JUSHA-C1210G/JUSHA-C1210/C1210G/C1210 LCD Monitor is intended to be used in displaying and viewing digital images, including standard and multi-frame digital mammography, for review, analysis, and diagnosis by trained medical practitioners. It is specially designed for displaying and viewing digital images, including standard and multi-frame digital mammography, for review, analysis, and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.
Device Story
Display system for medical imaging; inputs digital image data from medical workstations; transforms data via 16-bit grayscale LUT and high-resolution color-TFT LCD panel; outputs visual images for clinical review. Used in clinical settings by radiologists/physicians. Features ambient brightness adaptation, real-time DICOM automatic calibration, full-screen brightness equalization, and presence induction. Output allows clinicians to perform diagnostic review of mammography and tomosynthesis images; aids in clinical decision-making by providing high-fidelity, calibrated visualization of diagnostic data.
Clinical Evidence
Bench testing only. No clinical or animal studies were required. Bench testing included measurement of angular luminance response, luminance/chromaticity non-uniformity (TG18 guidelines), small-spot contrast ratio, temporal response, and luminance stability. Electrical safety and EMC testing confirmed compliance with IEC 60601-1 and IEC 60601-1-2.
Indicated for trained medical practitioners to display and view digital images, including standard and multi-frame digital mammography and breast tomosynthesis, for review, analysis, and diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Nanjing Jusha Display Technology Co., Ltd. % Donny Lee Certification Engineer 8A, Block 1. No. 301, Hanzhongmen Street Nanjing, Jiangsu 210036 CHINA
March 25, 2022
#### Re: K212231
Trade/Device Name: C1210G LCD Monitor, JUSHA-C1210G LCD Monitor, C1210 LCD Monitor, JUSHA-C1210 LCD Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: PGY Dated: January 24, 2022 Received: January 24, 2022
### Dear Donny Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D. Assistant Director Mammography Ultrasound and Imaging Software Branch Division of Radiological Health OHT7: Office of in vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212231
Device Name C1210G/JUSHA-C1210G/C1210/JUSHA-C1210 LCD Monitor
Indications for Use (Describe)
JUSHA-C1210G/JUSHA-C1210/C1210G/C1210 LCD Monitor is intended to be used in displaying and viewing digital images, including standard and multi-frame digital mammography, for review, analysis, and diagnosis by trained medical practitioners. It is specially designed for displaying and viewing digital images, including standard and multi-frame digital mammography, for review, analysis, and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CER 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CER 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K212231
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | May 26, 2021 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Nanjing Jusha Display Technology Co., Ltd |
| Contact Person: | Add: 301, 8F Block A, No.1, Nanjing International Service<br>Outsourcing Mansion, Hanzhongmen Street,Nanjing, 210036 China<br><br>Dongdong Li<br><br>Certification Manager<br><br>Nanjing Jusha Display Technology Co., Ltd<br><br>Tel: +86-25- 83305050<br><br>Fax: +86-25- 58783273 |
| Device Trade Name: | JUSHA-C1210G LCD Monitor, JUSHA-C1210 LCD Monitor,<br>C1210G LCD Monitor, C1210 LCD Monitor |
| Common/Usual Name: | 12MP Color LCD Monitor |
| Classification Name: | System, image processing, Radiological 21CFR 892.2050<br>PGY |
| Product Code: | |
| Predicate Device(s): | JUSHA JUSHA-M550G; K190848 |
| Device Description: | JUSHA-C1210G/JUSHA-C1210/C1210G/C1210 LCD Monitor is the<br>display system with the high resolution (4200×2800), high<br>luminance (800 cd/m²), and 16-bit grayscale (65536 level), built-in<br>DICOM standard LUT. In particular, JUSHA- C1210G has ambient<br>brightness adapt inside. In particular, JUSHA- C1210G has ambient<br>brightness adapting, real-time DICOM automatic calibration, full-<br>screen brightness equalization and presence induction system, with<br>these this display can automatic adjustment according to different<br>requirements in order to achieve the best results.<br><br>The product is consisted of the following components:<br><br>- 31" Color-TFT LCD Panel<br><br>- DMF2604AR0/main board/REV1.0<br><br>- JUSHA-C1210G LCD Monitor software<br><br>- Power Adapter |
