | Similar |
| Proximal OD (Max) | 0.082" | 0.083" (2.11mm) | Similar |
| Shaft | Progressively softer from proximal end to distal tip | Progressively softer from proximal end to distal tip | Similar |
| Tip Configuration | Single, straight flexible tip | Single, straight flexible tip | Similar |
| Guidewire<br>Compatibility | Can be navigated over<br>guidewire with maximum OD<br>of 0.038" | Can be navigated over<br>guidewire with maximum<br>OD of 0.035" | Testing<br>demonstrated<br>that the<br>difference does<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Packaging | Catheter in polyethylene hoop<br>attached to packaging card inside<br>PET / PE / Tyvek pouch inside<br>SBS carton | Catheter in polyethylene hoop<br>attached to packaging card<br>inside PET / PE / Tyvek pouch<br>inside SBS carton | Similar |
| Accessories | Introducer Tool | Introducer Peel Away Tool | Testing<br>demonstrated<br>that the<br>difference does<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Sterilization Method | Ethylene oxide (EO) gas | EO gas | Similar |
| | | | difference does<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
Table 2: Comparison between Minjie Catheter System and Arc™ Intracranial Support
Catheter and Arc™ Mini Intracranial Support Catheter.
Page 3 of 8
{6}------------------------------------------------
### Performance Data (Non-Clinical Testing)
### Sterilization and Shelf-Life
The packaged Minjie Catheter System is sterilized using a validated EO sterilization cycle. The sterilization cycle has been validated to ensure a sterility assurance level (SAL) of 106 in accordance with ISO 11135:2014, "Sterilization of health care products - Ethylene oxide - Requirements for the development, validation, and routine control Endotoxin testing was conducted in accordance with FDA guidance, "Pyrogen and Endotoxins Testing: Questions and Answers," issued in June 2012.
Aging studies for the Minjie Catheter System have established that the product remains functional, and packaging maintains sterility for up to twenty-four months. Aging studies for packaging integrity and product stability (device functionality) were performed and met all acceptance criteria.
## Biocompatibility Testing
Biocompatibility testing was performed in accordance with FDA guidance. "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part Page 4 of 8
{7}------------------------------------------------
1: Evaluation and testing within a risk management process"," issued on September 4, 2020, and ISO 10993-1:2018, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process," for the Minjie Catheter System categorized as a limited exposure (< 24 hours), external communicating device contacting circulating blood. Refer to Table 3 below for summary of the test results.
| Test Standard and Study<br>Name | Test Results | Conclusions |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Cytotoxicity per ISO 10993-<br>5; ISO Elution Method | The test article extract met the<br>requirements of the test because<br>the grade was less than a grade 2<br>(mild reactivity). | Non-cytotoxic |
| Sensitization per ISO 10993-<br>10; ISO Guinea Pig<br>Maximization Sensitization<br>Test | Animals tested with the test extract<br>did not show evidence of delayed<br>dermal contact sensitization. | Non-sensitizer |
| Intracutaneous Irritation per<br>ISO 10993-10; ISO<br>Intracutaneous Study in<br>Rabbits | Animals tested with the test extract<br>exhibited similar edema and<br>erythema scores compared to the<br>negative control. | Non-irritant |
| Acute Systemic Toxicity per<br>ISO 10993-11; ISO Acute<br>Systemic Toxicity Study in<br>Mice | No animals injected with the test article<br>showed a significantly greater<br>biological reaction than the animals<br>treated with the control article. | No acute systemic<br>toxicity |
| Material Mediated Pyrogen<br>per ISO 10993-11; USP<br>Rabbit Pyrogen Study,<br>Material Mediated | No single animal showed a<br>temperature rise of ≥ 0.5°C above its<br>baseline temperature. Total rise of<br>rabbit temperatures during three (3)<br>hour observation period was within<br>acceptable USP requirements. The<br>test article met the requirements for<br>the absence of pyrogens. | Non-pyrogenic |
| Hemocompatibility per<br>ISO 10993-4; ASTM<br>Hemolysis Study | The hemolytic index for the test article<br>in direct contact with blood was 0.0%,<br>and for test article extract was 0.7%. | Non-hemolytic |
| Hemocompatibility per ISO<br>10993-4; SC5b-9<br>Complement Activation<br>Assay | SC5b-9 Concentration of the test<br>article was not statistically different<br>than activated non-human serum<br>control. | Non-activator of the<br>complement system |
| Hemocompatibility per<br>ISO 10993-4; ASTM<br>Partial Thromboplastin<br>Time (PTT) with Sponsor<br>Provided Control | Inactivated partial thromboplastin time<br>test was conducted, and the results<br>showed no different compared to the<br>reference device, ENVOY ® Distal<br>Access (DA) Guiding Catheter,<br>K140080. | Non-activator of<br>coagulation |
| Platelet and Leukocyte<br>Counts (Platelet Adhesion) | The leukocyte and platelet counts that<br>were exposed to the negative control | Non-activator of<br>platelet and |
| per ISO 10993-4: 2017 | subject device and predicate device (ArcTM Intracranial Support Catheter, K150107) extracts did not show difference. | leukocyte |
Table 3: Summary of Biocompatibility Testing Results
{8}------------------------------------------------
## Bench Testing
A summary of the non-clinical bench testing performed for the Minjie Catheter System is presented in Table 4 below.
