Cross Protection ResQ300 Plus Surgical Face Mask

K212273 · Cross Protection (M) Sdn Bhd · FXX · Aug 17, 2022 · General, Plastic Surgery

Device Facts

Record IDK212273
Device NameCross Protection ResQ300 Plus Surgical Face Mask
ApplicantCross Protection (M) Sdn Bhd
Product CodeFXX · General, Plastic Surgery
Decision DateAug 17, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Cross Protection ResQ300 Plus Surgical Face Masks are intended to be used as isolation face mask and dental face mask. Cross Protection ResQ300 Plus Surgical Face Masks are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating personnel from transms, bodily fluids and particulate material.

Device Story

Pleated, multi-ply, single-use, non-sterile surgical face mask; covers nose and mouth; secured via ear loops or tie-on straps. Outer layer: 100% blue spun-bond polypropylene; middle layer: 100% white melt-blown polypropylene filter media; inner layer: 100% white spun-bond polypropylene or medical grade tissue paper. Includes malleable aluminum nosepiece. Used in OR, dental, and clinical settings by healthcare personnel and patients. Acts as physical barrier to reduce exposure to microorganisms, bodily fluids, and particulates. Protects both wearer and patient.

Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100-19, ASTM F2101, ASTM F2299, ASTM F1862, and 16 CFR 1610. Biocompatibility verified per ISO 10993-5 and ISO 10993-10.

Technological Characteristics

Multi-ply pleated mask; materials: spun-bond polypropylene, melt-blown polypropylene, medical grade tissue paper, malleable aluminum nosepiece. ASTM F2100 Level 1 performance. Biocompatibility: ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization). Non-sterile, single-use.

