K212107 · Intelis Instruments , Ltd. · GEI · Nov 10, 2021 · General, Plastic Surgery
Device Facts
Record ID
K212107
Device Name
Reverso
Applicant
Intelis Instruments , Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Reverso device is a non-invasive device intended for use in dermatological procedures requiring ablation and resurfacing of the skin.
Device Story
Tabletop computerized system; console with power supply, controller, LCD touch screen; handheld applicator with detachable single-use 80 or 160 pin tips. Delivers fractional RF energy (0.46 MHz) to skin; induces sub-necrotic heating of papillary and reticular dermis; triggers collagen remodeling. Used in hospital or clinic settings by physicians/dermatologists. Operator adjusts energy settings, tip pattern, and pulse profile via console interface. Output affects tissue ablation and coagulation patterns; benefits patient through skin resurfacing. Device is non-sterile; tips sterilized by user via autoclave.
Clinical Evidence
No clinical data. Bench testing and ex-vivo human tissue study performed. Ex-vivo study on human abdomen tissue compared Reverso to Venus Viva MD; histology confirmed equivalent ablation and coagulation patterns. Safety validated via electrical/mechanical (IEC 60601-1), EMC (IEC 60601-1-2), and usability (IEC 60601-1-6) testing.
Technological Characteristics
Fractional RF energy delivery; 0.46 MHz frequency; 10W max output. Materials: PC Makrolon 2458 (handle/tip body), Stainless steel 302 (pins). Dimensions: 31.8cm x 31.5cm x 31.5cm; Weight: 6kg. Connectivity: Standalone console. Sterilization: User-performed autoclaving for tips. Standards: ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-1-6.
Indications for Use
Indicated for adults requiring dermatological procedures for skin ablation and resurfacing.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 10, 2021
Intelis Instruments Ltd. % Amit Goren Regulatory Consultant A. Stein - Regulatory Affairs Consulting Ltd. 18 Hata'as Str., Suite 21 Kfar Saba, 4442518 Israel
Re: K212107
Trade/Device Name: Reverso Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: October 10, 2021 Received: October 13, 2021
Dear Amit Goren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K212107
Device Name Reverso
Indications for Use (Describe)
The Reverso is a non-invasive device intended for use in dermatological procedures requiring ablation and resurfacing of the skin.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
K212107
# 510(K) SUMMARY REVERSO DEVICE
# 510(k) Number K212107
#### Applicant Name:
| Company Name: | Intelis Instruments Ltd. |
|---------------|----------------------------------------|
| Address: | HaOman 12<br>Hadera, 3850172<br>Israel |
| Tel: | +972-528701672 |
| E-mail: | amit@asteinrac.com |
# Contact Person:
| Official Correspondent: | Amit Goren | |
|-------------------------|------------------------------------------------------------|--------------------|
| Company Name: | A. Stein - Regulatory Affairs Consulting Ltd. | |
| Address: | 18 Hata'as Str.<br>Kfar Saba 4442518<br>Israel | |
| | Tel: | +972-9-7670002 |
| | Fax: | +972-9-7668534 |
| | E-mail: | amit@asteinrac.com |
| Date Prepared: | November 10, 2021 | |
| Trade Name: | Reverso | |
| Classification Name: | CFR Classification section 878.4400;<br>(Product code GEI) | |
Classification: Class II Medical Device
#### Predicate Device:
The Reverso device is substantially equivalent to the following predicate device;
| Manufacturer | Device | 510(k) No. |
|---------------------------|---------------|------------|
| Venus Concept<br>USA Inc. | Venus Viva MD | K201164 |
{4}------------------------------------------------
K212107
### Device Description:
The Reverso device is a tabletop computerized system comprised a console (main unit) and Applicator with detachable single-use tips. The system delivers RF energy in a fractional manner to the skin of a treatment area for ablation and resurfacing of the skin. The Reverso device provides individual adjustments of energy settings, tip pattern, pulse profile to achieve appropriate efficiency and safety for each eligible patient.
The device console includes a power supply unit, controller card and LCD screen with touch panel.
The Reverso device Applicator is designed for use in dermatological procedures requiring ablation and resurfacing of the skin. The applicator is hand- held and ergonomically designed for the treatment requiring ablation and resurfacing of the skin. The applicator is connected to the system via a cable. When not in use, the applicator can be placed within the applicator holder positioned on the system console.
The Reverso Applicator is compatible with two types of fractional RF tip heads; 80 pin tip head and 160 pin tip head.
