AATMA™ is a medical imaqe processing library intended to produce derived data sets for use as input into radiation therapy treatment planning systems or other intermediate pre-treatment-planning applications. AATMA™ does not provide a user interface and is designed to be accessed through its application programming interface (API) by other devices. The data sets created by AATMA™ must be reviewed and validated by a qualified healthcare professional prior to clinical use.
Device Story
AATMA™ is a medical image processing library acting as an accessory to radiation therapy treatment planning systems. It functions as a computational engine accessed via API by other devices; it lacks a user interface and does not store input/output data or logs. The device uses pre-trained machine-learning convolutional neural networks to perform auto-segmentation on 3D CT image sets. It produces derived datasets in standard formats (e.g., DICOM) for use by other applications. Healthcare professionals must review and edit the generated contours within a treatment planning system before clinical use. The device benefits patients by automating the segmentation process, potentially increasing efficiency in radiation therapy planning workflows.
Clinical Evidence
No clinical or animal testing performed. Bench testing only. Performance validated using 3D CT image sets. Head & Neck model (trained on 66 sets) achieved average DICE coefficient of 0.84 on verification set. Male Pelvis model (trained on 205 sets) achieved average DICE coefficient of 0.93 on verification set. All validation and verification met defined acceptance criteria.
Technological Characteristics
Medical image processing library; machine-learning convolutional neural network architecture; supports standard DICOM image formats; TCP/IP networking; API-based integration; software-only device; validated for CT imaging.
Indications for Use
Indicated for use as a medical image processing library to produce derived data sets for radiation therapy treatment planning or intermediate pre-treatment-planning applications. Intended for use by qualified healthcare professionals who must review and validate all outputs prior to clinical application.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
Elekta Solutions AB % Anju Kurian, M.S., RAC Manager, Regulatory Affairs - Software 1450 Beale Street. Suite 205 SAINT CHARLES MO 63303
## Re: K212218
Trade/Device Name: Advanced Algorithms for Treatment Applications (AATMA™) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QKB, LLZ Dated: September 8, 2021 Received: September 16, 2021
# Dear Anju Kurian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Expiration Date: 06/30/2023
See PRA Statement below.
Form Approved: OMB No. 0910-0120
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K212218
### Device Name Advanced Algorithms for Treatment Management Applications (AATMA™)
Indications for Use (Describe)
AATMA™ is a medical imaqe processing library intended to produce derived data sets for use as input into radiation therapy treatment planning systems or other intermediate pre-treatment-planning applications. AATMA™ does not provide a user interface and is designed to be accessed through its application programming interface (API) by other devices. The data sets created by AATMA™ must be reviewed and validated by a qualified healthcare professional prior to clinical use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Elekta, a company that specializes in radiation therapy and neurosurgery solutions. The logo consists of a stylized circle with three smaller circles inside, followed by the word "Elekta" in a sans-serif font. The color of the logo is a teal blue.
# TRADITIONAL 510(K) SUMMARY (21 CFR § 807.92)
| I. | SUBMITTER | Elekta Solutions AB<br>Kungstensgatan 18 Box 7593<br>Stockholm, Stockholms lan [SE-01] SE SE10393 |
|-----|----------------------------------|---------------------------------------------------------------------------------------------------|
| | Contact: | Anju Kurian, M.S., RAC<br>Manager, Regulatory Affairs - Software |
| | Establishment<br>Registration #: | 3015232217 |
| | 510(k) Number: | K212218 |
| | Date Prepared: | 10/18/2021 |
| II. | DEVICE | |
| | Trade Name: | AATMA ™ (Advanced Algorithms for Treatment<br>Management Applications) |
| | Release Version #: | Release 1.0 |
| | Product Classification: | Class II |
| | Common Name: | Radiological Image Processing Software for Radiation<br>Therapy |
| | Classification Name: | Medical Image Management and Processing System |
| | Regulation Number: | 21 CFR § 892.2050 |
| | Product Code: | QKB/LLZ |
#### lll. PREDICATE DEVICE
Workflow Box by Mirada Medical (K181572)
#### IV. DEVICE DESCRIPTION
AATMA™ is an optional accessory to treatment planning systems and intermediate pre-treatment planning applications. The auto-segmentation algorithm in AATMA™ is based on machine-learning convolutional neural networks and includes pre-trained models that will be used to automatically segment image sets. The algorithm itself functions as a computational engine and does not store any input data, output data, or logs. The available models have been pre-trained on specific datasets that exhibit similar characteristics (e.g., body site and imaging modality).