Indications for Use

Indicated for operating room personnel, patients, and healthcare workers during surgical, medical, or dental procedures to protect against transfer of microorganisms, bodily fluids, and particulate material.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text. August 17, 2022 Cross Protection (M) Sdn Bhd Grace Tan Operations Manager 11, JLN Perindustrian5, JLN Meru Klang, Selangor 41050 Malaysia Re: K212273 Trade/Device Name: Cross Protection ResQ300 Plus Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: June 17, 2022 Received: June 17, 2022 Dear Grace Tan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Bifeng Qian M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212273 Device Name Cross Protection ResQ300 Plus Surgical Face Mask Indications for Use (Describe) The Cross Protection ResQ300 Plus Surgical Face Masks are intended to be used as isolation face mask and dental face mask. Cross Protection ResQ300 Plus Surgical Face Masks are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating personnel from transms, bodily fluids and particulate material. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| | | | | | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY: K212273 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92. ### l. SUBMITTER | Applicant | Cross Protection (M) Sdn Bhd | | | |----------------|-----------------------------------------------------------------------------------------------|-------|--------------------| | Address | 11, Jln Perindustrian 5, Batu 5 ½, Off Jln Hj Abdul<br>Manan, 41050 Klang, Selangor, Malaysia | | | | Phone | 603 3392 0000 | Fax | 603 3392 3867 | | Contact Person | Grace Tan | Title | Operations Manager | | Date Prepared | August 12, 2022 | | | ### II. PROPOSED DEVICE AND PREDICATE DEVICE | | New Device 807.92(a)(2) | Predicate Device 807.92(a)(3) | |-------------------------------------|-------------------------------|---------------------------------------------| | Proprietary Names/<br>510(k) Number | ResQ300 Plus / K212273 | Non-Sterile Surgical Face Mask /<br>K051291 | | Manufacturer | Cross Protection (M) Sdn Bhd. | A.R. Medicom Inc. | | Classification Code | FXX | FXX | | Device Classification | 2 | 2 | | Regulation Number | 878.4040 | 878.4040 | | Panel | General and Plastic Surgery | General and Plastic Surgery | | Device Classification Name | Mask, Surgical | Mask, Surgical | | Common Name | Surgical Apparel | Surgical Apparel | #### lll. DEVICE DESCRIPTION The Cross Protection® ResQ300 Plus Surgical Facemask is a pleated, multi-ply design, which is a nonsterile, single use, disposable device. The mask covers the nose and mouth and is secured to the face using the attached ear loops or attached ties. The outer layer is made of 100% blue spun-bond polypropylene. The middle layer is filter media composed of 100% white melt-blown polypropylene. The inner (patient contacting) layer is made of either 100% white spun-bond polypropylene or 100% white medical grade tissue paper. The ear attachments are either Ear Loop style. Ear Loops are made of elastic and the Tie-On straps are made of 100% white spun-bond polypropylene. {4}------------------------------------------------ The nosepiece is made of malleable aluminum. All materials used in the construction of the mask are being used in currently marketed devices. ResQ300 Plus model numbers: | Description | Product Code | |------------------------|--------------| | ResQ300 Plus (Tie On) | 2-406TO | | ResQ300 Plus (Earloop) | 2-406EL | #### IV. INDICATIONS FOR USE The Cross Protection ResQ300 Plus Surgical Face Mask is intended to be used as an isolation face mask, procedure mask, and dental face mask. Cross Protection ResQ300 Plus Surgical Face Masks are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the opersonnel from transfer of micro-organisms, bodily fluids and particulate material. ### V. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE | | Proposed Device | Predicate Device | Result | |-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | 510(K) Number | K212273 | K051291 | | | Device Name | ResQ300 Plus Surgical<br>Facemask | Non-Sterile Surgical Mask | | | Manufacturer | Cross Protection (M) Sdn Bhd | A.R. Medicom Inc. | | | Classification | Class 2 | Class 2 | Same | | Product Code | FXX | FXX | Same | | Regulation No. | 878.4040 | 878.4040 | Same | | Indications for Use | The Cross Protection® ResQ300<br>Plus Surgical Facemask is<br>intended to be used as isolation<br>face mask, procedure mask and<br>dental face mask.<br>It is intended to be worn by<br>operating room personnel<br>during surgical procedures<br>to protect both the surgical<br>patient and the operating<br>personnel from transfer of<br>micro-organisms, body<br>fluids and particulate<br>material. | The medical/surgical masks<br>are indicated as a protective<br>nose and mouth covering for<br>health care workers and<br>patients involved in medical<br>and surgical procedures.