The TheraFace microcurrent is a hand-held device for over-the-counter aesthetic purposes . The TheraFace microcurrent is indicated for facial stimulation.
Device Story
Hand-held device for over-the-counter aesthetic facial stimulation; utilizes internal lithium rechargeable battery. Device features electrode contactors applied to facial skin; delivers regulated biphasic pulsed electrical microcurrent. User adjusts output intensity via device buttons or mobile app via Bluetooth. Device includes automatic overload/no-load trip and automatic shut-off. Intended for home use by consumers; provides non-invasive electrical stimulation to facial muscles. Output parameters (voltage, current, frequency) are designed to be comparable to predicate devices to ensure safety and effectiveness. Device performance verified through bench testing, biocompatibility, and usability studies.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility (ISO 10993-5, ISO 10993-10), electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), wireless coexistence (AAMI TIR69, ANSI C63.27), and a usability engineering study (TG-Theraface microcurrent-013). No clinical trials were performed.
Technological Characteristics
Hand-held stimulator; PC plastic and stainless steel construction. Sensing/actuation: Biphasic pulsed electrical microcurrent (8.3 Hz, 60ms pulse width). Energy source: Internal lithium rechargeable battery. Connectivity: Bluetooth for mobile app control. Sterilization: Not specified. Software: Microprocessor-controlled with automatic overload/no-load trip and shut-off. Complies with IEC 60601-1, IEC 60601-1-11, and IEC 60601-2-10.
Indications for Use
Indicated for facial stimulation in adults for over-the-counter aesthetic purposes.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 29, 2021
Theragun, Inc. % Thomas Padula Vice President Regulatory Compliance Schiff & Company, Inc. 583 Mountain Avenue North Caldwell, New Jersey 07006
Re: K212238
Trade/Device Name: TheraFace microcurrent Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NFO Dated: July 15, 2021 Received: July 19, 2021
Dear Thomas Padula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212238
Device Name TheraFace microcurrent
Indications for Use (Describe)
The TheraFace microcurrent is a hand-held device for over-the-counter aesthetic purposes . The TheraFace microcurrent is indicated for facial stimulation.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) PREMARKET NOTIFICATION FOR THERAFACE MICROCURRENT THERAGUN, INC.
### 510(k) Summary (as required by 807.92)
### (1) SUBMITTER:
THERAGUN, Inc. 6100 Wilshire Blvd Suite 200 Los Angeles, CA 90048 Registration Number: 3012386142 FEI Number: 3012386142 Contact person: CJ Frederick - Director, Regulatory Compliance Telephone: 310-570-8341 Email: jaime@therabodycorp.com Date prepared: September 20, 2021
### Application Correspondent:
Contact Person: Thomas Padula Company: Schiff & Company, Inc. Address: 583 Mountain Avenue, North Caldwell, NJ 07006 Tel: 201-317-8810 Email: thomaspadula(@schiffandcompany.com
### (2) DEVICE NAME:
Trade Name: TheraFace microcurrent Common Name: Stimulator, Transcutaneous Electrical, Aesthetic Purposes Classification Name: Stimulator, Transcutaneous Electrical, Aesthetic Purposes Device Classification: Class II Review Panel: Neurology; General & Plastic Surgery Regulation Number: 21 CFR 882.5890 Product Code: NFO
| Sponsor | Heat In A Click |
|-----------------------|-------------------------|
| Device Name and Model | 2 Face / Face Evolution |
| 510(k) Number | K171821 |
| Product Code | NFO |
| Regulation Number | 882.5890 |
| Regulation Class | II |
(3) PREDICATE DEVICE(S): Substantial equivalence is based on following legally marketed devices.
### (4) DESCRIPTION OF THEDEVICE:
The TheraFace microcurrent device consist of main control unit and its attachment applicators for microcurrent output which can be control by the device control button or APP in cell phone via Bluetooth
{4}------------------------------------------------
connection. The device powered by one internal lithium rechargeable battery which can be charged by external battery charger.
The device has the electrode contractors for facial stimulation by applying an electrical micro current to face skin. The output waveform is formed of regulated Voltage of Biphasic pulse and provided with 3 levels of output intensity, which can be adjusted by user. The low-level electrical current pulse goes through the face muscle and cause face muscle contraction for facial stimulation.
### (5) INDICATIONS FOR USE:
The TheraFace microcurrent is a hand-held device for over-the-counter aesthetic purposes. The TheraFace microcurrent is indicated for facial stimulation.
### (6) COMPARISON WITH PREDICATEDEVICES:
Following table is a comparison of TheraFace microcurrent and predicate devices.
