The SimpleSENSE Platform is intended for use at home, or at a healthcare facility, under the direction of a licensed medical professional, to record, display and store the following physiological data: a) 2 leads of Electrocardiogram; b) Respiration rate measured through thoracic impedance; c) Heart Sounds; d) Activity including posture; and e) other validated data sources. The SimpleSENSE Platform is intended for use when the licensed medical professional decides to evaluate the physiologic signals of adult patients as an aid to diagnosis and treatment. The SimpleSENSE Platform is intended to be used by patients at rest and not performing any activities or movements. ECG recordings are indicated for the manual assessment of cardiac rhythm disturbances. The SimpleSENSE Platform does not is not intended for active patient monitoring (real-time). The SimpleSENSE Platform is not intended for use as life supporting equipment on high-risk patients such as critical care patients. The SimpleSENSE Platform is not intended for use in the presence of a pacemaker.
Device Story
SimpleSENSE Platform captures 2-lead ECG, respiration (thoracic impedance), heart sounds (MEMS microphone), and activity/posture (accelerometer) via sensors embedded in a wearable textile garment. Signal Acquisition Unit (SAU) stores data on microSD and transmits wirelessly via Bluetooth to a mobile app (iOS/Android). Data is forwarded to a web server for physician review. Used at home or healthcare facilities; operated by patients at rest. Physicians access data via web portal to evaluate cardiac rhythm and physiologic signals as an aid to diagnosis. Benefits include unobtrusive, efficient data capture for clinical assessment.
Clinical Evidence
No clinical data. Bench testing only. Verification focused on software/firmware updates for Android OS compatibility and web server integration.
Technological Characteristics
Wearable textile garment with integrated nanosensor electrodes (ECG, thoracic impedance) and MEMS microphone. SAU uses 24-bit resolution for ECG/impedance/heart sounds and 16-bit for accelerometer. Connectivity via Bluetooth, cellular, and Wi-Fi. Power: rechargeable Li-Ion battery. Data storage: removable 16GB microSD. Software: iOS/Android mobile apps and web server interface.
Indications for Use
Indicated for adult patients requiring evaluation of physiologic signals (ECG, respiration, heart sounds, activity/posture) as an aid to diagnosis and treatment. Intended for use at rest. Contraindicated for high-risk/critical care patients, life-support applications, and patients with pacemakers.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 22, 2021
Nanowear Inc. Venkatesh Varadan CEO 53 Boerum Place, Suite 3F Brooklyn, New York 11201
Re: K212160
Trade/Device Name: SimpleSENSE Platform Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter and Receiver Regulatory Class: Class II Product Code: DXH, DQD, DSB, BZQ, DPS Dated: August 19, 2021 Received: August 23, 2021
Dear Venkatesh Varadan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212160
Device Name SimpleSENSE Platform
#### Indications for Use (Describe)
The SimpleSENSE Platform is intended for use at home, or at a healthcare facility, under the direction of a licensed medical professional, to record, display and store the following physiological data: a) 2 leads of Electrocardiogram; b) Respiration rate measured through thoracic impedance; c) Heart Sounds; d) Activity including posture; and e) other validated data sources. The SimpleSENSE Platform is intended for use when the licensed medical professional decides to evaluate the physiologic signals of adult patients as an aid to diagnosis and treatment. The SimpleSENSE Platform is intended to be used by patients at rest and not performing any activities or movements. ECG recordings are indicated for the manual assessment of cardiac rhythm disturbances. The SimpleSENSE Platform does not is not intended for active patient monitoring (real-time). The SimpleSENSE Platform is not intended for use as life supporting equipment on high-risk patients such as critical care patients. The SimpleSENSE Platform is not intended for use in the presence of a pacemaker.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="margin-right:5px;"> </span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="margin-right:5px;"> </span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY PER 21 CFR §807.92
## 1. SUBMITTER'S INFORMATION
Nanowear Inc.