{4}------------------------------------------------
| | - Data Cable. | | |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| | The LCD Monitor is designed, tested, and will be manufactured in<br>accordance with both mandatory and voluntary standards: | | |
| | 1. IEC 60601-1:2012, EN 60601-1:2013, ANSI/AAMI<br>ES60601-1:2005+A1:2012+C1:2009+A2:2010,<br>CAN/CSA C22.2 NO.60601-1:14, Medical equipment<br>medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance. | | |
| | 2. IEC 60601-1-2 Edition 4:2014, EN 60601-1-2:2015,<br>CFR 47 FCC Part15 subpart B: 2017, Medical electrical<br>equipment - Part 1-2: General requirements for basic<br>safety and essential performance - Collateral standard:<br>Electromagnetic disturbances - Requirements and tests. | | |
| Intended Use: | JUSHA-C1210G/JUSHA-C1210/C1210G/C1210 LCD Monitor is<br>intended to be used in displaying and viewing digital images,<br>including standard and multi-frame digital mammography, for review,<br>analysis, and diagnosis by trained medical practitioners. It is specially<br>designed for displaying and viewing digital images, including<br>standard and multi-frame digital mammography, for review, analysis,<br>and diagnosis by trained medical practitioners. It is specially designed<br>for breast tomosynthesis applications. | | |
| Technology: | JUSHA-C1210G/JUSHA-C1210/C1210G/C1210 LCD Monitor is the<br>display system with the high resolution ( $4200\times2800$ ), high<br>luminance (800 cd/m²), and 16-bit grayscale (65536 level), built-in<br>DICOM standard LUT. In particular, JUSHA- C1210G has ambient<br>brightness adapt inside. In particular, JUSHA- C1210G has ambient<br>brightness adapting, real-time DICOM automatic calibration, full-<br>screen brightness equalization and presence induction system, with<br>these this display can automatic adjustment according to different<br>requirements in order to achieve the best results. | | |
| Determination of Substantial<br>Equivalence: | Summary of Non-Clinical Tests:<br>The LCD Monitor complies with voluntary standards as following: | | |
| | 1 IEC 60601-1:2012, EN 60601-1:2013, ANSI/AAMI ES60601-<br>1:2005+A1:2012+C1:2009+A2:2010, CAN/CSA C22.2<br>NO.60601-1:14, Medical equipment medical electrical equipment<br>- Part 1: General requirements for basic safety and essential<br>performance. | | |
{5}------------------------------------------------
| 2 | IEC 60601-1-2 Edition 4:2014, EN 60601-1-2:2015, CFR 47<br>FCC Part15 subpart B: 2017, Medical electrical equipment - Part<br>1-2: General requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic disturbances -<br>Requirements and tests |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | |
| | JUSHA-C1210G is substantially equivalent to JUSHA JUSHA-<br>M550G. JUSHA-C1210G employs the maximum resolution values<br>larger than that of JUSHA-M550G. Comparison table of the principal<br>characteristics of 2 devices is shown in the Product Comparison. |
| | |
| | The following quality assurance measures were applied to the<br>development of the system: |
| | • Risk Analysis<br>• Requirements Reviews<br>• Design Reviews<br>• Raw materials verification<br>• Testing on unit level (Module verification)<br>• Integration testing (System verification)<br>• Final acceptance testing (Validation)<br>• Performance testing (Verification)<br>• Safety testing (Verification) |
| | |
| Summary of Clinical Tests: | |
| | The subject of this premarket submission, LCD Monitor, did not<br>require clinical studies to support substantial equivalence. |
| | |
| | The proposed device is Substantially Equivalent (SE) to the predicate<br>device which is US legally market device. Therefore, the subject<br>device is determined as safe and effectiveness. |
| Conclusion: | Nanjing Jusha Display Technology Co., Ltd Considers the JUSHA-<br>C1210G LCD Monitor to be as safe, as effective, and performance is |
# 12.1 Product Comparison
{6}------------------------------------------------
This comparison identifies the similarities and differences of the proposed JUSHA-C1210G LCD Monitor device to the legally marketed predicate JUSHA JUSHA-M550G LCD Monitor device to which substantial equivalency is claimed.
