K212039 · Alcon Laboratories, Inc. · MSS · Aug 23, 2021 · Ophthalmic
Device Facts
Record ID
K212039
Device Name
CLAREON MONARCH IV IOL Delivery System
Applicant
Alcon Laboratories, Inc.
Product Code
MSS · Ophthalmic
Decision Date
Aug 23, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4300
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The CLAREON MONARCH IV IOL Delivery System is for implantation of qualified Alcon foldable IOLs. No unqualified lenses should be used with the CLAREON MONARCH IV IOL Delivery System.
Device Story
System comprises reusable titanium handpiece and single-use sterile cartridge; used for implanting foldable IOLs into eye. Clinician applies ophthalmic viscosurgical device (OVD) to cartridge lumen; loads IOL into cartridge; installs cartridge into handpiece. Handpiece plunger mechanism delivers IOL through cartridge nozzle into capsular bag. Used in surgical setting by ophthalmologists. Provides controlled, reliable IOL placement. Benefits patient by facilitating minimally invasive IOL implantation during cataract surgery.
Clinical Evidence
Bench testing only. Performance data included compatibility/delivery performance (ISO 11979-3), biocompatibility (ISO 10993-5, 10993-18), reprocessing/sterilization validation (ISO 17664, 17665, 14937), and human factors/usability engineering (IEC 62366-1).
Technological Characteristics
Materials: Titanium handpiece, single-use cartridge. Energy: Manual mechanical actuation. Sterilization: Steam sterilization (reusable handpiece). Standards: ISO 11979-3 (delivery performance), ISO 10993 (biocompatibility), ISO 17664/17665/14937 (reprocessing/sterilization), IEC 62366-1 (usability).
Indications for Use
Indicated for implantation of qualified Alcon foldable intraocular lenses (IOLs) into the capsular bag following natural crystalline lens removal.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Food and Drug Administration (FDA) and the Department of Health & Human Services. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The Department of Health & Human Services logo is a symbol with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" surrounding it.
August 23, 2021
Alcon Laboratories, Inc. Jwalitha Shankardas Associate Director, Global Regulatory Affairs 6201 South Freeway Fort Worth, Texas 76134
### Re: K212039
Trade/Device Name: CLAREON MONARCH IV IOL Delivery System Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I, reserved Product Code: MSS Dated: June 29, 2021 Received: June 30, 2021
### Dear Jwalitha Shankardas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Bennett Walker, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
K212039
Device Name
#### CLAREON MONARCH IV IOL Delivery System
Indications for Use (Describe)
The CLAREON MONARCH IV IOL Delivery System is for implantation of qualified Alcon foldable IOLs. No unqualified lenses should be used with the CLAREON MONARCH IV IOL Delivery System.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
This 510(k) summary document has been prepared in accordance with 21 CFR section 807.92.
# I. SUBMITTER
| Company: | Alcon Laboratories, Inc. |
|-------------------------|------------------------------------------------------|
| | 6201 South Freeway<br>Fort Worth, TX 76134-2099, USA |
| Primary Contact Person: | Jwalitha Shankardas, Global Regulatory Affairs |
| Phone: | 817-551-8582 |
| Email: | Jwalitha.shankardas @alcon.com |
| Date Prepared: | June 28, 2021 |
# II. DEVICE
| Trade Names: | CLAREON MONARCH IV IOL Delivery System |
|------------------------|-----------------------------------------------|
| Common Name: | Folders and Injectors, Intraocular lens (IOL) |
| Classification Name: | Folders and Injectors, Intraocular lens (IOL) |
| Device Classification: | Class I- 21CFR 886.4300 |
| Product Code: | MSS |
# III. PREDICATE DEVICE
| Trade Names: | Monarch III IOL Delivery System, Monarch III C cartridge,<br>Monarch III Handpiece (H4) |
|----------------------|-----------------------------------------------------------------------------------------|
| 510K Number | K112977 |
| FDA Clearance date | 03/27/2012 |
| Submitter | Alcon Laboratories, Inc. |
| Common Name: | Folders and Injectors, Intraocular lens (IOL) |
| Classification Name: | Folders and Injectors, Intraocular lens (IOL) |
{4}------------------------------------------------
| Device Classification: | Class I- 21CFR 886.4300 |
|------------------------|-------------------------|
| FDA Panel | Ophthalmic |
| Product Code: | MSS |
### IV. DEVICE DESCRIPTION
The CLAREON MONARCH IV IOL Delivery System consists of two parts. An autoclavable, reusable, titanium handpiece and a sterile, single-use cartridge are used for implanting foldable intraocular lenses into the eye following removal of the natural crystalline lens.
The system provides a controlled means to reliably place Alcon qualified intraocular lenses (IOLs) into the capsular bag when a qualified combination of handpiece, cartridge, Alcon foldable IOL, and ophthalmic viscosurgical device (OVD) is used.
The cartridge is loaded by inserting the IOL into the opening in the back of cartridge after OVD is applied to the inner lumen of the cartridge. The loaded cartridge is installed into the handpiece and the IOL is delivered through the cartridge nozzle.
## V. INDICATIONS FOR USE
The CLAREON MONARCH IV IOL Delivery System is for implantation of qualified Alcon foldable IOLs. No unqualified lenses should be used with the CLAREON MONARCH IV IOL Delivery System.
## VI. INDICATIONS FOR USE COMPARISON
The indications for use of the CLAREON MONARCHIV IOL Delivery System are the same as the predicate device.
## VII. TECHNOLOGICAL COMPARISON
The technological characteristics of the CLAREON MONARCH IV IOL Delivery System are equivalent to those of the predicate device, MONARCH III IOL Delivery System in terms of operating principle, technological characteristics, component materials, manufacturing process, biocompatibility, cleaning, steam sterilization method, performance, and FDA-recognized standards used for performance testing. The CLAREON MONARCHIV IOL Delivery System
{5}------------------------------------------------
has the same intended use, and is compatible with the same Alcon's Monarch cartridges as its predicate.
The CLAREON MONARCH IV IOL Delivery System is primarily differentiated from its predicate by minor design change optimizations to the handpiece housing, knob and plunger for the Clareon IOL design.
# VIII. NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY & CONCLUSIONS
The following performance data were provided in support of the substantial equivalence determination.
- Compatibility and delivery performance with qualified Monarch cartridges, Clareon IOLs . and OVDs in accordance with ISO 11979-3-12.
- Biocompatibility of the patient-contact aspects of the CLAREON MONARCH IV IOL . Delivery System in accordance with the intended use of the device (in accordance with ISO 10993-18-20 and ISO 10993-5-09).
- Reprocessing and sterilization over the claimed use life in accordance with ISO 17664-. 17, ISO 17665-1-06 and ISO 14937-09.
- Human factors and usability for safe and effective use by intended users in accordance with . IEC 62366-1-20 as well as the FDA guidance, Applying Human Factors and Usability Engineering to Medical Devices: Guidance for Industry and Food and Drug Administration Staff; 2016.
The technological characteristics that determine the functionality and performance of the CLAREON MONARCH IV IOL Delivery System is substantially equivalent to the predicate device. The data from the non-clinical tests demonstrate that the safety and effectiveness profile of the CLAREON MONARCH IV IOL Delivery System is equivalent to that of the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.