Portable Folding Electric Wheelchair (Model: DC01)
K212092 · Anhui Jbh Medical Apparatus Co., Ltd. · ITI · May 6, 2022 · Physical Medicine
Device Facts
Record ID
K212092
Device Name
Portable Folding Electric Wheelchair (Model: DC01)
Applicant
Anhui Jbh Medical Apparatus Co., Ltd.
Product Code
ITI · Physical Medicine
Decision Date
May 6, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Indications for Use
The wheelchair (Model: DC01) is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
Portable folding electric wheelchair (Model DC01) provides mobility for disabled/elderly users. Input: user manipulation of joystick controller. Operation: dual brushless DC motors drive rear wheels; intelligent electromagnetic brake system engages automatically upon joystick release. Output: motorized movement (speed/direction). Used in indoor/outdoor environments; operated by patient. Benefits: provides independent mobility. Safety features: electromagnetic braking, flame-retardant materials, carbon fiber frame. Device folds for storage/transport.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence is supported by bench testing, including ISO 7176 series (stability, braking, energy consumption, dimensions, speed, strength, climatic, obstacle climbing, power/control systems, EMC) and IEC 62133-2 (battery safety). Biocompatibility (cytotoxicity, sensitization, irritation) was verified per ISO 10993.
Technological Characteristics
Carbon fiber frame; dual 180W brushless DC motors; 24V 6Ah Li-ion battery; PU solid tires; joystick controller. Dimensions: 800mm turning radius. Connectivity: off-board charger (100-240Vac). Sterilization: N/A. Software: embedded control system. Standards: ISO 7176 series, IEC 60601-1-2, ISO 10993-1, IEC 62133-2.
Indications for Use
Indicated for disabled or elderly persons with limited mobility who require a seated position for indoor and outdoor transportation.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
May 6, 2022
Anhui JBH Medical Apparatus Co., Ltd % Ivy Wang Technical Manager Shanghai SUNGO Management Consulting Co., Ltd. Room 1401. No.1500. Century Ave. Pudong New District, Shanghai 200122 China
Re: K212092
Trade/Device Name: Portable Folding Electric Wheelchair, Model DC01 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: February 5, 2022 Received: February 10, 2022
Dear Ivy Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212092
Device Name
Portable Folding Electric Wheelchair (Model: DC01)
Indications for Use (Describe)
The wheelchair (Model: DC01) is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K212092 510(k) Summary
Name: Anhui JBH Medical Apparatus Co., Ltd
Address: No. 116 qicang Road, Mingguang City, Chuzhou, Anhui, China
Name of contact person: Yang Dandan
Tel: 0086-18936033123
Fax: 0086-550-8106688
Date prepared: 2021/06/30
Submission Correspondent:
Primary contact: Ms. Ivy Wang
Shanghai SUNGO Management Consulting Co., Ltd.
Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China
Tel: +86-21-58817802
Email: haiyu.wang(@sungoglobal.com
Secondary contact: Mr. Raymond Luo
Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China
Tel: +86-21-68828050
Email: fda.sungo@gmail.com
#### I. Device
Device trade name: Portable Folding Electric Wheelchair
Device model: DC01
Classification name: Powered wheelchair
Regulation class: 2
Regulation number: 21CFR 890.3860
Panel: Physical Medicine
Product code: ITI
### II. Predicate device
K113463, PL001 power wheelchair, SUZHOU KID MEDICAL APPLIANCE CO., LTD. page 1 of 9
{4}------------------------------------------------
### III. Device description
The device is Electric Wheelchair, model no. is DC01. An electric wheelchair is a four-wheeled personal mobile device with a complementary chair support system that is powered by two motors. The traveling speed is controlled by the motor, and the traveling direction is controlled by the passenger. This product is a device suitabled people with mobility difficulties and elderly people and it is intended to provide mobility to a disabled or elderly person limited to a seated position. The electric wheelchair can be travelled on flat and obstacle ground surface, and direction and speed of the wheelchair can be controlled by the passenger's hand with the help of the joystick. The device can be used to provide indoor and outdoor mobility at a certain distance but not allowed to be travelled on the road or highway.
The device consists of two parts: the electrical part and the wheelchair main body. The electrical part includes motor, electromagnetic brake system, battery box, controller and battery charger. The main parts of the wheelchair include front wheels, rear wheels, frame, armrest, seat and back cushion.
The device is powered by Li-ion Battery pack (24V 6Ah, 144Wh) with 20 Km (12.5 miles) range, which can be recharged by an off-board battery charger that can be plugged into an AC socket outlet (100-240V, 50/60Hz) when the device is not in use.
The patient can activate the controller handle (joystick) to control the speed and direction of the wheelchair movement. In addition, when the patient releases the joystick will return back to the central position and the wheelchair will be automatically stopped soon due to automatic intelligent electromagnetic brake system starts to work. Once the joystick is activated again move to other position, the wheelchair will be re-energized.
