← Product Code [LZA](/productcode/LZA) · K212148

# Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs (K212148)

_Ansell Healthcare Products, LLC · LZA · Mar 10, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212148

## Device Facts

- **Applicant:** Ansell Healthcare Products, LLC
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Mar 10, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: 45.4 minutes and Thiotepa: 23.5 minutes. Warning: Do not use with Carmustine or Thiotepa.

## Device Story

Non-sterile, powder-free, polychloroprene examination glove; ambidextrous; beaded cuff; finger-textured; blue. Designed for medical examination to prevent cross-contamination. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Used in clinical settings by healthcare professionals. Provides barrier protection; however, specific drugs (Carmustine, Thiotepa) show low breakthrough times, necessitating caution. Protects patient and examiner from contamination.

## Clinical Evidence

No clinical data. Bench testing only. Device evaluated for physical dimensions, tensile strength, elongation, freedom from holes, powder residue, and biocompatibility (ISO 10993 series). Chemotherapy permeation testing performed per ASTM D6978-05.

## Technological Characteristics

Material: Polychloroprene. Coating: Polyacrylic polymer. Design: Non-sterile, powder-free, ambidextrous, beaded cuff. Standards: ASTM D6977-19 (dimensions, physical properties), ASTM D5151-06 (holes), ASTM D6124-06 (powder), ASTM D6978-05 (chemotherapy). Biocompatibility: ISO 10993-5, 10, 11.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Microflex® Nitrile Patient Examination Gloves Blue Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate ([K210401](/device/K210401.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

March 10, 2022

Ansell Healthcare Products LLC Donald Cronk Associate Director 2301 Robb Drive Reno, Nevada 89523

Re: K212148

Trade/Device Name: Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: February 28, 2022 Received: March 1, 2022

Dear Donald Cronk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

encies. You must comply with all the Act's

Page 2

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K212148

Device Name

Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs

#### Indications for Use (Describe)

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: 45.4 minutes and Thiotepa: 23.5 minutes. Warning: Do not use with Carmustine or Thiotepa.

| Tested Chemotherapy drug & Concentration | Average Breakthrough Detection Time (Minutes) |
|------------------------------------------|-----------------------------------------------|
| Carmustine - 3.3 mg/ml                   | 45.4                                          |
| Cisplatin - 1.0 mg/ml                    | >240                                          |
| Cyclophosphamide - 20.0 mg/ml            | >240                                          |
| Dacarbazine - 10.0 mg/ml                 | >240                                          |
| Doxorubicin HCl - 2.0 mg/ml              | >240                                          |
| Etoposide - 20.0 mg/ml                   | >240                                          |
| Fluorouracil - 50.0 mg/ml                | >240                                          |
| Methotrexate - 25.0 mg/ml                | >240                                          |
| Paclitaxel - 6.0 mg/ml                   | >240                                          |
| Thiotepa - 10.0 mg/ml                    | 23.5                                          |
| Vincristine Sulfate - 1.0 mg/ml          | >240                                          |

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510K Summary

## 510(k) Number: K212148

### Submitter:

Ansell Healthcare Products LLC. 2301 Robb Drive Reno, NV 89523

## Contact Person(s):

Don Cronk Associate Director, Regulatory Affairs Phone: (775) 470-7106 Email: don.cronk@ansell.com

Jacob Ramirez Senior Coordinator, Regulatory Affairs Phone: (775) 624-8118 Email: jacob.ramirez@ansell.com

### Date Prepared:

2/23/2022

## Name of the Device:

| Trade Names:               | Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue)<br>Tested for Use with Chemotherapy Drugs |  |
|----------------------------|------------------------------------------------------------------------------------------------------------|--|
| Common Name:               | Patient Examination Glove                                                                                  |  |
| Classification Name:       | Patient Examination Glove                                                                                  |  |
| Classification Regulation: | 21 CFR 880.6250                                                                                            |  |
| Device Class:              | I                                                                                                          |  |
| Product Code:              | LZA, LZC                                                                                                   |  |
| Classification Panel:      | Non-powdered patient examination glove                                                                     |  |

### Legally Marketed Predicate Device:

| Company:       | Ansell Healthcare Products LLC                                                                                            |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Trade Names:   | Microflex® Nitrile Patient Examination Gloves Blue Colored Tested for<br>Use with Chemotherapy Drugs and Fentanyl Citrate |
| 510(k) Number: | K210401                                                                                                                   |
| Device Class:  | Class I                                                                                                                   |
| Product Code:  | LZA, LZC, QDO                                                                                                             |
| Device Name:   | Patient Examination Glove (21 CFR 880.6250)                                                                               |

{4}------------------------------------------------

## Device Description:

Microflex 73-743 Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs is a non-sterile, single use only, disposable, powder free examination gloves.

