K212037 · Sanctuary Health Sdn. Bhd. · MSC · Mar 28, 2022 · Obstetrics/Gynecology
Device Facts
Record ID
K212037
Device Name
SancDam Latex Oral Dam
Applicant
Sanctuary Health Sdn. Bhd.
Product Code
MSC · Obstetrics/Gynecology
Decision Date
Mar 28, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
The SancDam™ Latex Oral Dam is used as a barrier when engaging in oral/vaginal sex to help reduce the transmission of bodily fluids, harmful pathogens, and sexually transmitted infections.
Device Story
SancDam™ Latex Oral Dam is a rectangular, non-porous, natural rubber latex film; dimensions 155mm x 254mm x 0.07mm. Device acts as a physical barrier placed over vagina or anus during oral sexual contact to reduce transmission of bodily fluids, pathogens, and sexually transmitted infections. Available in four color/flavor variants (pink/strawberry, green/mint, yellow/vanilla, purple/grape). Intended for over-the-counter, single-use by individuals. Device provides a protective barrier; does not involve electronic components, algorithms, or software.
Clinical Evidence
Bench testing only. Testing included dimensional analysis, tensile strength, elongation at break, tearing force, freedom from holes, and packaging integrity per ISO 29942:2011. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity (ISO 10993-5), irritation (ISO 10993-10), sensitization (ISO 10993-10), and acute systemic toxicity (ISO 10993-11). All results were acceptable.
Technological Characteristics
Material: Natural rubber latex. Form factor: Rectangular film (155mm x 254mm x 0.07mm). Non-sterile, single-use. Testing standards: ISO 29942:2011 (mechanical properties), ISO 10993-1 (biocompatibility). No electronic, software, or energy-based components.
Indications for Use
Indicated for use as a barrier during oral/vaginal and oral/anal sex to reduce transmission of bodily fluids, pathogens, and sexually transmitted infections in individuals engaging in such sexual activity.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
March 28, 2022
Sanctuary Health Sdn. Bhd. Sumethi Selavaraju QAQC cum RA Deputy Manager No. 16, Persiaran Perindustrian Kanthan 1, Kawasan Perindustrian Kanthan Chemor, Perak 31200 Malaysia
Re: K212037
Trade/Device Name: SancDam™ Latex Oral Dam Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: MSC Dated: February 18, 2022 Received: February 24, 2022
Dear Sumethi Selavaraju:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212037
Device Name SancDam™ Latex Oral Dam
Indications for Use (Describe)
The SancDam™ Latex Oral Dam is used as a barrier when engaging in oral/vaginal sex to help reduce the transmission of bodily fluids, harmful pathogens, and sexually transmitted infections.
**Type of Use (Select one or both, as applicable)**
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Sanctuary Health. The word "Sanctuary" is written in red, and the word "HEALTH" is written in blue below it. To the right of the word "HEALTH" are two Chinese characters, also written in blue.
# 510(k) Summary
# a. Submitter and Device Information
| Category | Comments |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Sanctuary Health Sdn. Bhd.<br>No. 16, Persiaran Perindustrian Kanthan 1,<br>Kawasan Perindustrian Kanthan,<br>31200 Chemor, Perak Darul Ridzuan,<br>Malaysia.<br>Tel No: +605 - 201 2800<br>Fax No: +605 - 201 7800 |
| Correspondent Contact<br>Information | Sumethi Selvaraju, QAQC cum RA Deputy Manager<br>Tel: +605-201 2800 ext. 204<br>E-mail: regulatory@sanctuaryhealth.com.my<br>Add: No. 16, Persiaran Perindustrian Kanthan 1,<br>Kawasan Perindustrian Kanthan,<br>31200 Chemor, Perak Darul Ridzuan,<br>Malaysia. |
| Device Proprietary Name | SancDam™ Latex Oral Dam |
| Device Common Name | Latex Dam |
| Device Classification Number | 21 CFR 884.5300 |
| Device Classification Name | Condom |
| Device Class | II |
| Product Code | MSC (Barrier, Std, Oral Sex) |
| Predicate Device | |
|----------------------------------------------|-----------------------------------------------------------------------|
| Predicate Device Manufacturer | Line One Laboratories, Inc. |
| Predicate Device Common<br>Name | Latex Dam |
| Predicate Device Proprietary<br>Name | Line One Trust Dam |
| Predicate Device Premarket<br>Notification # | K091769 |
| Predicate Device Classification | 21 CFR 884.5300 - Condom |
| Predicate Device Class | II |
| Predicate Device Product Code | MSC (Barrier, Std, Oral Sex) |
| Design-related recall | The predicate device has not been subject to a design-related recall. |
#### b. Date Summary Prepared
March 24, 2022
{4}------------------------------------------------
# SancDam™ Oral Latex Dam - Traditional 510(k)
### c. Description of Device
The SancDam™ Latex Oral Dam are rectangular shaped, non-porous, intact polymer films made of natural rubber latex that are placed over the vagina or anus during oral sexual contact. The dam is used as a physical barrier to cover the external female genitalia or the anal area.
