Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs
K212148 · Ansell Healthcare Products, LLC · LZA · Mar 10, 2022 · General Hospital
Device Facts
Record ID
K212148
Device Name
Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs
Applicant
Ansell Healthcare Products, LLC
Product Code
LZA · General Hospital
Decision Date
Mar 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: 45.4 minutes and Thiotepa: 23.5 minutes. Warning: Do not use with Carmustine or Thiotepa.
Device Story
Non-sterile, powder-free, polychloroprene examination glove; ambidextrous; beaded cuff; finger-textured; blue. Designed for medical examination to prevent cross-contamination. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Used in clinical settings by healthcare professionals. Provides barrier protection; however, specific drugs (Carmustine, Thiotepa) show low breakthrough times, necessitating caution. Protects patient and examiner from contamination.
Clinical Evidence
No clinical data. Bench testing only. Device evaluated for physical dimensions, tensile strength, elongation, freedom from holes, powder residue, and biocompatibility (ISO 10993 series). Chemotherapy permeation testing performed per ASTM D6978-05.
Indicated for use as a disposable patient examination glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for use with specific chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Microflex® Nitrile Patient Examination Gloves Blue Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (K210401)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
March 10, 2022
Ansell Healthcare Products LLC Donald Cronk Associate Director 2301 Robb Drive Reno, Nevada 89523
Re: K212148
Trade/Device Name: Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: February 28, 2022 Received: March 1, 2022
Dear Donald Cronk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
encies. You must comply with all the Act's
Page 2
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K212148
Device Name
Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs
#### Indications for Use (Describe)
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: 45.4 minutes and Thiotepa: 23.5 minutes. Warning: Do not use with Carmustine or Thiotepa.
| Tested Chemotherapy drug & Concentration | Average Breakthrough Detection Time (Minutes) |
|------------------------------------------|-----------------------------------------------|
| Carmustine - 3.3 mg/ml | 45.4 |
| Cisplatin - 1.0 mg/ml | >240 |
| Cyclophosphamide - 20.0 mg/ml | >240 |
| Dacarbazine - 10.0 mg/ml | >240 |
| Doxorubicin HCl - 2.0 mg/ml | >240 |
| Etoposide - 20.0 mg/ml | >240 |
| Fluorouracil - 50.0 mg/ml | >240 |
| Methotrexate - 25.0 mg/ml | >240 |
| Paclitaxel - 6.0 mg/ml | >240 |
| Thiotepa - 10.0 mg/ml | 23.5 |
| Vincristine Sulfate - 1.0 mg/ml | >240 |
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510K Summary
## 510(k) Number: K212148
### Submitter:
Ansell Healthcare Products LLC. 2301 Robb Drive Reno, NV 89523
## Contact Person(s):
Don Cronk Associate Director, Regulatory Affairs Phone: (775) 470-7106 Email: don.cronk@ansell.com
Jacob Ramirez Senior Coordinator, Regulatory Affairs Phone: (775) 624-8118 Email: jacob.ramirez@ansell.com
### Date Prepared:
2/23/2022
## Name of the Device:
| Trade Names: | Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue)<br>Tested for Use with Chemotherapy Drugs | |
|----------------------------|------------------------------------------------------------------------------------------------------------|--|
| Common Name: | Patient Examination Glove | |
| Classification Name: | Patient Examination Glove | |
| Classification Regulation: | 21 CFR 880.6250 | |
| Device Class: | I | |
| Product Code: | LZA, LZC | |
| Classification Panel: | Non-powdered patient examination glove | |
### Legally Marketed Predicate Device:
| Company: | Ansell Healthcare Products LLC |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Trade Names: | Microflex® Nitrile Patient Examination Gloves Blue Colored Tested for<br>Use with Chemotherapy Drugs and Fentanyl Citrate |
| 510(k) Number: | K210401 |
| Device Class: | Class I |
| Product Code: | LZA, LZC, QDO |
| Device Name: | Patient Examination Glove (21 CFR 880.6250) |
{4}------------------------------------------------
## Device Description:
Microflex 73-743 Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs is a non-sterile, single use only, disposable, powder free examination gloves.