| Test Name | Test Method Summary | Results |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Catheter Visual<br>Inspection | Fully assembled devices were inspected with an<br>unaided eye for dents, kinks, cracks or other<br>damage or anomalies that may impact function of<br>device. | All devices met<br>acceptance criteria. |
| Coating<br>Integrity | Fully assembled devices were inspected under a<br>minimum 40X magnification before and after<br>simulated use and particulate testing, for coating<br>defects. | All devices met<br>acceptance criteria. |
| Particulate<br>Testing | The device was evaluated for particulate generation<br>under simulated use in a representative tortuous<br>anatomical model. | All devices met<br>acceptance criteria. |
| Simulated Use | The device was repeatedly navigated through a<br>tortuous benchtop model to assess compatibility<br>with accessories and interventional devices<br>including guidewire, guide catheter, microcatheter<br>and stent retriever, and assess device stability and<br>ability to navigate to the M1 and M2 segments of<br>the middle cerebral artery (MCA). | All devices met<br>acceptance criteria. |
| Coating<br>Frictional<br>Forces /<br>Durability | Device coating was evaluated for frictional forces,<br>and durability via repeated navigation through<br>simulated use test model. | All devices met<br>acceptance criteria. |
| Distal Tip<br>Buckling | Tip buckling was evaluated, including testing the<br>distal tip under compressive loads at 5 mm, 10 mm<br>and 20 mm from the distal tip, to evaluate stiffness<br>of the distal tip. | All devices met<br>acceptance criteria. |
| Kink Resistance | The device was wrapped around several mandrels of<br>clinically relevant diameters and inspected for<br>kinks. | All devices met<br>acceptance criteria. |
| Liquid Leakage<br>under pressure | The device was tested for liquid leakage under<br>pressure per ISO 10555-1, Annex C. | All devices met<br>acceptance criteria. |
| Hub Air<br>Aspiration Leak | The device was tested per ISO 10555-1, Annex<br>D for hub air aspiration leak. | All devices met<br>acceptance criteria. |
| Torque to<br>Failure | The device was tested in a simulated use test model<br>for full-length torque strength to determine the | All devices met<br>acceptance criteria. |
| Test Name | Test Method Summary | Results |
| | number of rotations to failure. | |
| Manual<br>Injection /<br>Peak Pressure | The device was tested for full-length, after<br>simulated use testing, using manual syringe<br>injection of worst-case contrast media, to verify<br>that device can withstand pressures that may be<br>generated during manual delivery of contrast<br>media. | All devices met<br>acceptance criteria. |
| Static / Dynamic<br>Burst | The device was tested under full-length static<br>conditions to burst using worst-case contrast media<br>per ISO 10555-1, Annex F, and compared against<br>burst pressures exhibited during manual syringe<br>injection of contrast media. | All devices met<br>acceptance criteria. |
| Luer Hub<br>Compatibility | Luer hub testing was performed per ISO 80369-7<br>and ISO 80369-20. | All devices met<br>acceptance criteria. |
| Dimensional<br>Inspection | The usable length, proximal and distal inner and<br>outer diameters were measured and recorded. | All devices met<br>acceptance criteria. |
| Lumen Patency | The total length of the device must allow to pass a<br>mandrel of the required size from the proximal hub<br>to distal tip. | All devices met<br>acceptance criteria. |
| Shaft Peak<br>Tensile Force | Shaft peak tensile strength was tested to failure at<br>the distal tip section, and each joint of the device,<br>per ISO 10555-1, Annex B test methods. | All devices met<br>acceptance criteria. |
| Hub Peak<br>Tensile Force | Hub-shaft joint peak tensile strength was tested to<br>failure, per ISO 10555-1, Annex B test methods. | All devices met<br>acceptance criteria. |
| Physician<br>Usability<br>Testing | The device was navigated through a tortuous<br>benchtop model to assess compatibility with<br>accessories, device stability, ability to inject saline,<br>and the user's ability to navigate to the M1 and M2<br>segments of the MCA. | All devices met<br>acceptance criteria. |
Table 4: Summary of Bench Testing Results
{9}------------------------------------------------
### Animal Testing
An animal study was performed to verify the safety and validate the functionality of the Minijie Catheter System in comparison to the predicate device, ArcTM Intracranial Support Catheter. The Minjie Catheter System yielded acceptable usability criteria ratings to validate functionality. As confirmed by angiographic and histological evaluations, the Minije Catheter System was verified to be safe for its intended use.
## Clinical Testing
No clinical testing was conducted. The differences in technological characteristics do not raise new questions of safety and effectiveness as demonstrated through nonclinical bench testing using well-established acceptable scientific methods.
{10}------------------------------------------------
## Conclusion
The Minjie Catheter System has the same intended use as the predicate device and the technological characteristics are similar to the predicate device. The differences do not raise new questions of safety and effectiveness. The information and testing presented in this 510(k) submission demonstrate that the Minjie Catheter System is substantially equivalent to the predicate device, Arc™ Intracranial Support Catheter, for use in the peripheral and neurovasculature.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.