Following are the Reverso device specifications:
| RF Max Output Power: | 10 Watt |
|--------------------------------|-----------------------------------------------------------------|
| RF Output Frequency: | 640[KHz] |
| Dimension: | 31.8cm W x 31.5cm D x 31.5cm H<br>[12.4" W x 12.3" D x 12.3" H] |
| Weight: | 6 Kg (13.2 lbs.) |
| Main Line Frequency (nominal): | 50-60 Hz |
| Input Voltage (nominal): | 100-240 VAC |
#### Intended Use/Indication for Use:
The Reverso device is a non-invasive device intended for use in dermatologic procedures requiring ablation and resurfacing of the skin.
{5}------------------------------------------------
#### Performance Standards:
The Reverso device has been tested and complies with the following voluntary recognized standards:
- ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 . and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- . IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- IEC 60601-2-2 Edition 6.0 2017-03 Medical electrical equipment Part ● 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.
- . IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
# Non-Clinical (Bench) Performance Data:
Performance bench tests were conducted to evaluate and compare the Reverso device RF performance specifications to the specific design requirements and to the RF performance specifications of the predicate device. The results of the bench test demonstrate that the Reverso device complies with the design requirements comprise similar RF output specifications as the predicate devices and therefore, is substantially equivalent to the predicate devices.
The safety and effectiveness of the Reverso device were further evaluated in an ex-vivo tissue testing for dermatological procedures requiring skin ablation and skin resurfacing. The study was conducted on human abdomen tissues following an abdominoplasty surgery and included a single treatment utilizing the Reverso device in its final, finished version. The predicate device; the Venus Viva MD device (Subject of K201164) was used for comparison utilized in
{6}------------------------------------------------
similar operation condition on similar ex-vivo tissues. Untreated tissue was used as a control. Treatment was followed by biopsy sampling of slices trimmed along the pin's penetration path and collection immediately stained by H&E staining to visualize the tissue ablation and coagulation patterns. The ex-vivo study results demonstrated that the Reverso device is safe for use and effective in achieving the specified indications of dermatological skin ablation and skin resurfacing. Furthermore, the histology results of the subject device tissue samples were compared with the histology results of the predicate device tissue samples. The comparison demonstrated an equivalency in ablation and coagulation characteristics.
# Animal Performance Data / Histology Data:
Not Applicable
# Clinical Performance Data:
Not Applicable
# Substantial Equivalence:
The Reverso device is substantially equivalent to the Venus Viva MD Device (manufactured by Venus Concept USA Inc. and the subject of K201164). The subject device and predicate device share the same underlying technology of fractional RF, same RF output specifications and same operational principals. A comparison table is provided below comparing the intended use and basic technological characteristics of the Reverso device to the intended use and basic technological characteristics of the predicate device.
{7}------------------------------------------------
| Technological<br>Characteristic | Reverso Device<br>Intelis Instruments<br>Ltd.<br>(Subject Device) | Venus Viva MD TM<br>Venus Concept USA Inc.<br>K201164<br>(Predicate Device) |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code,<br>Class | GEI<br>Class II | GEI<br>Class II |
| Regulation<br>Number | 21 CFR 878.4400 | 21 CFR 878.4400 |
| Indications for<br>Use | The Reverso is a non-<br>invasive device intended<br>for use in dermatological<br>procedures requiring<br>ablation and resurfacing<br>of the skin. | The Venus Viva MD is a non-<br>invasive device intended to be<br>used by aesthetic- related<br>physicians or dermatologists.<br><br>When used with the Viva MD<br>applicator, the Venus Viva MD<br>Device is intended for use in<br>dermatological procedures<br>requiring ablation and resurfacing<br>of the skin. |
| Anatomical Sites | Body parts requiring<br>treatment as specified in<br>the indication for use | idem |
| Target<br>Population | Adults requiring<br>treatment as specified in<br>the indication for use | idem |
| Environment<br>Used | Hospital or Clinic setting | idem |
| Energy Used /<br>Delivered | RF energy | idem |
| Design: | Fractional RF: Use of RF<br>energy delivered through a<br>matrix of multiple pin<br>electrodes allocated on the<br>applicator tip | idem |
| - Mechanism of<br>Action | Treatment is based on<br>fractional sub- necrotic<br>heating of papillary and<br>reticular dermis triggering<br>slow collagen remodeling. | idem |
| - Components | Tabletop Console<br>including :<br>Display panel Main CPU Power supply<br>Reverso applicator. The<br>applicator comprises the<br>applicator handle, trigger, | Tabletop Console including:<br>Display panel Main CPU Power supply RF Generator<br>Two optional applicators:<br>Viva MD applicator. The<br>applicator comprises the<br>applicator handle, trigger, |