Characteristics:
- . Ambidextrous with beaded cuff and straight fingers
- Finger-textured .
- . Blue colored
- Five (5) sizes – extra-small, small, medium, large, and extra-large
- . Tested against chemotherapy drugs
High levels of ozone will degrade rubber material of the glove; therefore, the glove should be protected from ozone in particular.
The glove is designed to meet the specifications of ASTM D6977-19, Standard Specification for Polychloroprene Examination Gloves for Medical Application.
## Indications for Use Statement:
Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: Carmustine: 45.4 minutes and Thiotepa: 23.5 minutes. Warning: Do not use with Carmustine or Thiotepa.
| TEST CHEMOTHERAPY DRUG AND<br>CONCENTRATION | AVERAGE MINIMUM BREAKTHROUGH DETECTION<br>TIME<br>(Sample 1,2,3) (Minutes) |
|---------------------------------------------|----------------------------------------------------------------------------|
| Carmustine (3.3 mg/ml) | 45.4 |
| Cisplatin (1.0 mg/ml) | >240 |
| Cyclophosphamide (20.0 mg/ml) | >240 |
| Dacarbazine (10.0 mg/ml) | >240 |
| Doxorubicin HCl (2.0 mg/ml) | >240 |
| Etoposide (20.0 mg/ml) | >240 |
| Fluorouracil (50.0 mg/ml) | >240 |
| Methotrexate (25.0 mg/ml) | >240 |
| Paclitaxel (6.0 mg/ml) | >240 |
| Thiotepa (10.0 mg/ml) | 23.5 |
| Vincristine Sulfate (1.0 mg/ml) | >240 |
## Tested chemotherapy drugs are as follows:
{5}------------------------------------------------
# Technological Characteristics:
| | Predicate | Subject Device | Comparison | |
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------|
| Trade name | Microflex® Nitrile Patient<br>Examination Gloves Blue Colored<br>Tested for Use with Chemotherapy<br>Drugs and Fentanyl Citrate | Non-Sterile Powder-Free<br>Polychloroprene Examination<br>Glove (Blue) Tested for Use with<br>Chemotherapy Drugs | Different | |
| Product Owner | Ansell Healthcare | Ansell Healthcare | Same | |
| Product Code | LZA, LZC, QDO | LZA, LZC | Different | |
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Same | |
| Regulatory Class | I | I | Same | |
| Regulation Name | Patient Examination Glove | Patient Examination Glove | Same | |
| Indications for use | A powder-free patient<br>examination glove is a disposable<br>device intended for medical<br>purposes that is worn on the<br>examiner's hand to prevent<br>contamination between patient<br>and examiner | A powder-free patient<br>examination glove is a disposable<br>device intended for medical<br>purposes that is worn on the<br>examiner's hand to prevent<br>contamination between patient<br>and examiner | Same | |
| Material<br>Composition | Synthetic nitrile rubber | Polychloroprene | Different | |
| Coating | Polyacrylic polymer inner coating<br>to aid donning | Polyacrylic polymer inner coating<br>to aid donning | Same | |
| Design | Non-sterile | Non-sterile | Same | |
| | Single use | Single use | Same | |
| | Powder-free | Powder-free | Same | |
| | Ambidextrous | Ambidextrous | Same | |
| | Beaded cuff | Beaded cuff | Same | |
| Color | Blue, Green, and Black | Blue | Different | |
| Performance | | | | |
| a. Dimensions | Meets ASTM D6319-10 requirements | Meets ASTM D6977-19 requirements | Similar | |
| b. Physical Properties | Meets ASTM D6319-10 requirements | Meets ASTM D6977-19 requirements | Similar | |
| c. Freedom from holes | Meets ASTM D6319-10 requirements of GI, AQL 2.5 | Meets ASTM D6977-19 requirements of GI, AQL 2.5 | Similar | |
