BioRoot Flow is a hydrophilic mineral root canal sealer supplied as a ready-to-use paste in a syringe. It utilizes a tricalcium silicate reaction with in situ water to seal the root canal. Used in dentistry by dental healthcare professionals in combination with gutta-percha points to fill root canals. The device ensures adaptation to the root canal, a tight interface with gutta-percha, and radiopacity. It is an implantable device intended for permanent root canal filling. The ready-to-use paste format eliminates the manual mixing required by previous powder/liquid systems, simplifying the clinical workflow for the dentist while maintaining equivalent chemical and physical performance to predicate sealers.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing (physical-chemical properties, setting time, sealing, adhesion) and biocompatibility testing (chemical characterization and toxicological risk assessment). Clinical performance is bridged to the primary predicate, BioRoot RCS, based on identical intended use, site of application, patient population, and mechanical/biological properties.
Technological Characteristics
Hydrophilic mineral-based paste. Composition: Tricalcium silicate, Zirconium oxide, Calcium carbonate, Propylene glycol, Polyvidone K90, Sepineo P600, Aerosil R1812S. Principle: Tricalcium silicate reaction with in situ water. Form factor: Syringe-delivered paste. Setting time: 1h to 6h. Non-sterile.
Indications for Use
Indicated for permanent root canal filling in combination with gutta-percha points in patients with inflamed or necrotic pulp, or following a retreatment procedure. For use in permanent and mature teeth by dental healthcare professionals.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 16, 2021
Septodont Greg Montgomery Business Operations Manager 205 Granite Run Drive, Suite 150 Lancaster, Pennsylvania 17601
Re: K212283
Trade/Device Name: BioRoot Flow 0.5g, BioRoot Flow 2g Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF Dated: July 19, 2021 Received: July 21, 2021
Dear Greg Montgomery:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212283
Device Name BioRoot Flow 0.5g, BioRoot Flow 2g
#### Indications for Use (Describe)
- · Permanent root canal filling in combination with gutta-percha points in case of inflamed or necrotic pulp.
- · Permanent root canal filling in combination with gutta-percha points following a retreatment procedure.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span> <b> <div style="display:inline-block; width:15px; height:15px;">☑</div> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <b> <div style="display:inline-block; width:15px; height:15px;">☐</div> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5. 510(K) Summary K212283
#### Septodont BioRoot™ Flow 0.5q. BioRoot™ Flow 2q (per 21 CFR 807.93)
#### 1. SUBMITTER/510(K) HOLDER
Septodont 58, rue du Pont De Créteil Saint-Maur des Fossés Cedex Phone: 331-49-76-71-16 Establishment Registration No: 9610964
Contact: Leslie Fillion, Regulatory Affairs Manager Contact Phone: 1-519-623-4800 Ext. 6354 Contact Email: Ifillion@septodont.com Date Prepared: July 13, 2021
## 2. DEVICE NAME
BioRoot™ Flow 0.5g, BioRoot™ Flow 2g Device Name: Device Classification Name: Resin, Root Canal Filling/Dental Cement Regulation Number: Primary: 872.3820, Secondary: 872.3250 Classification Product Code: KIF Device Class: Class II
## 3. PREDICATE DEVICE
#### Primary Predicate
| Manufacturer: | Septodont |
|------------------------------|-----------------------------------------|
| Device Name: | BioRoot TM RCS |
| Device Classification Name: | Resin, Root Canal Filling/Dental Cement |
| 510(k) Number: | K130601 |
| Regulation Number: | Primary: 872.3820, Secondary: 872.3250 |
| Classification Product Code: | KIF |
| Device Class: | Class II |
#### Secondary Predicate
| Manufacturer: | Innovative BioCeramix Inc |
|------------------------------|-----------------------------------------|
| Device Name: | iRoot SP |
| Device Classification Name: | Resin, Root Canal Filling/Dental Cement |
| 510(k) Number: | K080917 |
| Regulation Number: | Primary: 872.3820, Secondary: 872.3250 |
| Classification Product Code: | KIF |
| Device Class: | Class II |
{4}------------------------------------------------
## 4. DEVICE DESCRIPTION
BioRoot™ Flow 0.5g, BioRoot™ Flow 2g is hydrophilic mineral root canal sealer presented as a ready-to-use paste in a syringe. BioRoot™ Flow 0.5g, BioRoot™ Flow 2g is based on a tricalcium silicate reaction with in situ water to seal the root canal. The sealer ensures a good adaptation to the root canal, tight interface with gutta-percha point and an adequate radiopacity.