#### IV. Indication for use
The wheelchair (Model: DC01) is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Anhui JBH Medical<br>Apparatus Co., Ltd | SUZHOU KD Medical<br>Appliance Co. Ltd. | / |
| Proprietary<br>name, model | Electric Wheelchair, DC01 | Power wheelchair, PL001 | / |
| 510(k) number | K212092 | K113463 | / |
| Device<br>classification<br>name | Class II | Class II | Same |
| Classification | 21 CFR 890.3860 | 21 CFR 890.3860 | Same |
| Attribute<br>regulations | Subject device | Predicate device | Discussion/Conclusion |
| Product code | ITI | ITI | Same |
| Similarities | | | |
| Intended user | disabled people and<br>elderly people | disabled or elderly person | Same |
| Indication for<br>use | The wheelchair (Model:<br>DC01) is a motor driven,<br>indoor and outdoor<br>transportation vehicle with<br>the intended use to provide<br>mobility to a disabled or<br>elderly person limited to a<br>seated position. | They are motor driven,<br>indoor and outdoor<br>transportation vehicle with<br>the intended use to provide<br>mobility to a disabled or<br>elderly person limited to a<br>seated position. | Same |
| Use condition | indoor and outdoor use | indoor and outdoor use | Same |
| Number of<br>wheels | 4, including two front<br>wheels and two rear<br>wheels | 4, including two pivoting<br>casters and two rear drive<br>wheels | Same |
| Function of<br>wheels | Front wheels: driven<br>wheels suitable for<br>rotation, acceleration,<br>retrograde<br>Rear wheels: driving<br>wheels to control the speed<br>and direction | Two pivoting casters:<br>driven wheels suitable for<br>rotation, acceleration,<br>retrograde<br>two rear drive wheels:<br>driving wheels to control<br>the speed and direction | Same |
| Movement<br>control<br>method | By Joystick control | By Joystick control | Same |
| Driving<br>system | Direct drive on the rear<br>wheels | Direct drive on the rear<br>wheels | Same |
| Brake system | Smart Electromagnetic<br>brake | Intelligent regenerative<br>Electromagnetic brake | Same |
| Electronic<br>controller | Dual Drive Controller for<br>Brushless Motor | Brushless dual-drive<br>rocker controller | Same |
| Motor | Brushless DC motor<br>180W x 24 VDC x 2 pcs | Brushless DC motor; 24<br>VDC; 180 W; 2 pcs | Same |
| Armrest | PU | PU | Same |
| Max speed<br>forward | 3.75 mph (6 km/h) | Up to 6 km/h (3.75 mph),<br>variable | Same |
| Maximum<br>distance of<br>travel on the<br>fully charged<br>battery | 20 km | 20 km | Same |
| Turning<br>Radius | 800 mm | 31.5" (800 mm) | Same |
| Differences | | | |
| Frame design<br>and material | The frame of the<br>wheelchair is type capable<br>of front and rear close. The<br>main part of the frame can<br>be folded for saving space<br>and convenient storage and<br>transportation. The main<br>frame is made of carbon<br>fiber material. | The frame of the<br>wheelchair is type of<br>aluminum frame with front<br>and rear folding structure;<br>Up-and-down turnable<br>handrail, front and rear<br>adjustable armrest. The<br>main frame is made of<br>aluminum alloy material. | both of the wheelchair<br>can be folded in the<br>way of front and rear<br>folding type. |
| Minimum<br>braking<br>distance from | Forward: 0.5 m | Forward: 1.5 m | shorter braking<br>distance in the subject<br>device than the |
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
| maximum speed | | | predicate device, all relevant tests are performed according to standard ISO 7176-3, no safety and performance will be affected. |
| Battery | Lithium-ion, 24V6Ah, 6Ah x 24 VDC | Li-ion, Rechargeable; 24 VDC 20Ah | Same rated voltage, as for difference on power capacity, both batteries are tested according to standard IEC 62133-2, no differences of batteries between both devices will affect the safety and performance of the subject device. |
| Battery charger | Off-board charger<br>Input: 100-240Vac, 50/60Hz, 1.5 A;<br>output: DC 24V, 2A, | Off-board, Automatic Type<br>Input: 110-220 V / 50-60 Hz,<br>Output: 24 Vdc, 2A | More wide range of input voltage in the device which will not cause new safety and effectiveness concerns raised. |
| seat cushion/back cushion | linen cloth filled with PU foam | PU foam covered by nylon fabric cloth | Different surface material in contact with the driver, the biocompatibility evaluation is performed on both devices, such difference will not affect the safety and performance of the subject device |
| Maximum safe operational incline degree | 8 ° | 9 ° | Minor difference on safe operational incline degree will not cause new safety and effectiveness concerns are raised as both the static and dynamic stability under specific inclining degree have been evaluated according to standard ISO 7176 series. |
| Max loading weight | 120 kg (264 lbs) | 114 kg (251 lbs) | Minor difference on loading weight will not cause different performance as all performance tests are performed according to standard ISO 7176 series |
| Front wheel size/type | 7" x 1.75" /PU Solid tire | 6"x 2"/PU Solid tire | different size of driven wheel will not affect safety and performance of the subject device as all related stability tests are performed |
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
| Rear<br>wheel<br>size/type | 8"x1.95" /PU solid tires | 8" x 2.4"/PU Solid tire | according to standard<br>ISO 7176 series.<br>different size of driving<br>wheel will not affect<br>safety and performance<br>of the subject device as<br>all related stability tests<br>are performed<br>according to standard<br>ISO 7176 series. |
| Maximum<br>obstacle<br>climbing | 20 mm | 1.2" (30 mm) | less distance in the<br>obstacle climbing will<br>not impact the safety<br>and effectiveness of the<br>subject device. |
