Impertech Surgical Face Mask

K212046 · Impertech Industries , Ltd. · FXX · Nov 18, 2021 · General, Plastic Surgery

Device Facts

Record IDK212046
Device NameImpertech Surgical Face Mask
ApplicantImpertech Industries , Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateNov 18, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Impertech Surgical Face Mask are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluid. This is a single use, disposable device, provided non-sterile.

Device Story

Impertech Surgical Face Mask is a flat-pleated, three-layer disposable mask. Construction includes inner and outer spunbond polypropylene layers and a melt-blown fiber filter layer. Features elastic ear loops and a nose piece made of polypropylene-laminated soft annealed carbon steel wire. Used in clinical settings by healthcare personnel and patients to provide a physical barrier against microorganisms, body fluids, and particulates. Operates as a passive barrier; no active components. Provides ASTM F2100 Level 3 protection. Intended to reduce potential exposure to blood and body fluids during infection control practices.

Clinical Evidence

No clinical data. Bench testing only. Device met ASTM F2100 Level 3 requirements: Bacterial Filtration Efficiency ≥98%, Differential Pressure <6 mm H2O/cm², Sub-micron particulate filtration ≥98%, synthetic blood penetration resistance at 160 mm Hg, and Class I flammability. Biocompatibility testing per ISO 10993-1, -5, and -10 confirmed the device is non-cytotoxic, non-sensitizing, and negligibly irritating.

Technological Characteristics

Three-layer flat-pleated mask. Materials: Spunbond polypropylene (inner/outer), melt-blown fiber (filter), nylon/spandex ear loops, polypropylene-laminated soft annealed carbon steel wire nose piece. Dimensions: 175 mm x 95 mm. ASTM F2100 Level 3 compliant. Non-sterile, single-use, disposable. No software or electronic components.

