← Product Code [GCJ](/productcode/GCJ) · K212214

# HandX (K212214)

_Human Xtensions , Ltd. · GCJ · Sep 13, 2021 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212214

## Device Facts

- **Applicant:** Human Xtensions , Ltd.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Sep 13, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The HandX is intended to assist in the control of Human Xtensions laparoscopic instruments including needle holder, grasper and monopolar instruments, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, cutting and/or coagulation, during laparoscopic surgical procedures. The HandX monopolar instruments are connected by a standard cable to a standard electrosurgical power source. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use.

## Device Story

HandX is a handheld, electromechanically controlled laparoscopic device acting as a surgeon's hand extension. It consists of a reusable control interface (CI) handle and a disposable instrument (grasper, needle holder, or monopolar hook). The surgeon operates control buttons on the CI to translate hand movements into articulation of the instrument tip. The device supports two control modes: direct ('Ro-bo style') and inverted ('La-pro style'). Monopolar instruments connect to standard electrosurgical power sources. Used in operating rooms by trained physicians to enhance maneuverability and control during minimally invasive surgery. The device benefits patients by providing precise, articulated instrument control for complex surgical tasks.

## Clinical Evidence

No clinical data. Usability testing was conducted in accordance with IEC 62366-1:2015 and ANSI/AAMI HE75:2009/(R)2018 to evaluate the inverted control mode. Results demonstrated that potential safety-critical consequences are mitigated by design and clear instructional materials for both novel and experienced users.

## Technological Characteristics

Electromechanically controlled, software-driven handheld device. Comprised of a reusable control interface handle and single-use instruments (grasper, needle holder, monopolar hook). Features include articulation, direct and inverted control modes, and standard electrosurgical connectivity. Sterilization is required for reusable components per instructions. Connectivity is limited to standard electrosurgical cables.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- HX Device, Human Xtensions ([K173919](/device/K173919.md))
- Monopolar Hook, Human Xtensions ([K203603](/device/K203603.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

September 13, 2021

Human Xtensions Ltd. % Randy Prebula Partner Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW Washington, District of Columbia 20004

Re: K212214

Trade/Device Name: HandX Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, NAY, GEI Dated: July 15, 2021 Received: July 15, 2021

Dear Randy Prebula:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use Statement

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration            |
| <b>Indications for Use</b>              |

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement on last page

510(k) Number (if known) K212214

Device Name

HandX

Indications for Use (Describe)

The HandX is intended to assist in the control of Human Xtensions instruments including needle holder, grasper and monopolar instruments, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, cutting and/or coagulation, during laparoscopic surgical procedures. The HandX monopolar instruments are connected by a standard cable to a standard electrosurgical power source. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use.

Type of Use (Select one or both, as applicable)

Z Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "

FORM FDA 3881 (6/20)

Page 1 of 1 FDA

PSC Publishing Services (301) 443-6740 ER

{3}------------------------------------------------

### 510(k) SUMMARY

#### Human Xtensions Ltd.'s HandX

| Submitter:                                                                                                     | Human Xtensions Ltd.<br>4 Meir Ariel Street<br>Netanya, 4250574, Israel<br>Telephone: + (972) 77 36 30 300                                                                                                                                |
|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                                                                                                       | Randy J Prebula<br>Hogan Lovells US LLP<br>Columbia Square<br>555 Thirteenth Street, NW<br>Washington, DC 20004<br>Telephone: +1 202 637 5600<br>Direct: +1 202 637 6548<br>Fax: +1 202 637 5910<br>Email: randy.prebula@hoganlovells.com |
| Date Prepared:                                                                                                 | September 9, 2021                                                                                                                                                                                                                         |
| Subject Device:                                                                                                |                                                                                                                                                                                                                                           |
| Device Name:<br>Common Name:<br>Classification Name:<br>Product Codes:<br>Product Class:<br>Regulation Number: | HandX<br>HandX<br>Laparoscope, General & Plastic Surgery<br>GCJ, NAY, GEI<br>Class II<br>21 CFR 876.1500, 876.4400                                                                                                                        |
| Predicate Devices:                                                                                             | HX Device, Human Xtensions (K173919)<br>Monopolar Hook, Human Xtensions (K203603)                                                                                                                                                         |

## Purpose of the Special 510(k) Notice

The HandX is a modification to the prior versions of the device that were cleared under K173919 (HX Device) and K203603 (Monopolar Hook).

## Intended Use

The HandX is intended to assist in the control of Human Xtensions laparoscopic instruments including needle holder, grasper and monopolar instruments, for endoscopic manipulation of tissue, including grasping, approximation, ligation, suturing, cutting and/or coagulation, during laparoscopic surgical procedures. The HandX monopolar instruments are connected by a standard cable to a standard electrosurgical power source. It is intended to be used by trained physicians in an operating room environment in accordance with its Instructions for Use.

## Device Description

The Human Xtensions HandX is a handheld powered laparoscopic device. The HandX enables, by its attached instruments Grasper, Needle Holder and Monopolar Hook, the surgeon to perform a variety 

{4}------------------------------------------------

of minimally invasive surgeries. As the device is controlled by the surgeon's hand, it acts as a "surgeon's hand extension," enhancing maneuverability and control.

The HandX device is electromechanically controlled and includes both hardware and software. The device is comprised of two main parts: the Handpiece and the Instrument. The Handpiece consists of a reusable Control Interface (CI) handle and Handpiece Body that translates the surgeon's maneuvers and movements of their hands by means of control buttons to the instrument-articulating tip. The tip moves according to the CI's direction. Prior to use, a set of Power Cable, Finger Pads, Arc and Spacer are assembled on the Handpiece and the Handpiece is covered with a standard sterile cover (not provided with the HandX device). After use, the Instrument is disposed of, and the Handpiece is cleaned and disinfected per specifications in the Instructions for Use.