Contact: Venkatesh Varadan
53 Boerum Pl, Suite 3F
Brooklyn, NY 11201
United States
Phone: 718-637-4815
Date: June 30, 2021
# 2. CLASSIFICATION
The classification for the new device is shown in the table below:
| 21 CFR<br>Reference | Product<br>Code | Class | Trade Name | Classification Name |
|---------------------|-----------------|-------|-------------------------|--------------------------------------------------------------|
| §870.2920 | DXH | 2 | | Transmitters and Receivers,<br>Electrocardiograph, Telephone |
| §870.2770 | DSB | 2 | SimpleSENSE<br>Platform | Plethysmograph, Impedance |
| §868.2375 | BZQ | 2 | | Monitor, Breathing Frequency |
| §870.2340 | DPS | 2 | | Electrocardiograph |
| §870.1875 | DQD | 2 | | Stethoscope, Electronic |
### 3. PREDICATE DEVICE
The predicate device is:
- K201669 SimpleSENSE cleared on November 6, 2020, from Nanowear Inc. ●
# 4. INDICATIONS FOR USE
The SimpleSENSE Platform is intended for use at home, or at a healthcare facility, under the direction of a licensed medical professional, to record, display and store the following physiological data: a) 2 leads of Electrocardiogram; b) Respiration rate measured through thoracic impedance; c) Heart Sounds; d) Activity including posture; and e) other validated data sources. The SimpleSENSE Platform is intended for use when the licensed medical professional decides to evaluate the physiologic signals of adult patients as an aid to diagnosis and treatment. The SimpleSENSE Platform is intended to be used by patients at rest and not performing any activities or movements. ECG recordings are indicated for the manual assessment of cardiac rhythm disturbances. The SimpleSENSE Platform does not produce alarms and is not intended for active patient monitoring (real-time). The SimpleSENSE Platform is not intended for use as life supporting equipment on high-risk
{4}------------------------------------------------
patients such as critical care patients. The SimpleSENSE Platform is not intended for use in the presence of a pacemaker.
### 5. DEVICE DESCRIPTION
The Nanowear SimpleSENSE Platform is the next generation diagnostic monitoring technology that captures electrocardiographic (ECG) signals, respiration rate though thoracic impedance, heart sounds, activity including posture and movement with sensors embedded on a wearable textile garment. The signals are stored and wirelessly transmitted to a smartphone, which is then forwarded to a medical professional for review.
The garment is designed to be unobtrusive to everyday activity and provide an easier and more efficient means of capturing ECG data from patients. The device consists of a combination of:
- The SimpleSENSE Garment: an integrated network of nanosensor electrodes for measuring ECG and respiratory rate from thoracic impedance. A MEMS microphone for measuring heart sounds.
- The SimpleSENSE Signal Acquisition Unit (SAU): data acquisition, storage, and transmission to a phone running iOS or Android. An accelerometer to measure activity.
- The SimpleSENSE Mobile Application: mobile application to start/stop a recording, logging symptoms, and data transmission.
- . SimpleSENSE Web Server: allows initiation of a test, storage, and review of prescribed test data by a medical professional.
Image /page/4/Picture/11 description: The image shows a diagram of a medical monitoring system. The diagram shows a person wearing a chest strap, which is connected to a smartphone app. The smartphone app then sends data to a doctor who is monitoring the patient's health. The arrows indicate the flow of information between the different components of the system.
FIGURE 1: SIMPLESENSE PLATFORM CLINICAL MODEL OF DATA COLLECTION, TRANSMISSION AND REVIEW
#### 6. INTENDED USE
The Nanowear SimpleSENSE Platform is intended for use by licensed medical professionals or patients to record, store, and transfer electrocardiogram (ECG), respiratory rate, heart sounds, and activity including posture and movement.
{5}------------------------------------------------
### 7. CHARACTERISTICS
The table below delineates the similarities and differences between the cleared SimpleSENSE System, and the proposed SimpleSENSE Platform. The table includes comparisons of the system features of the devices. Any differences between the two designs have been marked with bold font which are further discussed below the table.