| d. Powder Residual | Meets ASTM D6319-10 requirements; Not more than 2.0mg/glove | Meets ASTM D6977-19 requirements; Not more than 2.0mg/glove | Similar | |
| e. Sterility | Non-sterile | Non-sterile | Same | |
| Ozone | Avoid Ozone | Avoid Ozone | Same | |
| Biocompatibility | Passes Primary Skin Irritation Test and Dermal Sensitization Test and Acute Systemic Toxicity Test and <i>In Vitro</i> Cytotoxicity Test | Passes Primary Skin Irritation Test and Dermal Sensitization Test and Acute Systemic Toxicity Test and <i>In Vitro</i> Cytotoxicity Test | Same | |
| Chemotherapy Claim | Microflex® Nitrile Patient Examination Gloves Blue Colored Tested for Use with Chemotherapy Drugs and Fentanyl Citrate<br><br>A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drug has an extremely low permeation time: Carmustine: 47.9 minutes. Warning; Do not use with Carmustine.<br><br>Tested chemotherapy drugs/concentration & Average Minimum Breakthrough Detection Time (Minutes) are as follows: | Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs<br><br>A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: Carmustine: 45.4 minutes and Thiotepa: 23.5 minutes. Warning; Do not use with Carmustine or Thiotepa.<br><br>Tested chemotherapy drugs/concentration & Average Minimum Breakthrough Detection Time (Minutes) are as follows: | Different | |
| | Carmustine (3.3 mg/ml) $47.9$ | Carmustine (3.3 mg/ml) $45.4$ | Different | |
| | Cyclophosphamide (20.0 mg/ml) $>240$ | Cyclophosphamide (20.0 mg/ml) $>240$ | Same | |
| | Doxorubicin HCI (2.0 mg/ml) $>240$ | Doxorubicin HCI (2.0 mg/ml) $>240$ | Same | |
| | Etoposide (20.0 mg/ml) $>240$ | Etoposide (20.0 mg/ml) $>240$ | Same | |
| Fluorouracil (50.0 mg/ml) | >240 | Fluorouracil (50.0 mg/ml) | >240 | Same |
| Methotrexate (25.0 mg/ml) | >240 | Methotrexate (25.0 mg/ml) | >240 | Same |
| Paclitaxel (6.0 mg/ml) | >240 | Paclitaxel (6.0 mg/ml) | >240 | Same |
| Thiotepa (10.0 mg/ml) | >240 | Thiotepa (10.0 mg/ml) | 23.5 | Different |
| Vincristine Sulfate (1.0 mg/ml) | >240 | Vincristine Sulfate (1.0 mg/ml) | >240 | Same |
| | | Cisplatin (3.3 mg/ml) | >240 | Different |
| | | Dacarbazine (10.0 mg/ml) | >240 | Different |
| Tested hazardous drug & Average<br>Minimum Breakthrough Detection Time<br>(Minutes) are as follows: | | | | Different |
| | Fentanyl Citrate Injection<br>(100 mcg/2 ml) | >240 | | |
{6}------------------------------------------------
{7}------------------------------------------------
The subject device meets the applicable requirements for patient examination gloves regarding dimensions and sizes, physical properties, freedom from holes, and powder residues as found in the following standards: ASTM D6977, ASTM D5151 and ASTM D6124.
# Non-Clinical Testing:
Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs has the following technological characteristics as compared to ASTM or equivalent standards:
| Test Methodology | Purpose | Acceptance Criteria | Results |
|---------------------------|------------------------------------------------------------------------------------------------|---------------------|------------------------------------------------------------------------|
| Physical Characteristics: | | | |
| Dimensions: | Standard Specification<br>for Polychloroprene<br>Examination Gloves for<br>Medical Application | ASTM D6977-19 | Meets ASTM D6977-19<br>requirements for length, width and<br>thickness |
| Length | | Minimum 230mm | Minimum 240mm |
| Palm width (mm) | | | |
| Size — XS | | $70 \pm 10$ | $75 \pm 5$ |
| Size — S…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.