K212198 · Plasmapp Co,., Ltd. · FRG · Oct 22, 2021 · General Hospital
Device Facts
Record ID
K212198
Device Name
Tyvek(R) Roll with CI for STERLINKTM Sterilizer
Applicant
Plasmapp Co,., Ltd.
Product Code
FRG · General Hospital
Decision Date
Oct 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in CHAMBER mode, is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERLINK™ sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 1 month post sterilization. The maximum load weight that can be placed in the Tyvek® roll is 3.97 pounds (1.8kg). The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer. The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in the follow 1 type: · Sterilization roll, Flat
Device Story
Sterilization packaging system for medical devices; consists of Tyvek® sheet and clear plastic film heat-sealed into flat rolls. Used by healthcare workers in clinical settings to enclose instruments before processing in STERLINK™ sterilizer. Features integrated chemical indicator bar (ISO 11140-1:2005) printed on Tyvek®; changes color from red to blue/lighter upon exposure to hydrogen peroxide vapor. Provides visual confirmation of exposure to sterilization process. Maintains sterility of enclosed items for up to 1 month post-sterilization. Protects patient by ensuring instrument sterility at point of use.
Clinical Evidence
No clinical data. Bench testing only. Performance validated per ISO 14937, ANSI/AAMI/ISO 11140-1, ASTM F1980-16, ASTM F1140/F1140M-13, ASTM F1886/F1886M-16, ASTM F1929-15, ASTM D5035-11, ASTM D882, ASTM F88, ASTM D1922-20, DIN 58953-6, ISO 10993-5, and various material compatibility standards (ASTM D638, D790, D256, etc.). All tests passed.
Technological Characteristics
Materials: Tyvek®, PET, PE, adhesive, and hydrogen peroxide-sensitive ink. Form factor: Flat sterilization roll. Principle: Passive barrier packaging with integrated chemical process indicator (ISO 11140-1:2005). Sterilization compatibility: Hydrogen peroxide vapor (STERLINK™ system). Standards: ASTM F1980-16 (shelf-life), ASTM F1929-15 (dye penetration), ASTM F88 (seal strength), ISO 10993-5 (biocompatibility).
Indications for Use
Indicated for healthcare workers to enclose medical devices for sterilization in the STERLINK™ sterilizer (CHAMBER mode) and maintain sterility for up to 1 month post-sterilization. Maximum load weight: 1.8kg.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Sterilization Pouch/Roll Made with Tyvek® (K180672)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 22, 2021
Consultant CardioMed Device Consultants LLC 1783 Forest Drive, Suite 254 Annapolis, Maryland 21401
## Re: K212198
Trade/Device Name: Tyvek® Roll with CI for STERLINK™ Sterilizer Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, JOJ Dated: September 20, 2021 Received: September 21, 2021
## Dear Candace Cederman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212198
Device Name Tyvek® Roll with CI for STERLINK™ Sterilizer
#### Indications for Use (Describe)
Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in CHAMBER mode, is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERLINK™ sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 1 month post sterilization.
The maximum load weight that can be placed in the Tyvek® roll is 3.97 pounds (1.8kg).
The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer.
The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in the follow 1 type: · Sterilization roll, Flat
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------|--|--|
| | | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 1. General Information
| Applicant/Submitter: | Plasmapp Co., Ltd.<br>BVC-111, 125, Gwahak-ro,<br>Yuseong-gu, Daejeon, 34141,<br>Rep. of Korea (South Korea)<br>Tel: +82 (0)42 716 2115 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Candace Cederman |
| Address: | CardioMed Device Consultants LLC<br>1783 Forest Drive<br>Suite 254<br>Annapolis, MD 21401<br>Tel: +1 410 674 2060 |
| Preparation Date: | October 12, 2021 |
# 2. Device Name and Code
| Device Trade Name: | Tyvek® Roll with CI for STERLINK™ Sterilizer |
|----------------------|-------------------------------------------------------------|
| Common Name: | Tyvek® Roll for VH2O2 Sterilizer |
| Classification Name: | 1) Sterilization Wrap<br>2) Sterilization Process Indicator |
| Product Code: | 1) FRG<br>2) JOJ |
| Regulation Number: | 1) 21 CFR 880.6850<br>2) 21 CFR 880.2800 |
| Classification: | Class II |
| Review Panel: | General Hospital |
{4}------------------------------------------------
# 3. Indications / Intended Use
Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in CHAMBER mode, is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERLINK™ sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 1 month post sterilization.
The maximum load weight that can be placed in the Tyvek® roll is 3.97 pounds (1.8kg).
The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer.
The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in the follow 1 type:
- . Sterilization roll, Flat
## 4. Device Description
Tyvek® Roll with CI for STERLINK™ Sterilizer is intended to contain medical devices to be terminally sterilized in the STERLINK™ sterilization system. The medical devices are inserted into the roll, sealed, and then sterilized in the STERLINK™ sterilization system. After completion of the sterilization process, the roll maintains sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices for up to 1 month post sterilization.