Characteristics:

- . Ambidextrous with beaded cuff and straight fingers
- Finger-textured .
- . Blue colored
- Five (5) sizes – extra-small, small, medium, large, and extra-large
- . Tested against chemotherapy drugs

High levels of ozone will degrade rubber material of the glove; therefore, the glove should be protected from ozone in particular.

The glove is designed to meet the specifications of ASTM D6977-19, Standard Specification for Polychloroprene Examination Gloves for Medical Application.

## Indications for Use Statement:

Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: Carmustine: 45.4 minutes and Thiotepa: 23.5 minutes. Warning: Do not use with Carmustine or Thiotepa.

| TEST CHEMOTHERAPY DRUG AND<br>CONCENTRATION | AVERAGE MINIMUM BREAKTHROUGH DETECTION<br>TIME<br>(Sample 1,2,3) (Minutes) |
|---------------------------------------------|----------------------------------------------------------------------------|
| Carmustine (3.3 mg/ml)                      | 45.4                                                                       |
| Cisplatin (1.0 mg/ml)                       | >240                                                                       |
| Cyclophosphamide (20.0 mg/ml)               | >240                                                                       |
| Dacarbazine (10.0 mg/ml)                    | >240                                                                       |
| Doxorubicin HCl (2.0 mg/ml)                 | >240                                                                       |
| Etoposide (20.0 mg/ml)                      | >240                                                                       |
| Fluorouracil (50.0 mg/ml)                   | >240                                                                       |
| Methotrexate (25.0 mg/ml)                   | >240                                                                       |
| Paclitaxel (6.0 mg/ml)                      | >240                                                                       |
| Thiotepa (10.0 mg/ml)                       | 23.5                                                                       |
| Vincristine Sulfate (1.0 mg/ml)             | >240                                                                       |

