Terragene Bionova® SCBI (BT96) is a self-contained biological indicator inoculated with viable 106 Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. BT96 has Super Rapid readout at 30 minutes at 60°C. Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio) incubate at 60°C and 37°C and read the Terragene Bionova® SCBI for fluorescent results at the times prescribed in the User Manuals. Terragene Chemdye® (CD42) is a chemical process indicator intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. The chemical indicator changes from red to yellow for CD42 to indicate that the conditions of the cycle have been met. Terragene Cintape® CT40 is a chemical process indicator tape intended for monitoring the efficacy of Vaporized Hydrogen Peroxide sterilization processes. The indicating tape changes from purple to green when exposure to vaporized hydrogen peroxide. The self-contained biological indicator and chemical processing indicators are intended for use with the STERLINK™ FPS-15s Plus when operating in chamber mode.
Device Story
System monitors vaporized hydrogen peroxide sterilization efficacy. Components: self-contained biological indicator (SCBI) with Geobacillus stearothermophilus spores; reader incubators; chemical indicator strips/tape. SCBI uses fluorescence for rapid readout (30 min). Chemical indicators provide visual color change (red-to-yellow or purple-to-green) upon exposure. Used in clinical settings; operated by healthcare personnel. Output: fluorescent signal (via reader) or color change (chemical indicators) confirms sterilization cycle conditions met. Benefits: ensures instrument sterility, reducing infection risk.
Clinical Evidence
Bench testing only. Performance validated per ANSI/AAMI/ISO 11138-1:2017 (biological indicators) and ANSI/AAMI/ISO 11140-1:2014 (chemical indicators). Tests included viable spore population, resistance characteristics, reduced incubation time, recovery protocols, visual readout stability, shelf life, and biocompatibility. All tests passed.
Technological Characteristics
SCBI: polypropylene tube, spore carrier, glass ampoule with culture medium. Chemical indicators: plastic strips/tape with indicator ink. Sensing: fluorescence (biological) and colorimetric (chemical). Incubators: 60°C/37°C heating. Standards: ANSI/AAMI/ISO 11138-1:2017, ANSI/AAMI/ISO 11140-1:2014.
Indications for Use
Indicated for monitoring efficacy of vaporized hydrogen peroxide sterilization processes in STERLINK™ FPS-15s Plus (chamber mode). Includes biological indicator (BT96) with Geobacillus stearothermophilus spores, reader incubators, and chemical indicators (CD42, CT40).
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
October 22, 2021
Plasmapp Co., Ltd. % Candace Cederman Consultant CardioMed Device Consultants LLC 1783 Forest Drive. Suite 254 Annapolis, Maryland 21401
Re: K212193
Trade/Device Name: Terragene Bionova SCBI (BT96), Terragene Bionova Reader Incubators (IC10/20FRLCD, Mini-bio), Terragene Chemdye (CD42), Terragene Cintape (CT40) Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC, JOJ Dated: September 20, 2021 Received: September 21, 2021
Dear Candace Cederman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212193
#### Device Name
Terragene Bionova® SCBI (BT96), Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio), Terragene Chemdye® (CD42), Terragene Cintape® (CT40)
#### Indications for Use (Describe)
Terragene Bionova® SCBI (BT96) is a self-contained biological indicator inoculated with viable 10% Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. BT96 has Super Rapid readout at 30 minutes at 60°C.
Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio) incubate at 60°C and 37°C and read the Terragene Bionova® SCBI for fluorescent results at the times prescribed in the User Manuals.
Terragene Chemdye® (CD42) is a chemical process indicator intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. The chemical indicator changes from red to yellow for CD42 to indicate that the conditions of the cycle have been met.
Terragene Cintape® CT40 is a chemical process indicator tape intended for monitoring the efficacy of Vaporized Hydrogen Peroxide sterilization processes. The indicating tape changes from purple to green when exposure to vaporized hydrogen peroxide.