| | Nanowear Inc.<br>SimpleSENSE System | Nanowear Inc.<br>SimpleSENSE Platform | |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| 510(k) Number | K201669 | To be assigned | |
| Indication<br>Statement | The SimpleSENSE System is intended for<br>use at home, or at a healthcare facility,<br>under the direction of a licensed medical<br>professional, to record, display and store<br>the following physiological data: a) 2<br>leads of Electrocardiogram; b)<br>Respiration rate measured through<br>thoracic impedance; c) Heart Sounds;<br>and d) Activity including posture.<br>The device is<br>intended for use when the<br>clinician decides to evaluate<br>the physiologic signals of adult patients<br>as an aid to diagnosis and treatment.<br>The SimpleSENSE System is intended to<br>be used by patients at rest and not<br>performing any activities or movements.<br>ECG recordings are indicated for the<br>manual assessment of cardiac rhythm<br>disturbances. The device<br>does not produce alarms and is not<br>intended for active patient monitoring<br>(real-time). The device is<br>not intended for use as life supporting<br>equipment on high-risk patients such as<br>critical care patients. The<br>device is not intended for use in the<br>presence of a pacemaker. | The SimpleSENSE Platform is intended<br>for use at home, or at a healthcare<br>facility, under the direction of a licensed<br>medical professional, to record, display<br>and store the following physiological<br>data: a) 2 leads of Electrocardiogram; b)<br>Respiration rate measured through<br>thoracic impedance; c) Heart Sounds; d)<br>Activity including posture; and e) other<br>validated data sources. The<br>SimpleSENSE Platform is intended for<br>use when the licensed medical<br>professional decides to evaluate the<br>physiologic signals of adult patients as<br>an aid to diagnosis and treatment. The<br>SimpleSENSE Platform is intended to be<br>used by patients at rest and not<br>performing any activities or movements.<br>ECG recordings are indicated for the<br>manual assessment of cardiac rhythm<br>disturbances. The SimpleSENSE Platform<br>does not produce alarms and is not<br>intended for active patient monitoring<br>(real-time). The SimpleSENSE Platform is<br>not intended for use as life supporting<br>equipment on high-risk patients such as<br>critical care patients. The SimpleSENSE<br>Platform is not intended for use in the<br>presence of a pacemaker. | |
| Product Code | DXH, DSB, BZQ, DPS, DQD | Same | |
| Acquired Data | • Electrocardiogram (EKG/ECG)<br>• Respiration Rate derived from thoracic<br>impedance<br>• Heart Sounds | Same | |
| Sensor<br>Technology | • Electrocardiogram (EKG/ECG): Textile-<br>based Nanosensors<br>• Respiration Rate: Textile-based<br>Nanosensors measure thoracic<br>impedance and respiration is derived<br>from thoracic impedance<br>• Heart Sound: Microelectromechanical<br>(MEMs) microphone | Same | |
| Signal<br>acquisition<br>method | • Electrocardiogram (EKG/ECG):<br>Standard Bipolar lead instrumentation<br>amplifier and Sigma-Delta Analog to<br>Digital Converter.<br>• Respiration rate: Thoracic Impedance<br>is measured using four-point probe<br>using low amplitude current applied to<br>the body and impedance measured<br>from voltage drop derived from<br>thoracic impedance.<br>• Heart Sound: solid state<br>Microelectromechanical (MEMs)<br>microphone embedded in garment<br>and located near Apex of heart. | Same | |
| Display Type | No on-device display. | Same | |
| Display<br>Requirement | User provided display hardware for a<br>healthcare professional to view the<br>recorded data. A general-purpose<br>PC/Laptop/Desktop may be used. | Same | |
| Power Source | Internally powered using Li-Ion<br>rechargeable battery | Same | |
| Internal<br>Memory/data | Removeable MicroSD card | Same | |
| Communication<br>Interface | Wireless transceiver using Bluetooth | Same | |
| Access to<br>recorded data | Data is transferred to the iPhone,<br>which is then shared with the healthcare<br>professional via email.<br><br>For redundancy, encrypted data is also<br>stored in the removeable storage<br>medium. | Data is transferred to the smart phone,<br>which is then shared with the healthcare<br>professional via the SimpleSENSE Web<br>Server.<br><br>For redundancy, encrypted data is also<br>stored in the removeable storage<br>medium. | |
| Characteristic | Category | New Device<br>SimpleSENSE Platform | Predicate<br>SimpleSENSE System |
| Frequency<br>response | ECG | 0.05 Hz - 65 Hz | Same |
| | Thoracic Impedance | 0.05 Hz - 65 Hz | Same |
| | Heart Sound | 0.05 Hz - 236 Hz | Same |
| | Accelerometer | 0-25 Hz | Same |
| Channels | ECG | 2 channels | Same |
| | Thoracic Impedance | 2 channels | Same |
| | Heart Sound | 1 channel | Same |
| | Accelerometer | 1 channel | Same |
| Resolution | ECG | 24-bit | Same |
| | Thoracic Impedance | 24-bit | Same |
| | Heart Sound | 24-bit | Same |
| | Accelerometer | 16-bit | Same |
| Sampling Rate | ECG | 200 Hz | Same |
| | Thoracic Impedance | 200 Hz…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.