| Technological<br>Characteristic | Reverso Device<br>Intelis Instruments<br>Ltd. | Venus Viva MDTM<br>Venus Concept USA Inc.<br>K201164 |
| | (Subject Device)<br>detachable electrode tip<br>(160 pin or 80 pin). | (Predicate Device)<br>detachable electrode tip<br>(160 pin or 80 pin).<br>● Diamondpolar Applicator<br>(irrelevant for this<br>510(K) substantial<br>equivalent discussion) |
| RF Performance<br>specifications | RF Frequency: 0.46 MHz •<br>Energy per pin:<br>62 mJ/pin (160 pin tip)<br>124 mJ/pin (80 pin tip) | idem |
| - System<br>Dimensions | 31.8cm W x 31.5cm D x<br>31.5cm H<br>[12.4" W x 12.3" D x 12.3"<br>H] | 38cm W x 40cm D x 40cm H<br>[14.8" W x 15.6" D x 15.6" H] |
| - Weight<br>Platform weight<br>Applicator weight | 6 Kg (13.2 lbs.)<br>Applicator - 0.3 Kg (0.66<br>lbs.) | 8 Kg (17.6 lbs.)<br>Applicator - N/A |
| Cable Dimensions: | 170 cm | idem |
| Materials | Applicator Handle: PC<br>Makrolon 2458<br>Tip outer body: PC<br>Makrolon 2458<br>RF pin electrodes:<br>Stainless steel 302. | idem |
| Standards Met | AAMI/ANSI ES 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2<br>IEC 60601-1-6 | idem |
| Biocompatibility | All materials are<br>biocompatible | idem |
| Compatibility with<br>Environment and<br>Other Devices | Reverso is compliant with<br>the IEC 60601-1-2 (EMC<br>Safety) standard | idem |
| Sterility | Provided non-sterile.<br>The detachable, single use<br>tip head is to be sterilized<br>by the user close to<br>treatment onset, using<br>autoclaving sterilization<br>technique. | idem |
| Electrical Safety | Power Requirements: | idem |
| Technological<br>Characteristic | Reverso Device<br>Intelis Instruments<br>Ltd.<br>(Subject Device) | Venus Viva MDTM<br>Venus Concept USA Inc.<br>K201164<br>(Predicate Device) |
| | The Reverso is compliant<br>with the IEC 60601-1<br>standard. | |
| Mechanical Safety | The Reverso is compliant<br>with the IEC 60601-1<br>standard. | idem |
| Chemical Safety | Not Applicable | Not Applicable |
| Thermal Safety | The Reverso is compliant<br>with the IEC 60601-1<br>standard. | idem |
| Radiation Safety | The Reverso is compliant<br>with the IEC 60601-1-2<br>(EMC Safety) standard. | Idem |
{8}------------------------------------------------
{9}------------------------------------------------
The indications for use and technological characteristics of the Reverso device are substantially equivalent to the indications for use and technological characteristics of the predicate, the Venus Viva MD device (and the subject of K201164).
The design of and the components in the Reverso device, including the console (with the power supply unit, controller and display user interface) and the applicator (with cable, connector to console, handle and tip head) are identical to the design and components found in the predicate device, with similar to identical dimensions and weight.
The RF performance specifications (RF frequency, RF energy per pin, RF energy level) of the Reverso device were shown to be identical to those of the predicate device as demonstrated in the bench performance testing. Furthermore, the Reverso device and the predicate device demonstrated similar thermal effect over ex-vivo human tissues following a single treatment performed under similar RF performance specifications and similar operation conditions.
All of the identified device related hazards are of low risk and are mostly related with transient symptoms of erythema and edema which are resolved 24-72 hours post treatment.
{10}------------------------------------------------
Therefore, it can be deduced that the Reverso device is as safe and effective as the predicate device, and that no new safety or effectiveness issues are raised in regard to the subject device functionality.
The safety features and compliance with safety standards in the subject device are similar to the safety features and compliance with safety standards found in the predicate device.
Patient contact materials are the same between the subject device and its predicate device. Any minor differences in the technological characteristics do not raise new safety or effectiveness concerns.
Furthermore, the Reverso device underwent performance testing, including software validation testing (provided in Section 16), Electrical and Mechanical Safety testing according to IEC 60601-1, Electromagnetic Compatibility testing according to IEC 60601-1-2 and Usability according to IEC 60601-1-6 (provided in Section 17), bench and ex-vivo testing to evaluate and compare the RF performance specifications and the thermal effect of the device on target tissues to that of the predicate device (provided in Section 18). These performance tests demonstrated that the minor differences in the device design and specifications meet the system requirements and do not raise new safety or effectiveness concerns.
Consequently, it can be concluded that the Reverso device is substantially equivalent to the predicate device, the Venus Viva MD, FDA cleared under 510(k) K201164, and therefore, may be legally marketed in the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.