V Comparison of technological characteristics with the predicate device
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
#### Summary of substantial equivalence discussion VI.
The DC01 electric wheelchair complied with the requirements of ISO 7176-1:2014, ISO 7176-2:2017, ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2018, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2008, ISO 7176-13:1989, ISO 7176-14:2008, ISO 7176-15:1996, ISO 7176-16:2012, ISO 7176-21:2009, ISO 7176-22:2014, ISO 7176-25:2013, IEC 60601-1-2: 2014, ISO 10993-1:2018, ISO10993-5:2009, ISO 10993-10:2010.
The intended uses for both devices are totally same. Mainframes of two devices are folded by way of front and rear close, although the locking structure for both devices are different from each other, the folding principle is the same. As for different frame materials used and minor difference on safe operational incline degree for both devices, considering all safety and performance tests are carried out on the subject device with favorable result, such difference will not affect the safety and performance of the subject device. The design principles of the controller and Driving system are the same, and both meet the requirements of the ISO 7176-14:2008. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3:2012. Maximum obstacle climbing and maximum speed are slightly different while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2:2017, ISO 7176-10:2008. The biocompatibility of the subject device is evaluated according to standard ISO 10993-1 and meet the requirements accordingly.
All seat cushion/back cushion and armrest are made of flame retardant material for both devices.
{8}------------------------------------------------
Therefore, the subject device is assured to be under the same safety level.
In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. The differences between the subject device and predicate device do not raise new issues of safety or effectiveness.
#### VII. Summary of non-clinical testing (Performance testing-bench)
The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination.
- . Risk Analysis developed in accordance with ISO 14971:2019.
- . Software evaluation
- ISO 7176-1:2014 Wheelchairs Part 1: Determination of static stability .
- . ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs
- ISO 7176-3:2012 Wheelchairs Part 3: Determination of effectiveness of brakes ●
- ISO 7176-4:2008 Wheelchairs Part 4: Energy consumption of electric wheelchairs and ● scooters for determination of theoretical distance range
- . ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space
- . ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs
- ISO 7176-7:1998 Wheelchairs Part 7: Measurement of seating and wheel dimensions .
- ISO 7176-8:2014 Wheelchairs Part 8: Requirements and test methods for static, impact and . fatigue strength
- . ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs
- ISO 7176-10:2008 Wheelchairs Part 10: Determination of obstacle-climbing ability of . electrically powered wheelchairs
- . ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies
- ISO 7176-13:1989 Wheelchairs Part 13: Determination of coefficient of friction of test .
{9}------------------------------------------------
surfaces.
- . ISO 7176-14:2008 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods
- . ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling.
- . ISO 7176-21:2009 Wheelchairs - Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters
- Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2014. ●
- IEC 62133-2:2017 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
## VIII. Biocompatibility of patient-contacting material
Biocompatibility of patient-contacting material are evaluated in accordance with ISO 10993-1: 2018. All parts of the wheelchair surface in contact with user skin are patient contact parts for the electric wheelchair. The biocompatibility tests are carried out on patient-contact materials, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2010) and irritation (ISO 10993-10:2010). The test results show all materials used are bio-safe. Below is the summary table:
| Component name | Material | Direct contact/<br>indirect contact | Contact<br>body | Contact<br>duration | Evaluation tests |
|---------------------------|---------------------------|-------------------------------------|------------------------|---------------------|---------------------------------------------|
| Seat cushion<br>(surface) | linen cloth | Direct | intact skin<br>surface | <24 Limited | Cytotoxicity<br>Sensitization<br>Irritation |
| | Back cushion<br>(surface) | | | | |
| Armrest | | PU (polyurethane) | | | |
| Frame | carbon fiber | Direct | intact skin<br>surface | <24 Limited | Cytotoxicity<br>Sensitization<br>Irritation |
| | | | | | |
| controller surface | plastic part | Direct | intact skin<br>surface | <24 Limited | Cytotoxicity<br>Sensitization |
| Joystick | plastic part | Direct | intact skin<br>surface | <24 Limited | Irritation |
# IX. Summary of clinical testing
No animal study and clinical studies are available for our device. Clinical testing was not required to
demonstrate the substantial equivalence of the electric wheelchair to its predicate device.
{10}------------------------------------------------
#### X. Conclusions
The differences between DC01 electric wheelchair and its predicate devices do not introduce a new intended use and do not raise new issues of safety and effectiveness. Verification and Validation testing demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended. From the results of nonclinical testing described, it can be concluded that the subject device electric wheelchair DC01 is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.