## Tested chemotherapy drugs are as follows:

{5}------------------------------------------------

# Technological Characteristics:

|                                                                                                     | Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comparison |           |
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------|
| Trade name                                                                                          | Microflex® Nitrile Patient<br>Examination Gloves Blue Colored<br>Tested for Use with Chemotherapy<br>Drugs and Fentanyl Citrate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Non-Sterile Powder-Free<br>Polychloroprene Examination<br>Glove (Blue) Tested for Use with<br>Chemotherapy Drugs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Different  |           |
| Product Owner                                                                                       | Ansell Healthcare                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Ansell Healthcare                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same       |           |
| Product Code                                                                                        | LZA, LZC, QDO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | LZA, LZC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Different  |           |
| Regulation Number                                                                                   | 21 CFR 880.6250                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 21 CFR 880.6250                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same       |           |
| Regulatory Class                                                                                    | I                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | I                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same       |           |
| Regulation Name                                                                                     | Patient Examination Glove                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Patient Examination Glove                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same       |           |
| Indications for use                                                                                 | A powder-free patient<br>examination glove is a disposable<br>device intended for medical<br>purposes that is worn on the<br>examiner's hand to prevent<br>contamination between patient<br>and examiner                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | A powder-free patient<br>examination glove is a disposable<br>device intended for medical<br>purposes that is worn on the<br>examiner's hand to prevent<br>contamination between patient<br>and examiner                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same       |           |
| Material<br>Composition                                                                             | Synthetic nitrile rubber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Polychloroprene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Different  |           |
| Coating                                                                                             | Polyacrylic polymer inner coating<br>to aid donning                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Polyacrylic polymer inner coating<br>to aid donning                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same       |           |
| Design                                                                                              | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same       |           |
|                                                                                                     | Single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same       |           |
|                                                                                                     | Powder-free                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Powder-free                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same       |           |
|                                                                                                     | Ambidextrous                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Ambidextrous                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same       |           |
|                                                                                                     | Beaded cuff                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Beaded cuff                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same       |           |
| Color                                                                                               | Blue, Green, and Black                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Blue                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Different  |           |
| Performance                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |            |           |
| a. Dimensions                                                                                       | Meets ASTM D6319-10 requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Meets ASTM D6977-19 requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similar    |           |
| b. Physical Properties                                                                              | Meets ASTM D6319-10 requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Meets ASTM D6977-19 requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similar    |           |
| c. Freedom from holes                                                                               | Meets ASTM D6319-10 requirements of GI, AQL 2.5                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Meets ASTM D6977-19 requirements of GI, AQL 2.5                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Similar    |           |
| d. Powder Residual                                                                                  | Meets ASTM D6319-10 requirements; Not more than 2.0mg/glove                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Meets ASTM D6977-19 requirements; Not more than 2.0mg/glove                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar    |           |
| e. Sterility                                                                                        | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same       |           |
| Ozone                                                                                               | Avoid Ozone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Avoid Ozone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same       |           |
| Biocompatibility                                                                                    | Passes Primary Skin Irritation Test and Dermal Sensitization Test and Acute Systemic Toxicity Test and <i>In Vitro</i> Cytotoxicity Test                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Passes Primary Skin Irritation Test and Dermal Sensitization Test and Acute Systemic Toxicity Test and <i>In Vitro</i> Cytotoxicity Test                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same       |           |
| Chemotherapy Claim                                                                                  | Microflex® Nitrile Patient Examination Gloves Blue Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate<br><br>A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drug has an extremely low permeation time: Carmustine: 47.9 minutes. Warning; Do not use with Carmustine.<br><br>Tested chemotherapy drugs/concentration & Average Minimum Breakthrough Detection Time (Minutes) are as follows: | Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs<br><br>A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: Carmustine: 45.4 minutes and Thiotepa: 23.5 minutes. Warning; Do not use with Carmustine or Thiotepa.<br><br>Tested chemotherapy drugs/concentration & Average Minimum Breakthrough Detection Time (Minutes) are as follows: | Different  |           |
|                                                                                                     | Carmustine (3.3 mg/ml) $47.9$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Carmustine (3.3 mg/ml) $45.4$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Different  |           |
|                                                                                                     | Cyclophosphamide (20.0 mg/ml) $>240$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Cyclophosphamide (20.0 mg/ml) $>240$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same       |           |
|                                                                                                     | Doxorubicin HCI (2.0 mg/ml) $>240$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Doxorubicin HCI (2.0 mg/ml) $>240$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same       |           |
|                                                                                                     | Etoposide (20.0 mg/ml) $>240$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Etoposide (20.0 mg/ml) $>240$                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same       |           |
| Fluorouracil (50.0 mg/ml)                                                                           | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Fluorouracil (50.0 mg/ml)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | >240       | Same      |
| Methotrexate (25.0 mg/ml)                                                                           | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Methotrexate (25.0 mg/ml)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | >240       | Same      |
| Paclitaxel (6.0 mg/ml)                                                                              | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Paclitaxel (6.0 mg/ml)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | >240       | Same      |
| Thiotepa (10.0 mg/ml)                                                                               | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Thiotepa (10.0 mg/ml)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 23.5       | Different |
| Vincristine Sulfate (1.0 mg/ml)                                                                     | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Vincristine Sulfate (1.0 mg/ml)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | >240       | Same      |
|                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Cisplatin (3.3 mg/ml)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | >240       | Different |
|                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Dacarbazine (10.0 mg/ml)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | >240       | Different |
| Tested hazardous drug & Average<br>Minimum Breakthrough Detection Time<br>(Minutes) are as follows: |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |            | Different |
|                                                                                                     | Fentanyl Citrate Injection<br>(100 mcg/2 ml)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |            |           |

{6}------------------------------------------------

{7}------------------------------------------------

The subject device meets the applicable requirements for patient examination gloves regarding dimensions and sizes, physical properties, freedom from holes, and powder residues as found in the following standards: ASTM D6977, ASTM D5151 and ASTM D6124.

# Non-Clinical Testing:

Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs has the following technological characteristics as compared to ASTM or equivalent standards:

| Test Methodology          | Purpose                                                                                        | Acceptance Criteria | Results                                                                |
|---------------------------|------------------------------------------------------------------------------------------------|---------------------|------------------------------------------------------------------------|
| Physical Characteristics: |                                                                                                |                     |                                                                        |
| Dimensions:               | Standard Specification<br>for Polychloroprene<br>Examination Gloves for<br>Medical Application | ASTM D6977-19       | Meets ASTM D6977-19<br>requirements for length, width and<br>thickness |
| Length                    |                                                                                                | Minimum 230mm       | Minimum 240mm                                                          |
| Palm width (mm)           |                                                                                                |                     |                                                                        |
| Size — XS                 |                                                                                                | $70 \pm 10$         | $75 \pm 5$                                                             |
| Size — S…

---

**Source:** [https://fda-staging.innolitics.com/device/K212148](https://fda-staging.innolitics.com/device/K212148)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).