TheraFace microcurrent is substantially equivalent in terms of the technological characteristics, features, specifications, materials, mode of operation and indications for use, to 2 Face Evolution - K 171821, cleared for marketing under510(K).
| Elements of<br>Comparison | Subject Device | Primary Predicate Device | Remark | Intended Use | The TheraFace microcurrent is a<br>hand-held device for over-the-counter<br>aesthetic purpose. The TheraFace<br>microcurrent is indicated for facial<br>stimulation. | 2 Face / Face<br>Evolution is a hand-held device for over the<br>counter aesthetic purposes.<br>(1) The EMS mode is indicated for facial<br>stimulation.<br>(2) The Photon mode: The red light is<br>intended for the treatment of periorbital<br>wrinkles and the blue light is intended<br>for the treatment of the mild to moderate<br>inflammatory acne. | SE<br>Note 1<br>(Only compared with<br>its EMS function of<br>K171821) |
|-------------------------------------------------|-------------------------------|--------------------------|--------|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Basic specification | | | | Apply parts | Face | Face | SE |
| Device Name and<br>Model | <i>TheraFace microcurrent</i> | 2 Face / Face Evolution | -- | Power Sources | Lithium battery :2x3.7V | DC 3.7V 2200mAh | SE |
| 510 (K) Number | Applying | K171821 | -- | Method of Line<br>Current Isolation | Battery Supply | Battery Supply | SE |
| Product Code | NFO | NFO | -- | Number of Modes for<br>Micro current<br>stimulation | l | 1 | SE |
| Regulation Number | 882.5890 | 882.5890 | -- | Number of Channels<br>for Micro current<br>stimulation | l | l | SE |
| Synchronous or<br>Alternating | N/A | N/A | SE | | | | |
| Regulated Current or<br>Regulated Voltage | Both | Both | SE | | | | |
| Software/Firmware/<br>Microprocessor<br>control | Yes | Yes | ਟਜ | | | | |
| Automatic Overload<br>Trip | Yes | Yes | ਟਜ | | | | |
{5}------------------------------------------------
{6}------------------------------------------------
| Automatic No-load<br>Trip | | Yes. | Yes | SE |
|-------------------------------------|------------------------------|-------------------------------------------------|---------------------------------------------------|--------------|
| Automatic Shut Off | | Yes. | Yes | SE |
| Patient Override<br>Control | | Yes | Yes | SE |
| Indicator<br>Display | On/Off<br>Status | Yes | Yes | SE |
| | Low<br>Battery | Yes | Yes | SE |
| | Voltage/<br>Current<br>Level | Yes | Yes | SE |
| Time Range | | Micro current 5 minutes | EMS Mode (5 minutes)<br>Photon Mode (5~7 minutes) | SE |
| Console weight | | 230g | 200g | SE<br>Note 2 |
| Waveform | | Pulsed Biphasic | Pulsed Biphasic, | SE |
| Shape | | Rectangular | Rectangular | SE |
| Maximum Output<br>Voltage (+/- 10%) | | 0.24V @ $500Ω$<br>1.0V @ $2kΩ$<br>5.0V @ $10kΩ$ | 0.31V @ $500Ω$<br>1.16V @ $2kΩ$<br>5.56V @ $10kΩ$ | SE<br>Note 3 |
| | | | | |
| Maximum output<br>Current | 500 μA @ 500 Ohm | 620μΑ @ 500Ω | SE | |
| | 500 μA @ 2k Ohm | 580μA @ 2kΩ | | |
| | 500 μA @ 10k Ohm | 556μΑ @ 10kΩ | Note 3 | |
| Frequency range | 8.3 Hz | 8.33 Hz | SE | |
| Pulse width range | 60ms | 60ms | SE | |
| Net Charge | 0 μC @ 500Ω | 0 μC @500 Ω | SE | |
| Max Phase charge | 29.4µC@ 500 Ohm | 26.3μC@ 500 Ohm | SE Note 4 | |
| Maximum Current<br>Density | 0.2mA/cm² @500Ω | 0.33 mA/cm²@ 500Ω | SE Note 5 | |
| Maximum Power<br>Density | 23.1μW/cm² @500Ω | 4.34 μW/cm² @ 500Ω | SE Note 5 | |
| ON time | 60 ms | 60 ms | SE | |
| OFF time | 60 ms | 60 ms | SE | |
| Contraction and<br>Relaxation time | 60 ms ON/OFF | 60 ms ON/OFF | SE | |
{7}------------------------------------------------
{8}------------------------------------------------
| Material of device and<br>construction | PC Plastic & Stainless Steel | ABS Plastic & Stainless Steel | SE Note 6 |
|----------------------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------|
| Bio-compatibility | All user directly contacting materials<br>are compliance with ISO10993-5,<br>ISO10993-10 | All user directly contacting materials are<br>compliance with ISO10993-5 and<br>ISO10993-10 requirements. | SE |
| Electrical Safety | Comply with IEC 60601-1<br>IEC 60601-1-11<br>IEC 60601-2-10 | Comply with: IEC 60601-1<br>IEC 60601-1-11<br>IEC 60601-2-10 | SE |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SE |
## Comparison in Detail(s):
Note 1:
There is multiple function operation mode in the predicate device, the subject device TheraFace microcurrent only has one micro-current operation mode. To this microcurrent mode, the subject device can be substantial equivalent to predicate device K171821.