The roll is printed with a chemical indicator bar that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer.
The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in one type as a Flat Sterilization roll. The sterilization roll is made from a Tyvek® sheet and a clear plastic film that are heat sealed together on opposite two sides. After being cut into a suitable length, the product to be sterilized is placed inside and the two open ends are heat sealed. The process indicator printed on the Tyvek® will exhibit a color change after the roll is exposed to hydrogen peroxide (H2O2).
## 5. Technical Characteristics in Comparison to Predicate Devices
The proposed subject device uses identical technology as the cited predicate devices and has the same intended uses. Based upon the overall performance characteristics for Tyvek® Roll with CI for STERLINK™ Sterilizer, Plasmapp Co., Ltd. believes there are no significant differences in
{5}------------------------------------------------
usage of its underlying technological principles between Tyvek® Roll with CI for STERLINK™ Sterilizer and the predicate device.
A summary of the technical characteristics of the subject device and predicate device can be found in the table below.
| | Subject Device | Predicate Device |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Sponsor | Plasmapp Co., Ltd. | Sigma Medical Supplies Corp. |
| Manufacturer | Sigma Medical Supplies Corp. | Same |
| Device Name | Tyvek® Roll with CI for STERLINK™ Sterilizer | Sterilization Pouch/Roll Made with Tyvek® |
| 510(k) Number | K212198 | K180672 |
| Device Classification Name | 1) Sterilization Wrap<br>2) Sterilization Process Indicator | Same |
| Classification Product Code | 1) FRG<br>2) JOJ | Same |
| Regulation Number | 1) 21 CFR 880.6850<br>2) 21 CFR 880.2800 | Same |
| Intended Use | Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in CHAMBER mode, is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERLINK™ sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 1 month post sterilization.<br><br>The maximum load weight that can be placed in the Tyvek® roll is 3.97lbs (1.8kg).<br><br>The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer.<br><br>The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in the follow 1 type:<br>• Sterilization roll, Flat | Sterilization Pouch/Roll Made with Tyvek are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® 100S Sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post sterilization.<br><br>The pouches and rolls are printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD® 100S Sterilizer.<br><br>The Sterilization Pouch/Roll Made with Tyvek® is offered in the follow 5 types:<br>• Self-sealing sterilization pouches<br>• Sterilization pouches, Flat<br>• Sterilization pouches, Gusseted<br>• Sterilization rolls, Flat<br>• Sterilization rolls, Gusseted |
| Pouch Types | The subject device includes a subset of the predicate device types:<br>Sterilization roll, Flat | Self-sealing sterilization pouches<br>Sterilization pouches, Flat<br>Sterilization pouches, Gusseted<br>Sterilization rolls, Flat<br>Sterilization rolls, Gusseted |
## Summary of Subject and Predicate Device Technical Characteristics
{6}------------------------------------------------
| Subject Device | | Predicate Device | | |
|--------------------------------------------------|----------|-----------------------------|---------------------------|-----------------|
| | | Sterilization Pouches, Flat | | |
| | | Model | Dimensions | |
| | | TYFP075200 | 75 mm × 200 mm | |
| | | TYFP075300 | 75 mm × 300 mm | |
| | | TYFP100200 | 100 mm × 200 mm | |
| | | TYFP100300 | 100 mm × 300 mm | |
| | | TYFP150255 | 150 mm × 255 mm | |
| | | TYFP150300 | 150 mm × 300 mm | |
| | | TYFP190330 | 190 mm × 330 mm | |
| | | TYFP200400 | 200 mm × 400 mm | |
| | | TYFP250450 | 250 mm × 450 mm | |
| | | TYFP300500 | 300 mm × 500 mm | |
| | | | Sterilization Rolls, Flat | |
| | | Model | Dimensions | |
| | | TYFR050070 | 50 mm × 70 M | |
| | | TYFR022f | 50.8 mm × 61 M | |
| | | TYFR075070 | 75 mm × 70 M | |
| | | TYFR100070 | 100 mm × 70 M | |
| | | TYFR042f | 101.6 mm × 61 M | |
| Device models<br>(Configurations<br>/Dimensions) | | TYFR150070 | 150 mm × 70 M | |
| | | Sterilization Roll, Flat | TYFR062f | 152.4 mm × 61 M |
| | Model | Dimensions | TYFR200070 | 200 mm × 70 M |
| | FR400100 | 400 mm × 100 M | TYFR082f | 203.2 mm × 61 M |
| | | TYFR250070 | 250 mm × 70 M | |
| | | TYFR102f | 254 mm × 61 M | |
| | | TYFR300070 | 300 mm × 70 M | |