The SancDam™ has a smooth surface on both sides. The size of the SancDam™ is 155 mm (± 5 mm) by 254 mm (± 5 mm) by 0.07 mm (± 0.02 mm).
The SancDam™ available in 4 different flavors and colors versions as listed below:
| Color | Flavor |
|--------|------------|
| Pink | Strawberry |
| Green | Mint |
| Yellow | Vanilla |
| Purple | Grape |
#### d. Indications for Use
The SancDam™ Latex Oral Dam is used as a barrier when engaging in oral/vaginal and oral/anal sex to help reduce the transmission of bodily fluids, harmful pathogens, and sexually transmitted infections.
#### e. Comparison of Indications for Use and Technological Characteristics with the Predicate Device
Table 1 below compares the technological characteristics of the subject device and the predicate device.
#### Table 1 - Comparison of Technological Characteristics of the Subject and Predicate Devices
| | SancDam™ Latex Oral<br>Dam<br>(Subject Device) | Line One Trust Dam<br>(Predicate Device) |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | SancDam Latex Oral Dam | Trust Dams |
| Manufacturer | Sanctuary Health Sdn. Bhd. | Line One Laboratories, Inc. |
| 510(k) | K212037 | K091769 |
| Regulatory Class | II | II |
| Regulation number | 21 CFR 884.5300 | 21 CFR 884.5300 |
| Product Code | MSC | MSC |
| Classification Name | Barrier, Std, Oral Sex | Barrier, Std, Oral Sex |
| Indications for Use | The SancDam™ Latex Oral<br>Dam is used as a barrier<br>when engaging in<br>oral/vaginal and oral/anal sex<br>to help reduce the<br>transmission of bodily fluids,<br>harmful pathogens, and<br>sexually transmitted<br>infections. | The Line One Trust Dam is<br>used as a barrier when<br>engaging in oral/vaginal sex<br>and oral/anal sex to help<br>reduce the transmission of<br>bodily fluids, harmful germs,<br>and sexually transmitted<br>diseases. |
| Design & dimensional<br>aspects | Thin, rectangular sheet with<br>smooth surface both sides<br>155 mm x 254 mm,<br>Thickness 0.07 mm | Thin, rectangular sheet with<br>smooth surface both sides<br>154 mm x 250 mm, Thickness<br>0.07 mm |
| Material | Natural rubber latex | Natural rubber latex |
| Lubricated | No | No |
| Color | Yellow, Pink, Purple, Green | Yellow, Red, Green |
| Fragrance/Flavor | Vanilla, Strawberry, Grape,<br>Mint | Banana, Strawberry, Mint |
| Mechanical<br>Properties | Tested in accordance with<br>ISO 29942:2011, including<br>the following tests:<br>• Tensile strength<br>• Elongation at break<br>• Tearing force<br>• Freedom from holes<br>• Tear resistance | Tested in accordance with<br>ASTM D3492:2008 |
| Sterile | No | No |
| Body Location Target<br>Area | Anal, vaginal, oral cavity<br>(mouth) | Anal, vaginal, oral cavity<br>(mouth) |
| Single Use | Yes | Yes |
| Shelf-life | 2 years | 5 years |
| Biocompatibility<br>Tested? | Yes | Yes |
{5}------------------------------------------------
# SancDam™ Oral Latex Dam – Traditional 510(k)
The intended use of the subject device, SancDam™ Latex Oral Dam, and the legally marketed predicate devices, Line One Trust Dam (K091769), are the same. Both the subject device and the predicate device are used as a barrier when engaging in oral/vaginal and oral/anal sex to help reduce the transmission of bodily fluids, harmful pathogens and sexually transmitted infections. The differences between the subject and predicate device include the
{6}------------------------------------------------
colors, flavors, and shelf life. However, these differences in technological characteristics do not raise different questions of safety or effectiveness.
# f. Nonclinical Testing
The following non-clinical testing was completed on the SancDam™ Latex Oral Dam:
- . Per ISO 29942:2011 Prophylactic dams -- Requirements and test methods
- Dimensional Testing O
- Tensile Testing and Elongation at Break O
- Tear Resistance and Tearing Force O
- Freedom from Holes and Visual Defects o
- O Packaging Integrity Testing
The bench testing was completed on aged and unaged samples to support the 2-year shelf life.
- Per the 2020 FDA guidance Use of International Standard ISO 10993-1 "Biological • evaluation of medical devices – Part 1: Evaluation and testing within a risk management process":
- Cytotoxicity (ISO 10993-5:2009) O
- Irritation Testing (ISO 10993-10:2010) O
- Sensitization (ISO 10993-10:2010) O
- Acute Systemic Toxicity (ISO 10993-11:2017) о
All test results were acceptable.
### g. Conclusion
The results from the non-clinical bench testing above demonstrate that the SancDam™ Latex Oral Dam are as safe and effective as the predicate device and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.