The HandX has nearly identical technological characteristics and principles of operation as the combined predicate devices. The only notable difference between the subject HandX and the combined predicates is the addition of the inverted control mode; whereas the predicate devices' software included only direct control ("Ro-bo style") for the Cl. the subject HandX provides direct control as well as inverted control ("La-pro style") as an additional, optional CI control mode. This difference does not raise new questions of safety or effectiveness, as it does not alter the fundamental therapeutic purpose of the device or its basic principles of operation (users control the inserted instruments in the same manner, just using one or another set of vectors). The new mode also does not change the key metrics for evaluating device safety and performance, and is supported by successful usability testing (as noted below).

# Performance Data

The company conducted usability testing of the inverted control mode in line with IEC 62366-1:2015 and ANSI/AAMI HE75:2009/(R)2018, as well as FDA's 2016 guidance document. The testing demonstrated that potential safety-critical consequences are reasonably mitigated by the design of the inverted control mode and its associated instructional materials, which were clear and easy to understand by both novel and experienced users.

# Conclusion

Based on the intended use, technological characteristics, principles of operation, and usability testing, it can be concluded that the HandX is substantially equivalent to its predicate devices.

{5}------------------------------------------------

# Substantial Equivalence (SE) Table

|                                    | Human Xtensions HandX (Modified<br>Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Human Extensions HX Device<br>(K173919)                                                                                                                                                                                                                                                                                                                                                                                               | Human Xtensions Monopolar Hook<br>(K203603)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison of<br>Subject to Predicates                                                              |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Intended Use                       | Intended for use in various laparoscopic<br>surgeries to perform a variety of surgical<br>functions, including grasping,<br>approximation, ligation, suturing, cutting<br>and/or coagulation                                                                                                                                                                                                                                                                                                                                                                                                                    | Intended for use in various<br>laparoscopic surgeries to perform a<br>variety of surgical functions,<br>including grasping, approximation,<br>ligation, and suturing.                                                                                                                                                                                                                                                                 | Intended for use in various laparoscopic<br>surgeries to perform a variety of surgical<br>functions, including grasping,<br>approximation, ligation, suturing, cutting<br>and/or coagulation                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                |
| Indications for Use                | The HandX is intended to assist in the<br>control of Human Xtensions laparoscopic<br>instruments including needle holder,<br>grasper and monopolar instruments, for<br>endoscopic manipulation of tissue,<br>including grasping, approximation,<br>ligation, suturing, cutting and/or<br>coagulation, during laparoscopic surgical<br>procedures. The HandX monopolar<br>instruments are connected by a standard<br>cable to a standard electrosurgical power<br>source. It is intended to be used by<br>trained physicians in an operating room<br>environment in accordance with its<br>Instructions for Use. | The HX Device is intended to assist<br>in the accurate control of HX<br>laparoscopic instruments including<br>needle holder and grasper, for<br>endoscopic manipulation of tissue,<br>including grasping, approximation,<br>ligation, suturing, during<br>laparoscopic surgical procedures. It<br>is intended to be used by trained<br>physicians in an operating room<br>environment in accordance with its<br>Instructions for Use. | The HandX is intended to assist in the<br>control of Human Xtensions<br>laparoscopic instruments including<br>needle holder, grasper and monopolar<br>instruments, for endoscopic<br>manipulation of tissue, including<br>grasping, approximation, ligation,<br>suturing, cutting and/or coagulation,<br>during laparoscopic surgical procedures.<br>The HandX monopolar instruments are<br>connected by a standard cable to a<br>standard electrosurgical power source.<br>It is intended to be used by trained<br>physicians in an operating room<br>environment in accordance with its<br>Instructions for Use. | Same - Identical to<br>K203603; only adds<br>Monopolar Hook to<br>indications cleared in<br>K173919 |
| Specialty                          | Laparoscopic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                |
| Mode of Operation                  | Electromechanically operated, software controlled                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                |
| Components                         | Handpiece and compatible instruments<br>(Grasper, Needle Holder, Monopolar<br>Hook); Arc, Pads, Spacer, Power Cable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Handpiece and compatible<br>instruments (Grasper, Needle<br>Holder); Arc, Pads, Spacer, Power<br>Cable                                                                                                                                                                                                                                                                                                                                | Instrument compatible with HandX<br>Handpiece (Monopolar Hook)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same as combined<br>predicates                                                                      |
| Grasper Type                       | Fenestrated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                       | N/A*                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same as K173919                                                                                     |
| Needle Holder Type                 | Straight - Single Action                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                       | N/A*                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same as K173919                                                                                     |
| Electrosurgical<br>Instrument Type | Monopolar hook                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                   | Monopolar hook                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same as K203603                                                                                     |
| Articulation                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                |
| Control Interface<br>Mode          | Direct control and inverted control                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Direct control                                                                                                                                                                                                                                                                                                                                                                                                                        | N/A*                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Adds inverted control<br>mode                                                                       |
| Single Use /<br>Reusable           | Handpiece: reusable<br>Instruments: single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Handpiece: reusable<br>Instruments: single use                                                                                                                                                                                                                                                                                                                                                                                        | Instrument: single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same as combined<br>predicates                                                                      |
| Sterile Components                 | Instruments, Arc, Finger Pads, Spacer,<br>and Power Cable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Instruments, Arc, Finger Pads,<br>Spacer and Power Cable                                                                                                                                                                                                                                                                                                                                                                              | Instrument (does not include other<br>components)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same as K173919                                                                                     |

*The K203603 filing only included the Monopolar Hook and did not re-include the entire system.

---

**Source:** [https://fda-staging.innolitics.com/device/K212214](https://fda-staging.innolitics.com/device/K212214)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