As a medical image processing library, AATMA™ is designed to produce derived datasets in standard formats (e.g., DICOM) that can be utilized by other applications. AATMA™ does not have a user interface and, as such, calling applications must
{4}------------------------------------------------
execute the auto-segmentation algorithms via AATMA™'s application programming interface (API).
AATMA™ must be used in conjunction with appropriate software to review and edit results generated automatically by the auto-segmentation alqorithm. A pre-treatment planning system or treatment planning system must be used to facilitate the review and edit of contours generated by the auto-segmentation algorithm within AATMA™.
#### V. INTENDED USE
AATMA™ is a medical imaqe processing library intended to produce derived data sets for use as input into radiation therapy treatment planning systems or other intermediate pre-treatment-planning applications. AATMA™ does not provide a user interface and is designed to be accessed through its application programming interface (API) by other devices. The data sets created by AATMA™ must be reviewed and validated by a qualified healthcare professional prior to clinical use.
#### VI. INDICATIONS FOR USE
AATMA™ is a medical image processing library intended to produce derived data sets for use as input into radiation therapy treatment planning systems or other intermediate pre-treatment-planning applications. AATMA™ does not provide a user interface and is designed to be accessed through its application programming interface (API) by other devices.
The data sets created by AATMA™ must be reviewed and validated by a qualified healthcare professional prior to clinical use.
#### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE
| Technological Characteristic | AATMA™<br>(Subject Device) | Workflow Box<br>Predicate Device<br>K181572 |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Automatic contouring of imaging data using<br>machine learning based models | ✓ | ✓ |
| No Graphical User Interface | ✓ | ✓ |
| View manipulation and Volume rendering – Not<br>Applicable | ✓ | ✓ |
| Image registration | N/A | ✓ |
| Reporting and Data Routing | N/A | ✓ |
| Supported modalities: Standard DICOM image<br>modality support | ✓<br>Subject device<br>validated with CT<br>images for image<br>processing. | ✓<br>Predicate device<br>validated with CT,<br>MR, DICOM<br>RTSTRUCT for<br>image processing. |
| TCP/IP Networking and Communication | ✓ | ✓ |
{5}------------------------------------------------
#### VIII. SUMMARY OF PERFORMACE TESTING (NON-CLINICAL)
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
AATMA™ is validated and verified against its user needs and intended use by the successful execution of planned performance, functional and algorithmic testing included in this submission. The results of performance, functional and algorithmic testing demonstrate that AATMA™ meets the user needs and requirements of the device, which are demonstrated to be substantially equivalent to those of the listed predicate device.
Verification and Validation for AATMA™ has been carried out in compliance with the requirements of CFR 21 Part 820 and in adherence to the DICOM standard.
Performance testing for two models – Head & Neck, Male Pelvis were conducted.
The Head & Neck model was trained on 66 unique clinical patient 3D CT image sets from a variety of institutions and equipment. A different set of six(6) patient CT image sets with expert contours were chosen for verification and the average DICE coefficient over all structures was determined to be 0.84 which met the defined acceptance criteria. A different set of 13 3D CT image sets were used for validation and these met the acceptance criteria as well.
The Male Pelvis model was trained on 205 unique patient 3D CT image sets from a global variety of institutions and equipment from patients undergoing RT. A different set of five (5) patient CT image sets with expert contours were chosen for verification and the average DICE coefficient over all structures was determined to be 0.93 which met the defined acceptance criteria. A different set of 20 3D CT image sets were used for validation and these met the acceptance criteria as well.
#### IX. SUMMARY OF PERFORMACE TESTING (CLINICAL)
No animal or clinical tests were performed to establish substantial equivalence with the predicate device. The performance data demonstrate that AATMA™ is as safe and effective and performs as well as the predicate device Workflow Box by Mirada Medical cleared under K181572.
#### SUBSTANTIAL EQUIVALENCE CONCLUSION X.
In conclusion, performance testing and device evaluations presented in this 510(k) demonstrates that AATMA™ is substantially equivalent to and performs at least as safely and effectively as the listed predicate device. AATMA™ meets the requirements for safety and effectiveness as applicable to radiological image processing software and does not introduce any new potential safety risks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.