| Attributes | Predicate Device | Proposed Device | Discussion of<br>Differences |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product | JUSHA JUSHA-M550G | JUSHA-C1210G LCD<br>Monitor | |
| 510(k) Number | K190848 | / | |
| Display Performance/Specifications | | | |
| Screen<br>technology | 21.3inches, Mono-TFT LCD<br>Panel | 31inches, Color-TFT LCD<br>Panel | C1210G is larger<br>than M550G |
| Viewing angle<br>(H, V) | Horizontal 170°,Vertical<br>170° | Horizontal 178°,Vertical<br>178° | C1210G is better<br>than M550G |
| Resolution | $2560 \times 2048/2048 \times 2560$ | $4200\times2800$ | C1210G is larger<br>than M550G |
| Display area | $422.4(H)\times377.92(V)$ mm | $676.9(H)\times459.7(V)$ mm | C1210G is larger<br>than M550G |
| Contrast Ratio | 1700:1 | 1500:1 | C1210G is<br>smaller than<br>M550G, but the<br>difference does<br>not make<br>difference to the<br>diagnosis. |
| Supplied<br>CURVE | DICOM\CIE\GAMMA2.2\G<br>AMMA2.4 | DICOM\CIE\GAMMA2.2\G<br>AMMA2.4 | Same |
| DICOM<br>calibrated<br>luminance | 1000cd/m² | 800cd/m² | C1210G is<br>smaller than<br>M550G, and<br>both of them<br>meet the<br>requirements of<br>AAPM<br>regarding breast<br>diagnosis. |
| Pixel Pitch | $0.165\times0.165$ mm | $0.1554\times0.1554$ mm | C1210G is better<br>than M550G |
| Backlight | LED | LED | Same. |
| DICOM LUT | 16-bit:65536 | 16-bit:65536 | Same. |
| Attributes | Predicate Device | Proposed Device | Discussion of<br>Differences |
| Product | JUSHA JUSHA-M550G | JUSHA-C1210G LCD<br>Monitor | |
| 510(k) Number | K190848 | / | |
| Luminance<br>calibration | Built in calibration sensor<br>provided | Built in calibration sensor<br>provided | Same |
| Video Signal Input | | | |
| Input signals | DVI standard 1.0,<br>DisplayPort 1.2a | DisplayPort 1.2a | The difference<br>only shows that<br>they have<br>different<br>input,has nothing<br>to do with the<br>display function. |
| Input<br>terminational | DVI-D×1,<br>DisplayPort×1 | DisplayPort×2 | The difference<br>only shows that<br>they have<br>different<br>input,has nothing<br>to do with the<br>display function. |
| Output signals | DisplayPort 1.2a | HDMI 2.0 | The difference<br>only shows that<br>they have<br>different output,<br>has nothing to do<br>with the display<br>function. |
| Output<br>Terminational | DisplayPort×1 | HDMI×1 | The difference<br>only shows that<br>they have<br>different output,<br>has nothing to do<br>with the display<br>function. |
| Display<br>controller | Off the shelf | Off the shelf | Same |
| Power Related Specification | | | |
| Power<br>Requirement | AC 100~240V 50~60Hz | AC 100~240V 50~60Hz | Same |
| Attributes | Predicate Device | Proposed Device | Discussion of<br>Differences |
| Product | JUSHA JUSHA-M550G | JUSHA-C1210G LCD<br>Monitor | |
| 510(k) Number | K190848 | / | |
| Power<br>Consumption/Save Mode | 55W/less than 0.5W | 150W/less than 0.5W | The differences<br>caused by<br>components used<br>in the LCD<br>Monitor. This<br>only shows the<br>power<br>consumption is<br>different,<br>nothing to do<br>with the display<br>function |
| Power<br>Management | DVI DMPM<br>DisplayPort 1.2a | DisplayPort 1.2a | The difference<br>only shows that<br>they have<br>different power<br>Management,<br>has nothing to do<br>with the display<br>function. |
| Miscellaneous Features/Specifications | | | |
| USB<br>Ports/standard | 1 upstream (endpoint),<br>2 downstream/ Rev. 2.0 | 1 upstream (endpoint),<br>2 downstream/ Rev. 3.0 | C1210G is better<br>than M550G. |
| Dimensions w/o<br>stand<br>(W×H×D) | Without stand:<br>$363mm×475mm×66mm$<br>With stand:<br>$363mm×635mm×238mm$ | Without stand:<br>$701.3 mm× 500 mm×$<br>86.8mm<br>With stand:<br>$701.3 mm× 589 mm×$<br>245mm | Different<br>housing design<br>due to the<br>different panel<br>size. |
| Attributes | Predicate Device | Proposed Device | Discussion of<br>Differences |
| Product | JUSHA JUSHA-M550G | JUSHA-C1210G LCD<br>Monitor | |
| 510(k) Number | K190848 | / | |
| Indication for<br>use | JUSHA-M550G/JUSHA-<br>M550/M550G/M550 LCD<br>Monitor is intended to be<br>used in displaying and<br>viewing digital images,<br>including standard and<br>multi-frame digital<br>mammography, for review,<br>analysis, and diagnosis by<br>trained medical practitioners.<br>It is specially de displaying<br>and viewing digital images,<br>including standard and<br>multi-frame digital<br>mammography, for review,<br>analysis, and diagnosis by<br>trained medical practitioners.<br>It is specially designed for<br>breast tomosynthesis<br>applications. | JUSHA-C1210G/JUSHA-<br>C1210/ C1210G/ C1210<br>LCD Monitor is intended to<br>be used in displaying and<br>viewing digital images,<br>including standard and<br>multi-frame digital<br>mammography, for review,<br>analysis, and diagnosis by<br>trained medical practitioners.<br>It is specially de displaying<br>and viewing digital images,<br>including standard and<br>multi-frame digital<br>mammography, for review,<br>analysis, and diagnosis by<br>trained medical practitioners.<br>It is specially designed for<br>breast tomosynthesis<br>applications. | Same |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
| Attributes | Predicate Device | Proposed Device | Discussion of<br>Differences |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.