| | | TYFR350070 | 350 mm × 70 M | |
| | | TYFR400070 | 400 mm × 70 M | |
| | | TYFR450070 | 450 mm × 70 M | |
| | | TYFR500070 | 500 mm × 70 M | |
| | | TYFR050100 | 50 mm × 100 M | |
| | | TYFR075100 | 75 mm × 100 M | |
| | | TYFR100100 | 100 mm × 100 M | |
| | | TYFR150100 | 150 mm × 100 M | |
| | | TYFR200100 | 200 mm × 100 M | |
| | | TYFR250100 | 250 mm × 100 M | |
| | | TYFR300100 | 300 mm × 100 M | |
| | | TYFR350100 | 350 mm × 100 M | |
| | | TYFR400100 | 400 mm × 100 M | |
| | | TYFR450100 | 450 mm × 100 M | |
| | | TYFR500100 | 500 mm × 100 M | |
| | | TYFR050200 | 50 mm × 200 M | |
| | | TYFR075200 | 75 mm × 200 M | |
| | | TYFR100200 | 100 mm × 200 M | |
| | | TYFR150200 | 150 mm × 200 M | |
| | | TYFR200200 | 200 mm × 200 M | |
| | | TYFR250200 | 250 mm × 200 M | |
| | | TYFR300200 | 300 mm × 200 M | |
| | | TYFR350200 | 350 mm × 200 M | |
| | | TYFR400200 | 400 mm × 200 M | |
{7}------------------------------------------------
| | Subject Device | Predicate Device |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material Composition | Tyvek®, PET, PE, Water, CH3COOH,<br>Alcohol, n-Heptane adhesive, Hydrogen<br>peroxide vapor Process Indicator Print Ink | Same |
| Sterilization Cycle | STERLINK™ FPS-15s Plus - Chamber mode<br>(overall cycle: 36 minutes) | STERRAD® 100S sterilization cycles:<br>54-minute Short cycle for most surgical instruments 72-minute Long cycle for flexible endoscopes and instruments with<br>longer lumen |
| Design Feature | Sterilization roll, Flat: This roll is made<br>from a Tyvek® and plastic film that are heat<br>sealed on opposite two sides. It will be cut<br>into the suitable length and the opened sides<br>will be heat-sealed. The indicators printed on<br>the Tyvek® are the same with the self-sealing<br>sterilization roll. | Same |
| Chemical Indicator<br>Device Design | The color of the Chemical Indicator changes<br>from red to blue (or lighter) when exposed to<br>hydrogen peroxide. | Same |
# 6. Summary of Non-Clinical Testing
The Tyvek® Roll with CI for STERLINK™ Sterilizer has the identical intended use and indication for use as the predicate devices. Testing of the roll material was previously conducted on the predicate K180672. The identical materials are used in the Tyvek® Roll with CI for STERLINK™ Sterilizer. The size of the subject device is within the range of sizes available for the predicate device.
The performance and safety test of subject device was evaluated using the STERLINK™ FPS-15s Plus system, operating in Chamber mode, following the standards listed in the table below.
| Performance Testing | | Standard | Result |
|--------------------------|-------------------------------------|---------------------------------------|--------|
| Sterilant<br>Penetration | Sterilization efficacy test | ISO 14937 | Pass |
| | CI of sterilization roll validation | ANSI/AAMI/ISO 11140-1 | Pass |
| Shelf-life | Internal pressurization test | ASTM F1980-16<br>ASTM F1140/F1140M-13 | Pass |
| | Visual inspection test | ASTM F1980-16<br>ASTM F1886/F1886M-16 | Pass |
| | Dye penetration test | ASTM F1980-16<br>ASTM F1929-15 | Pass |
| | Tensile strength of Tyvek® | ASTM F1980-16<br>ASTM D5035-11 | Pass |
| | Tensile strength of plastic film | ASTM F1980-16<br>ASTM D882 | Pass |
| | Seal strength | ASTM F1980-16<br>ASTM F88 | Pass |
{8}------------------------------------------------
| Tyvek® Roll with CI for STERLINK™ Sterilize |
|---------------------------------------------|
| 510(k) Summary - K212198 |
| Performance Testing | Standard | Result |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Tear resistance | ASTM F1980-16<br>ASTM D1922-20 | Pass |
| Microbial Barrier Test | ASTM F1980-16<br>DIN 58953-6 | Pass |
| CI of sterilization roll validation | ANSI/AAMI/ISO 11140-1 | Pass |
| Toxicological Properties<br>(Biocompatibility Tests) | ISO 10993-5 | Pass - Test article is non-<br>cytotoxic after the sterilization. |
| Residual sterilant on Tyvek® validation | Internal test standard | Pass |
| Material compatibility test | ASTM D638<br>ASTM E8/E8M-ae1<br>ASTM D790<br>ASTM E290-14<br>ASTM D256<br>ASTM E23-18<br>ASTM E1164<br>ASTM E313<br>ASTM D3985<br>ASTM F1249<br>Internal test standard | Pass |
# 7. Conclusions
The conclusions drawn from the non-clinical tests demonstrates that the subject device in 510(k) submission K212198, Tyvek® Roll with CI for STERLINK™ Sterilizer, is as safe, as effective, and performs as well or better than the legally marketed predicate device cleared under K180672.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.