Note 2:
Even there is console weight difference between subject device and predicate devices; but they all are portable medical device and comply with IEC60601-1 testing, such minor difference would not affect safety and effectiveness issue.
Note 3:(Maximum Output Voltage and Maximum Output Current):
The effect of micro current stimulation is determined by micro current output waveform and output current.
There is only minor difference about the output voltage and current between the subject device and the predicated devices, it can still obtain the same effect because our output voltage and output current are in the range which is between the value of K171821. Also, the subject device complies with IEC 60601-1, and IEC60601-2-10 which means we have proved its safety as well as the effectiveness comparing with the predicate devices. Therefore, the subject device and predicate devices are substantially
{9}------------------------------------------------
equivalence on output voltage and current.
Note 4: (Max phase charge)
Even there is minor difference on the max phase charge between subject device and predicate device, but the pulse waveform is Pulsed Biphasic, and its net charge is 0 µC;
And the micro current waveform parameters comply with IEC60601-2-10, so the difference on max phase charge would not raise safety issue; and it would not affect effectiveness for facial stimulation. Therefore, such minor difference would not affect safety and effectiveness issue.
Note 5 (Maximum current density and Maximum power density):
The effect of micro current stimulation on facial skin are determined by micro-current output waveform and output current. The maximum current density and power density is the parameters related to safety of medical device.
For maximum current density, there is only little difference on it between the subject device and the predicated device K171821, it is little smaller than the value of K171821, so it would not raise safety issue. And viewing from max current density value in the predicate device K171821, the value of subject device is in the range of max current density between predicate device K171821. Therefore, such minor difference would not affect safety and effectiveness issue.
For maximum power density, the value of maximum power density of subject device is larger than the predicate device. But the maximum power density meets with the maximum allowed value 0.25 (W/cm²) required in FDA guidance. Also, the subject device complies with IEC 60601-1, and IEC60601-2-10 which means we have proved its safety as well as the effectiveness comparing with the predicate devices. Therefore, such minor difference would not affect safety and effectiveness issue.
### Note 6
Even there is different for the material between subject device and predicate device, but its material complies with ISO10993-5, and ISO10993-10.
### Finial Conclusion:
The TheraFace microcurrent device has the same indications use and technology characteristics as the predicate devices. TheraFace microcurrent device is as safe, as effective, and performs as well as the predicate devices.
{10}------------------------------------------------
## 510(K) PREMARKET NOTIFICATION FOR THERAFACE MICROCURRENT THERAGUN, INC.
### (7) PERFORMANCE STSTANDARDS APPLIED:
A series of studies were completed to demonstrate the substantial equivalence of TheraFace microcurrent to the predicate devices. All testing was conducted in accordance with and in conformance to applicable device regulations and guidance. Results of all testing demonstrate the device is non-toxic, is comparable to other currently marketed devices and is substantially equivalent to legally marketed predicates and included:
### Biocompatibility
- ISO 10993-5:2009, biological evaluation of medical devices -- part 5: tests for in vitro cytotoxicity. (Biocompatibility).
- ISO 10993-10 :2010, biological evaluation of medical devices part 10: tests for irritation and skin sensitization. (Biocompatibility).
Electrical Safety and Electromagnetic Compatibility
- IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012 (or IEC 60601-1:2012 reprint) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC / EN 60601-1-2: 2014-02, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - collateral standard: Electromagnetic Compatibility.
- IEC 60601-1-11 :2015. Medical electrical equipment -- part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home health care environment.
- IEC 60601-2-10: 2012, Medical electrical equipment -- Part 2: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators.
TheraFace Micro-current Test Report, Theragun study TG-TheraFace-1010
AAMI TIR69:2017, ANSI C63.27-2017 Wireless Coexistence Test
Usability Engineering
Usability Study Report, Theragun study TG-Theraface microcurrent-013
### (8) CONCLUSION:
TheraFace microcurrent has the same indications for use and technology characteristics as the predicate devices. TheraFace microcurrent is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.