It is an implantable medical device intended to be used in dentistry. It is part of endodontic cements used as permanent root canal sealer. BioRoot™ Flow 0.5g. BioRoot™ Flow 2g is used in combination with gutta-percha points to fill the root canal.
# 5. INDICATION FOR USE/INTENDED USE
The intended use for BioRoot™ Flow 0.5g, BioRoot™ Flow 2g is:
- Permanent root canal sealer .
The indications for BioRoot™ Flow 0.5g, BioRoot™ Flow 2g are:
- Permanent root canal filling in combination with gutta-percha points in case of . inflamed or necrotic pulp.
- Permanent root canal filling in combination with gutta-percha points following . a retreatment procedure.
# 6. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICES
Septodont demonstrated that BioRoot™ Flow 0.5g. BioRoot™ Flow 2q is substantially equivalent to the predicate devices, BioRoot™ RCS (K130601, cleared 17-Oct-2013) and iRoot SP (K080917, cleared 09-Apr-2008), in regards to the indications and basic design principles. The indications, technological characteristics and overall design of the subject device are substantially equivalent to those of the predicate devices, with any differences limited to minor differences in design and performance that do not raise any new questions of safety or effectiveness.
# 7. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
BioRoot™ Flow 0.5g, BioRoot™ Flow 2g has undergone extensive bench testing to provide evidence that its physical-chemical properties are substantially equivalent to the predicate devices, BioRoot™ RCS and iRoot SP. All three (3) devices are provided non-sterile, have comparable setting time, sealing and adhesion.
Biocompatibility test results determined that considering all existing data resulting from the chemical characterization and associated toxicological risk assessment.
{5}------------------------------------------------
and from biological data, the biocompatibility profile of BioRoot™ Flow 0.5g, BioRoot™ Flow 2g can be considered acceptable when used as intended.
## 8. SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
Clinical equivalence between BioRoot™ Flow 0.5g, BioRoot™ Flow 2g and the primary predicate, BioRoot™ RCS, was demonstrated based on equivalence in terms of the intended use, indications for use, technical, mechanical and biological properties. Furthermore, BioRoot™ Flow 0.5g, BioRoot™ Flow 2g and BioRoot™ RCS are used for the same clinical indication, at the same site in the body (root canal of the tooth), are in contact with the same tissues (root canal dentine), in the same population (mature and permanent teeth) and have the same clinical performance in regards to the expected clinical effect for a specific intended purpose. Clinical evaluation data obtained using the equivalent device, BioRoot™ RCS, was used to support clinical evidence of BioRoot™ Flow 0.5g, BioRoot™ Flow 2g and can be extended to BioRoot™ Flow 0.5g, BioRoot™ Flow 2g in terms of clinical performance and clinical safety.
# 9. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS
The subject device, BioRoot™ Flow 0.5g, BioRoot™ Flow 2g, and its' predicate devices, BioRoot™ RCS and iRoot SP, have demonstrated to be substantially equivalent and have comparable chemical, physical and performance specifications. Based on the above information and the data provided, the subject device has demonstrated to be substantially equivalent to both the primary predicate device, BioRoot™ RCS (K130601, cleared 17-Oct-2013), and the secondary predicate device, iRoot SP (K080917, cleared 09-Apr-2008).
