← Product Code [GXN](/productcode/GXN) · K212058

# Longeviti ClearFit™ OTS Cranial Implants (K212058)

_Longeviti Neuro Solutions, LLC · GXN · Sep 8, 2021 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212058

## Device Facts

- **Applicant:** Longeviti Neuro Solutions, LLC
- **Product Code:** [GXN](/productcode/GXN.md)
- **Decision Date:** Sep 8, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5330
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Longeviti ClearFit OTS Cranial Implants are manufactured to correct and/or restore bony voids and/or defects of the cranium.

## Device Story

Longeviti ClearFit OTS Cranial Implants are fixed-size, preformed prosthetic cranioplasty plates; used to correct or restore cranial bony voids/defects. Implants are manufactured from polymethyl methacrylate (PMMA). Provided sterile; fixated to cranial bone using commercially available fasteners. Used by neurosurgeons in clinical settings. Output is a physical structural implant; provides mechanical protection and restoration of cranial contour. Benefits include restoration of cranial integrity.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including tensile (ASTM D638-19), flexural (ASTM D790-15), and impact (ASTM D4812-10) testing, all of which passed.

## Technological Characteristics

Material: Polymethylmethacrylate (PMMA). Form factor: Fixed-size cranial plates (9cm² - 37cm²), max thickness 5mm. Radiopacity: Radiolucent. Sterilization: Ethylene Oxide. Biocompatibility: Permanent (>30 days) tissue/bone contact. Standards: ASTM D638-19 (tensile), ASTM D790-15 (flexural), ASTM D4812-10 (impact).

## Regulatory Identification

A preformed nonalterable cranioplasty plate is a device that is implanted in a patient to repair a skull defect and is constructed of a material, e.g., stainless steel or vitallium, that cannot be altered or reshaped at the time of surgery without changing the chemical behavior of the material.

## Predicate Devices

- Longeviti ClearFit Cranial Implant ([K202901](/device/K202901.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 8, 2021

Longeviti Neuro Solutions, LLC Heather Hourihan, RAC Director of Regulatory Affairs 303 International Circle Suite 190 Hunt Valley, Maryland 21030

Re: K212058

Trade/Device Name: Longeviti ClearFit OTS Cranial Implants Regulation Number: 21 CFR 882.5330 Regulation Name: Preformed Nonalterable Cranioplasty Plate Regulatory Class: Class II Product Code: GXN, PJN Dated: July 21, 2021 Received: August 5, 2021

## Dear Heather Hourihan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212058

#### Device Name

Longeviti ClearFit OTS Cranial Implants

Indications for Use (Describe)

The Longeviti ClearFit OTS Cranial Implants are manufactured to correct and/or restore bony voids and/or defects of the cranium.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| LONGEVITI NEURO SOLUTIONS                          |                                                                                                                                                                                                                                                                                                                                                                                        |  |  |  |  |
|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| 510(k) Summary                                     |                                                                                                                                                                                                                                                                                                                                                                                        |  |  |  |  |
| Date Prepared:                                     | September 8, 2021                                                                                                                                                                                                                                                                                                                                                                      |  |  |  |  |
| Applicant:                                         | Longeviti Neuro Solutions<br>303 International Circle Suite 190<br>Cockeysville, Maryland 21030<br>(410) 527-1803                                                                                                                                                                                                                                                                      |  |  |  |  |
| Contact Person:                                    | Heather Hourihan<br>hhourihan@longeviti.com<br>Director of Regulatory Affairs, RAC<br>Longeviti Neuro Solutions                                                                                                                                                                                                                                                                        |  |  |  |  |
| SUBJECT DEVICE                                     |                                                                                                                                                                                                                                                                                                                                                                                        |  |  |  |  |
| Trade Name:                                        | Longeviti ClearFit™ OTS Cranial Implants                                                                                                                                                                                                                                                                                                                                               |  |  |  |  |
| Common Name:                                       | PMMA cranial implant                                                                                                                                                                                                                                                                                                                                                                   |  |  |  |  |
| Class:                                             | 2                                                                                                                                                                                                                                                                                                                                                                                      |  |  |  |  |
| Classification:                                    | 882.5330                                                                                                                                                                                                                                                                                                                                                                               |  |  |  |  |
| Procode:                                           | GXN, PJN                                                                                                                                                                                                                                                                                                                                                                               |  |  |  |  |
| Predicate Device:                                  | Longeviti ClearFit™ Cranial Implant K202901                                                                                                                                                                                                                                                                                                                                            |  |  |  |  |
| Device Description:                                | The Longeviti ClearFit™ OTS (off-the shelf) cranial implants are fixed size<br>prosthetic cranioplasty plates intended to correct and/or restore bony<br>voids and/or defects of the cranium.  The implants are manufactured from<br>polymethyl methacrylate materials. The devices are provided sterile and<br>can be fixated to cranial bone using commercially available fasteners. |  |  |  |  |
| Indication for Use:                                | The Longeviti ClearFit™ OTS Cranial Implants are manufactured to<br>correct and/or restore bony voids and/or defects of the cranium.                                                                                                                                                                                                                                                   |  |  |  |  |
| Comparison of<br>Technological<br>Characteristics: | There are no significant technological differences between the Longeviti<br>ClearFit™ Cranial Implant (K202901) and the Longeviti ClearFit™ OTS<br>Cranial Implant. Longeviti manufactures ClearFit™ OTS Cranial Implants<br>using the same raw materials and applying the same manufacturing<br>processes used for their ClearFit™ Cranial Implant.                                   |  |  |  |  |