<br>The masks are indicated in<br>any procedure or situation<br>where there is a risk of<br>microorganism, body fluid,<br>and particulate aerosol<br>transfer. | Similar | | | Proposed Device | Predicate Device | Result | | Over-the-Counter Use | Yes | Yes | Same | | Sterile | No | No | Same | | Single-Use | Yes | Yes | Same | | Dimensions | | | | | Body Length | 185mm (7.0") | 175mm (6.8") | Similar | | Body Width | 90.0mm (3.5") | 90.5mm (3.5") | Similar | | Ear Loop | 150mm ( 5.9") | 165mm (6.5") | Similar | | Ear Tie | 900mm (35.4") | N/A | Different | | Nose Piece Length | 94mm (3.7") | 105 mm (4.1") | Similar | | Description | Pleated, 3-ply rectangular face<br>masks with a shapeable<br>nosepiece and two earloops (or<br>ties) present, one on each side,<br>in order to hold mask in place. | Shingle pleats rectangular<br>face masks with a shapeable<br>nosepiece and two earloops<br>present, one on each side, in<br>order to hold mask in place. | Similar | | Inner Layer (Patient<br>Contacting) | Spun Bond Polypropylene or<br>medical grade tissue paper,<br>white color | Spun Bond Polypropylene or<br>medical grade tissue paper,<br>white color | Same | | Middle Layer (Filter<br>Media) | Melt-blown<br>Polypropylene, white<br>color | Melt-blown Polypropylene,<br>white color | Same | | Outer Layer | Spun Bond Polypropylene, blue<br>color | Spun Bond Polypropylene,<br>blue color | Same | | Nose Piece | Malleable Aluminum | Malleable Aluminum | Same | | Ear Attachments | | | | | Ear Loops | 100% Flat-type Cotton<br>and Elastic | Flat latex and fiberglass-free<br>elastic | Similar | | Ear Ties | Spun-bond Polypropylene | N/A | Different | | Biocompatibility Testing | | | | | Cytotoxicity ISO 10993-5 | Not cytotoxic | Not cytotoxic | Same | | Irritation ISO 10993-10 | Non-irritating | Non-irritating | Same | | Sensitization ISO 10993-10 | Non-sensitizing | Non-sensitizing | Same | | ASTM F2100 Level | 1 | 1 | Same | | Performance Testing -<br>Bench | | | | | ASTM F2101 Bacterial<br>Filtration Efficiency (BFE) | BFE @ 3.0 µm large Bacteria ≥<br>95% | BFE @ 3.0 µm large Bacteria<br>> 95% | Same | | | Proposed Device | Predicate Device | Result | | EN14683 Differential<br>Pressure (Delta P) | ΔP < 5.0 mm of H2O/cm2 | ΔP < 5.0 mm of H2O/cm² | Same | | ASTM F2299 Particulate<br>Filtration Efficiency (PFE) | PFE @ 0.1 µm large Latex<br>particles ≥ 95% | PFE @ 0.1 µm large Latex<br>particles ≥ 95% | Same | | ASTM F1862 Fluid<br>Resistance (Synthetic<br>Blood) | Resistant @ 80mmHg | Resistant @ 80mmHg | Same | | Textile Flammability<br>16 CFR 1610 | Class I | Class I | Same | | Performance Testing -<br>Animal | None | Unknown | Unknown | | Performance Testing -<br>Clinical | None | Unknown | Unknown | {5}------------------------------------------------ {6}------------------------------------------------ # Substantial Equivalence Discussion: The above comparison table shows that ResQ300 Plus Surgical Face Masks are similar in design, intended use, and function to the A.R. Medcom Surgical Face Masks. There are a few differences: 1. The ResQ300 Plus Surgical Face masks are slightly longer than the A.R. Medicom Surgical Face Masks in surface area (185mm vs 175mm), which provides better facial coverage for adult-size faces. 2. The ResQ300 Plus ear loops are slightly smaller (150mm vs 160mm) than the A.R. Medicom ear loops and provide a tighter, more secure, hold. 3. The ResQ300 Plus masks have a tie-on option that is not offered in the A.R. Medicom Surgical Face Mask, which provides comfort for people who may be uncomfortable with elastic loops on their ears. 4. The ResQ300 Plus nose piece is smaller (94mm vs 105mm), which makes fitting to the nose area more precise. All materials used in the ResQ300 Plus Surgical Face masks are used in currently marketed devices. Biocompatibility and performance bench-testing demonstrate that the ResQ300 Plus Surgical Face Masks meet FDA requirements and that there are no new concerns of safety and effectiveness. ### VI. SUMMARY OF NON-CLINICAL TESTNG The ResQ300 Plus Surgical Face Mask was tested for biocompatibility and performance as shown in the below tables. # Biocompatibility Testing CROSS PROTECTION RESQ300 PLUS {7}------------------------------------------------ The ResQ300 Plus Surgical Face Mask was tested according to the below ISO 10993 biocompatibility testing standards and test results show that the device presents no new safety risk when compared with the predicate device. | Test Name | Standard | Pass Criteria | Results | |-------------------------------------------------------------------------------|--------------|------------------------------------|------------------------------------| | Cytotoxicity – MEM Elution Test | ISO 10993-5 | 0 – None<br>1 – Slight<br>2 – Mild | Not cytotoxic –<br>Mild reactivity | | Primary Skin Irritation Test | ISO 10993-10 | Negligible: 0 to 0.4 | Negligible<br>irritant | | Maximization Test for Delayed-Type<br>Hypersensitivity in Hartley Guinea Pigs | ISO 10993-10 | 0 – No visible change | Non-sensitizing | ### Performance Testing – Bench The ResQ300 Plus Surgical Face Mask was tested according to the below performance standards and test results show that the device presents no new safety risk when compared with the predicate device. | Test Name | Standard | Pass Criteria | Results | |-------------------------------------------------------|---------------------------------------------------|--------------------------------------------------|---------| | Bacterial Filtration Efficiency (BFE) | ASTM F2100-19 Clause 9.1,<br>ASTM F2101-19 | Level 1: ≥ 95%<br>Level 2: ≥98<br>Level 3: ≥98 | Level 1 | | Differential Pressure (Delta P),<br>mmH20/cm² | ASTM F2101, EN14683 | Level 1: <5.0<br>Level 2: <6.0<br>Level 3: <6.0 | Level 1 | | Sub-micron particulate filtration<br>efficiency (PFE) | ASTM F2100-19 Clause 9.3,<br>ASTM F2299/F2299M-03 | Level 1: ≥ 95%<br>Level 2: ≥98%<br>Level 3: ≥98% | Level 1 | | Synthetic Blood Penetration<br>Resistance (mmHg) | ASTM F1862 | Level 1: 80<br>Level 2: 120<br>Level 3: 160 | Level 1 | | Flammability | ASTM F2100-19 Clause 9.5,<br>16 CFR 1610 | Class 1 | Class 1 | # Animal Testing Not applicable #### SUMMARY OF CLINICAL TESTING VII. Not applicable #### VIII. CONCLUSION The conclusions drawn from the non-clinical testing demonstrates that the Cross Protection ResQ300 Plus Surgical Face Masks are as safe, as effective, and perform as well as or better than the legally marketed predicate device cleared under K051291.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...