← Product Code [FXX](/productcode/FXX) · K212235

# Surgical Face Mask (K212235)

_Shanghai Yunqing Industrial Co., Ltd. · FXX · Dec 6, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212235

## Device Facts

- **Applicant:** Shanghai Yunqing Industrial Co., Ltd.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Dec 6, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. Surgical Face Masks are single use, disposable devices, provided non-sterile.

## Device Story

Surgical face mask (model YOD2001); flat-pleated, 3-layer design; outer/inner layers spun-bond polypropylene; middle layer melt-blown polypropylene; includes malleable polypropylene-coated iron nose piece and 20% spandex/80% polyester ear loops. Device worn by patients and healthcare personnel to provide physical barrier against microorganisms, body fluids, and particulates. Used in clinical settings for infection control. Performance meets ASTM F2100 Level 2 requirements for fluid resistance, bacterial/particulate filtration, and differential pressure. Single-use, disposable, non-sterile.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified per ASTM F2100-19 Level 2: Bacterial Filtration Efficiency (BFE) >98% (ASTM F2101), Particulate Filtration Efficiency (PFE) >98% (ASTM F2299), Fluid Resistance at 120 mmHg (ASTM F1862), Differential Pressure <6.0 mmH2O/cm² (EN 14683), and Flammability Class 1 (16 CFR Part 1610). Biocompatibility confirmed per ISO 10993-1:2018 (cytotoxicity, irritation, sensitization).

## Technological Characteristics

3-layer flat-pleated mask. Materials: spun-bond polypropylene (outer/inner), melt-blown polypropylene (middle). Nose clip: iron wire coated with polypropylene. Ear loops: 20% spandex/80% polyester. Dimensions: 173mm x 97mm. Meets ASTM F2100 Level 2 barrier specifications. Biocompatible per ISO 10993. Non-sterile.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face Mask ([K182514](/device/K182514.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

December 6, 2021

Shanghai Yunqing Industrial Co., Ltd. Ryan Li RA Manager Shanghai Mind-link Business Consulting Co., Ltd. Room A08, Floor 14th, No 699, Jiaozhou Road, Jingan District Shanghai, 200040 China

Re: K212235

Trade/Device Name: Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: October 28, 2021 Received: November 8, 2021

Dear Ryan Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number K212235

Device Name Surgical Face Mask

Indications for Use (Describe)

Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. Surgical Face Masks are single use, disposable devices, provided non-sterile.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(K) Summary-K212235

#### I. SUBMITTER:

Shanghai Yunqing Industrial Co., Ltd. 2nd buliding1818, shangsu road,fengxian district, Shanghai, China Contact Person: Hulin Title: QA Manager Tel: +86 15921265138 Email: HL@yqgroup.vip

Submission Correspondent: Ryan Li Email: ryan.li(@mind-link.net Tel:+86 13701581791

Summary prepared: October 26, 2021

#### II. DEVICE

Name of Device: Surgical Face Mask Regulation Number: 21 CFR 878.4040 Common Name: Surgical Mask Classification Name: Surgical Mask Regulatory Class: II Product Code: FXX

## III. PREDICATE DEVICE

Primary predicate device:

Predicate: K182514 Trade/Device Name: Surgical Face Mask Manufacturer: Xiantao Zhibo Non-Woven Products Co., Ltd Product Code: FXX Classification Name: Mask, surgical Regulation Number: 21 CFR 878.4040

### IV. DEVICE DESCRIPTION

{4}------------------------------------------------

Surgical Face Mask contains one model called YOD2001.

YOD2001 Surgical Face Mask is in white, barrier level 2, size 173mm*97mm, flat pleated and ear loop type.

YQD2001 Surgical Face Mask is a flat pleated type mask composed of three layers. The mask materials consist of an outer layer (spun-bond polypropylene), a middle layer, between the outer layers (melt-blown polypropylene), and an inner layer (spun-bond polypropylene). Each mask contains ear loops to secure the mask over the users' mouth and face and includes a malleable nose piece made of polypropylene to provide a firm fit over the nose.

## V. INDICATIONS FOR USE

Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. Surgical Face Masks are single use, disposable devices, provided non-sterile.

# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Surgical Face Masks are compared with the predicate device Surgical Face Mask (K182514). The results are shown below in the Technological Characteristics Comparison Table:

| DEVICE                                                                                      | Subject Device                                                                                                                                                                                                                                                                                                                             | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                       | Remark              |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
|                                                                                             | Surgical Face Mask                                                                                                                                                                                                                                                                                                                         | Surgical Face Mask (K182514)                                                                                                                                                                                                                                                                                                                                   |                     |
| Intended Use                                                                                | The Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. Surgical Face Masks are single | The surgical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided | Same                |
|                                                                                             | use, disposable devices, providednon-sterile.                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                |                     |
|                                                                                             | non-sterile.                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                |                     |
| Classification                                                                              | FXX                                                                                                                                                                                                                                                                                                                                        | FXX                                                                                                                                                                                                                                                                                                                                                            | Same                |
| Product Code                                                                                |                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                |                     |
| Model                                                                                       | Ear Loop                                                                                                                                                                                                                                                                                                                                   | Ear loop                                                                                                                                                                                                                                                                                                                                                       | Same                |
|                                                                                             | Materials                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                |                     |
| Outer Facing Layer                                                                          | Spun-bond<br>non-woven fabric                                                                                                                                                                                                                                                                                                              | polypropylene Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                          | Similar<br>Note 1   |
| Middle Layer                                                                                | Polypropylene<br>non-woven fabric                                                                                                                                                                                                                                                                                                          | melt-blown Melt blown polypropylene filter                                                                                                                                                                                                                                                                                                                     | Similar<br>Note 1   |
| Inner Facing Layer                                                                          | Spun-bond<br>non-woven fabric                                                                                                                                                                                                                                                                                                              | polypropylene Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                          | Similar<br>Note 1   |
| Nose Clip                                                                                   | Iron wire coated<br>polypropylene                                                                                                                                                                                                                                                                                                          | with Malleable aluminum wire                                                                                                                                                                                                                                                                                                                                   | Similar<br>Note 1   |
| Ear Loop/Headband                                                                           | 20% Spandex 80% Polyester                                                                                                                                                                                                                                                                                                                  | Polyester                                                                                                                                                                                                                                                                                                                                                      | Similar<br>Note 1   |
|                                                                                             | Design Features                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                |                     |
| Style                                                                                       | Flat pleated                                                                                                                                                                                                                                                                                                                               | Flat pleated                                                                                                                                                                                                                                                                                                                                                   | Same                |
| Multiple Layers                                                                             | 3 Layers                                                                                                                                                                                                                                                                                                                                   | 3 Layers                                                                                                                                                                                                                                                                                                                                                       | Same                |
| Single Use                                                                                  | Single use                                                                                                                                                                                                                                                                                                                                 | Single use                                                                                                                                                                                                                                                                                                                                                     | Same                |
|                                                                                             | Sterility                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                |                     |
| Sterile                                                                                     | Non-sterile                                                                                                                                                                                                                                                                                                                                | Non-sterile                                                                                                                                                                                                                                                                                                                                                    | Same                |
|                                                                                             | Dimension                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                |                     |
| Length Width                                                                                | 173mm X 97mm                                                                                                                                                                                                                                                                                                                               | 175mm X 95mm                                                                                                                                                                                                                                                                                                                                                   | Different<br>Note 2 |
| Technological Characteristics Product Barrier Specifications Per ASTM F2100 – Meets Level 2 |                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                |                     |
| Fluid Resistance<br>ASTM F1862                                                              | Pass at 120mmHg                                                                                                                                                                                                                                                                                                                            | Pass at 120mmHg                                                                                                                                                                                                                                                                                                                                                | Similar             |
| Particulate Filtration<br>Efficiency (PFE)<br>ASTM F2299                                    | Pass at >98%                                                                                                                                                                                                                                                                                                                               | Pass at >98%                                                                                                                                                                                                                                                                                                                                                   | Similar             |
| Bacterial Filtration<br>Efficiency (BFE)<br>ASTM F2101                                      | Pass at >98%                                                                                                                                                                                                                                                                                                                               | Pass at >98%                                                                                                                                                                                                                                                                                                                                                   | Similar             |
| Differential Pressure<br>(Delta P)<br>MIL-M-36954C                                          | Pass at <6 mmH2O/cm²                                                                                                                                                                                                                                                                                                                       | Pass at <6 mmH2O/cm²                                                                                                                                                                                                                                                                                                                                           | Similar             |
| Flammability<br>16 CFRPART 1610                                                             | Class 1 Non-Flammable                                                                                                                                                                                                                                                                                                                      | Class 1 Non-Flammable                                                                                                                                                                                                                                                                                                                                          | Similar             |
| Biocompatibility                                                                            |                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                |                     |
| Cytotoxicity                                                                                | Under the conditions of the study, the subject device extract was determined to be non-cytotoxic.                                                                                                                                                                                                                                          | Under the conditions of the study, the subject device extract was determined to be non-cytotoxic.                                                                                                                                                                                                                                                              | Same                |
| Irritation                                                                                  | Under the conditions of the study, the subject device non-polar and polar extracts were determined to be non-irritating.                                                                                                                                                                                                                   | Under the conditions of the study, the subject device non-polar and polar extracts were determined to be non-irritating.                                                                                                                                                                                                                                       | Same                |
| Sensitization                                                                               | Under the conditions of the study, the subject device non-polar and polar extracts were determined to be non-sensitizing.                                                                                                                                                                                                                  | Under the conditions of the study, the subject device non-polar and polar extracts were determined to be non-sensitizing.                                                                                                                                                                                                                                      | Same                |