{4}------------------------------------------------

| Tabular Comparison of Technological Characteristics: |                                                                                                                                                    |                                                                                                                                                                                |           |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Characteristics                                      | K212058<br>Longevity ClearFit TM<br>OTS Cranial Implant<br>(Subject Device)                                                                        | K202901<br>Longevity ClearFit TM<br>Cranial Implant<br>(Predicate Device)                                                                                                      | Comments  |
| Device Classification                                | Class II                                                                                                                                           | Class II                                                                                                                                                                       | Same      |
| Product Code                                         | GXN, PJN                                                                                                                                           | GXN, PJN                                                                                                                                                                       | Same      |
| Indications for Use                                  | The Longeviti ClearFit TM<br>OTS Cranial Implants<br>are manufactured to<br>correct and/or restore<br>bony voids and/or<br>defects of the cranium. | The Longeviti ClearFit TM<br>Cranial Implant is<br>designed and<br>manufactured<br>individually for each<br>patient to correct bony<br>voids and/or defects of<br>the cranium. | Similar   |
| Material                                             | Polymethylmethacrylate<br>(PMMA)                                                                                                                   | Polymethylmethacrylate<br>(PMMA)                                                                                                                                               | Same      |
| Design                                               | Fixed Size Implant                                                                                                                                 | Patient Specific Implant                                                                                                                                                       | Different |
| Dimensions (cm²)                                     | 9cm² - 37 cm²                                                                                                                                      | 25cm² - 411cm²                                                                                                                                                                 | Similar   |
| Maximum thickness                                    | 5mm                                                                                                                                                | 5mm                                                                                                                                                                            | Same      |
| Radiopacity                                          | Radiolucent                                                                                                                                        | Radiolucent                                                                                                                                                                    | Same      |
| Sterilization                                        | Ethylene Oxide                                                                                                                                     | Ethylene Oxide (EO) or<br>Vaporized Hydrogen<br>Peroxide (VHP)                                                                                                                 | Similar   |
| Biocompatibility                                     | Implant device with<br>permanent (>30 d)<br>contact with<br>tissue/bone                                                                            | Implant device with<br>permanent (>30 d)<br>contact with<br>tissue/bone                                                                                                        | Same      |

{5}------------------------------------------------

| Evaluations and<br>Results to Support<br>Substantial<br>Equivalence: | Performance testing completed for the ClearFit™ (K202901) is applicable to the<br>ClearFit™ OTS as Longeviti uses the manufacturing processes and materials<br>validated for the ClearFit™ Cranial Implant to produce the ClearFit™ OTS Cranial<br>Implant. Performance testing completed includes: |                                                                                                                                                  |         |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|---------|
|                                                                      | Method                                                                                                                                                                                                                                                                                              | Standard                                                                                                                                         | Results |
|                                                                      | Tensile                                                                                                                                                                                                                                                                                             | ASTM D638-19: Standard Test Method for Tensile<br>Properties of Plastics                                                                         | Pass    |
|                                                                      | Flexural                                                                                                                                                                                                                                                                                            | ASTM D790-15:<br>Standard Test Methods for Flexural Properties of<br>Unreinforced and Reinforced Plastics and Electrical<br>Insulating Materials | Pass    |
|                                                                      | Impact                                                                                                                                                                                                                                                                                              | ASTM D4812-10:<br>Standard Test Method for Unnotched Cantilever Beam<br>Impact Resistance of Plastics                                            | Pass    |
| Conclusion:                                                          | Based on the similarities of the intended use/indications for use, technological<br>and functional characteristics, and the results of the non-clinical performance<br>testing, the ClearFit OTS cranial implant is substantially equivalent to the legally<br>marked predicate device.             |                                                                                                                                                  |         |

---

**Source:** [https://fda-staging.innolitics.com/device/K212058](https://fda-staging.innolitics.com/device/K212058)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).