The self-contained biological indicator and chemical processing indicators are intended for use with the STERLINK™ FPS-15s Plus when operating in chamber mode.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span></span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 1. General Information
| Applicant/Submitter: | Plasmapp Co., Ltd.<br>BVC-111, 125, Gwahak-ro,<br>Yuseong-gu, Daejeon, 34141,<br>Rep. of Korea (South Korea)<br>Tel: +82 (0)42 716 2115 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Candace Cederman |
| Address: | CardioMed Device Consultants LLC<br>3168 Braverton Street<br>Suite 200<br>Edgewater, MD 21037<br>Tel: +1 410 674 2060 |
| Preparation Date: | October 7, 2021 |
# 2. Device Name and Code
| Device Trade Name: | Terragene Bionova® SCBI (BT96); Terragene Bionova® Reader<br>Incubators (IC10/20FRLCD, Mini-Bio); Terragene Chemdye®<br>(CD42); Terragene Cintape® (CT40) |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Self-Contained Biological Indicator, Self-Contained Biological<br>Indicator Incubator, Chemical Indicator |
| Classification Name: | Sterilization Process Indicator |
| Product Code: | FRC, JOJ |
| Regulation Number: | 21 CFR 880.2800 |
| Classification: | Class II |
| Review Panel: | General Hospital |
{4}------------------------------------------------
## 3. Device Description
Terragene® Bionova® BT96 Fluorescence Super Rapid Readout Biological Indicators are singleuse Self-Contained Biological Indicators (SCBIs) that consist of a polypropylene tube, a spore carrier, and a glass ampoule with a culture medium, enclosed with a colored cap. Each tube contains a population of Geobacillus stearothermophilus ATCC 7953 spores inoculated on a spore carrier, a plastic cap with holes and a barrier permeable to Plasma or Vaporized Hydrogen Peroxide. Each BT96 has a Process Indicator on label that changes from purple to green when exposed to hydrogen peroxide. The Bionova® BT96 Biological Indicators have been designed for monitoring of Vaporized Hydrogen Peroxide sterilization processes when used in conjunction with Bionova® IC10/20FR, IC10/20FRLCD or MiniBio Auto-Readers Incubators.
Chemdye® CD42 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of plastic strips printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes within loads, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items.
Cintape® CT40 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of a roll of self-adhesive plastic tape printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items. The adhesive component of the tape allows the adhesion to different types of packaging and wraps, such as cloth, paper and plastic.
## 4. Indications for Use / Intended Use
The revised indications for use are as follows:
Terragene Bionova® SCBI (BT96) is a self-contained biological indicator inoculated with viable 106 Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. BT96 has Super Rapid readout at 30 minutes at 60°C.
Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio) incubate at 60°C and 37°C and read the Terragene Bionova® SCBI for fluorescent results at the times prescribed in the User Manuals.
{5}------------------------------------------------
Terragene Chemdye® (CD42) is a chemical process indicator intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. The chemical indicator changes from red to yellow to indicate that the conditions of the cycle have been met.
Terragene Cintape® CT40 is a chemical process indicator tape intended for monitoring the efficacy of Vaporized Hydrogen Peroxide sterilization processes. The indicating tape changes from purple to green when exposure to vaporized hydrogen peroxide.
The self-contained biological indicator and chemical processing indicators are intended for use with the STERLINK™ FPS-15s Plus when operating in chamber mode.
# 5. Technical Characteristics in Comparison to Predicate Devices
The subject device uses identical technology as the cited predicate devices and has the same intended uses. Based upon the overall performance characteristics for the sterilization process indicators used with the STERLINK" sterilizer, Plasmapp Co., Ltd. believes that there are no significant differences in usage for the underlying technological principles between the subject devices and predicate devices.
| | Subject Device | Predicate Device |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor | Plasmapp Co., Ltd. | Terragene® S.A. |
| Device Name | Terragene Bionova® SCBI (BT96);<br>Terragene Bionova® Reader Incubators<br>(IC10/20FRLCD, Mini-Bio);<br>Terragene Chemdye® (CD42);<br>Terragene Cintape® (CT40) | Same |
| 510(k) Number | K212193 | K191021 |
| Manufacturer | Terragene® S.A. | Same |
| Device Classification<br>Name | Sterilization Process Indicator | Same |