| Characteristic | BioRoot™ Flow 0.5g,<br>BioRoot™ Flow 2g<br>(Subject Device) | BioRoot™ RCS<br>(Primary Predicate<br>Device - K130601) | iRoot SP<br>(Secondary Predicate<br>Device - K080917) | Similarities and<br>Differences between<br>Subject Device and<br>Predicate Devices |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Septodont | Septodont | Innovative BioCeramix<br>Inc. | N/A |
| Classification<br>device code | KIF | KIF | KIF | Same; Substantially<br>Equivalent |
| Classification<br>Name | Resin, Root canal filling | Resin, Root canal<br>filling | Resin, Root canal<br>filling | Same; Substantially<br>Equivalent |
| Class | Class II | Class II | Class II | Same; Substantially<br>Equivalent |
| Intended<br>Use | Permanent root canal<br>sealer | Permanent root canal<br>sealer | Permanent root<br>canal sealer | Same, Substantially<br>Equivalent |
| Characteristic | BioRoot™ Flow 0.5g,<br>BioRoot™ Flow 2g<br>(Subject Device) | BioRoot™ RCS<br>(Primary Predicate<br>Device - K130601) | iRoot SP<br>(Secondary Predicate<br>Device - K080917) | Similarities and<br>Differences between<br>Subject Device and<br>Predicate Devices |
| Indications | Permanent root canal<br>filling in combination<br>with gutta-percha points<br>in case of inflamed or<br>necrotic pulp.<br>Permanent root canal<br>filling in combination<br>with gutta-percha points<br>following a retreatment<br>procedure. | Permanent root canal<br>filling in combination<br>with gutta-percha<br>points in case of<br>inflamed or necrotic<br>pulp.<br>Permanent root canal<br>filling in combination<br>with gutta-percha<br>points following a<br>retreatment procedure.<br>BioRoot™ RCS is<br>suitable for use in<br>single cone technique<br>or cold lateral<br>condensation. | Permanent obturation<br>of the root canal<br>following vital pulp-<br>extirpation.<br>Permanent obturation<br>of the root canal<br>following removal of<br>infected or<br>necrotic pulp and<br>placement of<br>intracanal dressings. | iRoot SP - Similar,<br>indications statement<br>does not mention gutta-<br>percha points<br>BioRoot™ RCS –<br>Similar, does not<br>include the statement<br>regarding being suitable<br>for use in single cone<br>technique or cold lateral<br>condensation (due to<br>the change in design,<br>obturation is no longer<br>limited to specific<br>techniques). |
| Intended<br>Users | Dental healthcare<br>professional. | Dental healthcare<br>professional. | Dental healthcare<br>professional. | Same, Substantially<br>Equivalent |
| Intended<br>Patient<br>Population | Permanent and mature<br>teeth | Permanent and mature<br>teeth | Permanent and<br>mature teeth | Same, Substantially<br>Equivalent |
| Prescription/<br>over-the-<br>counter use | Prescription | Prescription | Prescription | Same, Substantially<br>Equivalent |
| Single-use/<br>multiple use | Multi-Use Device (only<br>the tips are considered<br>single use) | Single Use Device<br>(The bottle of powder<br>is multi-use, whereas<br>the single dose<br>containers are single<br>use) | Multi-Use | iRoot SP - Same;<br>Substantially Equivalent |
| Galenic<br>Forms<br>(Delivery<br>System) | Paste<br>(paste syringe) | Powder/Liquid<br>(powder liquid mixing) | Paste<br>(paste syringe) | iRoot SP - Same;<br>Substantially Equivalent |