Indications for Use

Indicated for use by patients and healthcare personnel to protect against transfer of microorganisms, body fluids, and particulate material; intended for infection control to reduce exposure to blood and body fluids. Single-use, disposable, non-sterile.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. November 18, 2021 Impertech Industries Ltd. Vaibhav Rajal Official Correspondent for Impertech Industries Ltd. mdi Consultants Inc 55 Northern Blvd. Suite 200 Great Neck, New York 11021 Re: K212046 Trade/Device Name: Impertech Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: October 14, 2021 Received: October 18, 2021 Dear Vaibhav Rajal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212046 Device Name Impertech Surgical Face Mask #### Indications for Use (Describe) The Impertech Surgical Face Mask are intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluid. This is a single use, disposable device, provided non-sterile. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| | | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Impertech Surgical Face Mask ASTM F2100 Level 3 Surgical Face Mask 510(k) Submission ### 510(k) SUMMARY The assigned 510(k) number is: K212046. ### 1. Submitter's Identification: Impertech Industries Ltd. 20 Derech Barkan St. Barkan Industrial Zone, Barkan, IL 4482000 Contact Person: Mohamad Khweis Quality Assurance Manager Position: Phone: +972-3-9367682 mohamadk@supergum.co.il Email: Date Prepared: November 15, 2021 ## 2. Name of the Device: | Trade Name: | Impertech Surgical Face Mask | |-----------------------|------------------------------| | Common Name: | Surgical Mask | | Classification Name: | Surgical Mask | | Regulation Number: | 21 CFR 878.4040 | | Classification Panel: | General Hospital | | Regulatory Class: | Class II | | Product Code: | FXX | ## 3. Information for the 510(k) Cleared Device (Predicate Device): | Trade/Device Name: | Premier Guard USA 3 Layer Ear Loop ASTM Level 3 Surgical<br>Face Mask | |--------------------|-----------------------------------------------------------------------| | 510(k) Number: | K202595 | | Common Name: | Mask, Surgical | | Regulation Number: | 21 CFR 878.4040 | | Regulation Name: | Surgical apparel. | | Regulatory Class: | Class II | | Product Code: | FXX | ### 4. Device Description: The device description for the Impertech Surgical Face Mask is in accordance with the FDA Guidance Document, Surgical Masks -Premarket Notification [510K)] Submissions issued on March 5, 2004. The Impertech Surgical Face Mask is a flat-pleated style mask with elastic ear loops to secure it over the user's mouth and face. The mask consists of three-layers. The inner facing layer is white and is manufactured from spunbond polypropylene (three layers of nonwoven polypropylene). The inner filter material is made of melt blown fiber. The outer facing layer is white and is manufactured from spunbond polypropylene (three layers of nonwoven polypropylene). The mask is a single use, disposable device, provided nonsterile. The mask is in accordance to ASTM F2100 Level 3. The proposed device is not made from natural rubber latex {4}------------------------------------------------ ## 5. Indications for Use: The Impertech Surgical Face Mask are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluid. This is a single use, disposable device, provided non-sterile. ## 6. Comparison to the 510(k) Cleared Devices (Predicate Devices): | Device<br>Characteristic | Proposed Device | Predicate Device | Comparison<br>Analysis | | |-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|--| | Product Name | Impertech Surgical Face Mask | Premier Guard USA 3 Layer<br>Ear<br>Loop ASTM Level 3 Surgical<br>Face Mask | Similar | | | 510(k) Reference | K212046 | K202595 | Similar | | | Product Owner | Impertech Industries Ltd. | Premier Guard USA LLC | Similar | | | Product Code | FXX | FXX | Same | | | Indications for Use | The Impertech Surgical Face<br>Mask are intended to be worn to<br>protect both patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential exposure<br>to blood and body fluids. This<br>is a single use, disposable<br>device(s) provided non-sterile. | The Premier Guard USA 3<br>Layer Ear Loop ASTM Level 3<br>Surgical Face Masks are<br>intended to be worn to<br>protect both patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential exposure<br>to blood and body fluids. This<br>is a single use, disposable<br>device(s) provided non-sterile. | Same | | | Regulation Number | 21 CFR 878.4040 | 21 CFR 878.4040 | Same | | | Mask Style | Flat Pleated, Ear Loops, 3<br>Layers | Flat Pleated, Ear Loops, 3<br>Layers | Same | | | Mask Color | White | Blue | Similar | | | Materials | | | | | | Nose Piece<br>(material) | Polypropylene laminated soft<br>annealed carbon steel wire | Polyethylene laminated soft<br>annealed carbon steel wire | Same | | | Ear Loops (material) | Nylon and Spandex | Nylon and Spandex | Same | | | Outer Facing Layer | Spunbond polypropylene | Spunbond polypropylene | Same | | | Middle Layer | Melt blown fiber | Melt blown fiber | Same | | | Inner Facing Layer | Spunbond polypropylene | Spunbond polypropylene | Same | | | Dimensions - Width | 175 mm | 175 mm | Same | | | Dimensions -<br>Length | 95 mm | 95 mm | Same | | | ASTM F2100 Level | Level 3 | Level 3 | Same | | | Biocompatibility | | | | | | Cytotoxicity | Under the conditions of the<br>study, the device is noncytotoxic | Under the conditions of the<br>study, the device is<br>noncytotoxic | Same | | | Sensitization | Under the conditions of the<br>study, the device is<br>nonsensitizing | Under the conditions of the<br>study, the device is<br>nonsensitizing | Same | | | Irritation | Under the conditions of the<br>study, the device is nonirritating | Under the conditions of the<br>study, the device is<br>nonirritating | Same | | | Prescription vs.<br>OTC | OTC | OTC | Same | | | Sterile vs. Non-<br>Sterile | Non-Sterile | Non-Sterile | Same | | | Disposable vs. Non-<br>Disposable | Disposable | Disposable | Same | | | Single Use vs.<br>Reusable | Single Use | Single Use | Same | | {5}------------------------------------------------ # Discussion of Similarities and Differences between subject device and predicate device: The subject Impertech Surqical Face Mask subject device surgical mask and the Premier Guard USA 3 Layer Ear Loop ASTM Level 3 Surgical Face Mask predicate device are same with all the parameters except for the color of the mask. The color of the subject device mask is without Color (white) whereas the predicate device is blue in color. The change in the mask between the subject device and the predicate device does not raise additional questions for safety and effectiveness. Performance testing including biocompatibility evaluation has been performed on the final finished device. None of the biocompatibility testing or performance testing results demonstrated any safety hazards or any design characteristics that violated the requirements set forth in the Guidance for Industry regarding Premarket Notification Submissions of Surgical Masks # 7. Non-Clinical Performance Testing The Impertech Surgical Face Mask was tested in accordance with the tests recommended in the FDA guidance document, Guidance for Industry and FDA Staff Surgical Masks – Premarket Notification [510(k)] Submission issued March of 2004. Based upon the quidance document the following testing has been performed. {6}------------------------------------------------ - Fluid Resistance per ASTM F1862 ● - Particulate Filtration Efficiency per ASTM F2299 . - Bacterial Filtration Efficiency per ASTM F2101 ● - Differential Pressure (Delta P) per EN 14683 ● - Flammability per 16 CFR 1610 ● | Performance Testing | | | |-----------------------------------------------------------------------------------------|--------------------|--------| | Surgical Mask Characteristics – ASTM F2100 | Level 3<br>Barrier | Result | | Bacterial filtration efficiency % | ≥98 | Pass | | Differential pressure, mm H2O/cm² | <6 | Pass | | Sub-micron particulate filtration efficiency at 0.1 micron % | ≥98 | Pass | | Resistance to penetration by synthetic blood, minimum press<br>in mm Hg for pass result | 160 | Pass | | Flammability | Class I | Pass | ## Performance Testing (Animal) The biocompatibility evaluation for the Impertech Surgical Face Mask was conducted in accordance with ANSI/AAMI/ISO 10993-1:2018 Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing within a Risk Management Process, as recognized by FDA. The Impertech Surgical Face Mask is classified as a surface contacting device. Specific biocompatibility tests were selected under the guidance of ISO 10993-1:2018 Annex A. | Biocompatibility Evaluation | | | | | |-----------------------------|------------------|--------------|-----------------------|------| | | Bilogical Effect | Standard | Results | | | 1 | Cytotoxicity | ISO 10993-5 | Not cytotoxic | Pass | | 2 | Sensitization | ISO 10993-10 | Non sensitizing | Pass | | 3 | Irritation | ISO 10993-10 | Negligibly irritating | Pass | ## 8. Clinical Performance Testing Not Applicable ## 9. Conclusions: The conclusions drawn from the nonclinical tests demonstrate that the devices are as safe, as effective, and perform as well as or better than the legally marketed predicate device. K202595 Premier Guard USA 3 Layer Ear Loop ASTM Level 3 Surgical Face Mask.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...