{5}------------------------------------------------

{6}------------------------------------------------

### Comparison in Detail(s):

## Note 1

Although the material including outer facing layer, middle layer, inner facing layer, nose piece, ear loops is different from the predicate device, the subject device meet the requirement of essential performance standard ISO 10993. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.

## Note 2

There is a slightly difference of the mask dimensions between the subject device and the predicate device. However, both the subject and predicate device meet the ASTM F2100-19 barrier protection level 2. Therefore, the differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.

#### VII. PERFORMANCE DATA

## Summary of Non-Clinical Performance Test

{7}------------------------------------------------

| Performance<br>Characteristics                                                  | Test Method                                                                                                                                                                            | Pass Criteria<br>For Level 2               | Test Results |
|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|--------------|
| Bacterial<br>Filtration<br>Efficiency                                           | ASTM F2101 Standard Method<br>for Evaluating the Bacterial<br>Filtration Efficiency (BFE) of<br>Medical Face Mask Materials,<br>Using a Biological Aerosol of<br>Staphylococcus aureus | ≥98%                                       | Pass         |
| Differential<br>Pressure<br>(Delta-P)                                           | EN 14683: 2019, Annex C<br>Medical face masks -<br>Requirements and test methods<br>according to ASTM F2100:2019                                                                       | <6.0 mm<br>H2O/cm²                         | Pass         |
| Sub-Micron<br>Particulate<br>Filtration<br>Efficiency<br>(PFE) at 0.1<br>micron | ASTM F2299 Standard Test<br>Method for Determining the<br>Initial Efficiency of Materials<br>Used in Medical Face Masks to<br>Penetration by Particulates<br>Using Latex Spheres       | ≥98%                                       | Pass         |
| Resistance to<br>Penetration by<br>Synthetic<br>Blood                           | ASTM F1862 Standard Test<br>Method for Resistance of<br>Medical Face Masks to<br>Penetration by Synthetic Blood<br>(Horizontal Projection of Fixed<br>Volume at a Known Velocity)      | Fluid resistant<br>claimed at 120<br>mm Hg | Pass         |
| Flammability                                                                    | 16 CFR Part 1610 Standard for<br>the Flammability of Clothing                                                                                                                          | Class 1                                    | Pass         |

## Biocompatibility Testing

Based on ISO 10993-1:2018, the subject device contacts intact skin and its contact duration is less than or equal to 24h. Therefore, the following tests for the subject device were conducted to demonstrate that the subject device is biocompatible and safe for its intend use:

- . In Vitro Cytotoxicity Test per ISO 10993-5: 2009 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
- Skin Sensitization Tests per ISO 10993-10: 2010 Biological evaluation of ● medical devices - Part 10: Tests for irritation and skin sensitization
- Skin Irritation Tests per ISO 10993-10: 2010 Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization

{8}------------------------------------------------

## Clinical Test Conclusion

No clinical study is included in this submission.

# VIII. CONCLUSION

The conclusion drawn from the nonclinical tests demonstrates that the subject device, Surgical Face Masks are as safe, as effective, and performs as well as or better than the legally marketed predicate device Surgical Face Mask (K182514).

---

**Source:** [https://fda-staging.innolitics.com/device/K212235](https://fda-staging.innolitics.com/device/K212235)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).