| Classification Product<br>Code | FRC (biological indicators)<br>JOJ (chemical indicators) | Same |
| Regulation Number | 21 CFR 880.2800 | Same |
| | Subject Device | Predicate Device |
| Sponsor | Plasmapp Co., Ltd. | Terragene® S.A. |
| Indications for Use | Terragene Bionova® SCBI (BT96) is a self-<br>contained biological indicator inoculated<br>with viable 106 Geobacillus<br>stearothermophilus bacterial spores and is<br>intended for monitoring the efficacy of<br>vaporized hydrogen peroxide sterilization<br>processes. BT96 has Super Rapid readout at<br>30 minutes at 60°C.<br>Terragene Bionova® Reader Incubators<br>(IC10/20FRLCD, Mini-Bio) incubate at<br>60°C and 37°C and read the Terragene<br>Bionova® SCBI for fluorescent results at the<br>times prescribed in the User Manuals.<br>Terragene Chemdye® (CD42) is a chemical<br>process indicator intended for monitoring<br>the efficacy of vaporized hydrogen peroxide<br>sterilization processes. The chemical<br>indicator changes from red to yellow for<br>CD42 to indicate that the conditions of the<br>cycle have been met.<br>Terragene Cintape® CT40 is a chemical<br>process indicator tape intended for monitoring<br> | Excerpted from K191021:<br>Terragene Bionova® SCBI (BT96) is a self-<br>contained biological indicator inoculated with<br>viable 106 Geobacillus stearothermophilus<br>bacterial spores and is intended for monitoring<br>the efficacy of vaporized hydrogen peroxide<br>sterilization processes. BT96 has Super Rapid<br>readout at 30 minutes at 60°C.<br>Terragene<br>Bionova® Reader<br>Incubators<br>(IC10/20FRLCD, Mini-Bio) incubate at 60°C<br>and 37°C and read the Terragene Bionova®<br>SCBI for fluorescent results at the times<br>prescribed in the User Manuals.<br>Terragene Chemdye® (CD40, CD42) is a<br>chemical process indicator intended for<br>monitoring the efficacy of vaporized hydrogen<br>peroxide sterilization processes. The chemical<br>indicator changes from purple to green for<br>CD40 and red to yellow for CD42 to indicate<br>that the conditions of the cycle have been met.<br>Terragene Cintape® CT40 is a chemical<br>process indicator tape intended for monitoring<br>the efficacy of Vaporized Hydrogen Peroxide<br>sterilization processes. The indicating tape<br>changes from purple to green when exposure<br>to vaporized hydrogen peroxide. |
| Intended Use: Cycles<br>Terragene Bionova® SCBI (BT96) | Vaporized Hydrogen Peroxide<br>Models<br>Cycle<br>BT96, CD42,<br>STERLINK™ FPS-15s<br>CT40<br>Plus - Chamber mode | Vaporized Hydrogen Peroxide<br>Models<br>Cycles<br>BT96, CD42,<br>STERRAD 100S (54<br>CT40<br>minutes)<br>STERRAD Standard and<br>Express Cycles<br>V-Pro Max and Sterizone<br>VP4 |
| Type of Biological<br>Indicator | Self-Contained | Same |
| Organism Spore,<br>Species, Strain | Geobacillus stearothermophilus ATCC 7953<br>spores inoculated on a strip (spore carrier) | Same |
| Viable Spore<br>Population | ≥ 106 | Same |
| Resistance<br>characteristics | D-value<br>Survival time/Kill window | Same |
| | Subject Device | Predicate Device |
| Sponsor | Plasmapp Co., Ltd. | Terragene® S.A. |
| Intended Sterilization<br>Cycles | STERLINK™ FPS-15s Plus - Chamber mode<br>(overall cycle: 36 minutes) | STERRAD® 100S (54-minute Short cycle, 72-<br>minute Long cycle), STERRAD® 100NX<br>(Standard Cycle - 47 minutes, Express Cycle<br>- 24 minutes), V-Pro® maX (Non Lumen Cycle<br>- 28 minutes, Flexible Cycle - 35 minutes,<br>Lumen Cycle - 55 minutes), Sterizone® VP4<br>(Cycle 1) |
| Shelf Life | 2 years | Same |
| | Terragene Chemdye® (CD42), Terragene Chemdye® (CT40) | |
| Intended Sterilization<br>Cycles | STERLINK™ FPS-15s Plus - Chamber mode<br>(overall cycle: 36 minutes) | STERRAD® 100S sterilization cycles:<br>• 54-minute Short cycle for most surgical<br>instruments<br>• 72-minute Long cycle for flexible<br>endoscopes and instruments with longer<br>lumen<br>STERRAD® 100NX® sterilization cycles:<br>• 24-minute express cycle for fast turnaround<br>endoscopes and many other instruments<br>• 47-minute standard cycle for most surgical<br>instruments<br>AMSCO® V-PRO maX cycles:<br>• 28-minute non-lumen cycle for most<br>surgical instruments<br>• 35-minute flexible cycle for single or dual<br>channel flexible endoscopes and any non-<br>lumen devices<br>• 55-minute lumen cycle for stainless steel<br>lumen instruments with a single, dual or<br>triple channel.<br>Sterizone VP4 cycle. |
| Device design | Strip, Tape | Same |
| Color Change upon<br>Exposure to H2O2 | CD42: red to yellow<br>CT40: purple to green | Same |
| Recommended Storage<br>Conditions | Dry place, away from sunlight, at temperature<br>between 10-30°C, 30-80% relative humidity.<br>Do not wet. Do not store close to sterilizing<br>agents. | Same |
| Shelf Life | 5 years…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.