| Characteristic | BioRoot™ Flow 0.5g,<br>BioRoot™ Flow 2g<br>(Subject Device) | BioRoot™ RCS<br>(Primary Predicate<br>Device - K130601) | iRoot SP<br>(Secondary Predicate<br>Device - K080917) | Similarities and<br>Differences between<br>Subject Device and<br>Predicate Devices |
| Ingredients<br>and<br>quantities | Paste (per 100g):<br>• Tricalcium silicate<br>(36.07g)<br>• Zirconium oxide<br>(30.00g)<br>• Calcium carbonate<br>(3.50g)<br>• Propylene glycol<br>(27.93g)<br>• Polyvidone K90<br>(1.00g)<br>• Sepineo P600<br>(1.00g)<br>• Aerosil R1812S<br>(0.50g) | Powder (per 100g):<br>• Tricalcium silicate -<br>Micronized C3S<br>(30.00g)<br>• Tricalcium silicate<br>(30.00g)<br>• Zirconium oxide<br>(35.00g)<br>• Polyvidone K90<br>(5.00g)<br>Liquid (per 100g):<br>• Dihydrated Calcium<br>Chloride (29.40g)<br>• Neomere Tech<br>(2.00g)<br>• Purified water<br>(68.60g) | Paste:<br>• Tricalcium silicate<br>• Dicalcium silicate<br>• Calcium<br>Phosphate<br>• Zirconium oxide<br>• Calcium hydroxide<br>• Solvent<br>• Gelifying Agent<br>• Stabilizer<br>Quantities are<br>unknown as this is<br>proprietary<br>information. | The ingredients<br>contained in<br>BioRoot™ Flow 0.5g,<br>BioRoot™ Flow 2g and<br>iRoot SP are similar and<br>determined to be<br>substantially equivalent.<br>BioRoot™ Flow 0.5g,<br>BioRoot™ Flow 2g<br>performance testing has<br>demonstrated<br>substantial equivalence<br>to BioRoot™ RCS.<br>BioRoot™ Flow 0.5g,<br>BioRoot™ Flow 2g has<br>similar ingredients to<br>BioRoot™ RCS with a<br>few additional<br>ingredients to allow for<br>the paste. The<br>quantities also differ<br>between the two<br>devices however once<br>reconstituted the<br>cements have the same<br>amount of Tricalcium<br>silicate and Zirconium<br>oxide (approx. 66%)<br>indicating an equivalent<br>composition.<br>Performance reports<br>have demonstrated<br>substantial equivalence<br>(refer to Section 18). |
| Containers | Premixed paste is<br>contained in a syringe | Powder is packaged in<br>a capsule; liquid is<br>packaged in a single<br>dose container | Premixed paste is<br>contained in a<br>syringe | iRoot SP - Same;<br>Substantially Equivalent<br>BioRoot™ RCS -<br>Different; container<br>closure system is<br>different due to different<br>delivery system<br>iRoot SP - Same;<br>Substantially Equivalent |
| Operations<br>(Mixing and<br>Dispensing) | No mixing required,<br>product is a ready-to-<br>use paste | Mixing: liquid added to<br>powder and mixed by<br>hand by the dentist. | No mixing required,<br>product is a ready-to-<br>use paste | iRoot SP - Same;<br>Substantially Equivalent |
| Characteristic | BioRoot™ Flow 0.5g,<br>BioRoot™ Flow 2g<br>(Subject Device) | BioRoot™ RCS<br>(Primary Predicate<br>Device - K130601) | iRoot SP<br>(Secondary Predicate<br>Device - K080917) | Similarities and<br>Differences between<br>Subject Device and<br>Predicate Devices |
| | | | | BioRoot™ RCS -<br>Different; mixing is<br>required as it requires<br>reconstitution of the<br>powder and liquid. |
| Consistency | Paste | Not Performed (not a<br>paste) | Paste | iRoot SP - Same;<br>Substantially Equivalent |
| Setting<br>Time | 1h to 6h | 55min to 300min | 4h | Performance testing<br>have demonstrated that<br>BioRoot™ Flow 0.5g,<br>BioRoot™ Flow 2g is<br>substantially equivalent<br>to BioRoot™ RCS and<br>iRoot SP sealer<br>regarding setting time. |
## Side-by-Side Comparison of BioRoot™ Flow 0.5q, BioRoot™ Flow 2q with BioRoot™RCS and iRoot SP
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.